The SonarMed AMS is intended to be used to assist in verifying placement of the ETT, to assist in detecting movement of the ETT tip, and to assist in detecting obstruction of the ETT. The SonarMed AMS is intended for use by qualified personnel to assist with artificial airway management for patients in an in-hospital setting (intensive care, operating room, and emergency department settings, as well as intra-hospital transport). The SonarMed AMS is to be used as an adjunct to normal clinical practice, and is not to be used as a stand-alone diagnostic system. It is intended for use with patients who use ET tube sizes from 6.5 mm to 9.0 mm weighing >35 kilograms.
Device Story
SonarMed AMS monitors endotracheal tube (ETT) status using sonic reflection technology. System includes a sterile, single-use plastic adapter placed in-line between the ventilator circuit and the ETT, and a portable digital monitor. The adapter contains a speaker and two microphones that acoustically communicate with the tube lumen. The monitor processes acoustic signals to determine ETT tip location relative to a clinician-established baseline, quantifies passageway diameter, detects ETT movement, and identifies obstructions (percent obstructed). Output is displayed on a color LCD as waveforms or graphics. Used by clinicians in hospital settings (ICU, OR, ED) to manage artificial airways. Provides real-time adjunctive data to assist clinical decision-making regarding tube position and patency, potentially reducing complications from ETT displacement or occlusion.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical (lab) performance testing.
Technological Characteristics
System consists of a plastic disposable ETT adapter and a portable digital monitor. Adapter includes a speaker, two microphones, and a cable. Monitor features an embedded processor, graphical display, user input interface, and serial communications interface. Operates via sonic reflection technology. Powered by external supply with battery backup. Connectivity via serial interface.
Indications for Use
Indicated for patients >35 kg requiring endotracheal tubes (ETT) sizes 6.5-9.0 mm. Used by qualified personnel in hospital settings (ICU, OR, ED, transport) to assist in verifying ETT placement, detecting ETT tip movement, and detecting ETT obstruction. Adjunctive use only.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
K092611
## SonarMed™ Airway Monitoring System
## 510(k) Summary Traditional 510(k) Premarket Notification Summary of Safety and Effectiveness
| Submitter<br>Information | SonarMed, Inc.<br>5513 West 74th St, Indianapolis, IN 46268<br>317-489-3161<br>866-853-3684 | | | APR 29 2010 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----|-------------|
| Contact<br>Person | Laura Lyons<br>Vice President Clinical, Quality & Regulatory Affairs<br>317-489-3161 ext. 208<br>866-853-3684 (fax) | | | |
| Date | August 24, 2009 | | | |
| Trade Name | SonarMedTM Airway Monitoring System | | | |
| Common<br>Name | Airway Monitoring System | | | |
| Product Code | OQU | | | |
| Classification<br>Number | 21 CFR 868.5730 | | | |
| Predicate<br>Devices | Eccovision Acoustic<br>Diagnostic Imaging<br>Pharyngometer | K011329 | BXQ | 868.1800 |
| | ETView Tracheoscopic<br>Ventilation Tube | K082420 | BTR | 868.5730 |
| | PosiTube Esophageal<br>Intubation Detection | K000045 | BTR | 868.5730 |
| | RhinoScan Acoustic<br>Rhinometry | K000406 | BXQ | 868.1800 |
| | Tidal Wave SP<br>Model 710/715 | K032971 | DQA | 870.2700 |
| Device<br>Description | The SonarMed AMS is comprised of a SonarMed Monitor (Monitor) that is used<br>in conjunction with a sterile, single-use SonarMed Adapter (Adapter) and<br>software that operates the Monitor and Adapter. The Monitor is powered from<br>an external power supply and has a battery backup. When in use, the<br>SonarMed Adapter is placed in-line between the ventilator circuit and the | | | |
Image /page/0/Picture/4 description: The image shows the word "SONARMED" in a stylized font. To the right of the word, there is a graphic of curved lines emanating from a central point, resembling sonar waves. The text and graphic are in black and white.
Page 1 of 3
{1}------------------------------------------------
proximal end of the endotracheal tube (ETT) of a patient who is connected to a ventilator.
