K122690 · Arrow International, Inc. · CAZ · Dec 21, 2012 · Anesthesiology
Device Facts
Record ID
K122690
Device Name
ULTRAQUIK PERIPHERAL NERVE BLOCK NEEDLE
Applicant
Arrow International, Inc.
Product Code
CAZ · Anesthesiology
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arrow UltraQuik® Peripheral Nerve Block Needle is indicated for the delivery of single-shot peripheral nerve block anesthesia or analgesia using medical imaging devices.
Device Story
Single-shot peripheral nerve block needle; consists of cannula, hub, and fluid extension tube with female luer fitting. Available in various gauges (20-24 Ga), lengths (35-150 mm), and tip styles (bevel, pencil point, Touhy). Features echogenic grooves on cannula to enhance visibility under ultrasound. Used in hospitals, sub-acute facilities, pain clinics, and physician offices by clinicians to assist in needle tip location during nerve block procedures. Device facilitates delivery of anesthetic/analgesic agents; ultrasound guidance aids accurate placement, potentially improving procedural success and patient outcomes.
Clinical Evidence
Bench testing only. Evaluated tensile strength, injection tubing leakage, needle sharpness, corrosion resistance, needle stiffness, and component usability after 5-year accelerated shelf-life. Comparative ultrasound imaging performed against predicate; clinical preference survey indicated clinicians found the echogenic design easier to recognize.
Technological Characteristics
Non-insulated metallic cannula with echogenic grooves; hub; fluid extension tube with female luer lock. Available in 20-24 Ga, 35-150 mm lengths. Tip styles: bevel, pencil point, Touhy. Sterile, single-use. No electrical components or software.
Indications for Use
Indicated for delivery of single-shot peripheral nerve block anesthesia or analgesia in patients requiring peripheral nerve block procedures. Contraindicated in patients with skin sepsis at insertion site, systemic sepsis, or pre-existing nerve injury, neuritis, or plexitis. Use caution in patients with bleeding tendencies or those receiving anticoagulants.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{0}------------------------------------------------
K122690
510(k) Summary Page 1 of 6 29-Nov-12 . .
### DEC 2 1 2012
Arrow International Inc. 2400 Bernville Road . Reading, PA 19605
Official Contact:
Paul Amudala Regulatory Affairs Specialist
Tel: 610-378-0131 Fax: 610-478-3179
Proprietary or Trade Name: ARROW UltraQuik Peripheral Nerve Block Needle
Common/Usual Name: Peripheral Nerve Block
Classification Name:
Product code - CAZ CFR 868-5140 - anesthetic conduction kit Class 2
Predicate Device:
A VID-NIT (Arrow) - StimuQuik™ PNB Needle - K014246
### Device Description:
The Arrow UltraQuik Peripheral Nerve Block (PNB) Needle is a single shot needle which is comprised of a cannula, hub, and fluid extension tube with female luer fitting. The single shot needles are available in various gauges, lengths, and point styles. These include:
- Gauge .
- 20 Ga (Touhy only) o
- 21 to 24 Ga 0
- . Lengths
- 35 mm to 150 mm 0
- Point style ●
- Bevel tip o
- Pencil point 0
- Touhy tip 0
- Cannula echogenicity
- The cannula is naturally echogenic but the addition of grooves on the cannula 0 enhance the echogenicity
The UltraQuik may be used with medical imaging, i.e., ultrasound to assist in needle tip location.
### Basic Components for Peripheral Nerve Block needle
A typical basic Peripheral Nerve Block ("PNB") needle includes the following components.
{1}------------------------------------------------
### 510(k) Summary Page 2 of 6 29-Nov-12
- Cannula with sharp tip .
- Hub
- . Fluid extension line with female luer connector
### Indications for Use:
The Arrow UltraQuik® Peripheral Nerve Block Needle is indicated for the delivery of singleshot peripheral nerve block anesthesia or analgesia using medical imaging devices.
### Patient Population:
Patients requiring peripheral nerve block procedures.
