K163109 · Oxyheal Medical Systems, Inc. · CBF · Mar 22, 2017 · Anesthesiology
Device Facts
Record ID
K163109
Device Name
OxyHeal 4000 Multiplace Hyperbaric Chamber Family
Applicant
Oxyheal Medical Systems, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Mar 22, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is to administer hyperbaric oxygen therapy (HBOT) to treat patients with any of the below listed indications. The following indications which are listed on the Undersea & Hyperbaric Medical Society (UHMS) web site: www.uhms.org.are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee. 1. Air or Gas Embolism 2. Carbon Monoxide Poisoning a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning 3. Clostridial Myositis and Myonecrosis (Gas Gangrene) 4. Crush Injury, Compartment Syndrome and Other Acute Ischemias 5. Decompression Sickness 6. Arterial Insufficiencies a. Central Retinal Artery Occlusion b. Enhancement of Healing in Selected Problem Wounds 7. Severe Anemia 8. Intracranial Abscess 9. Necrotizing Soft Tissue Infections 10. Osteomyelitis (Refractory) 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis) 12. Compromised Grafts and Flaps 13. Acute Thermal Burn Injury
Device Story
Multiplace hyperbaric chamber system for administering hyperbaric oxygen therapy (HBOT). System comprises a cylindrical pressure vessel for human occupancy (PVHO) with 1-3 compartments; capacity 4-24 patients. Operates at 3-6 ATA. Inputs include compressed air (rotary screw compressors) and breathing gases (O2, medical air, mixtures) via built-in breathing system (BIBS). Controlled by a central console with HMI touchscreen; operator (qualified hyperbaric chamber technician) monitors chamber pressure, O2/CO2 levels, temperature, and humidity. Includes fire suppression (deluge/hand line), environmental control, and patient entertainment systems. Output is a pressurized, controlled environment for clinical treatment. Affects clinical decision-making by providing controlled hyperbaric environment for indicated conditions; benefits patients through therapeutic oxygen delivery.
Clinical Evidence
Bench testing only. Verification included non-destructive testing (penetrant, ultrasonic, radiographic, magnetic particle), hydrostatic testing at 1.3x MAWP, fire suppression system testing, and first operational system testing (FOST) to verify pressurization/depressurization/ventilation rates and gas delivery accuracy. Software validation confirmed control functions met design specifications.
Technological Characteristics
Cylindrical pressure vessel (PVHO) constructed of SA 516 Grade 70 steel. Complies with ASME PVHO-1, NFPA 99, and ISO 14971. Operating pressure 3-6 ATA. Features rotary screw air compressors, fire deluge/hand line suppression, BIBS, and environmental control (temp/humidity). Control console uses HMI touchscreen and PLC. Connectivity includes CCTV and entertainment systems. Sterilization/cleaning per facility protocols for hyperbaric environments.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for conditions including air/gas embolism, carbon monoxide poisoning (with/without cyanide), gas gangrene, crush injury/compartment syndrome, decompression sickness, arterial insufficiencies (central retinal artery occlusion, problem wounds), severe anemia, intracranial abscess, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised grafts/flaps, and acute thermal burns.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
Predicate Devices
OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K152223)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2017
OxyHeal Medical Systems, Inc. Edward J. Chomas VP. Regulatory Affairs 3224 Hoover Ave National City, California 91950
Re: K163109
Trade/Device Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: Class II Product Code: CBF Dated: February 22, 2017 Received: February 23, 2017
Dear Edward Chomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and a name. The name is "Tina Kiang-S". The signature is to the left of the name and is illegible.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K163109
Device Name
OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family
### Indications for Use (Describe)
The following indications which are listed on the Undersea & Hyperbaric Medical Society (UHMS) web site: www.uhms.org are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee.
