UNIPLEX NANOLINE CANNULA, PLEXOLONG NANOLINE SET, STIMULONG NANOLINE SET
Applicant
Pajunk GmbH
Product Code
CAZ · Anesthesiology
Decision Date
Jan 12, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pajunk's electrically insulated anesthesia conduction needles -- UniPlex-NanoLine cannula -are intended for professional anesthetists to facilitate the localization of peripheral nerve bundles for transient delivery of anesthetics during regional anesthesia. Pajunk's electrically insulated anesthesia conduction needles are also provided within PlexoLong- and StimuLong and online sets intended for professional anesthetists to facilitate the transient delivery of anesthetics during regional anesthesia. The localization of the peripheral nerve bundle is facilitated via insulated needles and impulses from a neuro-stimulator. The electric current flows from the de-insulated needles and mights innervated by the stimulated nerve bundle will the de-insulated needle the suitable setting. Following nerve bundle localization, a catheter may be placed through the insulated anesthesia conduction needle close to the news for longer pain relief. After the needle has been withdrawn, the anesthesia conduction to to to to for to for the fielde nate that been withdra anesther into the negipt call relater allows for bolus injections of local anesthetics into the peripheral nervous space.
Device Story
Device consists of electrically insulated needles (UniPlex-NanoLine), conduction catheters, and adapters; used by professional anesthetists in clinical settings. Input: electrical impulses from external neuro-stimulator. Operation: needle acts as electrode to localize peripheral nerve bundles via current flow from de-insulated tip; catheter placed through needle for subsequent bolus delivery of anesthetics. Output: localized nerve stimulation and delivery of anesthetic agent. Benefit: facilitates regional anesthesia and provides longer-term pain relief via catheter placement. NanoLine coating replaces previous lacquer coating for improved material purity.
Clinical Evidence
Bench testing only. Biocompatibility testing performed with successful results. Design changes validated using protocols identical to those used for previously cleared predicate devices.
Technological Characteristics
Electrically insulated needles and catheter sets. Features NanoLine (Parylene) coating on needles. Components include luer hubs, stylets, and catheters. Materials are identical to predicate devices. Sterilization via established methods. No software or active electronic components; device is passive, used with external neuro-stimulator.
Indications for Use
Indicated for professional anesthetists to facilitate peripheral nerve bundle localization and transient delivery of local anesthetics during regional anesthesia in patients requiring regional anesthesia or pain relief.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
HDC's Insul-Cote PTM needles and CLA kit (K994059)
Submission Summary (Full Text)
{0}------------------------------------------------
K 053283
JAN 1 2 2006
NanoLine - Premarket Notification Submission
Image /page/0/Picture/3 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, black letters. The background of the logo is black and white.
# 510(k) Premarket Notification Submission: Summary of Safety and Effectiveness
## Date of Preparation: October 27, 2005
## Submitter Information/ production site:
Pajunk GmbH Karl-Hall-Straße 01 78187 Geisingen Germany Fon: +49(0)7704-9291-0 Fax: +49(0)7704-9291-610
#### Contact:
Christian Quaß, Regulatory Affairs Fon: +49(0)7704-9291-533 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com
### Device Information:
| Trade Names: | UniPlex NanoLine cannula,<br>PlexoLong-NanoLine set,<br>StimuLong-NanoLine set | | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------|--|--|--|
| Common Name: | Anesthesia conduction kit/ cannula | | | |
| Classification Name: | Anesthesia conduction kit | | | |
| Classification Reference: | 21 CFR 868.5140, April 1. 2003 | | | |
| Proposed Classification: | Regulatory Class: II | | | |
| Proposed Product Code: | CAZ | | | |
| Predicate Devices: | 1. K000722: Pajunk's Unipolar needles (UniPlex facet & Sprotte tip) | | | |
| | 2. K023218: Pajunk's Unipolar needles (UniPlex Tuohy tip) | | | |
| | 3. K033018: Pajunk's StimuLong set | | | |
| | 4. K013041, K023218 and K042979: Pajunk's PlexoLong set | | | |
| | 5. K994059: HDC's Insul-Cote PTM needles and CLA kit<br>(nerve stimulation), includes HDC's Neuro-Trace insulated | | | |
needle.
ﯿﻨﭧ
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for PAJUNK MEDIZINTECHNOLOGIE. The word "PAJUNK" is in large, bold, white letters against a black background. Below that, the word "MEDIZINTECHNOLOGIE" is in smaller, thinner letters.
## Device Description:
Pajunk's PlexoLong and StimuLong sets consist of a Pajunk unipolar needle, an open ended conduction catheter and a catheter adapter, as well as a filter. The coating of the needle will be chnaged to NanoLine.
Pajunk's Unipolar needles marketed separately as single shot needles under the trade name UniPlex have a lacquer coating, which will be changed to NanoLine coating,
Only the coating material of the needles is affected by this change.
