The FLIGHT 60 Ventilator is intended to provide continuous or intermittent mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult and pediatric (i.e., infant, child and adolescent) patients, greater than or equal to 5kg (11 lbs). The FLIGHT 60 Ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Device Story
Electrically powered, microprocessor-controlled ventilator providing continuous/intermittent mechanical ventilation. Supports ACMV (Volume, Pressure, PRVC), SIMV (Volume, Pressure, PRVC), PSV/SPONT (with Volume Guarantee), and Bi-Level (APRV) modes. Inputs include patient respiratory effort (pressure, flow, or time-triggered). Internal compressor provides gas source; exhalation valve with proportional solenoid provides PEEP. Powered by AC/DC or internal Li-Ion batteries (up to 12 hours). Features comprehensive alarm system (audible, visual, LED) for safety limit violations. Operated by qualified personnel in clinical, home, and transport environments. Output data informs clinicians of ventilation status and safety limit adherence, facilitating respiratory support and patient stabilization.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed the device meets design specifications and performance requirements per FDA guidance for ventilators. Compliance with recognized standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-12, and ASTM F 1246-91) was demonstrated.
Technological Characteristics
Electrically powered, microprocessor-controlled ventilator. Features internal compressor, proportional solenoid exhalation valve, and Li-Ion battery power. Connectivity includes internal/external communication boards. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 60601-2-12 (ventilators), ASTM F 1246-91 (home care ventilators).
Indications for Use
Indicated for adult and pediatric (infant, child, adolescent) patients ≥ 5kg (11 lbs) requiring continuous or intermittent mechanical ventilation. Restricted to use by qualified, trained personnel under physician direction in hospital, sub-acute, emergency room, home care, transport, and emergency response settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
# 510(K) SUMMARY [as required by section 807.92(c)] FLIGHT 60 Ventilator
## 510(k) Number K130171
### Date Prepared (revised):
April 7, 2014
## Applicant's Name:
Flight Medical Innovations Ltd. 13 Hamelacha Street North Industrial Park Lod 71520, Israel Tel: (972) 8-9235111 Fax:(972) 8-9236111 Company Contact:
Yanir Shaked, RAQA Manager yanirs@flight-medical.com
### Contact Persons:
Shoshana (Shosh) Friedman, RAC; and Bosmat Friedman 1914 J N Pease Pl., Charlotte, NC 28262 Address: Telephone: 704-430-8695 or 704-899-0092 704-899-0098 Fax: shosh@pushmed.com E-mail: bosmat@pushmed.com
### Trade Name:
FLIGHT 60 Ventilator
#### Common & Classification Name:
Continuous Ventilator
### Classification:
Class II; product code 73 CBK and NOU; regulation 21 CFR 868.5895
### Classification and Review Panel:
Anesthesiology
### Predicate Devices:
- Flight 60 Ventilator, cleared under K120726, manufactured by Fligh . Medical Innovations Ltd
- Vela Ventilator, cleared under K032451, manufactured by Bird Product ●
{1}------------------------------------------------
### Device Description:
The FLIGHT 60 Ventilator is an electrically powered, microprocessor controlle ventilator with the following types of ventilatory support: ACMV Volume Pressure or PRVC, SIMV Volume, Pressure or PRVC, PSV/SPONT mode wit Pressure Support and Volume Guarantee, Bi-Level (APRV). It can be pressur flow or time triggered; volume or pressure limited; time, pressure or flow cycled. Manual inflation is possible, and an emergency intake valve allows th patient to pull ambient air into the breathing circuit in the event of a complet loss of supply gas pressure.
The FLIGHT 60 may be powered by external power (100 - 240 VACS or 12 - 1 VDC) or by its two internal Li Ion rechargeable batteries, which power th ventilator for up to 12 hours when fully charged.
The electrical system is comprised of three primary boards: the Main boar (motherboard) which holds the majority of the electronics including the mai CPU and the display CPU, the Power board, which holds the power subsystem and internal communication functions, and the Communication board, which holds internal communication and external communication connectors.
The main component of the pneumatic system is an electrically controlle compressor (pump). This compressor provides a compressed gas source so n external air compressor is needed. Additionally, the exhalation valve activated by an electrically controlled proportional solenoid that provides built in PEEP.
A comprehensive alarm system is built-in to alert the user to violations of se safety limits. The alarm system alerts the care giver by activating the audib alarm, screen display and the LED indicator.
## Intended Use:
The FLIGHT 60 Ventilator is intended to provide continuous or intermitter mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult an pediatric (i.e., infant, child and adolescent) patients, greater than or equal to 5k (11 lbs).
The FLIGHT 60 Ventilator is a restricted medical device intended for use b qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, a well as for transport and emergency response applications..