- 1. Air or Gas Embolism
- 2. Carbon Monoxide Poisoning
- a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
- 3. Clostridial Myositis and Myonecrosis (Gas Gangrene)
- 4. Crush Injury, Compartment Syndrome and Other Acute Ischemias
- 5. Decompression Sickness
- 6. Arterial Insufficiencies
- a. Central Retinal Artery Occlusion
- b. Enhancement of Healing in Selected Problem Wounds
- 7. Severe Anemia
- 8. Intracranial Abscess
- 9. Necrotizing Soft Tissue Infections
- 10. Osteomyelitis (Refractory)
- 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
- 12. Compromised Grafts and Flaps
- 13. Acute Thermal Burn Injury
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 008- 510(k) Summary for
# OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family
Device Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family
### 1. Submission Sponsor
OxyHeal® Medical Systems, Inc. 3224 Hoover Avenue National City, CA 91950 Phone: 619.336.2022 Fax: 619.336.2017 Contact: W. T. 'Ted' Gurneé, President & CEO
## 2. Submission Correspondent
OxyHeal® Medical Systems, Inc. 3224 Hoover Avenue National City, CA 91950 Phone: 619.336.2022 Fax: 619.336.2017 Contact: Edward J. Chomas, VP Regulatory Affairs Email: echomas@oxyheal-international.com
## 3. Date Prepared
28 October 2016
## 4. Device Name
Trade/Proprietary Name: OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Common/Usual Name: Multiplace Hyperbaric Chamber Classification Name: Chamber, Hyperbaric Classification Regulation: 21 CBF 868.5470 Classification Panel: Anesthesiology Product Code: CBF Device Class: II
FDA Establishment Registration #: 1000519737
## 5. Predicate Device
OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family (K15223)
## 6. Device Description
The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family is comprised of a multiplace hyperbaric chamber of major subsystems that support the overall system operation, control, and monitoring.
{4}------------------------------------------------
The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber is a pressure vessel for human occupancy (PVHO) that is designed in a horizontally orientated cylindrical geometry. Chamber configurations vary based on the needs of the end user, and may be designed and manufactured in one (1), two (2), or three (3), compartment configurations. Patient capacities may range anywhere from four (4) to twenty-four (24) dependent on chamber size, number of compartments, or the direction provided by the customer to meet their needs. Lastly, maximum operating pressures range from 3ATA (~30psi) to 6ATA (~73.5psi), with each of the compartments designed to operate independently.
The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System is regulated by the same codes and standards as the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System, predicate device K152223.
- 1. NFPA 99 (2012 Edition): National Fire Protection Agency Standard for Health Care Facilities Chapter 14 - Hyperbaric Chambers
- FDA recognized consensus standard: (Recognition Number 1-67) .
- 2. ANSI/ASME PVHO-1 (2012 Edition): American National Standards Institute/American Society of Mechanical Engineers - Safety Standard for Pressure Vessels for Human Occupancy.
- FDA recognized consensus standard: (Recognition Number 1-78) ●
- 3. ISO 14971 (2012 Edition): International Standard Organization. Medical Devices Application of Risk Management to Medical Devices
- FDA recognized consensus standard: (Recognition Number 5-40) .
OxyHeal® Medical Systems, Inc. complies with the ASME/PVHO-1 the FDA recognized consensus code and standard requirements for materials, design, fabrication, and testing of the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber. This includes: overall PVHO design, joint design, welding, non-destructive examination (NDE), viewports, penetrations, material reinforcement, pressure relief devices, piping, electrical outfitting, inspections, testing, risk analysis, documentation, and marking (labeling).
The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System design also complies with the hyperbaric facilities requirements specified in the FDA recognized consensus standard NFPA 99 and satisfies the requirements for protection against electrical, explosive, and fire hazards and associated facilities used for medical procedures at gauge pressures within the ranges: 0psi to 100psi.
OxyHeal® Medical Systems, Inc.'s design control processes for the OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System are the same as those employed for the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System, predicate device K152223. These processes conform to the requirements of 21 CFR 820.30 and comply with the ISO 14971 FDA recognized consensus standard.
The OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chambers consists of the hyperbaric chamber itself and the major subsystems briefly described below. Each substantially equivalent to that which is contained in the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family predicate device (K152223).
