K220898 · Linde Gas & Equipment, Inc. · MRN · Apr 27, 2022 · Anesthesiology
Device Facts
Record ID
K220898
Device Name
NOxBOXi Nitric Oxide Delivery System
Applicant
Linde Gas & Equipment, Inc.
Product Code
MRN · Anesthesiology
Decision Date
Apr 27, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5165
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspirator lines of a patient undergoing nitric oxide therapy (iNO). The NOxBOXi Nitric Oxide Delivery System includes: · The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode. · Continuous monitoring and alarms for NO, O2 and NO2. • The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals. The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.
Device Story
The NOxBOXi is a nitric oxide (NO) delivery and monitoring system for patients undergoing iNO therapy. It takes NO gas from cylinders and O2 from a hospital supply as inputs. The head unit delivers NO in an intelligent, synchronous manner to the patient's inspiratory ventilator line. It features an integrated NOxMixer for backup continuous flow delivery of NO mixed with O2, useful during manual bagging or patient transit. The device continuously monitors NO, NO2, and O2 levels, providing alarms to ensure patient safety. Used in hospital settings by healthcare professionals, the system provides real-time gas concentration data to clinicians, allowing for precise titration of therapy. The device benefits patients by maintaining constant, controlled NO concentrations, reducing the risk of hypoxia or toxic NO2 exposure. This submission adds compatibility with additional ventilators and introduces a modified sample line accessory.
Clinical Evidence
No clinical data was required to support substantial equivalence. Evidence is based on bench testing, including verification of functional performance, gas delivery accuracy, and compatibility testing with additional ventilators.
Technological Characteristics
System includes head unit, NOxMixer, NOxFLOW sample line, feed hoses, regulators, and monitors. Delivers NO (0-80 ppm) and monitors NO, NO2, O2. Battery backup provides 4 hours of operation. Connectivity includes integration with various ventilators. Complies with IEC 60601-1, IEC 60601-1-2, ISO 80601-2-55, and ISO 14971. Software lifecycle follows IEC 62304.
Indications for Use
Indicated for healthcare professional use for delivery and monitoring of nitric oxide (NO) therapy in patients (currently neonates) undergoing inhaled nitric oxide therapy (iNO).
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 27, 2022
Linde Gas & Equipment Inc. % Olga Lewis Consultant, Quality and Regulatory Affairs Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K220898
Trade/Device Name: NOxBOXi Nitric Oxide Delivery System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: Class II Product Code: MRN, MRO, MRP, MRQ, CCL Dated: March 28, 2022 Received: March 28, 2022
Dear Olga Lewis:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220898
Device Name NOxBOXi Nitric Oxide Delivery System
### Indications for Use (Describe)
NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspirator lines of a patient undergoing nitric oxide therapy (iNO).
The NOxBOXi Nitric Oxide Delivery System includes:
· The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
· Continuous monitoring and alarms for NO, O2 and NO2.
• The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.
The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Special 510(k) Summary
## NOxBOXi Nitric Oxide Delivery System
### 1. Submission Sponsor
Linde Gas & Equipment Inc. (LG&E) 175 E Park Drive Tonawanda, NY 14150 USA Phone: (412) 874.3315 Dave Loflin Director of Quality and FDA Regulations
### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Olga Lewis; Consultant, Quality and Regulatory Affairs
## 3. Date
March 25, 2022
#### Device Identification 4.
| Trade/Proprietary Name: | NOxBOXi Nitric Oxide Delivery System |
|---------------------------------------------|------------------------------------------------------------------|
| Classification Names: | Nitric Oxide administration apparatus, back-up and gas analyzers |
| Common/Usual Name: | Nitric Oxide administration apparatus - primary |
| | Nitric Oxide administration apparatus - backup |
| | Nitric Oxide Analyzer |
| | Nitrogen Dioxide Analyzer |
| | Oxygen Gas Analyzer |
| Classification Regulation and Product Code: | 21 CFR 868.5165 - primary delivery system MRN |
| Additional Regulations and Product Codes: | 21 CFR 868.5165 - backup delivery system MRO |
| | 21 CFR 868.2380 – Nitric Oxide Analyzer MRP |
| | 21 CFR 868.2385 - Nitrogen Dioxide Analyzer MRQ |
| | 21 CFR 868.1720 - Oxygen Gas Analyzer CCL |
| Device Class: | Class II |
| Classification Panel: | Anesthesiology |
### 5. Legally Marketed Predicate Device
Primary Predicate Device: NOxBOXi Nitric Oxide Delivery System K201339 Additional Predicate Device: NOxBOXi Nitric Oxide Delivery System K171696
{4}------------------------------------------------
#### 6. Device Description
The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide and Oxygen levels in the inspiratorylimb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.
