The INOmax DSIR® Plus MRI delivery system is indicated for delivery of INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. The INOmax DSIR® Plus MRI is indicated for use only with MR Conditional ventilators validated to be compatible, as identified in the device labeling. The INOmax DSIR® Plus MRI is indicated for continuous integrated monitoring of inspired O2, NO2, and NO. The INOmax DSIR® Plus MRI is considered MR Conditional with the use of 1.5 Tesla and 3.0 Tesla static magnetic field scanners ONLY in areas where the field strength is less than 100 gauss. The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is a clinical 1.5 Tesla and 3.0 Tesla diagnostic imaging environment.
Device Story
Device delivers nitric oxide (NO) gas to neonates in MRI environments; utilizes dual-channel architecture. Channel 1: delivery CPU, flow controller, injector module; tracks ventilator waveforms for synchronized, proportional NO delivery. Channel 2: monitoring system with separate CPU, gas cells (NO, NO2, O2), and user interface. System includes battery for 6-hour operation and backup delivery mode (fixed 250 mL/min NO flow). Used in 1.5T/3.0T diagnostic imaging suites; operated by clinicians. Monitoring system triggers automatic NO delivery shutdown if NO concentration exceeds 100 ppm for 12 seconds. Device features GaussAlert sensors for MRI safety. Output allows clinicians to maintain therapeutic NO levels during imaging; benefits patient by enabling continuous therapy during diagnostic procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), alarm system testing (IEC 60601-1-8), and MRI-specific testing (ASTM F2052, F2119, F2503). Compatibility with respiratory devices verified via bench testing of O2 dilution, delivered pressures, delivery accuracy, and NO2 generation across multiple NO concentrations (0-80 ppm). Formative usability study conducted per ANSI/AAMI HE75.
Technological Characteristics
Dual-channel delivery and monitoring system. Components: Delivery unit, INOblender, cart, gas tanks. Hardware: Automatic locking casters, GaussAlert sensors. Connectivity: Standalone with integrated monitoring. Software: Version 3.1.2. MR Conditional (1.5T/3.0T, <100 gauss). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ASTM F2052, ASTM F2119, ASTM F2503.
Indications for Use
Indicated for delivery of INOMAX (nitric oxide) therapy gas to neonates via inspiratory limb of breathing circuit; provides constant NO concentration throughout inspired breath. Indicated for continuous monitoring of inspired O2, NO2, and NO. For use only with validated MR Conditional ventilators in 1.5T or 3.0T MRI environments where field strength is <100 gauss.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure, possibly representing the department's focus on health and human well-being. The figure is composed of three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 10, 2015
INO Therapeutics/Ikaria Mr. Robert Bovy Associate Director, Regulatory Affairs 6603 Femrite Drive Madison, WI 53718
Re: K143213
Trade/Device Name: INOmax DSIR® Plus MRI Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN, MRO, MRP, MRQ Dated: June 3, 2015 Received: June 4, 2015
Dear Mr. Bovy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Bovy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143213
Device Name INOmax DSIR Plus MRI
#### Indications for Use (Describe)
The INOmax DSIR® Plus MRI delivery system is indicated for delivery of INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant on of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. The INOmax DSIR® Plus MRI is indicated for use only with MR Conditional ventilators validated to be compatible, as identified in the device labeling.
The INOmax DSIR® Plus MRI is indicated for continuous integrated monitoring of inspired 02, NO2, and NO.
The INOmax DSIR® Plus MRI is considered MR Conditional with the use of 1.5 Tesla static magnetic field scanners ONLY in areas where the field strength is less than 100 gauss.
The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is a clinical 1.5 Tesla and 3.0 Tesla diagnostic imaging environment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(K) SUMMARY
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
## Submitter Information
| Date: | July 10, 2015 |
|---------------------------|---------------------------------------------------------------------------------------------|
| Company: | INO Therapeutics doing business as Ikaria<br>6603 Femrite Drive<br>Madison, Wisconsin 53718 |
| Contact Person: | Robert Bovy |
| | Associate Director, Device Regulatory Affairs |
| Email: | Robert.Bovy@Ikaria.com |
| Telephone: | 608-226-3450 |
| FAX: | 608-223-0804 |
| Secondary Contact Person: | David Trueblood |
| Email: | David.Trueblood@Ikaria.com |
| Telephone: | 608-395-3910 |
### ldentification of the Device
| Device Name: | INOmax DSIR® Plus MRI |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Nitric Oxide Administration Apparatus (primary)<br>Nitric Oxide Administration Apparatus, Back-up System<br>Nitric Oxide Analyzer<br>Nitrogen Dioxide Analyzer |
| Classification Name: | Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric<br>Oxide Backup Delivery |
| Device Classification: | Class II - 21 CFR 868.5165 |
| Product Code: | MRN (Primary), MRO, MRP, MRQ |
| Predicate Device(s) | K131686 |
{4}------------------------------------------------
# Description of Device
The INOmax DSIR® uses a "dual-channel" design to provide delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The specially designed injector module enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and a comprehensive alarm system. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shut down INOMAX® delivery if the monitored NO concentration exceeds 100 ppm for 12 consecutive seconds. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.
The INOmax DSIR Plus MRI incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax DSIR Plus MRI includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO, which along with user supplied 10 L/min of oxygen, provides 20 ppm of NO to a patient breathing circuit. The INOblender® can also be used for backup.
