K201026 · Hytek Medical, Inc. · CBI · Aug 10, 2021 · Anesthesiology
Device Facts
Record ID
K201026
Device Name
sOLVe Tube
Applicant
Hytek Medical, Inc.
Product Code
CBI · Anesthesiology
Decision Date
Aug 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5740
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The sOLVe Tube™ is used to isolate the left or the right lung of a patient for surgery, one lung ventilation or one lung anesthesia.
Device Story
sOLVe Tube is a single-use, dual-lumen endobronchial tube (DLT) kit for lung isolation and ventilation. Used in hospitals by healthcare professionals. Kit includes DLT with pre-installed pliable aluminum intubation stylet, dual bronchoscopy swivel adapter with safety clamp, endotracheal tube holder/bite block, suction catheters, and syringes. Distal tip fits in either left or right mainstem bronchus, allowing universal left/right-sided use. Features ridges on balloons to prevent slippage and dislodgement. Healthcare providers insert the tube to isolate a lung; output is the physical isolation and ventilation of the target lung. Benefits include simplified lung isolation procedures and enhanced safety via balloon design.
Clinical Evidence
No human clinical data. Evidence includes bench testing (ISO 5361, ISO 16628, ASTM F640, biocompatibility per ISO 10993/18562/ANSI/AAMI ST72) and animal study. Animal testing evaluated ease of insertion, balloon performance, and success of lung isolation/ventilation.
Technological Characteristics
Double-lumen endobronchial tube; materials: silicone, polypropylene. Sterilization: Ethylene Oxide. Features: tracheal/endobronchial balloons with anti-slip ridges, swivel connector, 10Fr suction catheter. Standards: ISO 5361 (cuff/tube performance), ISO 16628 (lumen diameter), ASTM F640 (radio-opacity), ISO 10993 (biocompatibility).
Indications for Use
Indicated for adult patients requiring lung isolation, one-lung ventilation, or one-lung anesthesia during surgery.
Regulatory Classification
Identification
A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
HyTek Medical, Inc. % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies LLC 2251 San Diego Ave Suite B-257 San Diego, California 92110
Re: K201026
Trade/Device Name: sOLVe Tube Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: Class II Product Code: CBI Dated: July 13, 2021 Received: July 15, 2021
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201026
Device Name sOLVe Tube™
Indications for Use (Describe)
The sOLVe Tube™ is used to isolate the left or the right lung of a patient for surgery, one lung ventilation or one lung anesthesia.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K201026
### DATE PREPARED
August 5, 2021
### MANUFACTURER AND 510(k) OWNER
HyTek Medical, Inc. 8741 Shirley Ave., Northridge, CA 91324, USA Telephone: Official Contact: Nir Hoftman, M.D., President
### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Michelle Rubin-Onur, Ph.D. Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 Email: akomiyama@acknowledge-rs.com Website: www.AcKnowledge-RS.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | sOLVe Tube™ |
|------------------------------|----------------------------------------------------------------------------|
| Common Name: | Tube, Tracheal/Bronchial, Differential Ventilation (w/wo connecter) |
| Regulation Number: | 21 CFR 868.5740 |
| Class: | Class II |
| Product Code: | CBI |
| Premarket Review: | Anesthesiology |
| Review Panel: | ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Device (DHT1C) |
#### PREDICATE DEVICE IDENTIFICATION
The sOLVe Tube" is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|----------------------------------------------|-------------------|
| K051522 | Silbroncho® Tubes / Fuji Systems Corporation | ✓ |
The predicate device has not been subject to a design related recall.
#### DEVICE DESCRIPTION
The Simple One Lung Ventilation for Everyone (sOLVe) Tube is a single use dual lumen endobronchial tube (DLT) intended for lung isolation and for lung ventilation by healthcare professionals (HCP) in hospitals. The sOLVe Tube can function as both a left-sided and a rightsided DLT. The distal tip of the tube will fit in either the left or right mainstem bronchus.
{4}------------------------------------------------
## 510(k) Summary
The sOLVe Tube is packaged and provided to the HCP as a kit. The kit includes the following components:
- 1. sOLVe Tube with pre-installed pliable aluminum 10Fr intubation stylet
- 2. Dual Bronchoscopy Swivel adapter with integrated safety clamp
- 3. Haider TubeGuard endotracheal tube holder/bite block with strap
- 4. Soft suction catheters
- 5. Two syringes (3 mL and 10 mL)
The sOLVE tube is intended for an adult population only.
# INDICATIONS FOR USE
The sOLVe Tube™ is used to isolate the left or the right lung of a patient for surgery, one lung ventilation or one lung anesthesia.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
HyTek believes that the sOLVe Tube is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design and dimensions and uses similar or identical materials as the device cleared in K051522. The subject device is uniquely designed so that it can function as both a left-sided and a right-sided DLT and includes additional safety features (i.e., ridges on the balloons to prevent balloon slippage that can lead to DLT dislodgement). The subject device has the same intended use and similar technological characteristics (maximum inflation volumes, suction catheter diameters, and design features) to the device cleared in K051522. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicate.
