MONOPLACE HYPERBARIC OXYGEN TREATMENT SYSTEM, MODEL BLKS-307
K060739 · Khrunichev State Research & Production Space Centr · CBF · Jul 21, 2006 · Anesthesiology
Device Facts
Record ID
K060739
Device Name
MONOPLACE HYPERBARIC OXYGEN TREATMENT SYSTEM, MODEL BLKS-307
Applicant
Khrunichev State Research & Production Space Centr
Product Code
CBF · Anesthesiology
Decision Date
Jul 21, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The conditions listed as appropriate for the use of the Hyperbaric Oxygen Therapy in the current edition of Undersea and Hyperbaric Medical Society (UHMS) Hyperbaric Oxygen Therapy Committee Report (1999) are as follows: Air or gas embolism. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning. Clostridial myositis and myonecrosis. Crush injury, compartment syndrome, and other acute traumatic ischemias. Decompression sickness. Enhanced healing of selected problem wounds. Exceptional blood loss anemia. Necrotizing soft tissue infections. Osteomyelitis (refractory). Delayed radiation injury (soft tissue and bony necrosis). Skin grafts and flaps (compromised). Thermal burns. Intracranial abscess.
Device Story
BLKS-307 is a monoplace hyperbaric oxygen chamber; hermetic cylindrical shell made of AMg6 aluminum alloy; used in clinical settings for hyperbaric oxygen therapy. Patient enters via retractable bed on portable gurney. System operates via pneumatic control panel atop chamber; allows variable pressurization/depressurization rates for patient comfort. Includes emergency decompression button for rapid session termination. Safety features: pressure-relief valves (set at 43.5 PSI), antistatic bracelet, and grounding cable to prevent electrostatic discharge. Healthcare provider monitors patient and controls pressure; output is pressurized oxygen environment. Benefits include treatment of specific ischemic, infectious, and wound-related conditions.
Clinical Evidence
Bench and clinical testing data provided. The device demonstrated safety and effectiveness comparable to the predicate. No specific clinical trial metrics (e.g., p-values, sensitivity) were detailed in the summary.
Technological Characteristics
Materials: AMg6 aluminum alloy (shell), SB-211 A 92024 (lock ring), carbon steel (base), acrylic glass. Dimensions: 100" L x 48" W x 70.9" H. Operating pressure: 1-4 ATA. Control: Pneumatic. Connectivity: Standard electronic patient vitals/environment parameter indication. Standards: ASME PVHO, NFPA 99, ISO 9001, ISO 13485.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for conditions including air/gas embolism, carbon monoxide poisoning, clostridial myositis/myonecrosis, crush injury/compartment syndrome/acute traumatic ischemias, decompression sickness, problem wound healing, exceptional blood loss anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
Predicate Devices
Khrunichev BLKS-303 MK Monoplace Hyperbaric System (K0011312)
Submission Summary (Full Text)
{0}------------------------------------------------
K060739
## JUL 21 2006
### EXHIBIT 2 510(k) Summary
Khrunichev State Research and Production Space Centre 18, Novozavodskaya 121309 Moscow,
February 10, 2006
### Contact: Gleb Mitinskiy
- l . Identification of the Device: Proprietary-Trade Name: BLKS-307 MONOPLACE HYPERBARIC OXYGEN TREATMENT SYSTEM Classification Names: CBF Hyperbaric Chamber Common/Usual Name: Hyperbaric Chamber
- 2. Equivalent legally marketed devices: Khrunichev BLKS-303 MK Monoplace Hyperbaric System, K0011312.
- 3. Indications for Use (intended use) The conditions listed as appropriate for the use of the Hyperbaric Oxygen Therapy in the current edition of Undersea and Hyperbaric Medical Society (UI·IMS) Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- . Air or gas embolism
- . Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- Clostridial myositis and myonecrosis .
- . Crush injury, compartment syndrome, and other acute traumatic ischemias
- Decompression sickness .
- . Enhanced healing of selected problem wounds
- . Exceptional blood loss anemia
- . Necrotizing soft tissue infections
- . Osteomyelitis (refractory)
- . Delayed radiation injury (soft tissue and bony necrosis)
- Skin grafts and flaps (compromised) .
- Thermal burns .
- . Intracranial abscess
- 4. Description of the Device: The system body is a hermetic cylindrical chamber, which consists of a welded hermetic shell and a quick-to-open cover, made of the AMg6 -ASME Code Case 2403 aluminum alloy. The cover lock ring is made of the SB-211 A 92024 material. The hyperbaric oxygen chamber is installed on a base, which is made of carbon steel, with four wheel castors attached to the base bottom. It can be moved within a treatment room if necessary. A control panel is located atop of on the pressure chamber, actuating pneumatic aggregates of the pressure system pneumatic system which are located under the chamber body. A patient's ingress/egress is performed by means of a retractable chamber bed that is wheeled out of the pressure chamber on a portable gurney, which is attached to the chamber at the time the patient
{1}------------------------------------------------
is placed in and out of the chamber. The pressure system overall dimensions BI.KS-307-"Khrunichev" without the gurney: length - 100 in, width - 48 in, height - 70.9 in. The pressure system total weight BLKS-307-"Khrunichev" - 1102 lbs. Chamber pressure control is achieved by means of knobs and tumbler switches of a pneumatic control panel located atop of the chamber. Both the rates of pressurization and depressurization are variable to provide patient's comfort while in confined space and with regards to his/her individual tolerance to pressure rate change. The panel is equipped with an emergency decompression push button, which enables fast treatment session termination and egress of a patient in case of an emergency situation. If pressure inside the chamber exceeds maximal operating level of 43.5 PSI, two pressure-relief valves installed on the pressure chamber shell will open. This provides for the patient and personnel safety and prevents the chamber structure damage. In order to exclude possible electric spark formation in an oxygen atmosphere in case of transition of an accumulated electrostatic potential from a patient to the chamber shell, an antistatic bracelet is provided. The bracelet is put on a patient hand or foot. The system is equipped with a grounding cable, which is connected to the treatment room grounding system.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and clinical testing indicates that the new device is as safe and effective as the predicate device.
