PERCUTANEOUS DUAL CANNULA TRACHEOSTOMY TUBE WITH LOW PRESSURE/LOW PROFILE CUFF AND DISPOSABLE INNER CANNULA
K963732 · Mallinckrodt Medical · JOH · Jun 26, 1997 · Anesthesiology
Device Facts
Record ID
K963732
Device Name
PERCUTANEOUS DUAL CANNULA TRACHEOSTOMY TUBE WITH LOW PRESSURE/LOW PROFILE CUFF AND DISPOSABLE INNER CANNULA
Applicant
Mallinckrodt Medical
Product Code
JOH · Anesthesiology
Decision Date
Jun 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is used to provide an artificial airway in order to assist in the treatment of a variety of respiratory diseases and airway management. The device is intubated into the patient's neck and trachea through an opening that is created with dilators that are supplied in a separate Percutaneous Dilational Tracheotomy Kit. The intubation is performed by first inserting an Introducer, also supplied with the Kit, into the Trach Tube and second by passing the Introducer over a Guide Wire and finally through the dilated opening. The device is secured in place with sutures or by using a strap around the patient's neck which is attached to an integral neck plate, thereby providing for a secure artificial airway for ventilation or anesthesia equipment. The Cuff is intended to manage Tidal Volume leakage during Positive Pressure Ventilation and to assist in managing aspiration of food, fluids and secretions into the patient's lung.
Device Story
Percutaneous Dual Cannula Tracheostomy Tube provides artificial airway for adult patients; used during percutaneous dilational tracheotomy. Device consists of outer cannula with cuff, inner cannula, neck plate, and inflation system. Input: manual insertion over guide wire using introducer into trachea. Operation: cuff inflated via luer valve/syringe to manage tidal volume leakage and prevent aspiration. Output: secure airway for ventilation or anesthesia. Used in clinical settings by physicians. Benefits: smooth distal tip and inverted proximal cuff bond reduce insertion force compared to predicate; provides secure airway for respiratory support.
Clinical Evidence
No clinical data. Bench testing only. Physical integrity of cuff-to-cannula bond strength demonstrated equivalent to predicate. Insertion force testing showed reduced force compared to predicate.
Technological Characteristics
Dual cannula tracheostomy tube. Materials: polyvinyl chloride (outer cannula, cuff, inflation line, pilot balloon, luer valve, inner cannula), polypropylene (obturator, 15mm connector), polycarbonate (neck plate). Features: snap-lock inner/outer cannula connection, 15mm male conical connector, low-pressure/low-profile cuff. Sterilization: same as predicate. No software or electronic components.
Indications for Use
Indicated for adult patients requiring an artificial airway for respiratory disease treatment or airway management. Intended for use as the initial airway immediately post-tracheotomy in conjunction with Percutaneous Dilational Tracheotomy procedures. Prescription use only.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
Shiley DCT Dual Cannula Tracheostomy Tube with Low Pressure Cuff and Disposable Inner Cannula (K811447)
Shiley Cuffed Single Cannula SCT Tracheostomy Tubes (K810106)
Submission Summary (Full Text)
{0}
K963732
SECTION 9; 510(k) SUMMARY OF SAFETY AND EFFICACY: (summary pg 1 of 3)
June 26, 1997
(1) Company Name
Mallinckrodt Medical, Inc.
Address, Telephone and Contact
Corporate Headquarters
Mallinckrodt Medical, Inc.
P.O. Box 5840
St. Louis, MO 63134
Mike Schoeck
(314) 895-2318
Manufacturing and Distribution Location
Mallinckrodt Medical, Inc.
18691 Jamboree Road
Irvine, CA 92712
Tony Wondka
(714) 622-5797
Date of submission
August 1996
(2) Trade Names
Shiley Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure / Lower Profile Cuff and Disposable Inner Cannula
Common or Usual Name
Tracheostomy Tube, Adult, with Cuff
Classification
Class II
(3) Substantially Equivalent Devices Registration numbers:
- Shiley DCT Dual Cannula Tracheostomy Tube with Low Pressure Cuff and Disposable Inner Cannula - 510(k) No. K811447.
- Shiley Cuffed Single Cannula SCT Tracheostomy Tubes - 510(k) No. K810106.
(4) Intended Use: The device is used to provide an artificial airway in order to assist in the treatment of a variety of respiratory diseases and airway management. The device is intubated into the patient's neck and trachea through an opening that is created with dilators that are supplied in a separate Percutaneous Dilational Tracheotomy Kit. The intubation is performed by first inserting an Introducer, also supplied with the Kit, into the Trach Tube and second by passing the Introducer over a Guide Wire and finally through the dilated opening. The device is secured in place with sutures or by using a strap around the patient's neck which is attached to an integral neck plate, thereby providing for a secure artificial airway for ventilation or anesthesia equipment. The Cuff is intended to manage Tidal Volume leakage during Positive Pressure Ventilation and to assist in managing aspiration of food, fluids and secretions into the patient's lung.
(5) Description and Physical Characteristics: The device includes an Outer Cannula with a Cuff. The Cuff is deflated or inflated via an Inflation System comprised of tubing communicating between the Cuff and a Luer Valve to which a syringe can be attached. A Pilot Balloon is adjacent to the Luer Valve to indicate the inflation status of the Cuff. A Neck Plate is attached to the Outer Cannula. An Inner Cannula is included with an integral 15mm male conical connector. The Inner and Outer Cannula are attached together by a snap-lock mechanism. The Outer Cannula distal end is designed to have smooth transitions in outer diameter in order to minimize the amount of resistance that would occur with abrupt transitions.