Using sonic reflection technology, signals from the Adapter allow the color LCD on the Monitor to display to a clinician:
| on the Monitor to display to a clinician. | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The baseline location of the ETT tip as established by the clinician Quantification of passageway diameter relative to the ETT diameter (for example a reading of less than 1 may indicate the tube is in a passageway that is smaller than itself, such as the esophagus) ETT movement relative to the baseline location ETT occlusion / obstruction information including percent obstructed |
| The clinician can choose whether to view information about the patient's airway<br>in either a waveform or graphic on the Monitor's color LCD. This information<br>should only be used in an adjunctive manner to assist with management of the<br>artificial airway of the patient. | |
| Intended Use | The SonarMed AMS is intended to be used to assist in verifying placement of<br>the ETT, to assist in detecting movement of the ETT tip, and to assist in<br>detecting obstruction of the ETT. |
| | The SonarMed AMS is intended for use by qualified personnel to assist with<br>artificial airway management for patients in an in-hospital setting (intensive care,<br>operating room, and emergency department settings, as well as intra-hospital<br>transport). |
| | The SonarMed AMS is to be used as an adjunct to normal clinical practice, and<br>is not to be used as a stand-alone diagnostic system. |
| | It is intended for use with patients who use ET tube sizes from 6.5 mm to 9.0<br>mm weighing >35 kilograms. |
| Comparison to<br>Predicate<br>Devices | The SonarMed Airway Monitoring System is similar or identical in technology,<br>intended use, performance and environments of use. |
| Technological<br>Characteristics | The SonarMed AMS consists of two components: a plastic disposable ET tube<br>Adapter and a portable digital Monitor. The SonarMed AMS Adapter contains a<br>speaker, two microphones, and a cable. The speaker and microphones, which<br>acoustically communicate with the tube lumen via ports, have electrical<br>connections running to the cable exiting the Adapter. The cable connects the<br>speaker and microphones to a portable digital Monitor. The Monitor contains an<br>embedded processor, a graphical display, a user input interface, and a serial<br>communications interface. Both the Eccovision Acoustic Diagnostic Imaging<br>Pharyngometer and the RhinoScan Acoustic Rhinometry use the same<br>technology. |
| | The Tidal Wave SP, ETView Tracheoscopic Ventilation Tube and the PosiTube<br>Esophageal Intubation Detection verify placement within the airway. The Tidal<br>Wave SP and the ETView Tracheoscopic Ventilation Tube detect movement of<br>the endotracheal tube. The Acoustic Diagnostic Imaging Pharyngometer and<br>the Tidal Wave SP detect obstruction in the airway. |
| Performance | The results of non-clinical (lab) performance testing demonstrate that the device |
Performance
Image /page/1/Picture/4 description: The image shows the word "SONARMED" in a bold, sans-serif font. The letters are evenly spaced and appear to be in all caps. To the right of the word, there is a graphic that resembles a radar or sonar signal, with curved lines emanating from a central point. The overall design is simple and modern.
Page 2 of 3
{2}------------------------------------------------
is safe and effective. of Non-Clinical
:
Clinical data and conclusions were not needed for this device.
End of document.
1. The state of the state
Image /page/2/Picture/3 description: The image shows the word "SONARMED" in a stylized font. To the right of the word is a graphic that resembles sound waves emanating from a point. The graphic is composed of several curved lines that get progressively larger as they move away from the point of origin.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure.
APR 2 9 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Laura Lyons Vice President, Clinical, Quality & Regulatory Affairs SonarMed, Incorporated 5513 West 74th Street Indianapolis, Indiana 46268
Re: K092611
Trade/Device Name: SonarMed™ Airway Monitoring System Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: OQU Dated: April 27, 2010 Received: April 28, 2010
Dear Ms. Lyons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2- Ms. Lyons
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Th for
Anthony Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number:
Device Name: SonarMed TM Airway Monitoring System
Indications For Use:
The SonarMed™ AMS is used to assist in verifying placement of the endotracheal tube (ETT), to assist in detecting movement of the ETT tip, and to assist in detecting obstruction of the ETT.
The SonarMed AMS is intended for use by qualified personnel to assist with artificial airway management for patients in an in-hospital setting (intensive care, operating room, and emergency department settings, as well as intra-hospital transport).
The SonarMed AMS is to be used as an adjunct to normal clinical practice, and is not to be used as a stand-alone diagnostic system.
It is intended for use with patients who use ET tube sizes from 6.5 mm to 9.0 mm, with a body weight of >35 kg.
L. Schullton
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K092611
AND/OR Over-The-Counter Use Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.