To allow airflow over the vocal cords for speaking function. To be used with a standard 15 mm connection on tracheostomy tubes.
Device Story
Shikani speaking valve is a passive, ball-type check valve; attaches to standard 15 mm tracheostomy tube connectors. Device allows patient to inhale through tracheostomy tube; ball mechanism closes during exhalation, directing airflow through vocal cords to facilitate phonation. Used in hospital, sub-acute, or home settings by tracheostomy patients. Healthcare providers or patients use the device to restore speech function. Output is mechanical airflow redirection; clinical decision-making involves assessing patient's ability to tolerate valve and maintain airway patency. Benefits include improved communication for tracheostomy patients.
Clinical Evidence
Bench testing only. Performance data from original submission K982128 demonstrated equivalence in flow resistance and weight. Biocompatibility testing performed according to ISO 10993; materials passed.
Technological Characteristics
Housing: Celcon C251-A Acetyl Copolymer; Ball: Nylon 6/6 Grade 2; Pin: 316 Stainless steel. Mechanical ball-type check valve. Standard 15 mm/22 mm connections per ISO 5356-1. Weight: 32 gm. No software or electronic components.
Indications for Use
Indicated for tracheostomy patients requiring airflow over vocal cords for speech. Contraindicated for use with inflated tracheostomy tube cuffs, unconscious patients, or patients unable to exhale around the tracheostomy tube/cuff and through the upper airway. Not for in-line ventilator use.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
{0}------------------------------------------------
K100995
### Premarket Notification 510(k) Section 5 - 510(k) Summary
Shikani Medical
| | 510(k) Summary<br>Page 1 of 3<br>7-Apr-10 | |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shikani Medical LLC<br>689 Pyle Road<br>Forest Hill, MD 21050 | Tel - 410-638-1066<br>Fax - 410-638-1413 | JUN 2 5 2010 |
| Official Contact: | Fred DeBaugh - Partner | |
| Proprietary or Trade Name: | Shikani speaking valve | |
| Common/Usual Name: | Speaking valve | |
| Classification Name: | Tube tracheostomy and tube cuff<br>JOH - 21 CFR 868.5800 | |
| Predicate Devices: | Pilling Weck (Teleflex) - Shikani-French speaking valve<br>K982128 | |
| | | |
| 510(k) Summary<br>Page 2 of 3<br>7-Apr-10 | | |
| Features | Predicate<br>K982128<br>Shikani-French | Proposed<br>Shikani speaking valve<br>(Shikani-French 15) |
| Intended use | To allow airflow over the vocal cords for<br>speaking function | To allow airflow over the vocal cords for speaking function |
| Comment | The intended use is identical as the devices are the same | |
| Indications for use | To allow airflow over the vocal cords for<br>speaking function. To be used only as an<br>attachment to CL Jackson improved trachea<br>tubes | To allow airflow over the vocal cords for speaking function. To be used with<br>a standard 15 mm connection on tracheostomy tubes |
| Comment | The indications for use are identical with exception that K982128 was for the model which was only to be attached to the CL<br>Jackson tracheostomy tube. Teleflex later introduced another model for use with standard tracheostomy tubes with 15 mm<br>connections. This model was considered a line extension and not requiring a modification to the existing 510(k). | |
| | As the proposed Shikani speaking valve is the same as the Shikani-French 15 speaking valve, the indications for use are<br>identical. | |
| Environment of Use | Hospital, Sub-acute Institutions, Home | Hospital, Sub-acute Institutions, Home |
| Comment | Identical | Identical |
| Patient Population | Tracheostomy patients | Tracheostomy patients |
| Comment | K982128 did not specify patient population, | but we intend for the patient population to be the same. |
| Contraindications | Not for use in-line with a ventilator | Not for use with an inflated tracheostomy tube cuff<br><br>Not for use with an unconscious patient<br><br>Not for use with patients without the ability to exhale air around the<br>tracheostomy tube and cuff and through the upper airway. |
| Comment | | Teleflex did not include contraindications in K982128, but a review of other speaking valves suggests that the proposed<br>contraindications would be appropriate to add to our labeling.<br><br>These contraindications are not new and do not come as a result of any new risk analysis, but are felt to be informative only. |
| | | |
| | Page 3 of 3<br>7-Apr-10 | |
| Features | Predicate | Proposed |
| | K982128 | Shikani speaking valve |
| | Shikani-French | (Shikani-French 15) |
| Standard 22 / 15 mm<br>Connections Per ISO 5356-1 | Yes - it can connect to the breathing circuit<br>but has a fitting only to be attached to the CL<br>Jackson tracheostomy tubes | Yes - standard fitting to connect to the breathing circuit and to a standard<br>tracheostomy tube with a 15mm connector. |
| Comment | Identical as they are the same device | Identical as they are the same device |
| Weight (gm) | 32 gm | 32 gm |
| Comment | Identical as they are the same device | Identical as they are the same device |
| Valve type | Ball | Ball |
| Comment | Identical as they are the same device | Identical as they are the same device |
| Attachment to tube | Attachment clip | Attachment clip |
| Comment | Identical as they are the same device | Identical as they are the same device |
| Pressure drop | 1.3 cm H2O @ 30 lpm 2.0 cm H₂O @ 40 lpm | 1.3 cm H2O @ 30 lpm 2.0 cm H₂O @ 40 lpm |
| Comment | Identical as they are the same device | Identical as they are the same device |