{5}------------------------------------------------
# 6.1 Compressed Air System.
The compressed air system consists of two (2) rotary screw compressors capable of producing pressurized air that is then stored in an air receiver, which in turn is used to pressurize the hyperbaric chamber. Air is filtered prior to entering the hyperbaric chamber, resulting in a breathable quality. Compressed Gas Association (CGA) Grade "E" air as required by NFPA 99. Details of the parts and components comprising the compressed air system are identified and described in Table 1, Sub-table 1-3, Compressed Air System.
# 6.2 Fire Suppression System.
The fire suppression system (FSS) consists of both a fire deluge system (primary) and hand line system (secondary). Water (potable) for both systems is stored in pressure vessels manufactured to ASME standards. The fire deluge system is activated in the event of a fire in the hyperbaric chamber, while the hand line system is activated manually. This complies with NFPA 99:2012, Standard for Health Care Facilities Chapter 14 - Hyperbaric Facilities. Details of the parts and components comprising the FSS are identified and described in Table 1, Sub-table 1-4, Fire Suppression System.
# 6.3 Oxygen Delivery System
An oxygen (O2) delivery system is the primary source for supplying O2 to patients' breathing hoods inside the chamber. OxyHeal® Medical Systems, Inc. (OMS) provides the end user with the requirements needed for an OxyHeal® 4000 hyperbaric chamber system, and it is the end user's responsibility for supplying a system meeting these requirements. Examples of two types of O2 delivery systems and associated piping are identified and described in Table 1-5, Oxygen Delivery Requirements.
## 6.4 Built-in Breathing System
The built-in breathing system (BIBS) is capable of supplying each individually seated patient with breathing gas via standard oxygen hoods or free-flow masks. Breathing gasses can be O2, medical air, or a gas mixture. Details of the parts and components comprising the BIBS are identified and described in Table 1. Sub-table 1-6. Built-in Breathing System (BIBS).
## 6.5 Environmental Control System
The environmental control system (ECS) is used to manage the temperature (heating and cooling) and relative humidity (RH) of the hyperbaric chamber. Details of the parts and components comprising the ECS are identified and described in Table 1, Sub-table 1-7, Environmental Control System (ECS).
# 6.6 Control Console
The control console serves as the central location where a qualified hyperbaric chamber technician (CHT) is capable of controlling and monitoring an OxyHeal® 4000 product family hyperbaric chamber system. The Human-Machine Interface (HMI) touch screen control system installed in the operator control console is the primary location from which a hyperbaric chamber operator is able to initiate and monitor patient hyperbaric oxygen therapy (HBOT) treatments. Manual back-up control systems are built into the system for control of pressurization and depressurization from both inside and outside the hyperbaric chamber in the event that the automatic feature is inoperable for any reason.
From the HMI touchscreen, the operator is also able to control the following:
{6}------------------------------------------------
- Administer BIBS gasses a.
- b. Analyze / monitor O2
- Analyze/ monitor carbon dioxide (CO2) [option] c.
- Analyze / monitor relative humidity inside the hyperbaric chamber d.
- Control and monitor the temperature in the hyperbaric environment e.
- f. Open and close doors in any hyperbaric chamber compartment
- Turn ON/OFF and adjust the intensity of hyperbaric chamber lighting; and g.
- h. Perform administrative functions.
The FSS is activated from the control console. A communications system is installed at the control console allowing a CHT to communicate to patients and inside attendants within the hyperbaric chamber. The equipment used for the analysis of chamber and BIBS line gasses is mounted at the control console. Chamber lighting is managed at the control console. The control console also contains equipment used to visually monitor patients inside of the hyperbaric chamber from a CCTV, and initiate and adjust patient audio and visual entertainment (radio, CD, DVD, and TV). Details of the parts and components comprising the control console are identified and described in Table 1-8, Control Console.