## Performance Data
FLIGHT 60 Ventilator meets all applicable device specification The requirements for performance testing as identified in the FDA reviews guidance for ventilators. Verification of compliance with recognized standard has been made to support use of the device for its intended use and in i intended environment. Additionally, comparison between the performance the revised Flight 60 Ventilator (subject of this submission) with its predica
{2}------------------------------------------------
| | Equivalence Table:<br>Trilogy100 Ventilator | Vela Ventilator | FLIGHT 60 Ventilator | FLIGHT 60-O₂ (100)<br>Ventilator |
|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | |
| | Respironics Inc.<br>K083526<br>CBK | Bird Products Corp<br>K032451<br>CBK | Flight Medical Innovations Ltd<br>K120726<br>CBK, NOU | Flight Medical Innovations Ltd<br>K130171<br>CBK, NOU |
| | The Respironics Trilogy 100<br>system provides continuous<br>or intermittent ventilatory<br>support for the care of<br>individuals who require<br>mechanical ventilation.<br>Trilogy 100 is intended for<br>pediatric through adult<br>patients weighing at least 5<br>kg (11 lbs.).<br>The device is intended to be<br>used in home,<br>institution/hospital, and<br>portable applications such as<br>wheelchairs and gurneys,<br>and may be used for both<br>invasive and non-invasive<br>ventilation. It is not intended<br>to be used as a transport<br>ventilator. | The TBird VELA Ventilator is intended<br>to provide continuous or intermittent<br>mechanical ventilatory support for the<br>care of individuals who require<br>mechanical ventilation. The ventilator<br>is a restricted medical device intended<br>for use by qualified, trained personnel<br>under the direction of a physician.<br>Specifically, the ventilator is applicable<br>for adult and pediatric patients<br>weighing at least 10 kg (22 lbs.), who<br>require the following general types of<br>ventilatory support, as prescribed by<br>an attending physician:<br>• Positive pressure ventilation<br>• Assist/Control, SIMV, CPAP modes of<br>ventilation<br>The ventilator is suitable for use in<br>institutional and transport settings. It<br>is not intended for use as an emergency<br>medical transport ventilator. | The FLIGHT 60 Ventilator is<br>intended to provide continuous<br>or intermittent mechanical<br>ventilation support for the care<br>of individuals who require<br>mechanical ventilation.<br>Specifically, the FLIGHT 60 is<br>applicable for adult and pediatric<br>(i.e., infant, child and adolescent)<br>patients, greater than or equal to<br>5kg (11 lbs).<br>The FLIGHT 60 Ventilator is a<br>restricted medical device<br>intended for use by qualified,<br>trained personnel under the<br>direction of a physician; it is<br>suitable for use in hospital, sub-<br>acute, emergency room, and<br>home care environments, as well<br>as for transport and emergency<br>response applications. | The FLIGHT 60 Ventilator is<br>intended to provide<br>continuous or intermittent<br>mechanical ventilation<br>support for the care of<br>individuals who require<br>mechanical ventilation.<br>Specifically, the FLIGHT 60 is<br>applicable for adult and<br>pediatric (i.e., infant, child and<br>adolescent) patients, greater<br>than or equal to 5kg (11 lbs).<br>The FLIGHT 60 Ventilator is a<br>restricted medical device<br>intended for use by qualified,<br>trained personnel under the<br>direction of a physician; it is<br>suitable for use in hospital,<br>sub-acute, emergency room,<br>and home care environments,<br>as well as for transport and<br>emergency response<br>applications. |
| Features | | | | |
| | | Yes | | Yes |
| | | Yes | | Yes |
| | | Yes | | Yes |
| | Yes | | | Yes |
| | Yes | | | Yes |
| | | | | |
·
:
Equivalence Table:
.
{3}------------------------------------------------
### Standards
FLIGHT 60 Ventilator has been tested and shown to be compliant with the fo standards:
| IEC 60601-1:1998<br>+A1:1991+A2:1995 | Medical electrical equipment - Part 1: General requirements for<br>safety and essential performance |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2007 | Medical electrical equipment General requirements for basic safety<br>essential performance. Collateral standard. Electromagnetic<br>compatibility. Requirements and tests |
| IEC 60601-1-8:2006 | Medical electrical equipment -- Part 1-8: General requirements<br>safety and essential performance -- Collateral standard: General<br>requirements, tests and guidance for alarm systems in medical<br>equipment and medical electrical systems |
| IEC 60601-2-12:2001 | Medical electrical equipment -- Part 2-12: Particular requirement<br>the safety of lung ventilators -- Critical care ventilators |
| ASTM F 1246-91 | Standard Specification for Electrically Powered Home Care Ventilators<br>Part 1-Positive-Pressure Ventilators and Ventilator Circuits |
### Conclusion
Verification and validation activities were conducted to establish the perfi characteristics of the FLIGHT 60 Ventilator. All testing demonstrated that the 60 Ventilator met required design verification criteria and has acceptable perfec when used in accordance with its labeling. The device's intended use, o principles, ventilation modes and performance parameters are comparable referenced predicate devices. Therefore, the FLIGHT 60 Ventilator is subs equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2014
Flight Medical Innovations, Ltd C/O Ms. Soshana Friedman, President Push-Med LLC 1914 J.N. Pease Place Charlotte, NC 28262
Re: K130171
Trade/Device Name: FLIGHT 60 Ventilator Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Class: II Product Code: CBK Dated: March 18, 2014 Received: March 19, 2014
Dear Ms. Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Friedman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/5/Picture/8 description: The image contains a logo or seal on the left side, which is abstract and difficult to decipher. To the right of the logo, there is text that reads "Tejashri Purohit-Sheth, M.D.," followed by "Clinical Deputy Director" and "DAGRID." The word "FOR" is located in the bottom right corner of the image.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):___K130171_
#### Device Name:
FLIGHT60® Ventilator
#### Indications for Use:
The FLIGHT 60 Ventilator is intended to provide continuous or intermittent mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult and pediatric (i.e., infant, child and adolescent) patients, greater than or equal to 5kg (11 lbs).
The FLIGHT 60 Ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Prescription Use X-(Per 21 CFR 801 Subpart D) OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harry -S 2014.04.08 15:03:12 -04'00'
FLIGHT 60 Ventilator - Page 4-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.