K061094 · Newport Medical Instruments, Inc. · CBK · Jul 28, 2006 · Anesthesiology
Device Facts
Record ID
K061094
Device Name
NEWPORT E500 WAVE VENTILATOR
Applicant
Newport Medical Instruments, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jul 28, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric, and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only. The intended environments include hospital-type, and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.
Device Story
Newport e500 Wave Ventilator is a pneumatically powered, microprocessor-controlled ventilator. It provides ventilatory support for infant, pediatric, and adult patients. Operators select ventilation modes (A/CMV, SIMV, SPONT) and breath types (volume control, pressure control, volume target pressure control) via front panel controls. The device includes a comprehensive alarm system for technical and patient-related safety limits. Used in hospitals (OR, ICU, general care) and sub-acute facilities, as well as for intra-hospital transport. It assists clinicians in managing patient respiration, providing continuous or non-continuous support to improve patient outcomes in critical care settings.
Clinical Evidence
Bench testing only. Comprehensive verification and validation testing was performed, including hardware, software, electrical safety, functional safety, EMC, packaging, and environmental testing. All results met pre-defined acceptance criteria.
Technological Characteristics
Pneumatically powered, microprocessor-controlled continuous ventilator. Supports infant/pediatric (>20 mL) to adult patients. Features A/CMV, SIMV, and SPONT modes. Alarms comply with ISO9703-1, ISO9703-2, and EN475. Connectivity includes front panel user interface controls.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring, provided the patient has a tidal volume of ≥ 20 mL. Prescription use only.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K061094
JUL 2 8 2006
# ATTACHMENT 3, page 22 510(k) SUMMARY
### A. Sponsor
Newport Medical Instruments, Inc.
1620 Sunflower Ave
Costa Mesa, California 92626
Telephone: (714) 427-5811
Fax: (714) 427-0839
Contact Person: Richard Waters
Vice President, Regulatory Affairs & Quality Assurance
### B. Date Prepared
4/18/2006
### ். Device Name
Trade Name: Newport e500 Wave Ventilator
Classification Name: Continuous Ventilator
### D. Device Description
The Newport e500 Wave Ventilator is a pneumatically powered, microprocessor controlled ventilator cleared through Premarket Notification.
Performance characteristics and clinical features support infant/pediatric (>20 mL) through adult patients.
Front panel controls allow trained operators to select ventilation controls under volume control, pressure control and volume target pressure control breath types in A/CMV, SIMV and SPONT modes.
A comprehensive alarm system is built-in to alert the user to violations of preset safety limits.
The alarms associated with the e500 meet or exceed standards of critical care ventilators and have been developed in compliance with ISO9703-1, ISO9703-2 and EN475. The alarms of the e500 span both technical (ventilator related) alarms and non technical alarms (patient related alarms).
### E. Intended Use
The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric, and adult patients with a
{1}------------------------------------------------
# ATTACHMENT 3, page 23
tidal volume of ≥ 20 mL. The device is prescription use only. The intended environments include hospital-type, and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.
## F. Cleared/Predicate Device
Cleared Device
Newport e500 Ventilator, K030780
Predicate Devices
Drager EvitaXL K051263, July 12th, 2005
Maquet Servo-I Ventilator, K041223, July 29th, 2004
Puritan-Bennett 840, K984535, December 28, 1998
### G. Summary of Substantial Equivalence
The Newport e500 Wave Ventilator is substantially equivalent to the cleared/predicate devices in intended use, physical characteristics, performance specifications and safety characteristics.
## H. Testing
Comprehensive verification and validation testing was performed with the Newport e500 Wave Ventilators including; hardware, software, electrical safety, functional safety, EMC, packaging, and environmental. All test results met pre-defined acceptance criteria.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three stripes forming its wings and body. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 8 2006
Mr. Richard Waters Vice President, Regulatory Affairs & Quality Assurance Newport Medical Instruments, Incorporated 1620 Sunflower Avenue Costa Mesa, California 92626
Re: K061094
Trade/Device Name: Newport e500 Wave Ventilator Regulation Number: 21 CFR 868,5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: July 12. 2006 Received: July 13, 2006
Dear Mr. Waters:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Mr. Waters
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Ching-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# ATTACHMENT 1, page 20 INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: _Newport e500 Wave Ventilator
Indications for Use:
The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric, and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only.
The intended environments include hospital, hospital-type, and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital transport includes patient transport within the hospital or hospital-type facility.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1_of 1_
(Posted November 13, 2003)
Aun. M. Wem
Sign-C)
n of Anesthesiology, General Hospital,
n Control, Dental Devices
KOCIO94
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.