The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXiwill deliver nitric oxide in a synchronous manner to a single patient.
An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with Oxygen (O₂) for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.
The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O3 & NO2 monitors, power supply, drainage syringe, Operating Manual & Technical Guide.
Optional accessories include 4 separate NOxKITs (22mm, 12 mm, & 10mm), one way valve for HFOV (High Frequency Oscillatory Ventilation), bagging kits (hyperinflation & selfinflating) and circuit reducers (22f - 15m).
This submission is for the addition of compatibility claims for specific ventilators and is not related to product changes. It also introduces a new accessory, a modified sample line. There are no changes to the indications for use of the product and there are no significant design changes.
#### 7. Indication for Use Statement
NOxBOXi Nitric Oxide DeliverySystem is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).
The NOxBOXi Nitric Oxide Delivery System includes:
- The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
- · Continuous monitoring and alarms for NO, O2 and NO2.
- The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.
The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to this material prior to use.
### 8. Substantial Equivalence Discussion
The following table compares the NOxBOXi to the predicate device with respect to intended use, technological characteristics and principles of operation.
{5}------------------------------------------------
| | Subject: NOxBOXi Nitric Oxide<br>Delivery System | Predicate: NOxBOXi Nitric Oxide<br>Delivery System | Comparison | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--|
| 510(k) Number | K220898 | K201339 | N/A<br>No Change | |
| Manufacturer | NOxBOX Ltd. | | No Change | |
| | Regulatory & Indications for Use | | | |
| Product Code | MRN, MRO, MRP, MRQ, CCL | | No Change | |
| Regulation Numbers | 21 CFR 868.5165, .2380, .2385, .1720 | | No Change | |
| Regulation Name | Nitric Oxide administration apparatus, back-up and gas analyzers | | No Change | |
| Indications for Use | NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare<br>professionals for the delivery and monitoring of a constant (user set)<br>concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the<br>inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).<br><br>The NOxBOXi Nitric Oxide Delivery System includes:<br>The NOxBOXi head unit, which delivers NO gas while in the intelligent<br>delivery mode.Continuous monitoring and alarms for NO, O2 and NO2.The integrated NOxMixer which provides a backup NO delivery<br>capability that is intended to deliver a continuous flow of NO, mixed<br>with O2, for iNO therapy and provides a continuous treatment option<br>during transit and transfer conditions within hospitals.<br>The NOxBOXi Nitric Oxide Delivery System must only be used in accordance<br>with the indications, contraindications, warnings and precautions described<br>in the nitric oxide drug packaging inserts and labeling (currently neonates).<br>Refer to this material prior to use. | | No Change | |
| | Technical | | | |
| NO administration | NO blended with O₂ in the patient's inhalation circuit | | No Change | |
| NO flow rate (sample flow<br>rate) | 225 ml/min | | No Change | |
| NO concentration<br>provided | 0.0 TO 80ppm | | No Change | |
| NO monitor | Yes | | No Change | |
| O2 monitor | Yes | | No Change | |
| Monitoring accuracy | NO & NO2-+/-2% or 0.2ppm | | No Change | |
| NO2 monitor & alarm | Yes | | No Change | |
| Battery Backup capability | 4 hours without AC power | | No Change | |
| Manual bagging & back<br>up system | NOxMIXER® | | No Change | |
| NO dosing range in<br>manual mode | 0 - 185ppm on 800ppm cylinders | | No Change | |
| NO dosing Accuracy in<br>manual mode | ± 20% or 2 ppm, whichever is the greater for NO doses from 5 - 80 ppm (800 ppm drug cylinder) and O2 flow rates of 5 - 14 L/min | | No Change | |
| | Subject: NOxBOXi Nitric Oxide<br>Delivery System | Predicate: NOxBOXi Nitric Oxide<br>Delivery System | Comparison | |
| | * +/-40% or 4 ppm, whichever is the greater; for NO doses from 0 to < 5 ppm<br>or > 80 to 185 ppm (800 ppm drug cylinder) and O2 flow rates of 2 to < 5<br>L/min or > 14 to 25 L/min | | | |
| NO flow in manual mode | Adjustable 50 – 600 mL/min of NO/N2 | | No Change | |
| O2 flow range in manual<br>bagging mode | 2 to 25 L/min of O2 | | No Change | |
| Oxygen inlet pressure | 3.5 – 4.5 bar | | No Change | |
| NO delivery pressure | 1.65bar from manual control valve | | No Change | |
| Manual bagging & back<br>up system | NOxMIXER® | | No Change | |
| Dimensions | 65 mm (W) X 185 mm (H) x 60.8 mm (D) | | No Change | |
| Pre-use set up time | Instant set-up | | No Change | |
| Monitoring during manual<br>bagging | Yes | | No Change | |
| Alarms active during<br>bagging | Yes | | No Change | |
| Stand alone vs Built-in | Built-in | | No Change | |