The technological characteristics of the subject device and the predicate device are summarized in the table below.
| | Subject Device | Predicate Device |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | sOLVe Tube™ | Silbroncho® Tubes |
| | K201026 | K051522 |
| Indications for Use | The sOLVe Tube™ is used to isolate<br>the left or the right lung of a patient<br>for surgery, one lung ventilation or<br>one lung anesthesia | The Silbroncho® is used to isolate the<br>left or the right lung of a patient for<br>surgery, one lung ventilation or one<br>lung anesthesia |
| Product Code | CBI | CBI |
| Regulation<br>Number | 21 CFR 868.5740 | 21 CFR 868.5740 |
| Technological characteristics | | |
| Design Features | • Double lumen shaft with stylet<br>• Tracheal balloon<br>• Endobronchial balloon<br>• Swivel connector with Carlens Y<br>adapter | • Double lumen shaft with stylet<br>• Tracheal cuff<br>• Endobronchial cuff<br>• Swivel connector with Carlens Y<br>adapter |
{5}------------------------------------------------
| | Subject Device | Predicate Device |
|----------------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| | sOLVe Tube™<br>K201026 | Silbroncho° Tubes<br>K051522 |
| Sizes | Device available in 1 size: 35Fr | Device available in 4 distinct sizes: 33Fr,<br>35Fr, 37Fr, and 39Fr |
| Tube O.D. | 13 mm | 33Fr: 9.5 – 12.3 mm<br>35Fr: 10.0 – 13.3 mm<br>37Fr: 10.5 – 14.3 mm<br>39Fr: 11.0 – 15.3 mm |
| Recommended<br>bronchoscope O.D. | Tracheal lumen: ≤ 6.0 mm<br>Bronchial lumen: ≤ 5.0 mm | Sizes 33 and 35Fr: ≤ 3.1 mm<br>Sizes 37 and 39Fr: ≤ 4.0 mm |
| Maximum Inflation<br>Volumes | Endobronchial balloon: < 5 mL<br>Tracheal balloon: < 10 mL | Endobronchial cuff: 4.5 – 7 mL<br>Tracheal cuff: < 10 mL |
| Suction Catheter<br>Diameter | 10Fr | 8Fr, 12Fr |
| Main Materials of<br>Composition | silicone, polypropylene | silicone, polyethylene |
| Sterilization | Sterile | Sterile |
| Sterilization<br>method | Ethylene Oxide | Ethylene Oxide |
## SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the sOLVe Tube.
The sOLVe Tube includes components that come in contact with the patient during use. The dual lumen component and the T-type suction catheters were evaluated as external communicating devices with limited duration (≤ 24 hours) with mucosal membrane. Biocompatibility testing included cytotoxicity testing per ISO 10993-5, irritation testing per ISO 10993-10, sensitization testing per ISO 10993-10, acute systemic toxicity testing per ISO 10993-11, VOC emissions and toxicity per ISO 18562-3, particulate matter analysis per ISO 18562-2, and bacterial endotoxin testing per ANSI/AAMI ST72.
The following tests were performed to demonstrate equivalence to the predicate device:
- . Performance testing (Bench)
- Determination of Cuff Diameter (per ISO 5361) O
- Resistance to Cuff Tube Collapse (per ISO 5361) O
- Resistance to Cuff Herniation (per ISO 5361) O
- Cuff Tracheal Seal (per ISO 5361) O
- Resistance to Kinking of Main Shaft and Distal Bronchial Shaft (per ISO 5361) O
- Cuff Compliance Leak Test (per ISO 5361) O
- Determination of Dual Lumen Tube Lumen Internal Diameter (per ISO 16628) O
- O Radio-opacity Test (ASTM F640)
- Balloon Bench Testing O
- Membrane Durability Test O
{6}------------------------------------------------
- Bronchoscope Insertion Test O
- о Double Clamp Operation
- O Cadaver Trial to test the Universal Design
- Performance Testing (Animal) ●
- Ease of Insertion O
- O Performance of the Tracheal and Bronchial Balloon
- Success of Lung Isolation and One Lung Ventilation o
The results of these tests support the substantial equivalence of the sOLVe Tube to the predicate device.
### CONCLUSION
The subject device and the predicate device have the same intended use, and based on the testing performed, including non-clinical performance testing (bench and animal), it can be concluded that the differences in technological features do not raise different questions of safety and effectiveness. The similar indications for use, technological characteristics, and performance characteristics for the proposed sOLVe Tube are assessed to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.