K171696 · Praxair Distribution, Inc. · MRN · Oct 2, 2018 · Anesthesiology
Device Facts
Record ID
K171696
Device Name
NOxBOXi Nitric Oxide Delivery System
Applicant
Praxair Distribution, Inc.
Product Code
MRN · Anesthesiology
Decision Date
Oct 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5165
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO). The NOxBOXi Nitric Oxide Delivery System Includes: - The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode. - Continuous monitoring and alarms for NO, O2, and NO2. - The integrated NOxMixer, which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals. The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.
Device Story
NOxBOXi delivers nitric oxide (NO) gas to patients undergoing iNO therapy; inputs include NO gas source and ventilator waveform data. Device uses intelligent delivery mode for synchronous NO administration; integrated NOxMixer provides backup continuous flow NO/O2 delivery for manual bagging or transit. Used in hospital settings (NICU, transport) by healthcare professionals. System monitors NO, O2, and NO2 levels in inspiratory limb; provides alarms for safety. Output allows clinicians to maintain precise NO concentrations; helps manage respiratory status in neonates. Device includes head unit, sample lines, regulators, and power supply.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety/EMC (ISO 60601), gas monitoring accuracy (ISO 80601-2-55), software verification/validation (ISO 62304), and usability testing (ISO 62366).
Technological Characteristics
Class II device. Components: head unit, NOxMixer, NOxFLOW sample line, regulators. Sensing: NO, O2, NO2 electrochemical/gas sensors. Connectivity: standalone. Sterilization: not specified. Software: Major level of concern. Complies with ISO 10993-1, ISO 60601-1, ISO 60601-1-2, ISO 80601-2-55.
Indications for Use
Indicated for neonates undergoing inhaled nitric oxide (iNO) therapy. Used by healthcare professionals for delivery and monitoring of constant NO concentrations and monitoring of NO2 and O2 in inspiratory ventilator lines.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
Predicate Devices
INO Therapeutics' INOmax DS Delivery System (K092545)
INO Therapeutics' INOblender Nitric Oxide Administration Back-up system (K122689)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Praxair Distribution, Inc Michael Skrjanc Executive Director, Quality, Regulatory Compliance and Product Safety 10 Riverview Drive Danbury, Connecticut 06810
Re: K171696
Trade/Device Name: NOxBOXi Nitric Oxide Delivery System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: Class II Product Code: MRN, MRP, MRQ, CCL, MRO Dated: October 1, 2018 Received: October 1, 2018
# Dear Michael Skrjanc:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd D. Courtney -
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171696
Device Name NOxBOXi Nitric Oxide Delivery System
#### Indications for Use (Describe)
The NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).
The NOxBOXi Nitric Oxide Delivery System Includes:
- · The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
- · Continuous monitoring and alarms for NO, O2, and NO2.
- The integrated NOxMixer, which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.
The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary NOxBOXi Nitric Oxide Delivery System K171696
#### 1. Submission Sponsor
Praxair Distribution, Inc. (PDI) 10 Riverview Dr. Danbury, CT 06810 Phone: 330.949.3324 Contact: Mike Skrjanc, Executive Director, Quality, Regulatory Compliance and Product Safety
### 2. Submission Correspondent
Michael E. Skrjanc Executive Director, Quality, Regulatory Compliance and Product Safety Praxair Distribution, Inc. 10 Riverview Drive Danbury, CT 06810
### 3. Date Revised
September 23, 2018
### 4. Device Identification
| Trade/Proprietary Name: | NOxBOXi Nitric Oxide Delivery System | |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Classification Names: | Nitric Oxide administration apparatus, back-up and gas<br>analyzers | |
| Common/Usual Name: | Nitric Oxide administration apparatus – primary<br>Nitric Oxide administration apparatus – backup<br>Nitric Oxide Analyzer<br>Nitrogen Dioxide Analyzer<br>Oxygen Gas Analyzer | |
| Classification Regulation<br>and Product Code: | 21 CFR 868.5165 - primary | MRN |
| Additional Product Codes: | 21 CFR 868.5165 – backup<br>21 CFR 868.2380 – Nitric Oxide Analyzer<br>21 CFR 868.2385 - Nitrogen Dioxide Analyzer<br>21 CFR 868.1720 – Oxygen Gas Analyzer | MRC<br>MRP<br>MRC<br>CCL |
| Device Class:<br>Classification Panel: | Class II<br>Anesthesiology | |
{4}------------------------------------------------
# 5. Legally Marketed Predicate Device(s)
There are two predicate devices chosen for this submission.