The anesthesia conduction catheters and the packaging materials are the same as those used for Pajunk's PlexoLong sets cleared for market by FDA under 510(k) numbers K013041, K023218 and K042979.
The NanoLine coating is equivalent to HDC's Insul-Cote PTM needles and CLA kit (nerve stimulation) marketed under K994059. (In fact NanoLine is a better material because of a higher grade of purity. For a precise discussion of this topic see section 15 "Biocompatibility" of this submission of a special 510(k)).
The contract sterilizer other than a company name change (was IBA Griffith Micro Science, and now is Sterigenics) and the stepally hame change (was film inford Sterice, and
este sets.
#### Intended Use:
Pajunk's electrically insulated anesthesia conduction needles -- UniPlex-NanoLine cannula -are intended for professional anesthetists to facilitate the localization of peripheral nerve bundles for transient delivery of anesthetics during regional anesthesia.
Pajunk's electrically insulated anesthesia conduction needles are also provided within PlexoLong- and StimuLong and online sets intended for professional anesthetists to facilitate the transient delivery of anesthetics during regional anesthesia.
The localization of the peripheral nerve bundle is facilitated via insulated needles and impulses from a neuro-stimulator. The electric current flows from the de-insulated needles and mights innervated by the stimulated nerve bundle will the de-insulated needle the suitable setting.
Following nerve bundle localization, a catheter may be placed through the insulated anesthesia conduction needle close to the news for longer pain relief. After the needle has been withdrawn, the anesthesia conduction to to to to for to for the fielde nate that been withdra
anesther into the negipt call relater allows for bolus injections of local anesthetics into the peripheral nervous space.
## Technology Characteristics:
Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong- & StimuLong -NanoLine sets – have the same technological characteristics as the predicate devices identified above.
Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong-NanoLine set & StimuLong-NanoLine set – are equivalent (in fact they are the same) in design, physical dimensions, luer hub, stilett, metal and plastics materials of needles and catheter, filter and connect of there, there, materials of needles and
catheter, filter and connector, as well as packaging to Pajunk's PlexoLong sets cleared for
· ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "PAJUNK" in large, bold, white letters against a black background. The letters are slightly distorted, with some pixelation and jagged edges, giving the image a somewhat rough or textured appearance. The overall impression is that of a logo or brand name, possibly for a product or company.
MEDIZINTECHNOLOGIE
market by FDA under 510(k) numbers K013041, K023218 and K042979. The NanoLine coating material is substantial equivalent to HDC's Insul-Cote PTM needles and CLA hit (nerve stimulation) marketed under K994059.
Biocompatibility testing of Pajunk's anesthesia conduction needles & sets – UniPlex-NanoLine cannula, PlexoLong- & StimuLong -NanoLine sets -- has been accomplished with best results.
## Summary of Design Control Activities
Changes that could affect performance of the cannula or result in a risk to the patient or the user were validated using the same protocols used to validate cleared devices. Results can be found in the summary of Risk-Analysis and in the Validation Reports of Section 09.
#### Conclusion:
The electrically insulating coating on the cannula is only a change in material for Pajunk GmbH products. The comparison between the predicate devices and the proposed devices demonstrates that the proposed devices are safe and effective when used according to the instructions for use supplied with the devices, as well as substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### JAN 1 2 2006
Mr. Christian Quaß Manager Regulatory Affairs Pajunk GmbH Karl-Hall Strasse 1 78187 Geisingen GERMANY
Re: K053283
Trade/Device Name: Pajunk Anesthesia Conduction NanoLine Coated Needles Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ, BSP Dated: December 21, 2005 Received: December 27, 2005
Dear Mr. Quaß:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Quaß
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Ours
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Special Premarket Notification Submission
## NanoLine
Image /page/5/Picture/2 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in a smaller, sans-serif font.
## Section 04
#### Indications for use
510(k) Number: K053283
Device Name: Pajunk anesthesia conduction NanoLine coated needles
Indications for Use:
The change of coating material from lacqueur to Parylene does not affect the Indications for use of the predicated devices marketed under
K042979 (Pajunk Plexolong Sets)
K033018 (Pajunk Stimulong Plus Catheter Sets)
K023218 (Pajunk Plexolong Sets)
K013041 (Pajunk Plexolong Anesthesia Sets)
K999722 (Unipolar Needles with Standard and Sprotte-tip)
The change in coating has no effect on the Indications for use of these devices already cleared for market.
Prescription Use AND/OR (Per 21 CFR 801.109)
Over-The-Counter Use
(21 CFR 801 Subpan C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clues
K.053283
Page 1 of 1
| PMN | |
|------------|--|
| CQ | |
| 2006-01-05 | |
| P 1 of 1 | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.