(6) Materials: The device consists of an Outer Cannula, Cuff, Inflation Line, Pilot Balloon, Luer Valve and Inner Cannula all constructed with biocompatible polyvinyl chloride, an obturator and 15mm male connector both constructed with biocompatible polypropylene, and a neck plate constructed with biocompatible polycarbonate.
Mallinckrodt Medical, Inc.
510(k) Submission; Notification of Labeling Change and Modification to
Shiley DCT Cuffed Adult Tracheostomy Tube
Pg 11 of 28
{1}
SECTION 9, Continued; 510(k) SUMMARY OF SAFETY AND EFFICACY: (summary pg 2 of 3)
## (7) Differences from Predicate Devices referenced herein:
The device is not different in safety or efficacy from the predicate device(s). The design differences between the new PERC device and the Shiley 'DCT' predicate device are related to a relatively smooth tip for the PERC compared to the DCT.
### Similarities to DCT and SCT devices:
1) Dual Cannula System (same as DCT)
2) Inner Cannula in its entirety (same as DCT)
3) Outer Cannula length (same as DCT)
4) Outer Cannula ID and OD (same as DCT)
5) Cuff and Inflation System components (same as DCT)
6) Cuff attachment to Outer Cannula (same as DCT and SCT)
7) Inflation Line attachment to Outer Cannula (same as DCT)
8) Intended Use of device (same as DCT)
9) Sterilization Method (same as DCT and SCT)
### Differences from DCT device:
#### Physical Characteristics:
1) The Outer Cannula includes a distal tip section that has smooth gradual transitions in diameter. This is accomplished by eliminating the lip caused by the rim of the distal cuff bond. In addition, the proximal Cuff bond is inverted to reduce the effective diameter of this section of the device. These characteristics are intended to reduce the amount of force necessary to insert the device into the patient.
#### Labeling and User Instructions:
1) The device instruction sheet will include references to Cuff Resting Diameter, Cannula Angle, and Proximal and Distal Extensions as suggested in the ISO and CEN standards.
2) A statement will be added to the packaging indicating that the new device is compatible with Introducers provided with Percutaneous Dilational Tracheotomy Kits.
3) A Caution statement indicating that physicians using this device with a Percutaneous Dilational Tracheotomy Procedure should be properly instructed on the procedure prior to use.
4) A Caution statement indicating that prior to intubation of this device, that the preceding Percutaneous Dilational Tracheotomy Procedure be performed with visual confirmation of needle and/or guide catheter placement from inside the trachea with a bronchoscope or similar manner.
5) A statement will be added that states sterile technique should be used when inserting the device into the patient since the intubation will be performed in conjunction with a Percutaneous Dilational Tracheotomy Procedure.
6) A statement will be added to describe a method of deflating the device prior to removal from the patient to taper the cuff in the direction of removal.
Mallinckrodt Medical, Inc.
510(k) Submission; Notification of Labeling Change and Modification to Shiley DCT Cuffed Adult Tracheostomy Tube
Pg 12 of 28
{2}
SECTION 9, Continued; 510(k) SUMMARY OF SAFETY AND EFFICACY: (summary pg 3 of 3)
(8) Test Summary:
1) Non-Clinical Tests:
Physical integrity of the Cuff-to-Cannula bond strength of the new PERC device will be demonstrated to be equivalent to the predicate DCT device.
Insertion force of the new PERC device will be less than that of the predicate DCT device.
This data will be available from the manufacturer upon request.
2) Clinical Tests:
It is not required to conduct clinical tests on the new PERC product due to its similarity to the predicate DCT product and other competitor's predicate products. Moreover, we estimate that approximately 4000 DCTs are currently used in this application annually. The acceptance of the predicate DCT in this application eliminates the need to validate the acceptance of the new PERC product in this application because of the minor nature of the improvements.
3) Conclusion:
The similarity to the predicate products in conjunction together with the extensive usage of the predicate product in Percutaneous Dilational Tracheotomy applications, demonstrate that the new PERC product is safe and effective and will perform equal to or better than currently marketed predicate devices.
Mallinckrodt Medical, Inc.
510(k) Submission; Notification of Labeling Change and Modification to
Shiley DCT Cuffed Adult Tracheostomy Tube
Pg 13 of 28
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 26 1997
Mr. James E. Keller
Mallinckrodt Medical, Inc.
675 McDonnell Boulevard
P.O. Box 5840
St. Louis, Missouri 63134
Re: K963732
Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure/Low Profile Cuff and Disposable Inner Cannula
Regulatory Class: II (two)
Product Code: 73 JOH
Dated: May 8, 1997
Received: May 12, 1997
Dear Mr. Keller:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. James E. Keller
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number (if known): K963732
Device Name: Percutaneous Dual Cannula Tracheostomy Tube
## Indications For Use:
1. This device is used to provide an artificial airway in order to assist in the treatment of a variety of respiratory diseases and airway management for adult patients.
2. The device is to be used by the order of a physician.
3. The device is primarily intended for use in conjunction with Percutaneous Dilational (or Dilatational) Tracheotomy and is inserted into the patient using the Introducer provided with the Percutaneous Dilational Kit.
4. The device is intended to be used as the initial artificial airway immediately post tracheotomy.
5. There is no "new" intended therapeutic use of this device to existing devices marketed.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.