| Materials | Housing - Celcon C251-A Acetyl<br>Copolymer Ball - Nylon 6/6 Grade 2 Pin -<br>316 Stainless steel | Housing - Celcon C251-A Acetyl Copolymer Ball - Nylon 6/6 Grade 2 Pin<br>316 Stainless steel |
| Comment | Identical as they are the same device | Identical as they are the same device |
### Device Description
The Shikani speaking valve consists of a plastic housing with a captured ball that acts as a check valve. Once attached to the tracheostomy tube connector the speaking valve allows the patient to inhale through the tracheostomy tube and exhale across the vocal cords which facilitates speech or phonation.
### Indications for Use
To allow airflow over the vocal cords for speaking function. To be used with a standard 15 mm connection on tracheostomy tubes.
### Environment of Use
Home, Hospital, Sub-acute Institutions
#### Performance Testing
Performance testing which was used in the original submission K982128 was:
- Resistance to flow comparison to predicates demonstrated that the values were . equivalent
- Weight comparison to predicates demonstrated that the values were equivalent .
- Biocompatibility according to ISO 10993 was performed and the materials passed .
The proposed device meeting and passed all the performance testing as outlined above and thus demonstrated substantial equivalence.
{1}------------------------------------------------
Premarket Notification 510(
Shikani Medica
{2}------------------------------------------------
Premarket Notification 5100
Section 5 – 510(k) Summary
Differences between Other Legally Marketed Predicate Device
There are no differences between the proposed device and the predicate device as they are the same device
{3}------------------------------------------------
# Premarket Notification 510(k)
Section 9 – Declarations of Conformity and Summary Reports
Form Approved: OMB No 0910-0120; Expiration Date: 5/31/10
Department of Health and Human Services
Food and Drug Administration
**STANDARDS DATA REPORT FOR 510(k)s**
(*To be filled in by applicant*)
This report and the Summary Report Table are to be completed by the applicant when submitting a 510(k) that refer-
ences a national or international standard. A separate report is required for each standard referenced in the 510(k).
TYPE OF 510(K) SUBMISSION
| <span style="vertical-align:middle;">☑</span> Traditional | <span style="vertical-align:middle;">☐</span> Special | <span style="vertical-align:middle;">☐</span> Abbreviated |
|-----------------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------|
|-----------------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------|
| STANDARD TITLE | ISO 5356-1:2004. Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and sockets. |
|----------------|---------------------------------------------------------------------------------------------------------|
|----------------|---------------------------------------------------------------------------------------------------------|
Please answer the following questions
| Is this standard recognized by FDA? | <span>☑</span> Yes <span>☐</span> No |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| FDA Recognition number | # 1-62 |
| Was a third party laboratory responsible for testing conformity of the device to this standard identified in the 510(k)? | <span>☐</span> Yes <span>☑</span> No |
| Is a summary report describing the extent of conformance of the standard used included in the 510(k)?<br>If no, complete a summary report table. | <span>☑</span> Yes <span>☐</span> No |
| Does the test data for this device demonstrate conformity to the requirements of this standard as it pertains to this device? | <span>☑</span> Yes <span>☐</span> No |
| Does this standard include acceptance criteria?<br>If no, include the results of testing in the 510(k). | <span>☑</span> Yes <span>☐</span> No |
| Does this standard include more than one option or selection of tests?<br>If yes, report options selected in the summary report table. | <span>☐</span> Yes <span>☑</span> No |
| Were there any deviations or adaptations made in the use of the standard?<br>If yes, were deviations in accordance with the FDA supplemental information sheet (SIS)? | <span>☐</span> Yes <span>☑</span> No |
| Were deviations or adaptations made beyond what is specified in the FDA SIS?<br>If yes, report these deviations or adaptations in the summary report table. | <span>☐</span> Yes <span>☑</span> No |
| Were there any exclusions from the standard?<br>If yes, report these exclusions in the summary report table. | <span>☐</span> Yes <span>☑</span> No |
| Is there an FDA guidance that is associated with this standard?<br>If yes, was the guidance document followed in preparation of this 510(k)? | <span>☐</span> Yes <span>☑</span> No |
| Title of guidance: | |
<sup>1</sup> The formatting convention for the title is: [SDC] [numeric identifier
of standard] [date of publication]
<sup>2</sup> Authority [21 U.S.C. 360d], www.fda.gov/cdrh/standards/stdsprog.html
<sup>3</sup> http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/
search.cfm
<sup>4</sup> The summary report should include: any adaptations used to adapt
to the device under review (for example, alternative test methods);
choices made when options or a selection of method or attribute,
deviations from the standard; requirements not applicable to the
device; and the name and address of the test laboratory or
certification body involved in conformance assessment to this
standard. The summary report includes information on all standards
utilized during the development of the device.