{7}------------------------------------------------
# Comparison Table 1. OxyHeal® 4000 Cylindrical Multiplace Hyperbaric Chamber System Product Family Compared to the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family
| 1-1 Regulatory Info | | | |
|----------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Manufacturer | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® 4000 Family<br>Summary Comparison to<br>Predicate (K152223) |
| Trade Name | OxyHeal® 4000 Family<br>(K163109) | OxyHeal® 5000 Family<br>(K152223) | |
| 510(k) Number | K163109 | K152223 | Unique K numbers for each<br>product family have been<br>assigned. |
| Product Code | CBF | CBF | Identical |
| Regulation Number | 21 CFR 868.5470 | 21 CFR 868.5470 | Identical |
| Regulation Name | Hyperbaric Chamber | Hyperbaric Chamber | Identical |
| Indications for use: | As defined in the<br>Hyperbaric Oxygen<br>Therapy Committee Report,<br>dated 2008 | As defined in the<br>Hyperbaric Oxygen<br>Therapy Committee<br>Report, dated 2008 | Identical |
| 1-2 Hyperbaric Chamber | | | |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | OxyHeal® Medical Systems, Inc. | OxyHeal® Medical Systems, Inc. | OxyHeal® 4000 Family Summary Comparison to Predicate (K152223) |
| Trade Name | OxyHeal® 4000 Family (K163109) | OxyHeal® 5000 Family (K152223) | |
| Hyperbaric Chamber Code Design | 1. ASME: Boiler and Pressure Code<br>2. ASME PVHO-1: Safety Standard for Pressure Vessels for Human Occupancy.<br>FDA consensus standard recognition no. 1-78. | 1. ASME: Boiler and Pressure Code<br>2. ASME PVHO-1: Safety Standard for Pressure Vessels for Human Occupancy.<br>FDA consensus standard recognition no. 1-78. | Regardless of PVHO geometry:<br>Rectangular: OxyHeal® 5000 [non-uniform distribution of pressure (worst case)] Cylindrical: OxyHeal® 4000 [uniform distribution of pressure] Codes and standards defining requirements for PVHO materials, design, fabrication, and testing are identical. |
| Calculations | ASME Section VIII, Div. 1 | ASME Section VIII, Div. 1 | Identical |
| 1-2 Hyperbaric Chamber | | | |
| Manufacturer<br>Trade Name | OxyHeal® Medical<br>Systems, Inc.<br>OxyHeal® 4000 Family<br>(K163109) | OxyHeal® Medical<br>Systems, Inc.<br>OxyHeal® 5000 Family<br>(K152223) | OxyHeal® 4000 Family<br>Summary Comparison to<br>Predicate (K152223) |
| Joint Design<br>- | Design and fabrication<br>shall be in accordance<br>with specified Divisions of<br>Section VIII of the ASME<br>Code and the following<br>requirements of<br>ASME/PVHO-1, para. 1-<br>7.1 (a) through (c) | Design and fabrication shall<br>be in accordance with<br>specified Divisions of<br>Section VIII of the ASME<br>Code and the following<br>requirements of<br>ASME/PVHO-1, para. 1-<br>7.1 (a) through (c) | Identical |
| ASME PVHO-1<br>-<br>Material | SA 516 Grade 70 | SA 516 Grade 70 | Identical |
| Stress Allowance<br>- | Per ASME Section II | Per ASME Section II | Identical |
| Non-Destructive<br>-<br>Examinations<br>(NDE) | NDE shall be performed in<br>accordance with<br>ASME/PVHO-1, para. 1-<br>7.3 | NDE shall be performed in<br>accordance with<br>ASME/PVHO-1, para. 1-<br>7.3 | Identical |
| Hydrostatic Test<br>- | Hydrostatic testing to be<br>performed in accordance<br>with ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-99 @ 1.3<br>times max. allowable<br>working pressure. | Hydrostatic testing to be<br>performed in accordance<br>with ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-99 @ 1.3 times<br>max. allowable working<br>pressure. | Identical |