### Environment of Use:
The environment of use is - hospital, sub-acute facilities, pain clinics, physician offices
### Contraindications:
Pre-existing nerve injury, neuritis or plexitis are relative contraindications for use of peripheral nerve blocks. These conditions should be considered prior to needle insertion. Skin sepsis in the area where the needle placement is planned and systemic sepsis are relative contraindications. Extreme care should be taken in patients with bleeding tendencies or patients receiving anticoagulants.
### Comparison to Predicates
The UltraQuik PNB Needle for single shot non-stimulating nerve block procedures is viewed as substantially equivalent to the predicate device because:
### Indications -
The intended uses are identical and the indications for use are identical except for the nonstimulating feature. Use of a non-stimulating PNB needle is common and not a new procedure. Equivalent to predicate - K014246 -- AVID-NIT StimuQuik PNB Needle.
### Technology -
The components of the proposed UltraQuik PNB Needle are identical to the predicate K014246, A VID-NIT StimuQuik PNB Needle and there is no new technology or materials.
### Materials -
The materials are identical to the predicate, K014246 - AVIV-NIT StimuQuik PNB Needle.
### Environment of Use -
The proposed environments of use are identical to the predicate K014246 - AVID-NIT StimuQuik PNB Needle. They are - hospital, sub-acute facilities, pain clinics, and physician offices.
{2}------------------------------------------------
### 510(k) Summary Page 3 of 6
29-Nov-12
### Patient Population -
Patients requiring peripheral nerve block procedures. Identical to the predicate, K014246 -AVID-NIT StimuQuik PNB Needle.
### Non-clinical Testing Summary -
The components of the proposed device are identical to the predicate K014246, except the addition of echogenic grooves. We have performed bench testing after 5 year accelerated shelflife. This testing included:
- Tensile strength for regions of the needle not intended for insertion into the body ●
- . Injection tubing to needle hub
- Injection tubing to Luer Lock connector
- Tensile strength for the needle cannula to needle hub .
- . Needle sharpness
- Injection tubing leakage
- Resistance to leakage during aspiration or vacuum .
- Corrosion resistance of metallic components .
- Needle Stiffness
- Luer Lock connector .
### Summary of Test Results
| Description | Results |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile Strength<br>For regions not<br>Inserted in body | Pass / fail criteria force to break is > 15 N<br>Injection Tubing to needle hub - mean 27 N<br>Injection Tubing to Luer Lock - mean 27 N |
| Tensile Strength<br>Needle cannula to<br>Hub joint | Pass / fail criteria force to break > 40 N<br>All have a mean > 99 N |
| Needle Sharpness | Pass / fail criteria - none<br>Penetration force - all have a mean >3.17 N |
| Injection leakage | Pass / fail criteria - no leak at 50 psi for 30 sec<br>No leakage observed |
| Resistance to<br>leakage | Pass / fail criteria - no internal leakage at - 25 mmHg for 120 sec<br>No leakage observed |
| Corrosion<br>Resistance | Pass / fail - no visible sign of corrosion<br>No observed corrosion |
{3}------------------------------------------------
### 510(k) Summary Page 4 of 6 29-Nov-12
| Needle Stiffness | Pass / fail criteria - deflection < 0.5 mm ·<br>Deflection - all have a mean < 0.5 mm |
|-------------------------|---------------------------------------------------------------------------------------|
| Needle Markings | Pass / fail - permanent markings at 10 mm<br>Cannula are so marked |
| Component<br>Capability | Pass / fail - usability testing<br>Results were favorable |
### Imaging and Echogenic property evaluation:
Comparative ultrasound imagines were performed on the proposed design vs. the predicate. In addition a clinical preference survey was done with clinicians and all found the echogenic design to be easier to recognize.