<sup>5</sup> The supplemental information sheet (SIS) is additional information
which is necessary before FDA recognizes the standard. Found at:
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/
search.cfm
<sup>6</sup> The online search for CDRH Guidance Documents can be found at:
www.fda.gov/cdrh/guidance.html
FORM FDA 3654 (9/07) Page 1
{4}------------------------------------------------
# Premarket Notification 510(k) Section 9 – Declarations of Conformity and Summary Reports
EXTENT OF STANDARD CONFORMANCE
SUMMARY REPORT TABLE
| STANDARD TITLE | |
|----------------|--|
|----------------|--|
| CONFORMANCE WITH STANDARD SECTIONS* | | |
|-----------------------------------------------------------------------------------------|---------------------------------------|-------------------------------------------------------------------|
| SECTION NUMBER | SECTION TITLE | CONFORMANCE? |
| | TYPE OF DEVIATION OR OPTION SELECTED* | <div> ✓ Yes ☐ No ☐ N/A </div> |
| Conformance was not tested as the device is unchanged, but this standard is referenced. | | |
| DESCRIPTION | | |
| JUSTIFICATION | | |
| SECTION NUMBER | SECTION TITLE | CONFORMANCE? |
|----------------|---------------------------------------|------------------|
| | TYPE OF DEVIATION OR OPTION SELECTED* | ☐ Yes ☐ No ☐ N/A |
| DESCRIPTION | | |
| JUSTIFICATION | | |
| SECTION NUMBER | SECTION TITLE | CONFORMANCE? |
|----------------|---------------------------------------|------------------|
| | TYPE OF DEVIATION OR OPTION SELECTED* | ☐ Yes ☐ No ☐ N/A |
| DESCRIPTION | | |
| JUSTIFICATION | | |
* For completeness list all sections of the standard and indicate whether conformance is met. If a section is not applicable (N/A)
an explanation is needed under "justification." Some standards include options, so similar to deviations, the option chosen needs
to be described and adequately justified as appropriate for the subject device. Explanation of all deviations or description of
options selected when following a standard is required under "type of deviation or option selected," "description" and "justifica-
tion" on the report. More than one page may be necessary.
* Types of deviations can include an exclusion of a section in the standard, a deviation brought out by the FDA supplemental
information sheet (SIS), a deviation to adapt the standard to the device, or any adaptation of a section.
Paperwork Reduction Act Statement
Public reporting burden for this collection of information is estimated to average 1 hour per response, including the
time for reviewing instructions, searching existing data sources, gathering and maintaining the data needed, and
completing and reviewing the collection of information. Send comments regarding this burden estimate or any other
aspect of this collection of information, including suggestions for reducing this burden, to:
Center for Devices and Radiological Health
1350 Piccard Drive
Rockville, MD 20850
An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information
unless it displays a currently valid OMB control number.
FORM FDA 3654 (9/07)
Page 2
FORM FDA 3654 (9/07)
.
Page 2
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 2 5 2010
Shikani Medical, LLC C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
Re: K100995
Trade/Device Name: Shikani Speaking Valve Regulation Number: 21 CFR 868.5800 Regulation Name: Tracheostomy Tube and Tube Cuff Regulatory Class: II Product Code: JOH Dated: April 7, 2010 Received: April 9, 2010
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2- Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Cintan On
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
KI00995 (To be assigned)
Shikani speaking valve Device Name:
Indications for Use:
To allow airflow over the vocal cords for speaking function. To be used with a standard 15 mm connection on tracheostomy tubes.
Prescription Use XX or (Part 21 CFR 801 Subpart D)
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schuttman
(Division Sign-Üff) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K100993 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.