| Hyperbaric<br>Chamber System<br>Fire Safety Design | NFPA 99, Chapter 14 -<br>Hyperbaric Facilities.<br>FDA consensus standard<br>recognition no. 1-76. | NFPA 99, Chapter 14 -<br>Hyperbaric Facilities. FDA<br>consensus standard<br>recognition no. 1-76. | Regardless of PVHO<br>geometry:<br>• Rectangular: OxyHeal®<br>5000<br>• Cylindrical: OxyHeal®<br>4000<br>This standard defining<br>requirements for protection<br>against electrical, explosive,<br>and fire hazards in<br>hyperbaric facilities is<br>identical |
| Operating Pressure | 3.0ATA – 6.0ATA | 3.0ATA – 6.0ATA | Identical |
| Operating<br>Temperature | 50°F – 125°F | 50°F – 125°F | Identical |
| Design Temperature | 50°F – 125°F | 50°F – 125°F | Identical |
| 1-2 Hyperbaric Chamber | | | |
| Manufacturer | OxyHeal® Medical Systems, Inc. | OxyHeal® Medical Systems, Inc.<br>OxyHeal® 4000 Family Summary Comparison to Predicate (K152223) | |
| Trade Name | OxyHeal® 4000 Family (K163109) | OxyHeal® 5000 Family (K152223) | |
| Design Pressure | 30psig – 75psig | 30psig – 75psig | |
| Design Life | 90,000 cycles or 60 years, which ever happens first | 90,000 cycles or 60 years, which ever happens first | Identical |
| Hydrostatic Pressure | 39psi - 97.5psi | 39psi - 97.5psi | Identical |
| Inspection Authority | Independent 3rd Party ASME Authorized Inspector (AI).<br>Affix ASME Stamp on chamber data plate | Independent 3rd Party ASME Authorized Inspector (AI).<br>Affix ASME Stamp on chamber data plate | Identical<br>Reference FDA recognized consensus standard: ASME PVHO-1 (Recognition Number 1-78) |
| Weight (lbs.) | 12,600lbs to 100,000lbs | 15,000lbs to 120,000lbs | Substantially equivalent |
| Dimensions | For all compartments, the following min/max apply | For all compartments, the following min/max apply | Substantially equivalent |
| Main Compartment (Lock) (ML) | Min: 60" Diameter x 10'L<br>Max: 120" Diameter x 20'L | Min: 8' W x 7' H x 10'L<br>Max: 11'W x 8'H x 20'L | |
| Transfer Compartment (Lock) (EL) | | Flat Heads, Rectangular Shell, and Rectangular Door Frames. | |
| Inner Compartment (Lock) (IL) | Semi-elliptical or flat heads, cylindrical or semi-cylindrical shell, and circular or rectangular door frames | | |
| Total Volume | From 567 ft3 to 2434 ft3 | 600 ft3 to 2,600 ft3 | Substantially equivalent |
| Medical Lock | Cylindrical<br>Min: 10 inch diameter<br>Max: 16 inch diameter | Cylindrical<br>Min: 10 inch diameter<br>Max: 16 inch diameter | Identical |
| Main Doorway Size | Rectangular Door<br>Min: 39.5" W x 65" H<br>Max: 47.75" W x 75.5" H | Rectangular Door<br>Min: 44" W x 80" H<br>Max: 52" W x 80" H | Substantially equivalent |
| 1-2 Hyperbaric Chamber | | | |
| Manufacturer | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® 4000 Family<br>Summary Comparison to<br>Predicate (K152223) | |
| | OxyHeal® Medical<br>Systems, Inc. | | |
| Trade Name | OxyHeal® 4000 Family<br>(K163109) | | |
| | OxyHeal® 5000 Family<br>(K152223) | | |
| | Min: 33.5" D<br>Max: 90" D | | |
| Penetrators | Maximum of 30<br>Penetrations of 2" x 12"<br>blocks. | Identical | |
| Viewports<br>(PVHO-1) | Minimum: One (1) per<br>hyperbaric chamber.<br>Maximum: Six (6) per<br>compartment.<br>Minimum: 8" Diameter,<br>Maximum: 30" Diameter. | Identical | |
| Compartment Relief | One (1) ASME certified<br>pressure relief valve per<br>compartment.<br>30 psig to 75 psig | Identical | |
| Compartment Drain | Minimum One (1) manual<br>drain in each compartment | Identical | |
