INVACARE TPOC PORTABLE CONCENTRATOR, MODEL TPOC100
Applicant
Invacare Corporation
Product Code
CAW · Anesthesiology
Decision Date
May 21, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Invacare TPOC Portable Oxygen Concentrator is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare TPOC Portable Oxygen Concentrator can be used in a home, institution, vehicle, and various mobile environments. The Invacare TPOC Portable Oxygen Concentrator does not nor is it intended to sustain or support life.
Device Story
Portable oxygen concentrator (TPOC) for patients with respiratory disorders; provides supplemental oxygen via nasal cannula. Operates via molecular sieve and pressure swing adsorption; ambient air filtered and compressed; nitrogen removed by synthetic zeolite sieve beds; concentrated oxygen (87-95.6%) delivered to patient. Dual-mode operation: Pulse Delivery Mode (breath-detected delivery) and Continuous Flow Mode (up to 3 LPM). Integrated heated Zirconia oxygen sensor monitors purity; provides patient warnings/alarms. Used in home, institution, vehicle, or mobile environments. Supports humidification and oxygen bleed-in for PAP/Bi-Level devices during continuous flow. Powered by AC (120 VAC) or DC (12 VDC) or internal Li-ion battery. Healthcare providers use output to manage patient oxygenation; device improves patient mobility and access to supplemental oxygen.
Clinical Evidence
Bench testing only. Performance testing conducted per risk analysis to validate modifications (continuous flow, oxygen sensor, battery). Results demonstrated that predetermined acceptance criteria were met.
Technological Characteristics
Molecular sieve (synthetic zeolite) with pressure swing adsorption. Microprocessor-controlled valves. Integrated heated Zirconia oxygen sensor. Flow: 0.5-3.0 LPM continuous; 1-5 pulse settings. Power: 100-240 VAC, 50/60 Hz or 11-16 VDC. Battery: Li-ion, 16.8V, 15.6 Ah. Dimensions: ≤ 18.0 lbs. Connectivity: None specified. Software: Microprocessor-based control of flow and alarms.
Indications for Use
Indicated for patients with respiratory disorders requiring supplemental oxygen. Not intended to sustain or support life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Invacare logo. The logo is in black and white and features the word "INVACARE" in bold, white letters against a black oval background. The oval has a curved line above and below the wordmark.
. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
### 510(K) SUMMARY FOR THE INVACARE TRANSPORTABLE OXYGEN CONCENTRATOR (TPOC)
This summary of 510(k) safety and effectiveness information is being supplied in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92
The assigned 510(k) number is_ 1( ひ ¥ 3 つく
Date: December 11, 2008
Submitted by: Invacare Corporation Registration No. 1525712 One Invacare Way Elyria, Ohio 44035-4190
Telephone: 440-326-6356 Fax: 440-326-3458
Contact Person: Mr. Carroll Martin
Trade Name: TPOC
Model: TPOC100
Common Name: Oxygen Concentrator
Classification Name: Generator, Oxygen, Portable per 21 CFR 868.5440
Legally Marketed Predicate Device(s): Invacare Flyer Oxygen Concentrator; K0719028, Dec 12, 2007.
Device Description: The Invacare TPOC is to be used by patients with respiratory disorders who require supplemental oxygen. The device can be used in the home, an institution, vehicle and various mobile and outdoor environments. The device is not intended to sustain or support life. The device is typically used with a nasal cannula to direct oxygen from the device to the patient.
When set for Continuous Flow Mode, the TPOC can deliver up to 3 LPM continuous flow in 0.5 LPM increments. While in Pulse Delivery Mode, when the demand for oxygen is detected, the oxygen is delivered through pulsed flow with pulse flow settings of 1 through 5. The oxygen concentration level of the output gas ranges from 87% to 95.6% in all modes.
The modified TPOC contains an integrated oxygen sensor to provide added monitoring of the output gas purity, as well as additional patient warnings and alarms based on those levels. The oxygen sensor process uses a heated Zirconia element as the method of measurement. This is the exact same sensor process used in previously cleared oxygen concentrators such as the Invacare Platinum 5 Oxygen Concentrator, K020386.
**INVACARE CORPORATION**
One Invacare Way P.O. Box 4028 Elvria, OH 44036-2125 USA 440-329-6000 Fax: 440-366-1803 www.invaca
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Invacare logo. The logo is in black and white and features the word "INVACARE" in a bold, sans-serif font. The word is set inside of a black oval shape with white lines above and below the word. The logo is simple and recognizable.
