K151404 · Atos Medical AB · JOH · Feb 12, 2016 · Anesthesiology
Device Facts
Record ID
K151404
Device Name
ProTrach Dualcare
Applicant
Atos Medical AB
Product Code
JOH · Anesthesiology
Decision Date
Feb 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5800
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
ProTrach DualCare is a combined Speaking Valve and Heat and Moisture Exchanger (HME) intended for spontaneously breathing tracheostomized patients using a tracheostomy tube with a deflated cuff, or a tracheostomy tube without cuff. In HME-mode the device conditions inhaled air by retaining heat and moisture from the exhaled air. By turning the lid of the Speaking Valve into speaking mode air is re-directed to enable speech. The entire device is for single patient use and the HME-part is for single use.
Device Story
Combined speaking valve and heat and moisture exchanger (HME) for tracheostomized patients; connects to tracheostomy tube. Device features two modes: HME mode (conditions inhaled air via moisture/heat retention) and speaking mode (one-way valve redirects exhaled air through vocal folds for speech). User rotates lid to switch modes. Used in hospitals, ICU, sub-acute care, and home settings; operated by patients or caregivers. Eliminates need for manual finger occlusion of tracheostomy tube. Benefits include restored speech capability and airway humidification. Single patient use; HME component is single-use.
Clinical Evidence
Bench testing only. Verified moisture loss, airflow resistance, leakage, dead space, attachment/detachment forces, and durability. Conducted simulated use testing and usability studies. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, genotoxicity, and leachables/extractables) performed per ISO 10993 series with acceptable results.
Technological Characteristics
Materials: Calcium chloride-treated polyurethane foam (HME), silicone and plastic (valve). Interfaces: 15mm or 22mm standard connectors. Operating principle: Mechanical one-way valve and moisture-retaining filter. Biocompatibility: ISO 10993 compliant for permanent (>30 days) mucosal contact. No electronic components or software.
Indications for Use
Indicated for spontaneously breathing tracheostomized patients (adults and pediatric >10kg) using a tracheostomy tube with a deflated cuff or no cuff. Contraindicated for unresponsive/sedated patients, those with inflated cuffs, foam-filled cuffed tubes, severe airway obstruction, thick/copious secretions, or severely reduced lung elasticity.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2016
Atos Medical AB Elin Algotsson Regulatory Affairs Manager Kraftgatan 8 Horby, SE-242 22 Sweden
Re: K151404
Trade/Device Name: ProTrach DualCare Regulation Number: 21 CFR 868.5800 Regulation Name: Tracheostomy Tube and Tube Cuff Regulatory Class: Class II Product Code: JOH Dated: January 13, 2016 Received: January 15, 2016
Dear Elin Algotsson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151404
Device Name ProTrach DualCare
#### Indications for Use (Describe)
ProTrach DualCare is a combined Speaking Valve and Heat and Moisture Exchanger (HME) intended for spontaneously breathing tracheostomized patients using a tracheostomy tube with a deflated cuff, or a tracheostomy tube without cuff.
In HME-mode the device conditions inhaled air by retaining heat and moisture from the exhaled air.
By turning the lid of the Speaking Valve into speaking mode air is re-directed to enable speech.
The entire device is for single patient use and the HME-part is for single use.
Patient Population: For spontaneously breathing tracheostomized patients (adults and pediatric patients greater than 10kg in weight) using a tracheostomy tube with a deflated cuff, or a tracheostomy tube without cuff.
Environment of Use: Hospitals, ICU, sub-acute care institutions and home.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Non Confidential Summary
12 February 2016 Page 1 of 4
# Section 5 – 510(k) Summary
| Atos Medical AB<br>P.O Box 183<br>SE-242 22<br>Hörby Sweden | Tel - 011-46-415 198 00<br>Fax - 011-46-415 198 98 |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Elin Algotsson - Regulatory Affairs Manager |
| Proprietary or Trade Name: | ProTrach DualCare |
| Common/Usual Name: | Speaking Valve |
| Classification | Class II - 21 CFR 868.5800 |
| Classification Name/Code: | JOH - Tracheotomy Tube and Tube Cuff |
| Device: | ProTrach DualCare |
| Predicate Devices: | Fogless International AB - SPIRO Speaking Valve - (K122932)<br>Passy-Muir Inc.- Passy Muir Speaking Valves – (K944451) (for reference)<br>Engineered Medical Systems - TrachVox – (K013728) (for reference)<br>Atos Medical - Provox FreeHands HΜΕ – (Κ022125) (for reference) |
## Device Description:
The Speaking Valve, the HME DigiTop and the HME DigiTop O2 (sold separately) are all used to enable speaking. They will therefore be referenced to as "speaking devices".
