K070594 · Pulmonetic Systems, Inc. · CBK · Aug 1, 2007 · Anesthesiology
Device Facts
Record ID
K070594
Device Name
PALMTOP VENTILATOR, MODELS PTV-8, PTV-10
Applicant
Pulmonetic Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Aug 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
PALMTop PTV Models 8/10 are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). They are suitable for service in homecare, hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. PALMTop Ventilators are restricted medical devices intended for operation by trained personnel under the direction of a physician and in accordance with all applicable state laws and regulations. Federal law (USA) restricts the sale of this device except by or on the order of a physician.
Device Story
PALMTop PTV Models 8/10 are portable ventilators providing invasive or non-invasive positive pressure support. Device uses internal flow generator, eliminating need for external compressed gas. Inputs include patient circuit pressure and optional FIO2/oximetry data. System supports multiple modes: CPAP, Pressure Support, SIMV, Control, Assist/Control, and PRVC. Features include leak compensation, apnea backup, and synchronized nebulization. Operated by trained personnel in home, hospital, or transport environments. Output includes real-time waveforms, loops, trending histograms, and monitored parameters (PIP, PEEP, tidal volume, minute volume). Clinicians use data to adjust ventilation settings, optimizing patient respiratory support and gas exchange. Benefits include high-performance ventilation in a lightweight, mobile form factor.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing verified that the PALMTop Models PTV 8/10 meet specified performance requirements and are substantially equivalent to the predicate devices.
Indicated for patients requiring respiratory support or mechanical ventilation weighing at least 5 kg (11 lbs). Suitable for invasive or non-invasive continuous or intermittent positive pressure ventilation in home, hospital, and transport settings. Restricted to use by trained personnel under physician direction.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
# 510(K) SUMMARY
#### SUBMITTER
Pulmonetic Systems, Inc. 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
AUG - I 2007
| Contact Person: | Robert C. Samec | Telephone |
|-----------------|-----------------|-----------|
| | (763) 398-8305 | |
| | (763) 398-8400 | Facsimile |
#### DEVICE / TRADE NAME
| Trade Name: | PALMTop PTV Models 8/10 |
|----------------------|----------------------------------------------|
| Common Name: | Ventilator |
| Classification Name: | Ventilator, Continuous (Respirator) 868.5895 |
#### SUBMISSION DATE
Submission Date: August 1, 2007
#### DESCRIPTION
The PALMTop PTV Models 8/10 are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). They are suitable for service in homecare, hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively.
The PTV platform provides for the following features:
- . High performance ventilation in a small lightweight package
- A range of maneuvers including Inspiratory and Expiratory hold . and synchronized Nebulization (touch screen models only)
- . Internal flow generator, allowing the PTV to operate without an external compressed gas source
- . Four distinct breath types offering a range of treatment options. CPAP with optional Pressure Support 1, SIMV2, Control, Assist /
{1}------------------------------------------------
Control
- Configurable Apnea back-up ventilation .
- NPPV3 ventilation, providing an alarm package suitable for . mask ventilation of patients whose needs do not extend to life support
- Volume Control, Pressure Control and PRVC4 ventilation modes t
- Adjustable alarm settings including High Peak Pressure. Low . Peak Pressure, Low Minute Volume, and Apnea
- Oxygen blending from a high-pressure oxygen source (optional) . or low-pressure oxygen bleed-in
- Lockable front panel controls .
- A wide range of monitors including Breath Rate (f), I:E Ratio . (Measured), MAP, Exhaled Minute Volume (VE), PEEP, PIP and Tidal Volume (Vte)
- Dynamic waveform and loop displays .
- . Graphic trending histogram display
- Real-time patient circuit pressure display .
- . Variable termination options for Pressure Support breaths, including maximum inspiratory time and percentage of peak flow
- . Selectable Percentage of Peak Flow termination for Pressure Control breaths
- Leak Compensation to improve patient triggering when a circuit . leak is present
- . Operation from a variety of power sources including AC power. removable battery and external DC power sources
- . Optional FIO2 sensor package.
- . Optional Oximetry module
### INTENDED USE
PALMTop PTV Models 8/10 are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). They are suitable for service in homecare, hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively.
<sup>1</sup> Continuous Positive Pressure Ventilation
<sup>2</sup> Synchronized Intermittent Mandatory Ventilation
<sup>3</sup> Non-invasive Positive Pressure Ventilation
<sup>4</sup> Pressure Regulated, Volume Controlled
{2}------------------------------------------------
PALMTop Ventilators are restricted medical devices intended for operation by trained personnel under the direction of a physician and in accordance with all applicable state laws and regulations.
## Federal law (USA) restricts the sale of this device except by or on the order of a physician.
#### Summary of Testing and Validation
Performance testing (Bench testing) verified that the Palmtop Models PTV 8/10 meet specified performance requirements and that the devices are substantially equivalent to the predicate devices listed below.
| Predicate Device | 510(k) Clearance | Manufacturer |
|------------------------------------------------------------|--------------------|--------------------------------------------------------------------------------------------------------------|
| LTV 1200<br>Ventilator | K060647 | VIASYS<br>Healthcare/Pulmonetic<br>Systems, Inc.<br>17400 Medina Road.<br>Suite 100<br>Minneapolis, MN 55447 |
| LTV 1000<br>Ventilator | K981371<br>K984056 | VIASYS<br>Healthcare/Pulmonetic<br>Systems, Inc.<br>17400 Medina Road,<br>Suite 100<br>Minneapolis, MN 55447 |
| iVent 201<br>Portable<br>Ventilator with<br>Pulse Oximeter | K021981 | Versamed Corporation<br>2 Blue Hill Plaza<br>Pearl River, NY 10965 |
| TBird VELA<br>Ventilator | K032451 | VIASYS Healthcare/Bird<br>Products Corporation<br>1100 Bird Center Drive<br>Palm Springs, CA 92262 |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three lines forming its body and wings, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 1 2007
Pulmonetic Systems, Incorporated C/O Mr. Robert C. Samec Vice President- Quality Assurance/Regulatory Affairs Division of VIASYS Healthcare 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
Re: K070594
Trade/Device Name: PALMTop PTV Models 8/10 Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: July 26, 2007 Received: July 27, 2007
Dear Mr. Samec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Samec
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clues
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K070594 510(k) Number (if known):
Device Name: PALMTop PTV Models 8/10
Indications For Use:
Indications for Use
PALMTop PTV Models 8/10 are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in homecare, hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively.
PALMTop Ventilators are restricted medical devices intended for operation by trained personnel under the direction of a physician and in accordance with all applicable state laws and regulations.
Federal law (USA) restricts the sale of this device except by or on the order of a physician.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ch (Division Sign-Off)
(acting B.C.) Division of Anesthesinlogy, General Hospital Infection Control, Dental Devices
510(k) Number: K070594
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.