| Can be used as a back-up<br>function | Yes | | No Change | |
| Back-up accuracy | ± 20% or 2 ppm, whichever is the greater for NO doses from 5 - 80 ppm (800 ppm drug cylinder), * +/-40% or 4 ppm, whichever is the greater; for NO<br>doses from 0 to < 5 ppm or > 80 to 185 ppm (800 ppm drug cylinder) | | No Change | |
| | Ventilator Compatibility | | | |
| Compatible Ventilators | Various models from the following<br>manufacturers:<br>● Bio-Med Devices<br>● Bunnel<br>● Carefusion<br>● Carefusion / SensorMedics<br>● Drägerwerk<br>● Fisher & Paykel Healthcare<br>● Hamilton Medical<br>● IMT Medical (Vyaire)<br>● Maquet (Getinge)<br>● Newport (Covidien)<br>● Percussionaire<br>● Philips Respironics<br>● Puritan Bennett (Covidien)<br>● Smiths Medical | Various models from the following<br>manufacturers:<br>● (N/A)<br>● Bunnel<br>● Carefusion<br>● Carefusion / SensorMedics<br>● Drägerwerk<br>● Fisher & Paykel Healthcare<br>● Hamilton Medical<br>● (N/A)<br>● Maquet<br>● Newport (Covidien)<br>● (N/A)<br>● Philips Respironics<br>● Puritan Bennett (Covidien)<br>● Smiths Medical | Equivalent;<br>testing<br>shows no<br>new<br>questions<br>raised<br>regarding<br>safety and<br>effectiveness | |
## Comparison of Characteristics With Changes From Device Cleared in K201339
{6}------------------------------------------------
### Non-Clinical and Usability Performance Data 9.
{7}------------------------------------------------
NOxBOXi has been verified and validated to ensure that it meets its functional, performance, and usability specifications and requirements. The device has been tested in compliance to international standards and US FDA guidance documents including the following:
- FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of . medical devices – Part 1: Evaluation and testing within a risk management process" (K171696)
- . FDA guidance "Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer" including accuracy of NO delivery, response of NO delivery to external perturbations and user changes, purity of NO drug delivery, acceptable / minimal production of NO2, NO analyzer accuracy, NO2 analyzer accuracy, and compatibility testing of ventilators listed in the product labeling (K171696, K201339 and subject submission)
- . FDA guidance "Content of Premarket Submissions for Software Contained in Medical Devices" (K171696)
- FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices." (K171696)
- ISO 14971: Medical Devices – Application of Risk Management to Medical Devices (K171696, K201339 and subject submission)
- ISO 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process (K171696)
- ISO 10993-5: Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity (K171696)
- . ISO 10993-10: Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization (K171696)
- IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance (K171696)
- IEC 60601-1-2: Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests (K171696)
- IEC 62366: Medical Devices - Application of Usability Engineering to Medical Devices (K171696)
- ISO 80601-2-55: Medical Electrical Equipment – Part 2-55: Particular Requirements for the Basic Safety and Essential Performance of Respiratory Gas Monitors (K171696)
- IEC 62304: Medical Device Software – Software Life Cycle Processes (K171696)
- ISO 15223-1: Medical Devices - Symbols to be Used With Medical Device Labels, Labelling, and Information to be Supplied – Part 1: General Requirements (K171696, K201339 and subject submission)
Additionally, gases delivered through the NOxBOXi system were analyzed for the presence of volatile organic compounds (VOC) and particulate matter. Measurement of the resulting VOC concentrations resulted in levels that were three orders of magnitude below OSHA permissible exposure levels. Particulate testing determined that the gas delivered by the NOxBOXi system contained particulate levels well below the EPA's maximum limits for total suspended particulates.
{8}------------------------------------------------
Testing for this submission was limited to the aspects that could be affected by including compatibility with additional ventilators and introduction of a new accessory. No additional usability testing was conducted for this submission. No clinical testing was required to support substantial equivalency of this medical device.
### 10. Conclusions/Statement of Substantial Equivalence
This submission is for the addition of compatibility claims for specific ventilators and introduction of a new accessory. This submission is not related to product changes. There are no changes to the indications for use of the product. The above-described non-clinical data support the substantial equivalence of the device and the compatibility with additional ventilators. The NOxBOXi passed all testing and no different questions of safety or effectiveness were raised. The information provided within this premarket notification supports substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.