# Intended Use
The INOmax DSIR® Plus MRI delivery system is indicated for delivery of INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. The INOmax DSIR® Plus MRI is indicated for use only with MR Conditional ventilators validated to be compatible, as identified in the device labeling.
The INOmax DSIR® Plus MRI is indicated for continuous integrated monitoring of inspired O2, NO2, and NO.
The INOmax DSIR® Plus MRI is considered MR Conditional with the use of 1.5 Tesla and 3.0 Tesla static magnetic field scanners ONLY in areas where the field strength is less than 100 gauss.
The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is a clinical 1.5 Tesla and 3.0 Tesla diagnostic imaging environment.
# Technological Characteristics
All revisions of INOmax DSIR® utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the INOblender®, a stand/cart and the NO gas tanks. In this revision of the INOmax DSIR®, the cart hardware, the NO delivery cable and tubing, the gas sampling tubing, the software and the labeling have been updated.
{5}------------------------------------------------
# Determination of Substantial Equivalence
The modifications to INOmax DSIR® for the INOmax DSIR® Plus MRI device included hardware modifications to the cart, an update of the software to version 3.1.2 and modified labeling. The INOmax DSIR® Plus MRI has the same intended therapeutic effect and patient population as the cleared INOmax DSIR® predicate device. All features are the same except those described in the table below.
| Feature /<br>Specification | INOmax DSIR® - K131686 | INOmax DSIR® Plus MRI with four<br>respiratory care devices |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indicated Use<br>Environment | The device is not classified as MR<br>Conditional<br><br>The primary targeted clinical setting<br>is the Neonatal Intensive Care Unit<br>(NICU) and secondary targeted<br>clinical setting is the transport of<br>neonates | The device is classified as MR<br>Conditional with the use of 1.5 Tesla<br>and 3.0 Tesla static magnetic field<br>scanners<br><br>The primary targeted clinical setting<br>is a clinical 1.5 Tesla and 3.0 Tesla<br>diagnostic imaging environment |
| Cart Caster<br>Hardware | (4) manually locking casters | (2) automatically locking casters<br>with disengagement lever on cart<br>handle |
| Cart-Mounted<br>Gauss Alarms | None | (2) GaussAlert model 9957-A-<br>01828-B (100 Gauss) |
| User Convenience<br>Features | Wizards are available on the device<br>display for pre-use checkout, gas<br>cell calibration, use of backup NO<br>delivery mode and troubleshooting<br>of alarms | These same wizards are available<br>with the addition of a wizard for<br>device setup in the MRI suite |
| Labeling for<br>compatibility with<br>respiratory care<br>devices | A variety of transport, neonatal,<br>adult/pediatric, high frequency and<br>anesthesia ventilators, nasal CPAP<br>and nasal high flow cannulas | Compatible respiratory care devices<br>include:<br>Bio-Med MVP-10/MRI<br>Nasal Cannula<br>Maquet Servo-i MR Environment<br>Option<br>Pulmonetics LTV Series 1200 MR<br>Conditional |
## Comparison to Predicate Device
{6}------------------------------------------------
# Summary of Nonclinical Tests
The following quality assurance measures were applied to the modification of the system:
- . Risk analysis
- Formative usability study ●
- Requirements reviews
- . Design reviews
- Testing on unit level (module verification)
- Integration testing (system verification) ●
- . Performance testing (verification)
Support for the substantial equivalence of the INOmax DSIR® Plus MRI was provided as a result of risk management and testing which included electrical safety, performance and software tests. This testing includes conformity to the FDA recognized consensus standards and voluntary standards as follows:
- . IEC 60601-1:2005 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2007 - General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests
- . IEC 60601-1-8:2006 - General requirements for basic safety and essential performance -Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- . ASTM F2052-06 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- . ASTM F2119-07 – Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
- . ASTM F2503-13 - Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment
A formative usability study was conducted per ANSI/AAMI HE75 Human Factors Engineering, Design of Medical Devices, to aid in the development of hardware and software requirements and verification confirmed the INOmax DSIR® Plus MRI meets its system level requirements and that the new/modified features function as specified.
To confirm compatibility with the new respiratory care devices were set up and calibrated according to the manufacturer's recommendations and tested using the settings established for each respiratory care device test. The INOmax DSIR® was set up and calibrated according to the manufacturer's recommendations.
Five INOmax DSIR® settings were used [0 (baseline), 5. 10. 20. 40. and 80 ppm] for each mode of ventilation, as well as the Backup mode.
The measured values on the INOmax DSIR® were also recorded along with any anomalies found.
{7}------------------------------------------------
The testing concluded four requirements necessary for the INOmax DSIR® and the four respiratory care devices to be compatible:
- O2 dilution ●
- . Effect on delivered pressures
- INOmax DSIR® delivery accuracy .
- . NO2 generation
# Testing Conclusion:
The INOmax DSIR® performed within published specifications when used with each of the ventilators in both primary and backup delivery.
# Summary of Clinical Tests
The subject of this premarket submission, INOmax DSIR® Plus MRI, did not require clinical studies to support substantial equivalence.
# Conclusion
INO Therapeutics/Ikaria considers the INOmax DSIR® Plus MRI to be as safe and as effective as the predicate device, with performance substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.