The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively to patients who require the following types of ventilator support - Positive Pressure Ventilation, delivered invasively (via endotracheal tube or tracheotomy tube) or non-invasively (via mask or nasal prongs) - Assist/Control, SIMV, or Spontaneous modes of ventilation
Device Story
Dual-microprocessor-based critical care ventilator; provides continuous invasive/non-invasive positive pressure ventilation. Inputs: medical oxygen, compressed air (internal/external sources). Operation: touch-screen controlled; delivers precise breaths (tidal volumes 2mL–2500mL) via Assist/Control, SIMV, or Spontaneous modes. Used in hospitals/institutions by clinicians. Output: ventilatory support; capnography monitoring; trending data. Benefits: precise respiratory support for critically ill patients; intra-hospital transport capability.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including human factors/usability, software verification, system integration, electrical safety/EMC, and endurance testing of the internal air compressor.
Technological Characteristics
Dual-microprocessor architecture; touch-screen interface. Supports invasive/non-invasive ventilation. Features: capnography, trending, internal air compressor. Standards: IEC 60601-1, IEC 60601-1-2, EN ISO 80601-2-12, EN ISO 80601-2-55, IEC 62304, IEC 62366, IEC 60601-1-8, AIM 7351731.
Indications for Use
Indicated for neonatal, pediatric, and adult patients requiring respiratory support or mechanical ventilation; minimum weight 0.3kg. Suitable for invasive or non-invasive positive pressure ventilation in hospital or intra-hospital transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Puritan Bennett 980 Series Ventilator System (K131252)
Reference Devices
Puritan Bennett 840 Series Ventilator System (K151252)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health and Human Services logo. The logo on the right is the Food and Drug Administration logo. The FDA logo is in blue and contains the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
October 27, 2017
Covidien Kelsey Lee Regulatory Affairs Manager 6135 Gunbarrel Ave Boulder, Colorado 80301
Re: K162738
Trade/Device Name: Puritan Bennett 980 Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous ventilator Regulatory Class: Class II Product Code: CBK Dated: October 19, 2017 Received: October 20, 2017
Dear Kelsey Lee:
This letter corrects our substantially equivalent letter of October 27, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
## Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162738
Device Name Puritan Bennett 980 Series Ventilator System
#### Indications for Use (Describe)
The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively to patients who require the following types of ventilator support
- Positive Pressure Ventilation, delivered invasively (via endotracheal tube or tracheotomy tube) or non-invasively (via mask or nasal prongs)
- Assist/Control, SIMV, or Spontaneous modes of ventilation
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a 510k Summary for the Puritan Bennett™ 980 Series Ventilator System.
| Submitted By: | Covidien<br>6135 Gunbarrel Avenue<br>Boulder, CO 80301 |
|-----------------------------------|----------------------------------------------------------------------------------------|
| Date: | October 27, 2017 |
| Contact Person: | Kelsey Lee<br>Regulatory Affairs Manager<br>(303) 305-2760 |
| Proprietary Name: | Puritan Bennett™ 980 Series Ventilator System |
| Common Name: | Ventilator, Continuous, Facility Use |
| Device Classification Regulation: | 21 CFR 868.5895 – Class II |
| Device Product Code & Panel: | CBK |
| Predicate Devices: | Puritan Bennett™ 980 Series Ventilator System<br>(K131252) |
| Reference Device: | Puritan Bennett™ 840 Series Ventilator System<br>(K151252)<br>Maquet Servo-U (K151814) |
{4}------------------------------------------------
#### Device Description
The Puritan Bennett 980 Series Ventilator System is a dual-microprocessor-based, touch-screen controlled; critical care ventilator intended to provide continuous ventilation for neonate to adult patients who require either invasive ventilation or non-invasive ventilation. It can be used in hospitals and institutions and for intra –hospital transport applications with access to the appropriate services.
The ventilator system offers features for patient comfort while delivering sensitive, precise breaths to critically ill patients. The product ventilates Neonatal, Pediatric, and Adult patients with predicted body weights from 0.3 kg, and with tidal volumes for mandatory volume controlled breaths from 2 mL to 2500 mL.
#### Indications for Use/Intended Use
The subject Puritan Bennett 980 Series Ventilator System has similar indications for use as the predicate Puritan Bennett 980 Series Ventilator System:
The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air compressor or external air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively or non-invasively to patients who require the following types of ventilator support
- Positive Pressure Ventilation, delivered invasively (via endotracheal tube or tracheotomy tube) or noninvasively (via mask or nasal prongs)
- Assist/Control, SIMV, or Spontaneous modes of ventilation
The subject indications have been clarified to include the use of the optional internal air compressor or external air sources of gas to deliver between 21% and 100% oxygen.
{5}------------------------------------------------
#### Technological Characteristics Comparison
The subject Puritan Bennett 980 Ventilator System has the same intended population, principles of operation and fundamental technology as the predicate Puritan Bennett 980 Ventilator System. Additionally, the subject device is considered a derivative of the predicate device in terms of the software and hardware modifications. The subject device has the following similar technological characteristics as the predicate Maquet SERVO-U:
- -Capnography Monitoring
- -Exhalation Flow Sensor labelling
- -Trending Option
- -Pendant Mount Configuration
- -Optional Internal Air Compressor
#### Substantial Equivalence – Non-Clinical Evidence
Substantial equivalence was shown through the following verification &validation:
- Human Factors/Usability & Design Validation ●
- Software Verification
- System Integration
- Controls
- Electrical Safety/EMC
- Internal Air Compressor testing, including endurance and VOCs
- Standards Compliance testing
- o AIM 7351731, rev 2.0
- o IEC 60601-1-2, 300 edition
- IEC ISO 60601-1, 30 edition o
- o EN ISO 80601-2-55:2011
- o EN ISO 80601-2-12: 2011
- IEC 62366: 2007 O
- O IEC 60601-1-8: 2006
- IEC 62304: 2006 O
The results of these non-clinical verifications & validations demonstrate that the subject Puritan Bennett™ 980 Series Ventilator System can be considered substantially equivalent to the predicates.
#### Substantial Equivalence – Clinical Evidence
N/A - Clinical evidence was not necessary to show substantial equivalence
#### Substantial Equivalence - Conclusions
Substantial equivalence is shown similar technological characteristics, intended use, principles of operation and verification and validation. The subject Puritan Bennett™ 980 Series Ventilator System has software and hardware enhancements to maintain the intended performance of the device.
No new questions of safety and effectiveness have been raised because of these changes and addition of features. From the evidence presented in the Premarket Notification, the subject device can be considered substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.