- 1. K092545 INO Therapeutics' INOmax DS Delivery System
- 2. K122689 INO Therapeutics' INOblender Nitric Oxide Administration Back-up system.
Note that both the functionality of the INOmax DS Delivery System and INOblender NO administration backup unit are integrated into the NOxBOXi Nitric Oxide Delivery System.
# 6. Device Description
The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide and Oxygen levels in the inspiratory limb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.
The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXi will deliver nitric oxide in a synchronous manner to a single patient.
An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with Oxygen (O₂) for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.
The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O₂ & NO2 monitors, power supply, drainage syringe, Operating Manual & Technical Guide.
Optional accessories include 4 separate NOxKITs (22mm, 12 mm, & 10mm), one way valve for HFOV (High Frequency Oscillatory Ventilation), bagging kits (hyperinflation & selfinflating) and circuit reducers (22f – 15m).
{5}------------------------------------------------
# 7. Indication for Use Statement
NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).
# The NOxBOXi Nitric Oxide Delivery System includes:
- The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode. ●
- Continuous monitoring and alarms for NO, O2 and NO2.
- . The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.
The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to this material prior to use.
# 8. Substantial Equivalence Discussion
The following table compares the NOxBOXi to the predicate device with respect to intended use, technological characteristics and principles of operation.
| Comparison of Characteristics | | | |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Praxair Distribution, Inc. | INO Therapeutics | Comparison |
| Trade Name | <i>NOxBOXi</i> Nitric Oxide Delivery<br>System | INOmax DS | |
| 510(k) Number | K171696 | K092545<br>K122689 | N/A |
| Product Code | MRN, MRO, MRP, MRQ | MRN, MRO, MRP, MRQ | Identical |
| Regulation<br>Number | 21CFR 868.5165, .2380, .2385 | 21 CFR 868.5165, .2380, .2385 | Identical |
| Regulation Name | Nitric Oxide administration<br>apparatus, back-up and gas<br>analyzers | Nitric Oxide administration<br>apparatus, back-up and gas<br>analyzers | Identical |
| Indications for<br>Use | <i>NOxBOXi</i> Nitric Oxide<br>Delivery System is intended<br>for use by healthcare<br>professionals for the<br>delivery and monitoring of<br>a constant (user set)<br>concentration of nitric<br>oxide (NO) and the<br>monitoring of NO2 and O2<br>in the inspiratory ventilator | The INO Therapeutics INOmax<br>DS delivers INOmax (nitric<br>oxide of inhalation) therapy<br>gas into the inspiratory limb<br>of the patient breathing<br>circuit in a way that provides a<br>constant concentration of<br>nitric oxide (NO), as set by the<br>user, to the patient<br>throughout the inspired<br>breath. It uses a specially | Similar, both devices<br>provide controlled<br>concentrations of NO<br>into the inspiratory<br>ventilation circuit,<br>contain monitors and<br>alarms for NO, O2 and<br>NO2.<br><br>The <i>NOxBOXi</i> battery<br>functionality is |
| Manufacturer | Praxair Distribution, Inc. | INO Therapeutics | Comparison |
| Trade Name | <i>NOxBOXi</i> Nitric Oxide Delivery System | INOmax DS | |
