MULTIPLACE HYPERBARIC SYSTEM, MODEL MMI HYPERBARIC CHAMBER SYSTEM
K030418 · Makai Marine Industries, Inc. · CBF · Jan 29, 2004 · Anesthesiology
Device Facts
Record ID
K030418
Device Name
MULTIPLACE HYPERBARIC SYSTEM, MODEL MMI HYPERBARIC CHAMBER SYSTEM
Applicant
Makai Marine Industries, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Jan 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is the expressed, intended use of the Makai Marine Industries, Inc. MMI Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use. The UHMS is the Is ratther medical organization chartered with setting the standards of care defining the propriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit. The conditions listed as appropriate for the use of (HBO) in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows: 1. Air or gas embolism 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning 3. Clostridial myositis and Myonecrosis (Gas Gangrene) 4. Crush injury, compartment syndrome, and other acute traumatic ischemias 5. Decompression sickness 6. Enhanced healing of selected problem wounds 7. Exceptional blood loss anemia 8. Necrotizing soft tissue infections 9. Osteomvelitis (refractory) 10. Delayed radiation injury (soft tissue and bony necrosis) 11. Skin grafts and flaps (compromised) 12. Thermal burns 13. Intracranial abscess The Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities. Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operations of all related equipment such as the hyperbaric chamber.
Device Story
Class A multiplace hyperbaric chamber; treats up to 20 patients; max pressure 6 ATA (73.5 psig). Input: compressed air for pressurization; 100% oxygen for treatment via oral-nasal masks or head tents. Operation: single operator controls pressure via adjustable valves; low-voltage intercommunication system connects patients to operator. Used in medical facilities; operated by trained medical personnel. Features: large circular windows; rectangular doors for gurneys/wheelchairs; penetrator plates for IV lines/monitoring leads. Benefit: provides controlled hyperbaric environment for validated medical conditions; improves patient handling safety for nonambulatory patients.
Clinical Evidence
No clinical data. Device relies on established medical conditions for hyperbaric oxygen therapy as defined by the Undersea and Hyperbaric Medical Society (UHMS) 1999 Committee Report.
Technological Characteristics
Class A multiplace hyperbaric chamber; 11-40 ft length, 42-180 inch diameter. Materials/construction per ANSI/ASME Boiler and Pressure Vessel Code Section VIII Div 1 and ASME-PVHO-1. Pressurization via compressed air; treatment gas 100% oxygen. Includes low-voltage patient intercommunication system per NFPA 99. Mechanical pressure control via adjustable valves.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for conditions including air/gas embolism, carbon monoxide poisoning, gas gangrene, crush injury/compartment syndrome, decompression sickness, problem wounds, exceptional blood loss anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
Predicate Devices
Perry Sigma MP Multiplace Hyperbaric Chamber (K930748)
Tampa Hyperbaric Enterprises' Multiplace Hyperbaric Chamber (K960389)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K030418" in a handwritten style. The text is oriented diagonally, with the "K" at the bottom left and the "8" at the top right. The handwriting appears to be in black ink on a white background, and the letters are somewhat bold and slightly distorted.
JAN 2 9 2004
# Makai Marine Industries, Inc.
## EXHIBIT "A"
510(k) NOTIFICATION SUMMARY (Per 21 CFR 807.9)
Prepared: 27 January 2003
TRADE NAME: Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System COMMON NAME OF DEVICE: Multiplace Hyperbaric Chamber CLASSIFICATION: 73 CBF, 21 CFR 868.5470 ESTABLISHMENT REGISTRATION NUMBER: 1063240
#### CLAIMED PREDICATE DEVICE(S):
Perry Sigma MP Multiplace Hyperbaric Chamber - K930748
Gulf Coast Hyperbarics Multiplace Hyperbaric Chamber - K950957
Reimers T Class Hyperbaric Facility - K95438
Tampa Hyperbaric Enterprises' Multiplace Hyperbaric Chamber - K960389
#### ADDRESS OF MANUFACTURER:
Makai Marine Industries, Inc. P.O. Box 272039 Boca Raton, FL 33427-2039 730 S. Deerfield Avenue # 8 & 9 Deerfield Beach, FL 33441
CONTACT PERSON:
Marc Kaiser (305) 987-2247, Fax (305) 856-8795
{1}------------------------------------------------
#### EXECUTIVE SUMMARY
The Undersea and Hyperbaric Medical Society (UHMS) defines hyperbaric oxygen therapy as Inc Onection and Ityperoat pressures higher than atmospheric in a hyperbaric chamber. breating to the National Fire Protection Association (NFPA), hyperbaric chambers for humans are classified into two categories: Class A (multi-occupant) and Class B (single occupant). The Makai Marine Industries Hyperbaric Chamber System is a Class A multiplace hyperbaric chamber system designed to treat up to 20 patients to a maximum treatment pressure of 6 Atmospheres Absolute (ATA) or 73.5 pounds per square inch gauge (psig). The chamber uses compressed air as the pressurization gas and 100% oxygen as the hyperbaric treatment gas.
The Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System is intended to be Procured and used by physicians to treat a variety of medical conditions that respond to produces and and as a los by of the Medical Society (UHMS) produces a list of medical conditions that have been identified for the appropriate primary or adjunctive use of hyperbaric oxygen. These approved conditions include: air or gas embolism; carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning; clostridial myositis and myonecrosis (gas gangrene); crush injury, compartment syndrome and other acute traumatic ischemias; decompression sickness; enhanced healing of selected problem wounds; exceptional blood loss anemia; necrotizing soft tissue infections; osteomyelitis (refractory), delayed radiation injury (soft tissue and bony necrosis); compromised skin flaps and grafts; thermal bums; and, intracranial abscess. Aggressive research into the beneficial effects of hyperbaric oxygen, when appropriately applied, will result in additional medical conditions being added to the list of indications by the UHMS.
The Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System is designed and will be fabricated in accordance with the requirements of the ANSI/ASME Boiler and Pressure Vessel Code, Section VIII, Division 1, Pressure Vessels; ANSVASME-PVHO-1 (American Society of Mechanical Engineers Pressure Vessels for Human Occupancy); and, NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities. The overall external length of the chamber can range from 11 feet to 40 feet. Its internal diameter can range from 42 inches to 180 inches. When combined, these features allow for the state of the art, hands on treatment of one (1) to 20 ambulatory patients in comfort. Large circular windows create a more open atmosphere and help reduce patient claustrophobia. Pressurization is provided by compressed air with 100% oxygen, administered to the patient by using properly fitting oral-nasal masks or head tents.
A low-voltage patient intercommunication system designed and installed in accordance with NFPA 99, Chapter 20 and provides communications between the patients in the chamber and the outside chamber operator. It also provides patients with audio program content from external sources such as TV's, cassette players, radios, etc. The system consists of a master station mounted on the chamber operator's control console that contains all of the controls and connection points.
Single operator chamber pressure control is achieved via simple adjustable control valves providing infinite control.
Multiple penetrators and/or penetrator plate is provided in the vessel wall to allow user supplied intravenous lines, medical monitoring leads, etc., to be used as required. Where applicable a large rectangular door(s) allows a normal size patient gurney or wheelchair to be
{2}------------------------------------------------
used to transport nonambulatory patients without having to transfer the patient to a smaller transport device. This feature greatly improves patient handling safety. MMI has concluded that the general design approach, method of pressure control, and intended use of the Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System is substantially equivalent to the Perry Industries, the (innif) fryperoats Multiplace Hyperbaric Chamber (K930748); the Tampa Baronical Berrices brighternaric Chamber (K960389); the Gulf Coast Hyperbarics, Inc. Hyperbaric Enterprises Tryportant System (K950957); the Reimers Engineering T Class Multiplace Tryperbaric Treamean bys proposing them as predicate devices for the MMI Hyperbaric Chamber System.
#### INTENDED USE:
It is the expressed, intended use of the Makai Marine Industries, Inc. MMI Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use. The UHMS is the Is ratther medical organization chartered with setting the standards of care defining the propriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit.
The conditions listed as appropriate for the use of (HBO) in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- 1. Air or gas embolism
- 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- 3. Clostridial myositis and Myonecrosis (Gas Gangrene)
- 4. Crush injury, compartment syndrome, and other acute traumatic ischemias
- 5. Decompression sickness
- 6. Enhanced healing of selected problem wounds
- 7. Exceptional blood loss anemia
- 8. Necrotizing soft tissue infections
- 9. Osteomvelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
The Makai Marine Industries, Inc. (MMI) Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities. Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operations of all related equipment such as the hyperbaric chamber.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three parallel lines curving upwards and to the right, with a wavy line underneath.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 9 2004
Mr. Marc R. Kaiser Makai Marine Industries, Inc. P.O. Box 27-2039 Boca Raton, FI. 33427-2039
Re: K030418
Trade/Device Name: Model MMI Multiplace Hyperbaric Chamber Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: November 4, 2003 Received: November 10, 2003
Dear Mr. Kaiser:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becases is see is substantially equivalent (for the referenced above and have determined are to tegally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate conmicree prior to thay 20, 1978, clossified in accordance with the provisions of Amendinents, on to devices that have book (Act) that do not require approval of a premarket the Federal Food, Drug, and Cosment I 100 may, therefore, subject to the general approval approvision (1 hirvest. The general controls provisions of the Act include controls provisions of the Fre. - The Litting of devices, good manufacturing practice, lequirements for aihitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (PMA), It may be subject to such adam size and Regulations, Title 21, Parts 800 to 898. In the And your device can oc found in the Could receined.
addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Marc R. Kaiser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the rest or mily with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirentle, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you dontact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Inter of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General IIospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): #K030418
#### Device Name: MMI Hyperbaric Chamber System
Indications For Use:
The conditions listed as appropriate for the use of (HBO) in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- 1. Air or gas embolism
- 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- 3. Clostridial myositis and Myonecrosis (Gas Gangrene)
- 4. Crush injury, compartment syndrome, and other acute traumatic Ischemias
- 5. Decompression sickness
- 6. Enhanced healing of selected problem wounds
- 7. Exceptional blood loss anemia
- 8. Necrotizing soft tissue infections
- 9. Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
signature
Anesthesiology Genera
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.