PRECISION PORTABLE OXYGEN CONCENTRATION MODEL 4150
Applicant
Precision Medical, Inc.
Product Code
CAW · Anesthesiology
Decision Date
Jul 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Precision Medical EasyPulse portable oxygen concentrator (model PM4150) is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device can be used in a home, institution, vehicle and for transportable use.
Device Story
The EasyPulse POC (Model PM4150) is a lightweight, battery-powered device providing supplemental oxygen to patients requiring therapy. It utilizes Vacuum Swing Adsorption (VSA) technology with synthetic zeolite molecular sieves to extract oxygen from ambient air. The device features an electronic conserving mechanism that monitors the user's breath rate to deliver a precise pulsed dose of oxygen across 5 settings. It is intended for use in homes, institutions, vehicles, and mobile settings. Power is supplied via an internal non-removable lithium-ion battery, an external AC/DC power supply, or a 12V DC automobile adapter. The device automatically charges the internal battery when connected to external power. It includes various alarms for system monitoring, including low battery, breath detection, and temperature alerts. The device is operated by the patient or caregiver; output is delivered via a nasal cannula. By providing portable, on-demand oxygen, the device facilitates patient mobility and adherence to prescribed oxygen therapy.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electromagnetic compatibility (EN 60601-1-2), mechanical/electrical safety (UL 60601-1, IEC 60601-1), vibration/shock (IEC 60068 series), and performance verification (oxygen concentration 87-95%, bolus volume accuracy, battery runtime, trigger sensitivity, and environmental extremes).
Technological Characteristics
Materials: Synthetic zeolite sieve bed. Principle: Vacuum Swing Adsorption (VSA). Energy: Internal 14.4V Li-ion battery, 100-240 VAC or 11-16 VDC external power. Dimensions: 10.1" x 6.5" x 4.5". Weight: 6.8 lbs. Connectivity: None. Software: Microprocessor-controlled valves for pulse flow delivery. Standards: EN 60601-1-2, UL 60601-1, IEC 60601-1, IEC 60068-2-6/27/64.
Indications for Use
Indicated for patients requiring supplemental oxygen therapy. Suitable for use in home, institutional, vehicle, and mobile environments. Not intended for life support or life sustaining use.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
# K103563
# 510(k) Summary Precision Medical, Inc. Portable Oxygen Concentrator
JUL - 1 2011
**Submitter Information**
| Submitter | Precision Medical, Inc.<br>300 Held Drive<br>Northampton, Pa.<br>18067 |
|--------------------------|----------------------------------------------------------------------------|
| FDA registration number: | 2523148 |
| Contact | James Parker<br>Quality Assurance Manager |
| Tel: | (610)-262-6090 Extensions 228 |
| Fax: | (610)-262-6080 |
| Preparation Date: | June 29, 2011 |
| Device Name | Precision Medical EasyPulse portable oxygen concentrator<br>(Model PM4150) |
| Proprietary Name: | Precision Medical EasyPulse portable oxygen concentrator<br>(Model PM4150) |
| Common Name: | Portable Oxygen Generator |
| Regulation Number: | 21CFR 868.5440 |
| Product Code: | CAW<br>Class II Device |
#### Predicate Device Equivalence
Precision Medical, Inc. is claiming substantial equivalence to the Invacare Flyer Model XPO100 (510K 071928)
## Intended Use
The Precision Medical EasyPulse portable oxygen concentrator (model PM4150) is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device can be used in a home, institution, vehicle and for transportable use.
{1}------------------------------------------------
### Description of device
The EasyPulse Portable Oxygen Concentrator (POC), model PM4150, was designed to The Lasyl uise I ordain onygen equiring supplemental oxygen. The EasyPulse POC uses provide moonly to patients requiring from the electronic conserving device technology to oxygen concentrator with a lightweight, battery powered oxygen concentrator. The device provide the pation while a nghivoriging on various mobile environments. A nasal can be used in the nome, mistration, xxygen from the device to the patient. The device is not intended to be life supporting or life sustaining.
