K033779 · Pulmonox Medical, Inc. · MRN · Jan 26, 2004 · Anesthesiology
Device Facts
Record ID
K033779
Device Name
VIANOX DELIVERY SYSTEM, MODEL II
Applicant
Pulmonox Medical, Inc.
Product Code
MRN · Anesthesiology
Decision Date
Jan 26, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5165
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ViaNOx Delivery System is a nitric oxide administration device intended to add nitric oxide to gases that are to be breathed by a patient, and to be used in conjunction with a ventilator or other breathing system. The ViaNOx includes a nitric oxide, nitrogen dioxide, and oxygen monitor intended to monitor the concentration of these gases in respiratory gas mixtures during administration of nitric oxide.
Device Story
ViaNOx Delivery System (Model II) administers pharmaceutical-grade NO/N2 into patient breathing circuits. Device continuously measures inspiratory limb flow to adjust injected NO/N2 flow rate, maintaining constant gas concentration during and between breaths. System includes cart, gas plumbing, manual delivery component for resuscitators, control panel, and NO delivery module. Used in clinical settings by healthcare providers to manage nitric oxide therapy. Output provides real-time monitoring of NO, NO2, and O2 levels. Modifications include single-cylinder plumbing, configurable cart caddy for small cylinders, and oxygen supply alarm for manual delivery. Device supports clinical decision-making by ensuring precise, steady-state delivery of nitric oxide to patients on respiratory support.
Clinical Evidence
Bench testing only. Testing performed in accordance with FDA Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer (Jan 24, 2000).
Technological Characteristics
Nitric oxide administration apparatus; includes NO, NO2, and O2 monitoring. Features single or dual cylinder gas supply plumbing, configurable cart (caddy system), and manual delivery system with oxygen supply alarm. Compatible with various ventilators including Bio-Med Devices MVP 10. Class II device.
Indications for Use
Indicated for patients requiring nitric oxide administration via a ventilator or breathing system. Monitors NO, NO2, and O2 concentrations in respiratory gas mixtures.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
K033779
Image /page/0/Picture/1 description: The image shows the logo for PulmoNOx Medical Inc. The logo features the word "PulmoNOx" in a bold, sans-serif font, with the "NO" in a smaller font size and slightly raised. To the left of the word is a stick figure with arms raised. To the right of the word is a symbol that looks like a sun. Below the word is the text "PulmoNOx Medical Inc." in a smaller font.
Suite 200, 10835 - 120 Street, Edmonton, Alberta, Canada, TSH 3P9 Telephone # (780) 451-3660 Fax # (780) 452-0169
### 510(K) SUMMARY
JAN 2 6 2004
2003-12-02
Contact:
Paula Tomat Pulmonox Medical Inc. Suite 200, 10835-120 Street Edmonton. AB. Canada T5H 3P9
Telephone: 780-451-3660 Fax: 780-452-0169
Device Name: ViaNOx Delivery System™ (ViaNOx-ds™)
Common Names: Nifric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer
Predicate Device: ViaNOx Delivery System™
#### Device Description:
The ViaNOx Delivery System controls the delivery of pharmaceutical grade NO/N2 into the breathing gas stream passing down the inspiratory limb of a patient circuit. The injected flow of NO/Nz is controlled to maintain a steady concentration of NO/Nz within the inspiratory limb at all times, both during and between breaths. Constant concentration operation is accomplished by continuously measuring the flow in the inspiratory limb and adjusting the injected NO/N2 flow rate accordingly. The measure and adjust process is very rapid, and thus provides essentially immediate tracking of changes in the inspiratory flow rate and pattern.
The device consists of a cart, a gas plumbing connecting the gas supply to the device, a manual NO delivery system for use with a user supplied manual resuscitator and oxygen supply. a control panel, an NO Delivery Module and the main unit which houses the electronics and most of the software and to which all other components connect.
#### Intended Use:
"The ViaNOx Delivery System is a nitric oxide administration device intended to add nitric oxide to gases that are to be breathed by a patient, and to be used in conjunction with a ventilator or other breathing system. The ViaNOx includes a nitric oxide, nitrogen dioxide, and oxygen monitor intended to monitor the concentration of these gases in respiratory gas mixtures during administration of nitric oxide."
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called PulmoNOx Medical Inc. The logo features a stylized drawing of a person with their arms raised, next to the word "PulmoNOx" in a bold font. Above the word "PulmoNOx" is a small circle, possibly representing the sun or a molecule. Below the word "PulmoNOx" is the text "PulmoNOx Medical Inc."
