MODIFICATION TO LTV 1000 VENTILATOR/BREATHING CIRCUITS
K040790 · Pulmonetic Systems, Inc. · CBK · Jun 3, 2004 · Anesthesiology
Device Facts
Record ID
K040790
Device Name
MODIFICATION TO LTV 1000 VENTILATOR/BREATHING CIRCUITS
Applicant
Pulmonetic Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jun 3, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support: - Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). - Assist/Control, SIMV, or CPAP modes of ventilation. - Breath types including Volume, Pressure Control and Pressure Support. The ventilator is suitable for use in institutional, home and transport settings,
Device Story
LTV 1000 is a continuous ventilator providing mechanical ventilatory support; supports adult and pediatric patients (≥5 kg). Device delivers positive pressure ventilation (invasive/non-invasive) using Assist/Control, SIMV, or CPAP modes; supports Volume, Pressure Control, and Pressure Support breath types. Modification adds commercially available heated wire breathing circuit limbs (Allegiance Healthcare) to provide heated/humidified air under physician order. Used in institutional, home, and transport settings by trained clinicians. Output is mechanical ventilation; healthcare providers monitor patient status and ventilator settings to adjust therapy. Benefits include reliable respiratory support across diverse clinical environments.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of the modified breathing circuit components to predicate devices.
Technological Characteristics
Continuous ventilator; supports adult/pediatric patients ≥5 kg. Features heated wire breathing circuit limbs (Allegiance Healthcare) for heated/humidified air. Modes: Assist/Control, SIMV, CPAP. Breath types: Volume, Pressure Control, Pressure Support. Connectivity: compatible with external humidifiers (e.g., Fisher & Paykel MR 730/850).
Indications for Use
Indicated for adult and pediatric patients weighing at least 5 kg (11 lbs.) requiring continuous or intermittent mechanical ventilatory support via invasive (ET tube) or non-invasive (mask) positive pressure ventilation. Modes include Assist/Control, SIMV, or CPAP. Restricted to use by qualified, trained personnel under physician direction.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
JUN - 3 2004
K040790
### 510(K) SUMMARY 1.
#### SUBMITTER 1.1
Pulmonetic Systems, Inc. 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
| Contact Person: | Robert C. Samec | | |
|-----------------|-----------------|--|--|
| (763) 398-8305 | Telephone | | |
| (763) 398-8400 | Facsmilie: | | |
#### 1.2 DEVICE / TRADE NAME
LTV 1000 Ventilator/Breathing circuits Trade Name:
Common Name: Ventilator
Classification Name: Ventilator, Continuous (Respirator) 868.5895
#### 1.3 SUBMISSION DATE
Submission Date: March 26, 2004
#### 1.4 DESCRIPTION
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in institutional, home and transport settings, and is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via . mask).
- Assist/Control, SIMV, or CPAP modes of ventilation. .
- Breath types including Volume, Pressure Control and Pressure Support. .
The modification intended to be cleared by this submission is:
The addition of commercially available heated wire breathing circuit inspiratory/expiratory limbs manufactured and distributed by Allegiance Healthcare Corporation (K000697), as an option to the ventilator breathing circuits specified for use
{1}------------------------------------------------
commercially available humidifier with heated wire controller* to provide heated and humidified air under a physicians order/supervision.
*Fisher & Paykel Models: MR 730 (K913368) MR 850 (K020332)
#### 1.5 INTENDED USE
The LTV ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively . (via mask).
- Assist/Control, SIMV, or CPAP modes of ventilation. .
- Breath types including Volume, Pressure Control and Pressure Support. .
The ventilator is suitable for use in institutional, home and transport settings,
#### 1.6 EQUIVALENCE TO PREDICATE DEVICE(S)
The LTV 1000 Ventilator listed modifications are substantially equivalent to the following listed devices:
| Predicate Device | 510(k) Clearance | Manufacturer |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| LTV 1000 Ventilator<br>with breathing<br>circuits. | K981371 - Initial clearance for<br>Institutional and Transport<br>settings.<br>K984056 - Homecare settings.<br>K002881 - Enhancements.<br>K010608 - Lap Top Monitor.<br>K032226 - 5 kg Patient Application. | Pulmonetic Systems, Inc.<br>Colton, CA/Mpls., MN |
| Allegiance Airlife<br>Heated Ventilator<br>Breathing Circuits | K000697 | Allegiance Healthcare<br>Corporation |
The LTV 1000 Ventilator with breathing circuits, previously cleared for homecare use and for institutional and transport settings, is now being submitted for clearance with the listed modifications.
The table on the following page compares the modification of the LTV 1000 breathing circuits to the previously cleared LTV 1000 ventilator/breathing circuits and the Allegiance Airlife Heated Wire Breathing Circuits. Allegiance Healthcare Corporation, the manufacturer of the Airlife Heated Wire
{2}------------------------------------------------
Breathing Circuits is the selected supplier for the Heated Wire Breathing Circuit inspiratory/expiratory limbs to be utilized with the LTV 1000 ventilator.
The LTV 1000 ventilator/breathing circuits with the modification listed are substantially equivalent to the EP 1000 ventilator/breathing circuits (K981371) and the Allegiance Airlife Heated the products = r = 1000 (K000697), intended for use with commercially available ventilators.