The EasyPulse POC provides pulsed dose delivery to the user through a selection of 5 settings. Setting 1 delivers the lowest amount of oxygen and setting 5 provides the settings. Setting I don't read to rewelvolumes for each setting and oxygen concentration are listed in the device specifications. Oxygen minute volume is controlled electronically are ristou in and conner's breath rate. By monitoring the user's breath rate the bolus by monitoring are about the patient the same amount of oxygen per minute.
The EasyPulse POC offers the user multiple options to power the device. The device I he Lasyl uise I OC Offers ion battery, not accessible by the patient, that will provide mendes an meetime minute for outsely setting chosen. Stated battery runtimes are val ying full thires dopendent upon are external AC/DC power supply is included with the Insted in the AC/DC power supply allows the user to power the device from a standard 110 VAC outlet. An external DC/DC automobile adapter is also provided with the 110 VAC outler. The external DC/DC automobile adapter allows the user to power the device from a standard 12V automobile accessory jack. When connected to either the AC/DC power a standard 12 v automobile adapter, the device will default to charging the supply of external DOD C auce is in the OFF position. If the unit is turned on while micrhar battery when the abovee as ice will simultaneously run (provide oxygen) and charge the internal battery.
{2}------------------------------------------------
# Specifications of Device
r
: ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・
| Dimensions | Height: 10.1" (25.4 cm)<br>Width: 6.5" (16.5 cm)<br>Depth: 4.5" (11.3 cm) |
|-----------------------------------|-------------------------------------------------------------------------------------------------|
| Weight | Device: 6.8 lbs (3.08 kg) |
| Altitude | Up to 9000 ft (2743 m) |
| Storage Temperature | 41°F to 104°F (5°C to 40°C) |
| Operating Temperature | -2°F to 140°F (-20°C to 60°C) |
| Relative Humidity | Up to 95% Non-Condensing |
| Oxygen Concentration | 87% to 95% |
| Oxygen Minute Volume | Setting 1: 240 +/-15% cc/min |
| | Setting 2: 380 +/-15% cc/min |
| | Setting 3: 520 +/-15% cc/min |
| | Setting 4: 660 +/-15% cc/min |
| | Setting 5: 780 +/-15% cc/min |
| Power | AC Power Adapter: 100-240 VAC (1.0 A @ 120 VAC)<br>DC Power Adapter: 11-16 VDC (5.0 A @ 12 VDC) |
| Battery Duration<br>(approximate) | Setting 1: 4.5 hrs |
| | Setting 2: 3.2 hrs |
| | Setting 3: 2.5 hrs |
| | Setting 4: 1.8 hrs |
| | Setting 5: 1.5 hrs |
| Sound (@ Setting 2) | 44 dBA |
| Cannula | Maximum 7ft cannula |
4.2
{3}------------------------------------------------
# Predicate Comparison
:
| Feature/Specification | | Precision Medical | Invacare |
|-------------------------------------|--------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | | The Precision Medical EasyPulse POC (model<br>PM4150) is intended to provide supplemental<br>oxygen to persons requiring oxygen therapy.<br>The Precision Medical POC can be used in a<br>home, institution, vehicle and various mobile<br>environments. | The Invacare XP02 is intended to be used by patients with<br>respiratory disorders who require supplemental oxygen. It<br>supplies a high concentration of supplemental oxygen and is<br>used with a nasal cannula to channel oxygen from the<br>concentrator to the patient. The Invacare XP02 can be used in<br>a home, institution, vehicle and various mobile environments.<br>The Invacare XP02 does not, nor is it intended to sustain or<br>support life. |
| Method by which oxygen is released | | Molecular Sieve (mechanical) | Molecular Sieve (mechanical) |