## Comparison of Technological Characteristics
Other than the items listed in the table below, all other technological characteristics between the original and the new, modified device remain unchanged.
| Comparison<br>of... | Modified | Original |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Configuration | Single cylinder gas supply plumbing. | Dual cylinder gas supply plumbing. |
| | Large or small cylinder may be<br>utilized. | Large cylinder use only. |
| | Cart may be separated into two<br>parts to release "caddy" for small<br>cylinder use separate from large<br>cylinder and large cylinder stand. | Cart not configurable into separate<br>components. |
| Manual<br>delivery system | Alarm for inadequate oxygen<br>supply. | Rely on user to ensure adequate<br>oxygen supply. |
| Compatible<br>Ventilators | Bio-Med Devices MVP 10 added to<br>list of compatible ventilators | As per original submission |
#### Non-Clinical Performance Data
Non-clinical testing for the ViaNOx Delivery System was completed in accordance with the Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitroaen Dioxide Analyzer released by the FDA on January 24, 2000 as applicable to this modification only. All testing was performed as recommended where applicable and where not applicable, or where testing deviated from the recommendations, an explanation as to how the ViaNOx Delivery System met safety and efficacy concerns was documented.
#### Conclusions
Based on the non-clinical testing performance and the comparison to the predicate, the ViaNOx Delivery System is safe for use and is substantially equivalent to the predicate.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 6 2004
Ms. Paula Tomat QA/RA Manager Pulmonox Medical Incorporated Suite 200, 10835-120 Street Edmonton, Alberta T5H 3P9 CANADA
Re: K033779
Trade/Device Name: ViaNOx Delivery System, Model II Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric oxide administration apparatus Regulatory Class: II Product Code: MRN, MRP, MRP, MRQ Dated: January 7, 2004 Received: January 8, 2004
Dear Ms. Tomat:
We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your Section 910(t) process.
referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the do hearketed predicate devices marketed in
indications for use stated in the enclosure) to legally marketed predical Device indications for use stated in the cholosary to regainsment date of the Medical Device interstate comments, or to tvay 20, 1978, and stars with the provisions of Amendments, or to devices that have been roomshie approval of a premaikely
the Federal Food, Drug, and Cosmetic Act (Act) that do rice, subject to the neperal the Federal Food, Drug, and Cosmene Act (Prec) and acet the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provisions of the rec. "The genting of devices, good manufacturing practice, requirements for anilations against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III ( 11 your device is classified (see above) into chirolas Existing major regulations affecting.
(PMA), it may be subject to such additional controls. Existing II, Reris 800 to (PMA), it may be subject to such additional controllations, Title 21, Parts 800 to 898. In
your device can be found in the Code of Federal Regulations, Title 21, in the Fede your device can be found in the Code of Federal regariners.
addition, FDA may publish further announcements concerning your device in the Federal Register.
. Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA s Issualice of a substance vith other requirements
mean that FDA has made a determination that your devices with other Federal agencies mean that FDA has made a decemmanon mat-your as recerered by other Federal agencies.
{3}------------------------------------------------
Page 2 – Ms. Paula Tomat
You must comply with all the Act's requirements, including, but not limited to: registration You must comply with an the Act 3 requirements art 801); good manufacturing practice.
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and listing (21 CFR Fall 807), laboling (21 CFR Party (21 CFR Part 820); and if requirements as set form in the quality systems (Sections 531-542 of the Act);
applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); applicable, the electionic product radiation you to begin marketing your device as described.
21 CFR 1000-1050. This letter will allow you to begin marketing four started 21 CFR 1000-1030. This letter will and you to organismanning of substantial equivalence
in your Section 510(k) premarket notification. The FDA finding of substantial equival in your Seculon 510(k) premarted nounceases "callts in a classification for your of your device to ermits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), If you desire specific ativice for your device of (301) 594-4646. Also, please note the regulation please contact the Office of Comphanes as (2010) - 11:50 Part 807.97). You
entitled, "Misbranding by reference to premarket notification" (210) - the Art from the may obtain other general information on your responsibilities under the Act its toll for may obtain other general miorination on Journational and Consumer Assistance at its toll-free Division of Sinas -2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K033779
Device Name: ViaNOx Delivery System, Model II
#### Indications For Use:
The ViaNOx Delivery System is a nitric oxide administration device intended to add nitric oxide to gases that are to be breathed by a patient, and to be to add nithe oxiao to gases that ator or other breathing system. The ViaNOx includes a nitric oxide, nitrogen dioxide, and oxygen monitor intended to monitor the concentration of these gases in respiratory gas mixtures during administration of nitric oxide.
Prescription Use __ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
and the same of the same of the same of the many of the many of the many of the many of the many of the many of the many of the many of the many of the market of the many of
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence/pf CDRH, Office of Device Evaluation (ODE)
th
Page 1 of ______
Division Sign-Off Division of Anesthesiology, General H Infection Control, Dental Dev
510(k) Number: K033777
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.