The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs.), who require the following general types of ventilatory support, as prescribed by an attending physician: - Positive pressure ventilation . - Assist/Control, SIMV, CPAP modes of ventilation . The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator
Device Story
TBird VELA Ventilator provides mechanical ventilatory support via turbine gas delivery system and microprocessor control. Inputs include patient pressure and flow data; system utilizes pressure transducers and analog-to-digital converters for monitoring. Device operates in institutional and transport settings; managed by qualified, trained personnel. Output includes ventilation delivery and graphical display of patient data (manometer, alarm setpoints, message characters). Healthcare providers use output to adjust ventilation parameters; device supports patient respiratory function. System includes flow delivery, exhalation, safety, and inspiratory hold subsystems. Power management handles AC, internal battery, and external battery sources.
Clinical Evidence
Bench testing only. Performance testing verified that the TBird VELA Ventilator meets its performance requirements and is substantially equivalent to legally marketed devices.
Indicated for adult and pediatric patients weighing at least 10 kg (22 lbs.) requiring continuous or intermittent mechanical ventilatory support, including positive pressure ventilation, Assist/Control, SIMV, and CPAP modes. Not for emergency medical transport.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
APR 2 2002
K020746
| Manufacturing Site: | Bird Products Corporation<br>1100 Bird Center Drive<br>Palm Springs, CA 92262 | 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS | |
|-------------------------------------------|-------------------------------------------------------------------------------|--------------------------------------------|---------------------------|
| Contact: | Tom Gutierrez (760) 778-7255 (phone); (760) 778-7274 (fax) | | |
| Summary Date | August 11, 2000 | | |
| Device Trade Name: | TBird VELA Ventilator | | |
| Device<br>Common/Classificati<br>on Name: | Classification name: 868.5895 Continuous Ventilator, 73 CBK | | |
| Establishment<br>Registration Number | 2021710 | | |
| Device Class: | Class II | | |
| Classification Panel: | Anesthesiology | | |
| Predicate Device: | The predicate devices are: | | |
| | 1 K950484 | TBIRD AVS | Bird Products Corporation |
| | 2 K920113 | Bird Graphics Monitor | Bird Products Corporation |
| | 3 K973646 | Bird Sentry Blender | Bird Products Corporation |
| | 4 K911336 | Bird Nebulizer Synchronizer | Bird Product Corporation |
Device Description:
The TBird VELA Ventilator employs a turbine gas delivery system along with microprocessor control to provide support for pediatric to adult patients. Capable of delivering clinically advanced modes of ventilation like Pressure Support with an internal battery or AC power the TBird VELA Ventilator has an extensive patient range.
The TBird VELA Ventilator ventilator pneumatic system is an electromechanical system comprised of four major subsystems, each containing several components. These systems include the flow delivery system, the exhalation system, the safety system and the inspiratory hold valve.
This electromechanical system controls all inspiratory flow to the patient. The exhalation system controls the flow of gas from the patient's lungs during the exhalation phase. The mechanical safety system ensures that the patient can breath spontaneously from room air and that the patient pressure is limited to a maximum value in the event of a ventilator malfunction. When activated, the inspiratory hold valve blocks flow between the flow delivery system and the patient.
The TBird VELA Ventilator ventilator electronic system is comprised of several subsystems, each containing numerous components. These subsystems include the Graphical User Interface System, the Power System, the Main Controller System, Exhalation, and Flow Delivery systems. The Display System is comprised of three Alarm Setpoint Displays, seven Control Setpoint Displays, up to forty-eight Message Display characters, up to twentyfive Discrete Indicators, and a bargraph style Manometer. The Main Controller System is comprised of three Pressure Transducers, an Analog-to-Digital Converter, two Digital-to-Analog Converters, the Input-Output Processor, Solenoid Valves, and the Watchdog and Hardware Fault Monitors. The Power System conditions and controls energy from the AC line input, the internal battery, and the optional external battery pack. When energy is available from the AC line, the ventilator operates from this source, as well as recharging the internal battery.
{1}------------------------------------------------
## Intended Use:
The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs.), who require the following general types of ventilatory support, as prescribed by an attending physician:
- Positive pressure ventilation .
- Assist/Control, SIMV, CPAP modes of ventilation .
The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator
Substantial Equivalence The TBird VELA Ventilator is the same device as the TBird AVS Ventilator, which was cleared for market under 510(k) K950484. The name of the device was changed to the TBird VELA.
Modifications to the TBird VELA Ventilator are associated with this submittal
- . Graphical User Interface (GUI) touch-screen/display
- t OXYGEN monitor
- . Nebulizer high pressure gas source.
- Exhalation flow sensor was also modified to incorporate a direct hard wire . connection
The modified TBird VELA Ventilator have the following similarities to those which previously received 510(k) concurrence:
- . have the same indicated use,
- . use the same ventilation operating principle,
- incorporate the same basic ventilator design with the exception of modifications . described in this submittal.
- . incorporate the same basic electronic control system with the exception of modifications described within this submittal.
- . are manufactured and packaged utilizing the same basic processes.
- In summary, the TBird VELA Ventilator described in this submission is, in our opinion, . substantially equivalent to the predicate device(s).
## Summary of Testing and Validation:
Performance testing verified that the TBird VELA Ventilator meets it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States.
D3
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure with outstretched arms, surrounded by a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top and left side of the circle.
APR 2 2 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tom Gutierrez Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262
Re: K020746
TBird VELA Ventilator Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II (two) Product Code: 73 CBK Dated: February 29, 2002 Received: April 1, 2002
Dear Mr. Guitierrez:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becaused the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 use surved in the encreativent of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de rices that hat to been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may ; therefore, maxis of the Act include requirements for annual registration, listing of general voltarely pro resuring practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it 11 your de roo to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Tom Gutierrez
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 toast of advised that I DPT of issuaires complies with other requirements of the Act that I Dr has made a aown regulations administered by other Federal agencies. You must or any I catal statutes and regarients, including, but not limited to: registration and listing (21 Comply with an the rice of requirements, is, good manufacturing practice requirements as set CFK i at 607); laboring (21 CFR Part 820); and if applicable, the electronic rorth in the quality by bections (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This lotter will and h yourse FDA finding of substantial equivalence of your device to a legally prematicated predicated. The Pro macinssification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advices for in vitro diagnostic devices), please contact the Office of additionally 21 OF F Far 8646. Additionally, for questions on the promotion and advertising of Compinance at (301) 591-1010 ffice of Compliance at (301) 594-4639. Also, please note the your devices, preads "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation chitited, Tribbiananing oresponsibilities under the Act may be obtained from the Other general mionmation on your new and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication For Use
510 (k) Number (if known):
Page _________________________________________________________________________________________________________________________________________________________________________
Device Name: ___ TBird VELA Ventilator
Indication For Use:
The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical The TDIrd VEEX Ventifator is intonecise players mechanical ventilation. The ventilatory Support for the ours of marmatic or use by qualified, trained personnel ventifator is a restificied modiour dovice iffically, the ventilator is applicable for adult and under the direction of a priyelland 10 kg (22 lbs.), who require the following general pediatho putionto weighing as prescribed by an attending physician:
- Positive pressure ventilation ●
- Assist/Control, SIMV, CPAP modes of ventilation .
The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) Number 4020746
| Prescription Use | <div style="display:inline-block; vertical-align:middle;">✓</div> | OR | Over-The-Counter Use _________ |
|----------------------|-------------------------------------------------------------------|----|--------------------------------|
| (Per 21 cfr 801.109) | | | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.