#### Speaking Valve and HME 15 / 22
The reusable Speaking Valve is used with a single use 15 mm or 22 mm Heat and Moisture Exchanger (HME). The HME is placed so that it prevents direct contact between the Speaking Valve and the airways. This prevents the Speaking Valve from being clogged or soiled by mucus.
The Speaking Valve has two modes: speaking mode and HME mode.
In speaking mode, a flexible membrane is positioned in the airflow openings and acts as a one way valve. It opens during inhalation so the patient can inhale through the device. During exhalation, the membrane remains closed and the air is re-directed through the upper airways and the vocal folds. Thereby the patient is able to speak. In speaking mode the inhaled air does not get conditioned since the exhaled air goes out through the upper airways.
In HME mode the membrane is moved out of the airflow so that the patient both inhales and exhales through the device. The inhaled air is conditioned by the heat and moisture that is retained from the exhaled air in the impregnated HME mode, speaking is not possible. The device is switched between the modes by rotating the lid of the Speaking Valve until it clicks into the desired position.
#### HME DigiTop and HME DigiTop O2
The HME DigiTop and the HME DigiTop O2 enable use of the HME without the Speaking Valve, and can manually be occluded to enable speaking.
{4}------------------------------------------------
# Non Confidential Summary
12 February 2016 Page 2 of 4
## Indications for Use:
ProTrach DualCare is a combined Speaking Valve and Heat and Moisture Exchanger (HME) intended for spontaneously breathing tracheostomized patients using a tracheostomy tube with a deflated cuff, or a tracheostomy tube without cuff.
In HME-mode the device conditions inhaled air by retaining heat and moisture from the exhaled air.
By turning the lid of the Speaking Valve into speaking mode air is re-directed to enable speech.
The entire device is for single patient use and the HME-part is for single use.
Patient Population: For spontaneously breathing tracheostomized patients (adults and pediatric patients greater than 10kg in weight) using a tracheostomy tube with a deflated cuff, or a tracheostomy tube without cuff.
#### Environment of Use:
Environments of use include; Hospitals, ICU, sub-acute care institutions and home.
## Substantial Equivalence Comparison to Predicates
There are no significant differences between the ProTrach DualCare compared to the predicate device in terms of indications, materials, design and operating principles as summarized below.
| Attributes of predicate (SPIRO<br>Speaking Valve) | ProTrach DualCare<br>HME 15 Regular | ProTrach DualCare<br>HME 15<br>XtraMoist | ProTrach DualCare<br>HME 22 |
|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use:<br>For spontaneous breathing tracheotomy<br>patients. | Same | Same | Same |
| Environment of use:<br>Hospitals, ICU, sub-acute care<br>institutions and home | Same | Same | Same |
| Patient population:<br>Conscious patients, trained to use the<br>speech valve by qualified staff. | Same (Intended for spontaneously breathing tracheostomized<br>patients using a tracheostomy tube with a deflated cuff, or a<br>tracheostomy tube without cuff. Contraindicated for<br>unresponsive or sedated patients.) | Same (Intended for spontaneously breathing tracheostomized<br>patients using a tracheostomy tube with a deflated cuff, or a<br>tracheostomy tube without cuff. Contraindicated for<br>unresponsive or sedated patients.) | Same (Intended for spontaneously breathing tracheostomized<br>patients using a tracheostomy tube with a deflated cuff, or a<br>tracheostomy tube without cuff. Contraindicated for<br>unresponsive or sedated patients.) |
| Technology and Operating principle:<br>Eliminates the necessity of finger<br>occlusion for the patient with a<br>tracheostomy tube. | Same | Same | Same |
| Incorporated speech valve | Same | Same | Same |
| Provides heat and humidity (moisture<br>loss: 28 mgH2O/l air) | Better humidification:<br>22 mgH2O/l air | Better<br>humidification:<br>20 mgH2O/l air | Better<br>humidification:<br>21 mgH2O/l air |
| Dead space: 5.6 ml | 4.05 ml (dead space<br>does not raise concern<br>because the dead space is<br>less than predicate) | 4.05 ml (dead<br>space does not<br>raise concern<br>because the dead<br>space is less than<br>predicate) | 4.62 ml (dead<br>space does not<br>raise concern<br>because the dead<br>space is less than<br>predicate) |
| Dimensions (mm): | | | |
| Outer diameter: 26.9 | 27.6 | 27.6 | 27.6 |
| Inner diameter: 25.5 | 20.8-22.9 | 20.8-22.9 | 21.1-22.9 |
| Non Confidential Summary<br>Page 3 of 4 | | | |
| Height (contributing to dead space) 11.8 | 11.9 | 11.9 | 13.35 |
| Height (total): 26.9 | 26.1 | 26.1 | 15.2 |
| Can provide supplemental oxygen<br>through oxygen connector | Same | Same | Same |
| Intended to connect to tracheostomy<br>tube of spontaneously breathing patient | Same | Same | Same |