| Process by which oxygen is released | | Pressure/Vacuum Swing Absorption | Pressure Swing Absorption |
| Sieve Bed Material | | Synthetic Zeolite | Synthetic Zeolite |
| Software/Hardware | | Analog & digital electronics w/ microprocessor | Analog & digital electronics w/ microprocessor |
| Flow control | | Microprocessor controlled valves | Microprocessor controlled valves |
| Weight | | 6.8 lbs w/ internal battery | < 7.0 lbs w/ internal battery |
| | Height | 10.0" | 10.0" |
| Outside<br>Dimensions | Width | 6.5" | 7.0" |
| | Depth | 4.5" | 4.3" |
| Oxygen % | | 87% to 95% all flow rates | 87% to 95.6% all flow rates |
| Flow Rates | | Pulse Flow (Minute Volume)<br>Tolerance +/-15%<br>Setting 1: 260 cc/min<br>Setting 2: 380 cc/min<br>Setting 3: 520 cc/min<br>Setting 4: 660 cc/min<br>Setting 5: 760 cc/min | Pulse Flow (Minute Volume)<br>Tolerance Not Available<br>Setting 1: 240 cc/min<br>Setting 2: 380 cc/min<br>Setting 3: 480 cc/min<br>Setting 4: 600 cc/min<br>Setting 5: 700 cc/min |
| Operating Temperature | | 41 to 104°F (5 to 40°C) All power sources | 41 to 95°F (5 to 35°C) All power sources<br>95 to 104°F (35 to 40°C) continuous use w/ AC or DC adapter<br>95 to 104°F (35 to 40°C) limited use w/ internal battery |
| Storage Temperature | | -2 to 140°F (-20 to 60°C) | -2 to 140°F (-20 to 60°C) |
| Operating Humidity | | Up to 95% RH non-condensing | Up to 95% RH non-condensing |
| Storage Humidity | | Up to 95% RH non-condensing | Up to 95% RH non-condensing |
| Operating Altitude | | Up to 9000 ft | Up to 10000 ft |
| Sound Level (@ Setting 2) | | 44 dBA | 45 dBA |
| Power options | | AC/DC Power supply<br>Input: 100-240 VAC , 50/60 Hz<br>Output: 18 VDC, 60W<br>DC/DC Power supply<br>Input: 11-16 VDC , 50/60 Hz<br>Output: 18 VDC, 60W | AC/DC Power supply<br>Input: 100-240 VAC , 50/60 Hz<br>Output: 18 VDC, 60W<br>DC/DC Power supply<br>Input: 11-16 VDC , 50/60 Hz<br>Output: 18 VDC, 60W |
| Battery | | Non-removable<br>Lithium-Ion - 14.4 VDC, 5.0 Ah | Non-removable<br>Lithium-Ion - 14.8 VDC, 5.2 Ah |
| System Alarms | | Low Battery Warning<br>Low Battery Alarm<br>No Breath Detect Alarm<br>Excessive Breath Rate Alarm<br>Compressor Alarm (Abnormal Motor Current)<br>Battery Temperature Alarm<br>Motor Temperature Alarm<br>Pressure Alarm<br>Fan Failure Alarm<br>Motor Connection Alarm<br>Battery Connection Alarm<br>Performance Alarm (Monitor system pressures) | Low Battery Warning<br>Low Battery Alarm<br>Battery Discharged Alarm<br>No Breath Detect Alarm<br>Breath Rate Over Capacity Alarm<br>System Too Hot/Cold for Start Alarm<br>System Too Hot/Cold Running Alarm<br>Battery Too Hot/Cold<br>Stuck Button Alarm<br>Operating Alarm (Abnormal Operating Conditions)<br>Compressor Alarm (Abnormal Compressor Conditions)<br>System Alarm (Abnormal System Conditions) |
43
{4}------------------------------------------------
## Comparison Technology
The only significant difference relative to the predicate is the predicate device uses a PSA The only sightincan unforched form. For the Precision Medical POC opted for use of a (ITCSSure Swing Adsorption) og Adsorption) cycle. Both cycle types are proven VI SA ( vacuum 1 result o wing ther devices on the market. Precision Medical felt a technology and used in Tarrous es in reducing complexity of the pneumatic circuit and reducing power consumption.
The predicate comparison table shows the EasyPulse POC is similar to the Invacare The predicate comparison acto the wo do use, use similar methods for extracting oxygen AI O2. Dour deviece have similar environmental specifications, similar oxygen purity and provide similar power options.