| Parameters | Units | Khrunichev<br>BLKS-303MK<br>K011312. | Khrunichev<br>BLKS-307 |
|-----------------------------------------------------------------------|-----------------------------------|--------------------------------------|---------------------------------|
| Internal Diameter | In (mm) | 28.7 (730) | 42.1 (1080) |
| Length | In (mm) | 93.7 (2380) | 90.4 (2300) |
| Internal Volume | Ft3 (m3) | 26.9 (0.95) | 56.5 (1.6) |
| Weight | Lbs (kg) | 770 (320) | 1102 (500) |
| Operating Pressure | ATA (Bar) | 1 - 4 | 1 - 4 |
| Compression / Decompression<br>Rate | ATA/min<br>(Bar/min) | 0.05 - 0.25 | 0.05 - 0.25 |
| Emergency Decompression<br>Time from 4 to 1 ATA | Sec | 60 | 90 |
| Maximum Oxygen Consumption | | | |
| Isopression (constant<br>pressure) and<br>decompression | l/min | 50 | 52.5 |
| Blowing and compression | l/min | 400 | 400 |
| One hour treatment | l (m3) | 6000 (6) | 9000 (9) |
| Lifetime | One-hour<br>treatment<br>sessions | 10,000 | 10,000 |
| Control System | | Pneumatic | Pneumatic |
| Chamber Type | | Monoplace | Monoplace |
| Electronic Patient Vitals and<br>Environment Parameters<br>Indication | | Optional | Standard |
| Chamber Door Lock | | Manual | Pneumatic |
| Material | | Aluminum alloy/acrylic<br>glass | Aluminum alloy/acrylic<br>glass |
| Intercom Device | | Standard | Standard |
#### 6. Substantial Equivalence Chart
{2}------------------------------------------------
| Parameters | Units | Khrunichev<br>BLKS-303MK<br>K011312. | Khrunichev<br>BLKS-307 |
|-------------------------------------------------------|-------|------------------------------------------------|--------------------------|
| Built-In-Breathing System | | Optional | Optional |
| Standards and Regulatory<br>Documents<br>Compliant to | | FDA 510(k) | ASME |
| | | ASME | PVHO |
| | | PVHO | NFPA 99 |
| | | NFPA 99 | ISO 9001 |
| | | ISO 9001 | ISO 13485 |
| | | ISO 13485 | TUV |
| | | TUV | National Pressure |
| | | National Pressure<br>Vessels Code | Vessels Code |
| | | National Ministry of<br>Healthcare Certificate | National Ministry of |
| | | Healthcare Certificate | National Fire Protection |
| | | National Fire Protection<br>Code | Code |
#### 7. Conclusion
After analyzing both bench and clinical testing data, it is the conclusion of Khrunichev State Research and Production Space Centre that the BLKS-307 MONOPLACE HYPERBARIC OXYGEN TREATMENT SYSTEM is as safe and effective as the predicate device, and has few technological differences, thus it substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.
JUL 2 1 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Khrunichev State Research & Production Space Center C/O Mr. Daniel Kamm Regulatory Consultant Engineer Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K060739
Trade/Device Name: BLKS-307 Monoplace Hyperbaric Oxygen Treatment System Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: CBF Product Code: II Dated: June 15, 2006 Received: June 23, 2006
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 -Mr. Kamm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K06
Device Name: BLKS-307 MONOPLACE HYPERBARIC OXYGEN TREATMENT SYSTEM
Indications For Use:
The conditions listed as appropriate for the use of the Hyperbaric Oxygen Therapy in the current edition of Undersea and Hyperbaric Medical Society (UHMS) Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- Air or gas embolism .
- Carbon monoxide poisoning and carbon monoxide poisoning complicated by . cyanide poisoning
- Clostridial myositis and myonecrosis .
- Crush injury, compartment syndrome, and other acute traumatic ischemias .
- Decompression sickness .
- Enhanced healing of selected problem wounds .
- Exceptional blood loss anemia .
- Necrotizing soft tissue infections .
- . Ostcomvelitis (refractory)
- Delayed radiation injury (soft tissue and bony necrosis) .
- Skin grafts and flaps (compromised) .
- Thermal burns .
- Intracranial abscess .
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mhald Mhalal for Annbrahan. 7/21/06
Number K060739
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.