K122715 · Resmed Germany, Inc. · MNT · Jun 17, 2013 · Anesthesiology
Device Facts
Record ID
K122715
Device Name
STELLAR 150
Applicant
Resmed Germany, Inc.
Product Code
MNT · Anesthesiology
Decision Date
Jun 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.
Device Story
Stellar 150 is a pressure-controlled ventilator for non-dependent, spontaneously breathing patients. Device uses a microprocessor-controlled blower to generate airway pressure via a single-limb vented circuit. Supports CPAP, Bi-level, and volume-assured pressure support (iVAPS) modes. Used in hospital, home, or mobile (wheelchair) settings; operated by clinicians or patients. This submission adds an optional Remote Alarm accessory, which provides audible and visual alerts triggered by the ventilator. Integration requires a dedicated hardware connector, internal electronics, and software support. Healthcare providers use the Remote Alarm to monitor ventilator status remotely, facilitating timely intervention during alarm conditions.
Clinical Evidence
Bench testing only. Verification testing included mechanical and environmental testing (IEC 60601-1-11:2010), electrical safety (IEC 60601-1:2005), and EMC (IEC 60601-1-2:200). Regression analysis identified changed features; end-to-end testing confirmed performance. Compatibility and verification testing for the new Remote Alarm accessory demonstrated adherence to predetermined acceptance criteria.
Indicated for non-dependent, spontaneously breathing adult and pediatric patients (≥13 kg) with respiratory insufficiency or failure, with or without obstructive sleep apnea. Suitable for non-invasive or invasive (with ResMed Leak Valve) ventilation in stationary or mobile settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
# Stellar 150 Traditional 510(k) Premarket Notification
# JUN 1 7 2013
# 510(k) Summary - Stellar 150
Date prepared
Submitter
#### August 31, 2012
Sandra Grünwald Director Quality Management and Regulatory Affairs ResMed Germany Inc. Fraunhoferstr. 16 82152 Martinsried Germany Tel: +49 89 9901 -1561 Fax: +49 89 9901 -1441
| Official contact | Jim Cassi<br>Vice President, Quality Assurance Americas<br>ResMed Corp.<br>9001 Spectrum Center Blvd.,<br>San Diego CA 92123<br>USA<br>Tel: +1 858-836-6081<br>Fax: +1 858-836-5522 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary name | Stellar 150 |
| Common name | Continuous ventilator |
| Classification | 21 CFR 868.5895<br>Product code MNT<br>Ventilator, continuous, minimal ventilatory support, facility use |
| Predicate Devices | Stellar 150 (K113640)<br>Respironics Trilogy series Remote Alarm (K111610) |
| Reason for submission | New Remote Alarm accessory |
{1}------------------------------------------------
#### Intended Use
The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.
#### Device Description
The Stellar 150 is a pressure controlled ventilator using a single limb vented circuit, product code MNT, and is substantially equivalent to the already marketed Stellar 150 device (K113640). The device contains a microprocessor controlled blower that generates the required airway pressure. CPAP and Bi-level modes are implemented and the device is suitable for patients weighing above 30 lbs (13 kg) for CPAP and Bi-level modes. The device also includes a volume assured pressure support mode (iVAPS), indicated for patients above 66 lbs (30 kg).
This submission updates the Stellar 150 to include a new Remote Alarm (optional) accessory. The Remote Alarm generates an audible and visual signal when an alarm is triggered on the ventilator. For the connection of this new Remote Alarm, the Stellar 150 requires an additional connector on the back of the device as well as electronics and software support for this additional port. This new Remote Alarm triggers this 510[k].
#### Performance Data
The Stellar 150 remains unchanged from the predicate Stellar 150 (K113640) in terms of technology as well as in terms of the operation modes. Design and system verification testing in regards to mechanical and environmental testing according to IEC 60601-1-11:2010, electrical safety testing according to IEC 60601-1:2005, EMC testing according to IEC 60601-1-2:200, etc. was performed and shows that the device is substantially equivalent to the predicate device (K113640).
A rearession analysis identified the features which were changed and which test cases needed to be repeated. End-to-end testing on the appropriate test cases has been performed. The rests as well as the existing data from the K113640 submission, demonstrate that the Stellar 150 meets the predetermined acceptance criteria and is substantially equivalent to the predicate device (K113640).
The Remote Alarm was compared to the Respironics Trilogy series Remote Alarm (K111610). The result of this comparison is that the Remote Alarm is substantially equivalent to the Trilogy Remote Alarm (K111610).
The compatibility of Stellar 150 with the Remote Alarm was tested and confirmed. In addition verification testing for the Remote Alarm was performed and demonstrated that it met the predetermined acceptance criteria.
#### Substantial Equivalence
The Stellar 150 is substantially equivalent to the previously cleared predicate device Stellar 150 (K113640).
- > · Same indication for use
- Same operating principle s
- A Same technologies
- Similar manufacturing process A
The new optional Stellar accessory, the Remote Alarm, is substantially equivalent to the Remote Alarm of the Respironics Trilogy series in terms of operating principle and technology. (K111610).
{2}------------------------------------------------
#### Conclusion
The indications for use, technological characteristics, and principles of operation are similar to the predicate devices. Performance data demonstrate that the new device Stellar 150 is substantially equivalent to our predicate device Stellar 150 (K113640) and the Remote Alarm is substantially equivalent to the Respironics Trilogy series Remote Alarm (K111610).
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2013
ResMed Germany, Inc. Mr. Jim Cassi Vice President, Quality Assurance Americas ResMed Corporation 9001 Spectrum Center Boulevard SAN DIEGO CA 92123
Re: K122715
Trade/Device Name: Stellar 150 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNT Dated: June 4, 2013 Received: June 6, 2013
Dear Mr. Cassi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce.prior-to-May.28,-1976, the enactment date of the Medical-Device-Amendments.-or-todevices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/4/Picture/7 description: The image shows a letter closing with the phrase "Sincerely yours,". Below the closing is a signature and the name "Tejashri Purohit-Sheth, M.D.". The text "Clinical Deputy Director DAGRID" is also present. The word "FOR" is located at the bottom right of the image.
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
RESMED
## Indications for Use
1122715 510(k) Number (if known):
Device Name:
Stellar 150
Indications for Use:
The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
Page 1 of 1
Anya C.
Harry FIL
Digitally signed by Anya C. Harry
DN:C=US, O=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Anya C. Harry,
0.9.2342.19200300.100.1.1=0011315590
Date: 2013.06.13 09:01:05 -04'00'
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K122715
August 31, 2012
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.