{4}------------------------------------------------
## 510(k) Summary
## Predicate Comparative Table
.
| Features | Proposed<br>UltraQuik PNB Needle | Predicates<br>K014246 – AVID-NIT – StimuQuik PNB Needle |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification name | Anesthetic Conduction Kit | Anesthetic Conduction Kit |
| Product Code | CAZ<br>868.5140 | CAZ<br>868.5140 |
| Intended Use | Single Shot nerve block anesthetic administration | Single Shot nerve block anesthetic administration |
| Indications for use | The Arrow UltraQuik® Peripheral Nerve Block Needle is<br>indicated for the delivery of single-shot peripheral nerve block<br>anesthesia or analgesia using medical imaging devices. | The AVID-NIT Nerve Stimulation Needles consist of an<br>insulated hollow needles intended for locating peripheral<br>nerves by electrical stimulation, and for the single shot<br>administration of a local anesthetic drug. These needles are<br>to be used with battery powered peripheral nerve locators<br>only. These needles' are specifically not intended for nerve<br>stimulation for purposes other than nerve location. These<br>needles are specifically not intended for neurolytic<br>ablation. |
| Environment of Use | Hospital, sub-acute facilities, pain clinics, physician offices | Hospital, sub-acute facilities, pain clinics, physician offices |
| Patient Population | Patients requiring peripheral nerve block procedures | Patients requiring peripheral nerve block procedures |
| Contraindications | Pre-existing nerve injury, neuritis or plexitis are relative<br>contraindications for use of peripheral nerve blocks. These<br>conditions should be considered prior to needle insertion.<br>Skin sepsis in the area where the needle placement is<br>planned and systemic sepsis are relative contraindications.<br>Extreme care should be taken in patients with bleeding<br>tendencies or patients receiving anticoagulants. | Pre-existing nerve injury, neuritis or plexitis are relative<br>contraindications for use of peripheral nerve blocks. These<br>conditions should be considered prior to needle insertion.<br>Skin sepsis in the area where the needle placement is<br>planned and systemic sepsis are relative contraindications.<br>Extreme care should be taken in patients with bleeding<br>tendencies or patients receiving anticoagulants. |
| Basic components | Non-insulated cannula with hub<br>Fluid extension tube with luer lock fitting | Insulated cannula with insulated coating and hub<br>Fluid extension tube with luer lock fitting<br>Lead wire for connection to a nerve stimulator |
| Features | Proposed<br>UltraQuik PNB Needle<br>Component Design and Specifications | Predicates<br>K014246 - AVID-NIT - StimuQuik PNB Needle |
| Cannula | | |
| Gauges / dimensions | 20 Ga for Touhy<br>21 - 24 Ga | 21 - 24 Ga |
| Cannula length | 35 - 150 mm | 35 - 150 mm |
| Tip styles | Beveled<br>Pencil point<br>Touhy | Beveled<br>Pencil point |
| Cannula | Smooth<br>Grooved | Smooth |
| Cannula insulated | No, not required | Yes |
| Cannula markings | Yes | Yes |
| Fluid Extension Tube | with female luer lock fitting | with female luer lock fitting |
| Lead wire | No, this is not a feature of the proposed device | Yes for connection to nerve stimulator |
| Echogenic Properties | Yes | Yes |
| Provided Sterile | Yes | Yes |
| Shelf-life | 5 years | 5 years |
Page 5 Page 5
K122690 Page 14 of 39
{5}------------------------------------------------
## 510(k) Summary
# Substantial Equivalence Conclusion
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predical
have been found to substantially
.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### December 21, 2012
Mr. Paul Amudala Regulatory Affairs Specialist Arrow International, Incorporated 2400 Bernville Road READING PA 19605
Re: K122690
Trade/Device Name: UltraQuik Peripheral Nerve Block Needle Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ Dated: November 29, 2012 Received: December 3, 2012
Dear Mr. Amudala: · ·
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
### Page 2 - Mr. Amudala
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthong La m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use Statement
Page 1 of 1
510(k) Number:
. K122690
Device Name:
UltraQuik Peripheral Nerve Block Needle
Indications for Use:
The Arrow UltraQuik® Peripheral Nerve Block Needle is indicated for the delivery of single-shot peripheral nerve block anesthesia or analgesia using medical imaging devices.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
### Lester W. Schultheis Jr 2012.12.20 1,2:32:25 -05'00'
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number: k122690
K122690 Page 9 of 39
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.