With the addition of continuous flow, the TPOC is now capable of supporting two features commonly with the addition of ocultified no ne nextrators; usc with a humidification and use in conjunction with PAP, Bi-Level and other such devices as an oxygen bleed-in. The use of this equipment is only during continuous flow mode, as indicated in the User's Manual. odulphileit is only aaning othium battery pack, standard power options include switching power Anong with a rents from AC power outlet (120 VAC/ 60 Hertz nominal) and from accessory DC outlets typically found in a mobile vehicle type environment (12 VDC nominal).
The Invacare TPOC uses a molecular sieve and pressure swing adsorption methodology to produce the The in theater Tr C abber a nair enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient in specific volumes during the inhalation portion of a detected breath in pulse mode. In continuous flow mode, the operation is the same with the exception that the oxygen is delivered continuously, without the need for breath detection.
The basic technology of the Invacare TPOC is equivalent to its predicate device.
Intended Use: The Invacare TPOC is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is weed with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare TPOC can be used in a home, institution, vehicle, and various mobile environments. The Invacare TPOC does not nor is it intended to sustain or support life.
| Features | Invacare TPOC100 | Invacare Flyer |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K0719028 |
| Date Cleared | TBD | 12/12/2007 |
| Intended Use | The Invacare TPOC is intended to be<br>used by patients with respiratory<br>disorders who require supplemental<br>oxygen. It supplies a high concentration<br>of supplemental oxygen and is used with<br>a nasal cannula to channel oxygen from<br>the concentrator to the patient. The<br>Invacare TPOC can be used in a home,<br>institution, vehicle and various mobile<br>environments. The Invacare TPOC does<br>not nor is it intended to sustain or<br>support life. | The Invacare Flyer is intended to be<br>used by patients with respiratory<br>disorders who require supplemental<br>oxygen. It supplies a high<br>concentration of supplemental oxygen<br>and is used with a nasal cannula to<br>channel oxygen from the concentrator<br>to the patient. The Invacare Flyer can<br>be used in a home, institution, vehicle<br>and various mobile environments. The<br>Invacare Flyer does not nor is it<br>intended to sustain or support life |
| Method by which<br>Oxygen is Released | Molecular sieve (mechanical) | Molecular sieve (mechanical) |
Substantial Equivalence:
VACARE CORPORATIC
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Invacare logo. The logo is in black and white and features the word "INVACARE" in a bold, sans-serif font. The word is placed inside of a black oval shape. The logo is simple and recognizable.
| Process by which<br>Oxygen is Released | Pressure swing adsorption | Pressure swing adsorption |
|----------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|
| Sieve Bed Material | Synthetic zeolite | Synthetic zeolite |
| Software/<br>Hardware | Analog and digital electronics with<br>microprocessor | Analog and digital electronics with<br>microprocessor |
| Flow Control | Microprocessor controlled valves | Microprocessor controlled valves |
| Weight | ≤ 18.0 lbs. (without battery module or<br>cart) | ≤ 7.0 lbs. (with internal battery) |
| Oxygen Purity | 87% minimum at all flow rates | 87% minimum at all flow rates |
| Oxygen Sensor | Yes | No |
| Flow Rates | Continuous flow rates of 0.5 - 3.0 LPM<br>(0.5 LPM resolution) | No Continuous flow rate |
| | Pulse Flow - 1, 2, 3, 4, 5 | Pulse Flow - 1, 2, 3, 4, 5 |
| Power Options | AC power supply<br>100-240 V, 50/60 Hz;<br>DC power supply 11-16 V | AC power supply<br>100-240 V, 50/60 Hz;<br>DC power supply 11-16 V |
| Battery | Li-ion, 16.8 Volt, 15.6 Ah | Li-ion, 16.8 Volt, 5.2 Ah |
, 花
やらなくないとい
Performance Testing: The performance testing conducted as a result of the modifications to the Invacare Flyer, as required by the risk analysis, was performed and the results demonstrated that the predetermined acceptance criteria were met.
INVACARE CORPORATION
One Invacare Way P.O.Box 4028 Elyria, OH 44036-2125 USA 410-329-6000 Fax: 440-366-1803 www.invacare.com
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Carroll L. Martin, RAC Regulatory Manager Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036-2125
Re: K083751
Trade/Device Name: Invacare TPOC Portable Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: April 30, 2009 Received: May 4, 2009
MAY 2 1 2009
Dear Mr. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2- Mr. Martin
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Runner
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known): |کر 3 7 3 7 3 7 /
## Device Name: Invacare TPOC Portable Oxygen Concentrator
Indications for Use: The Invacare TPOC Portable Oxygen Concentrator is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare TPOC Portable Oxygen Concentrator can be used in a home, institution, vehicle, and various mobile environments. The Invacare TPOC Portable Oxygen Concentrator does not nor is it intended to sustain or support life.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k 083751
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.