| Finish - Chamber | Sandblasted and finished<br>with 2-part high quality<br>epoxy paint with glossy<br>finish. | Identical | |
| Capacity Main<br>Compartment | 4 Patients Up to 24<br>Patients | Identical | |
| Hyperbaric Chamber Interior | | | |
| Lighting Subsystem | LED lights<br>Min: 4<br>Max: 15<br>Chamber illumination<br>complies with paragraph<br>14.2.3 of NFPA 99 (FDA<br>consensus std. recognition<br>no. 1-67). | Identical | |
| BIBS with<br>Overboard Dump | Four (4) to Twenty-Four<br>(24), on demand gas<br>delivery. | Identical | |
| Hoods with<br>Overboard Dump | Four (4) to Twenty-Four<br>(24), 1-100LPM delivery | Identical | |
| | <b>1-2 Hyperbaric Chamber</b> | | |
| Manufacturer | OxyHeal® Medical Systems, Inc. | OxyHeal® Medical Systems, Inc. | OxyHeal® 4000 Family Summary Comparison to Predicate (K152223) |
| Trade Name | OxyHeal® 4000 Family (K163109)<br>flow for hoods: 40-lpm. | OxyHeal® 5000 Family (K152223)<br>flow for hoods: 40-lpm. | |
| Depth Measurement | Digital with analog backup | Digital with analog backup | Identical |
| Pressure Transmitter | Ranges:<br>Minimum = 0 psig<br>Maximum = 75 psig | Ranges:<br>Minimum = 0 psig<br>Maximum = 75 psig | Identical |
| Temperature Sensor | 75°F ± 5°F | 75°F ± 5°F | Identical |
| Relative Humidity Sensor | 0% - 100% | 0% - 100% | Identical |
| Television (TV) System | 24VDC LED TV Monitors. Customer specified. Two (2) or four (4) per treatment compartment | 24VDC LED TV Monitors. Customer specified. Two (2) or four (4) per treatment compartment | Identical |
{8}------------------------------------------------
### 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{9}------------------------------------------------
{10}------------------------------------------------
flow meters. Minimum
flow meters. Minimum
{11}------------------------------------------------
| nsmitter | Minimum = 0 psig<br>Maximum = 75 psig | Minimum = 0 psig<br>Maximum = 75 psig | |
|-----------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------|
| perature Sensor | 75°F ± 5°F | 75°F ± 5°F | Identical |
| ative Humidity<br>sor | 0% - 100% | 0% - 100% | Identical |
| vision (TV)<br>em | 24VDC LED TV<br>Monitors. Customer<br>specified.<br>Two (2) or four (4) per<br>treatment compartment | 24VDC LED TV Monitors.<br>Customer specified.<br>Two (2) or four (4) per<br>treatment compartment | Identical |
| 1-3 Compressed Air System | | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Manufacturer | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® 4000 Family<br>Summary Comparison to<br>Predicate (K152223) |
| Trade Name | OxyHeal® 4000 Family<br>(K163109) | OxyHeal® 5000 Family<br>(K152223) | |
| Air Compressors | Qty. two (2) rotary screw<br>Operating range:<br>Rated Power:<br>208V ±10%, 3-ph, 60Hz -<br>460V ±10%, 3-ph, 60Hz<br>Max Working Pressure:<br>125psig - 217psig<br>Air Delivery<br>155 CFM - 288 CFM | Qty. two (2) rotary screw<br>Operating range:<br>Rated Power:<br>208V ±10%, 3-ph, 60Hz -<br>460V ±10%, 3-ph, 60Hz<br>Max Working Pressure:<br>125psig - 217psig<br>Air Delivery<br>155 CFM - 288 CFM | Identical |
| Air Receiver | Ranges: Qty. 1 – Qty.5<br>depending on space and<br>location within client<br>facility. | Ranges: Qty. 1 – Qty.5<br>depending on space and<br>location within client<br>facility. | Substantially equivalent |
| - Design Code | ASME Boiler and Pressure<br>Code | ASME Boiler and Pressure<br>Code | Identical |
| - Calculations | ASME Section VIII, Div. 1 | ASME Section VIII, Div. 1 | Identical |