# Summary of Performance Testing
| Electromagnetic Compatibility Testing Performed per: | |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| EN 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirements for safety<br>Collateral standard: Electromagnetic compatibility |
| Mechanical and Electrical Safety Testing Performed per: | |
| UL 60601-1, 1st Edition 2006-04-26 | Medical electrical Equipment, Part 1: General Requirements for Safety |
| CAN/CSA-C22.2 No.601.1-M90, 2005 | Medical electrical Equipment, Part 1: General Requirements for Safety |
| IEC 60601-1:1998, A1:1991, A2:1995 | Medical electrical Equipment, Part 1: General Requirements for Safety |
| IEC 60068-2-6 | Sinusoidal Vibration |
| IEC 60068-2-27 | Shock/Bump |
| IEC 60068-2-64 | Random Vibration |
| VOC and Particulate Testing Performed per: | |
| EPA PM₂.₅ | Particulate Matter |
| ASTM D5466 | Standard Test Method for Determination of Volatile Organic Chemicals<br>in Atmospheres (Canister Sampling Methodology) |
| 21 CFR 801.415 | Ozone Levels |
| EPA NAAQS - Carbon Monoxide | Carbon Monoxide Levels |
| OSHA Permissible Exposure Limits | Carbon Dioxide Levels |
| Performance Testing: | |
| 738-2 Weight and Dimensions | Measure overall weight and size of device |
| 738-3 Operating Temperature | Verify performance at extremes of operating temperature range |
| 738-4 Storage Temperature | Subject packaged device to extremes of storage temperature range |
| 738-5 Barometric Pressure Range | Verify performance at maximum altitude |
| 738-6 Functionality | Verify basic functions of test units |
| 738-8 Outlet Temperature | Measure gas outlet temperature under max operating temperature |
| 738-9 Oxygen Concentration | Measure O2 concentration after operating for 1 hour |
| 738-10 Mean Oxygen Concentration | Measure O2 concentration over 9 hour period |
| 738-11 Bolus Volume | Measure O2 minute volume for each setting, at every breath rate from<br>15-35 breaths/min |
| 738-12 Basic Performance | Measure bolus volume @ setting 5 over 9 hour period |
| 738-13 Internal Battery Charging | Measure O2 concentration and bolus volume for each setting |
| 738-14 Sound Level | Measure battery runtimes for each setting and battery charge time |
| 738-15 Trigger Delay/Sensitivity | Measure sound of device while operating |
| 738-16 ISTA Drop Test | Measure trigger delay and trigger sensitivity<br>Packaging drop test |
4.4.
{5}------------------------------------------------
## Conclusions
In Summary, Precision Medical, Inc. has demonstrated that the Precision Medical, Inc. III Sulminary, I recision Modical, the nas actively to the combined testing and analysis of results provides assurance that the device meets the specifications and is safe and effective for the intended use.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Precision Medical Inc. C/O Mr. James Parker Quality Assurance Manager 300 Held Drive Northhamption, Pennsylvania 18067
JUL - 1 2011
Re: K103563
Trade/Device Name: Precision Medical Easy Pulse Portable Oxygen Concentrator (Model PM4150) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Concentrator Regulatory Class: II Product Code: CAW Dated: June 27, 2011 Received: June 28, 2011
Dear Mr. Parker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 -- Mr. Parker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division-of-Postmarket-Surveillance: -
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Antion O.n
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation · Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### Portable oxygen concentrator 510K 103563
## 510 (k) number (if known) K103563 Device Name: Precision Medical EasyPulse portable oxygen concentrator (Model PM4150)
#### Indications for use:
The Precision Medical EasyPulse portable oxygen concentrator (model PM4150) is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device can be used in a home, institution, vehicle, and for transportable use.
(Please do not write below this line- continue on another page if needed) Concurrence of CDRH, office of device Evaluation (ODE)
Prescription Use: X Or (Per 21 CFR 80
Over the counter use (Optional Format 1-2-9)
(Division Sign-Oil) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K115663
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.