| | lines of a patient undergoing nitric oxide therapy (iNO).<br><br><b>The <i>NOxBOXi</i> Nitric Oxide Delivery System includes:</b><br>The <i>NOxBOXi</i> head unit, which delivers NO gas while in the intelligent delivery mode. Continuous monitoring and alarms for NO, O₂ and NO₂. The integrated <i>NOxMixer</i> which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O₂, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals. The <i>NOxBOXi</i> Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to this material prior to use. | designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.<br><br>The INOmax DS provides continuous integrated monitoring of inspired O₂, NO₂ and NO, and a comprehensive alarm system.<br><br>The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.<br><br>The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250ml/min of NO, which along with user supplied 10L/min of oxygen, provides 20ppm in the gas flow to a patient's breathing circuit. It may also use the INOblender for backup.<br><br>The target population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and the secondary targeted clinical setting is the transport of neonates. | explained elsewhere in the labeling.<br><br>The <i>NOxBOXi</i> system allows adjustable NO gas flow rates for the backup delivery mode.<br><br>Indications for both devices follow the respective drug labeling for nitric oxide (currently indicated for use with neonates). |
| NO administration | NO blended with O₂ in the patient's inhalation circuit | NO blended with O₂ in the patient's inhalation circuit | Identical |
| NO flow rate<br>(sample flow rate) | 225 ml/min | 250 ml/min | Minor difference, The <i>NOxBOXi</i> system meets its measuring accuracy requirements with a lower sample flow rate from the |
| Manufacturer<br>Trade Name | Praxair Distribution, Inc.<br><i>NOxBOXi</i> Nitric Oxide Delivery<br>System | INO Therapeutics<br>INOmax DS | Comparison |
| | | | circuit. |
| NO concentration<br>provided | 0.0 TO 80ppm | 0-80ppm (800ppm cylinder) | Identical |
| NO monitor | Yes | Yes | Identical |
| O₂ monitor | Yes | Yes | Identical |
| Monitoring<br>accuracy | NO & NO₂ - +/- 2% or 0.2ppm | +/-20% or 2ppm (whichever is<br>greater) | Different, however,<br>substantially<br>equivalent. The<br>difference is that the<br>NOxBOXi system<br>measures with more<br>accurate specifications.<br><br>The NOxBOXi uses a<br>closed loop system<br>automatically adjusting<br>the dose to achieve a<br>set point within +/-2%<br>or 0.2 ppm (whichever<br>is greater). |
| NO₂ monitor &<br>alarm | Yes | Yes | Identical |
| Battery Backup<br>capability | 4 hours without AC power | 6 hours without AC power | Different, however,<br>acceptable for<br>emergency use<br>situations. The battery<br>is only to be used as a<br>backup in event of<br>facility power loss.<br>Most hospitals have<br>generators to provide<br>backup power. In a<br>worst case scenario,<br>the unit can be<br>switched to manual<br>override for gas<br>delivery without<br>monitoring (purely<br>mechanical). |
| Complies with ISO<br>10993-1 | Yes | Yes | Identical |
| Electrical Safety<br>Testing (ISO<br>60601-1 3rd ed)<br>Passed | Yes | Yes | Earlier versions of the<br>predicate device were<br>compliant with this<br>standard. |
| EMC Testing (ISO<br>60601-1-2) | Yes | Yes | Earlier versions of the<br>predicate device were<br>compliant with this<br>standard. |
| Manufacturer | Praxair Distribution, Inc. | INO Therapeutics | Comparison |
| Trade Name | NOX <i>BOXi</i> Nitric Oxide Delivery<br>System | INOmax DS | |
| Manual bagging &<br>back up system | NOxMIXER® | INOblender® | |
| NO dosing range in<br>manual mode | 0 - 185ppm on 800ppm cylinders | 5-80ppm on 800ppm NO; 2.5-<br>40ppm on 400ppm NO | Different, however,<br>substantially<br>equivalent. The<br>NOXBOXi has an<br>expanded user<br>selectable dose range. |
| NO dosing<br>Accuracy in<br>manual mode | ± 20% or 2 ppm, whichever is the<br>greater for NO doses from 5 - 80<br>ppm (800 ppm drug cylinder) and<br>O2 flow rates of 5 - 14 L/min<br>* +/-40% or 4 ppm, whichever is<br>the greater; for NO doses from 0<br>to < 5 ppm or > 80 to 185 ppm<br>(800 ppm drug cylinder) and O2<br>flow rates of 2 to < 5 L/min or > 14<br>to 25 L/min | +/- 20% indicated or 2 ppm,<br>whichever is greater | Different, however,<br>substantially<br>equivalent. The<br>NOxBOXi has an<br>expanded user<br>selectable dose range.<br>Similar accuracy for the<br>limited INOBlender<br>range. |
| NO flow in manual<br>mode | Adjustable 50 – 600 mL/min of<br>NO/N2…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.