| - Joint Design | Joint design and fabrication<br>shall be performed in | Joint design and fabrication<br>shall be performed in | Identical |
| 1-3 Compressed Air System | | | |
| Manufacturer | OxyHeal® Medical<br>Systems, Inc.<br>OxyHeal® 4000 Family<br>(K163109) | OxyHeal® Medical<br>Systems, Inc.<br>OxyHeal® 5000 Family<br>(K152223) | OxyHeal® 4000 Family<br>Summary Comparison to<br>Predicate (K152223) |
| Trade Name | accordance with UW-12<br>and Table UW-12 of<br>ASME Boiler and Pressure<br>Vessel Code, Division, I,<br>Section VIII | accordance with UW-12<br>and Table UW-12 of<br>ASME Boiler and Pressure<br>Vessel Code, Division, I,<br>Section VIII | Identical |
| Non-<br>-<br>Destructive<br>Examinations<br>(NDE) | Radiographic examination<br>and ultrasonic examination<br>shall be performed in<br>accordance with UW-51 of<br>ASME Boiler and Pressure<br>Vessel Code, Division, I,<br>Section VIII | Radiographic examination<br>and ultrasonic examination<br>shall be performed in<br>accordance with UW-51 of<br>ASME Boiler and Pressure<br>Vessel Code, Division, I,<br>Section VIII | Identical |
| | Ultrasonic examination<br>shall be performed in<br>accordance with UW-53<br>and Appendix 12 of ASME<br>Boiler and Pressure Vessel<br>Code, Division, I, Section<br>VIII | Ultrasonic examination<br>shall be performed in<br>accordance with UW-53<br>and Appendix 12 of ASME<br>Boiler and Pressure Vessel<br>Code, Division, I, Section<br>VIII | Identical |
| Hydrostatic<br>-<br>Test | Hydrostatic testing to be<br>performed in accordance<br>with ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-99 @ 1.3 times<br>maximum allowable<br>working pressure. | Hydrostatic testing to be<br>performed in accordance<br>with ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-99 @ 1.3 times<br>maximum allowable<br>working pressure. | Identical |
| Pressure Vessel<br>Relief | One (1) pressure relief<br>satisfying the requirements<br>of ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-125.<br>Rated at 200 psig | One (1) pressure relief<br>satisfying the requirements<br>of ASME Boiler and<br>Pressure Vessel Code,<br>Section UG-125.<br>Rated at 200 psig | Identical |
| Filters | | | Identical |
| Liquid Separator<br>- | Removes moisture from air<br>using centrifugal force | Removes moisture from air<br>using centrifugal force | Identical |
| Prefilter Filter<br>- | 5 micron element | 5 micron element | Identical |
| Coalescing Filter<br>- | 1 micron element | 1 micron element | Identical |
| Active Carbon<br>-<br>Filter | Charcoal element | Charcoal element | Identical |
| 1-3 Compressed Air System | | | |
| Manufacturer | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® 4000 Family<br>Summary Comparison to |
| Trade Name | OxyHeal® 4000 Family<br>(K163109) | OxyHeal® 5000 Family<br>(K152223) | Predicate (K152223) |
| Particulate Filter | 3 micron element | 3 micron element | Identical |
| High Efficiency<br>Filter | .01 micron element | .01 micron element | Identical |
| Piping | ASTM B88 | ASTM B88 | Identical |
| Testing | OMS Factory Acceptance<br>Test (FAT) | OMS Factory Acceptance<br>Test (FAT) | Substantially Equivalent |
{12}------------------------------------------------
{13}------------------------------------------------
| 1-4 Fire Suppression System | | | | |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | OxyHeal® Medical<br>Systems, Inc. | OxyHeal® Medical<br>Systems, Inc.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.