{3}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE SUMMARY TABLE
| LTV 1000 with Modification | LTV 1000<br>(Predicate Device)<br>(K981371) | Allegiance Airlife Heated Wire<br>Breathing Circuit Limb(s)<br>(K000697) Predicate Device | Discussion of Differences and<br>Similarities |
|-------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model 15090-001 Heated Circuit<br>Assy., 6', SPU, Adult w/PEEP | Model 10820 Circuit Assy., 6', SPU,<br>Adult w/PEEP | RT600-850 Adult Inspiratory Limb,<br>Heated, 6'. | Airlife Adult Heated Wire Inspiratory<br>Limb replaces PSI Adult Inspiratory<br>Limb. |
| Model 15090-002 Heated Circuit<br>Assy., 6', SPU, Adult PEEPless | Model 10822 Circuit Assy., 6', SPU,<br>Adult PEEPless | RT600-850 Adult Inspiratory Limb,<br>Heated, 6'. | Airlife Adult Heated Wire Inspiratory<br>Limb replaces PSI Adult Inspiratory<br>Limb. |
| Model 15090-003 Heated Circuit<br>Assy., 6', SPU, Pediatric w/PEEP | Model 10821 Circuit Assy., 6', SPU.<br>Pediatric w/PEEP | RT609-856 Pediatric Inspiratory Limb,<br>Heated, 6'. | Airlife Pediatric Heated Wire<br>Inspiratory Limb replaces PSI Pediatric<br>Inspiratory Limb. |
| Model 15090-004 Heated Circuit<br>Assy., 6' SPU, Pediatric PEEPless | Model 10823 Circuit Assy., 6', SPU,<br>Pediatric PEEPless | RT609-856 Pediatric Inspiratory Limb,<br>Heated, 6'. | Airlife Pediatric Heated Wire<br>Inspiratory Limb replaces PSI Pediatric<br>Inspiratory Limb. |
| Model 15091-001 Heated Circuit<br>Assy., 5', SPU, Adult w/ PEEP | Model 10820 Circuit Assy., 6', SPU,<br>Adult w/PEEP | RT500-853 Adult<br>Inspiratoty/Expiratory Limbs, Heated,<br>5'. | Airlife Adult Heated Wire<br>Inspiratory/Expiratory Limb replaces<br>PSI Adult Inspiratory/Expiratory Limb.<br>Length reduced from 6' to 5'. |
| | | | Airlife Adult Heated Wire<br>Inspiratory/Expiratory Limb replaces<br>PSI Adult Inspiratory/Expiratory Limb.<br>Length reduced from 6' to 5'. |
| | | | Airlife Pediatric Heated Wire<br>Inspiratory/Expiratory Limb replaces<br>PSI Pediatric Inspiratory/Expiratory<br>Limb. Length reduced from 6' to 5'. |
| Model 15091-004 Heated Circuit<br>Assy., 5', SPU, Pediatric PEEPless | Model 10823 Circuit Assy., 6', SPU,<br>Pediatric PEEPless | RT509-852 Pediatric<br>Inspiratoty/Expiratory Limbs, Heated,<br>5'. | Airlife Pediatric Heated Wire<br>Inspiratory/Expiratory Limb replaces<br>PSI Pediatric Inspiratory/Expiratory<br>Limb. Length reduced from 6' to 5'. |
| Model 15092-001 Assy. Heated<br>Inspiratory Limb Assy., 6', SPU,<br>Pediatric | Model 10823 Circuit Assy., 6', SPU,<br>Pediatric PEEPless | RT609-856 Pediatric Inspiratory Limb,<br>Heated, 6'. | Airlife Pediatric Heated Wire<br>Inspiratory Limb replaces PSI Pediatric<br>Inspiratory Limb. |
| Model 15092- 002 Assy. Heated<br>Inspiratory Limb Assy., 6', SPU, Adult | Model 10820 Circuit Assy., 6', SPU,<br>Adult w/PEEP | RT600-850 Adult Inspiratory Limb,<br>Heated, 6'. | Airlife Adult Heated Wire Inspiratory<br>Limb replaces PSI Adult Inspiratory<br>Limb. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three overlapping wings, symbolizing health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
UN - 3 2004
Mr. Robert C. Samec Vice President, Regulatory Affairs Pulmonetic Systems, Incorporated 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
Re: K040790
K040770
Trade/Device Name: Modification to LTV 1000 Ventilator/Breathing Circuits Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Il Product Code: CBK Dated: May 24, 2004 Received: May 25, 2004
Dear Mr. Samec:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bectoris in the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for tise stated in the encrosure) to regally management date of the Medical Device American's on to commerce prior to May 26, 1970, the enated with the provisions of the Federal Food, Drug, devices that have been recuire in assee approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approvisions of the general controls provisions of the Y ou may, ulciently, market the devices, obly of the courements for annual registration, listing of general controls provisions of the fice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it If your device is classified (secure) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations liDA m may be subject to such additional controls. Extremgions be found in the Code of Pouchal Ingerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that 1127 s issuates or our device complies with other requirements of the Act that FIA has made a decemmances and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. Robert C. Samec
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Ventilator, Continuous (Respirator)
Indications For Use:
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).
- Assist Control, SIMV, or CPAP modes of ventilation. -
The ventilator is suitable for use in institutional, home, or transport settings.
Prescription Use X
Over-The-Counter Use AND/OR
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aung Suulum
these sintray, General Hospital. Infection Control, Dental Devices 510(k) Number:
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.