| Intended for single patient use | Same | Same | Same |
| HME exchanged at least every 24 h. | Same | Same | Same |
| Prescription use | Same | Same | Same |
| Standard 15/22 mm connectors | Same (15 mm connector) | Same (15 mm connector) | Same (22 mm connector) |
| Speech valve one-way closed position,<br>always open for inhalation | Same | Same | Same |
| Allows exhalation to atmosphere freely | Similar for all versions of the device. The DualCare allows<br>the user to switch between speaking mode and HME mode by<br>turning the lid of the device. In speaking mode the exhaled<br>air is directed via the upper airways and vocal folds whereas<br>in HME mode the exhaled air is directed out of the<br>tracheostoma. | | |
| Contraindications:<br>Unconscious and/or Comatose Patients | Same | Same | Same |
| Inflated Tracheostomy Tube Cuff | Same | Same | Same |
| Foam Filled Cuffed Tracheostomy Tube | Same | Same | Same |
| Severe Airway Obstruction Which May<br>Prevent Sufficient Exhalation | Same | Same | Same |
| Thick and Copious Secretions | Same | Same | Same |
| Severely Reduced Lung Elasticity That<br>May Cause Air Trapping | Same | Same | Same |
| This Device Is Not intended For Use With<br>Endotracheal Tubes | Same | Same | Same |
| Materials:<br>Materials HME: Calcium chloride<br>treated polyurethane foam | Same. The materials<br>included have been<br>assessed for<br>biocompatibility in<br>accordance with the<br>ISO 10993 series and<br>have demonstrated<br>acceptable results. | Same. The<br>materials<br>included have<br>been assessed<br>for<br>biocompatibility<br>in accordance<br>with the ISO<br>10993 series<br>and have<br>demonstrated<br>acceptable<br>results. | Same. The<br>materials<br>included have<br>been assessed<br>for<br>biocompatibility<br>in accordance<br>with the ISO<br>10993 series<br>and have<br>demonstrated<br>acceptable<br>results. |
| Materials Speaking Valve: Silicone and<br>plastic | Same. The materials<br>included have been<br>assessed for<br>biocompatibility in<br>accordance with the<br>ISO 10993 series and<br>have demonstrated<br>acceptable results. | Same. The<br>materials<br>included have<br>been assessed<br>for<br>biocompatibility<br>in accordance<br>with the ISO<br>10993 series<br>and have<br>demonstrated<br>acceptable<br>results. | Same. The<br>materials<br>included have<br>been assessed<br>for<br>biocompatibility<br>in accordance<br>with the ISO<br>10993 series<br>and have<br>demonstrated<br>acceptable<br>results. |
| | results. | results. | |
| Clinical claims in labeling and<br>supporting evidence:<br>Make the function of speech easier. | Same | Same | Same |
| The filter and Speaking valve are not<br>affected by secretions during coughing. | Same in speaking<br>mode. In HME mode<br>the filter acts as a<br>barrier preventing the<br>speaking valve from<br>being affected by<br>secretion. | Same in<br>speaking mode.<br>In HME mode<br>the filter acts as<br>a barrier<br>preventing the<br>speaking valve<br>from being<br>affected by<br>secretion. | Same in<br>speaking mode.<br>In HME mode<br>the filter acts as<br>a barrier<br>preventing the<br>speaking valve<br>from being<br>affected by<br>secretion. |
{5}------------------------------------------------
{6}------------------------------------------------
12 February 2016
Page 4 of 4
Non Confidential Summary
The differences noted in the table above are addressed by the testing summarized below:
## Summary of Non-Clinical Testing:
The following tests were performed to verify that ProTrach DualCare met applicable safety and performance requirements.
The main focus areas of testing were:
- . Verification of the moisture loss, airflow resistance, leakage test and dead space.
- . Verification of attachment and detachment forces in the different interfaces including the durability of the devices.
- Simulated use of all DualCare products and their Instructions for use.
- Verification of DualCare Warning label durability.
- . Verification of DualCare products' function after drop test, climate testing, aging and transport.
- . Biocompatibility: The proposed device was classified as a Surface device with indirect contact with mucosal membranes via air. Duration: Permanent (>30 days). The following tests were conducted in accordance with the ISO 10993 series and rendered acceptable biocompatibility results:
- cytotoxicity. ●
- sensitization. .
- intracutaneous reactivity,
- acute systemic toxicity ●
- genotoxicity
- Leachables & Extractables quantitative and qualitative testing ●
In addition, risk management according ISO 14971 and simulated use testing (usability study) has been conducted.
## Substantial Equivalence Conclusion:
Based on the comparison information and non-clinical performance testing Atos Medical AB concludes that the ProTrach DualCare is as safe, as effective, and performs as well as or better than the predicate. ProTrach DualCare is therefore considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.