K173825 · Trudell Medical International · CAF · May 2, 2018 · Anesthesiology
Device Facts
Record ID
K173825
Device Name
MC 300R* Nebulizer
Applicant
Trudell Medical International
Product Code
CAF · Anesthesiology
Decision Date
May 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The nebulizer is intended to be used with pediatic (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician. The device is designed to aerosolize prescribed medication by a patient in the hospital, clinic or home care environment. The nebulizer is a single patient use device.
Device Story
Small volume jet nebulizer; operates on Venturi principle. Compressed air/oxygen source drives flow through converging nozzle; high-velocity stream creates vacuum; liquid medication drawn from reservoir into airstream; mixture impacts rigid surface to generate aerosol droplets. Used in hospital, clinic, or home environments by patients under physician care. Single-patient use device. Output is aerosolized medication for inhalation; healthcare providers use output to deliver respiratory treatments; benefits patient via effective lung delivery of prescribed drugs.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including aerosol characterization (total mass, output rate, fine particle fraction, MMAD, GSD) using Albuterol, Ipratropium Bromide, and Budesonide, and comprehensive biocompatibility testing per ISO 10993 standards.
Technological Characteristics
Pneumatic jet nebulizer; plastic molded construction. Operates via Venturi principle using compressed air or oxygen (4-8 LPM). Components: mouthpiece, nebulizer top, nozzle cover, nebulizer bottom. Biocompatibility testing per ISO 10993-3, 5, 10, 11, 12, 17, 18. Dry gas pathway testing for VOCs, PM2.5, and inorganic gases.
Indications for Use
Indicated for pediatric (ages 2+) and adult patients under the care of a licensed healthcare provider or physician for the administration of prescribed aerosolized medication in hospital, clinic, or home settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. Food & Drug Administration' in blue text.
May 2, 2018
Trudell Medical International Marianne Tanton Director, Quality and Regulatory Affairs 725 Third Street London, Ontario N5V 5G4 Canada
Re: K173825
Trade/Device Name: MC 300*R Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: March 29, 2018 Received: April 2, 2018
Dear Marianne Tanton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173825
Device Name MC 300*R Nebulizer
Indications for Use (Describe)
The nebulizer is intended to be used with pediatic (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician. The device is designed to aerosolize prescribed medication by a patient in the hospital, clinic or home care environment. The nebulizer is a single patient use device.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Prepared: 01-May-2018
### 1. Submitter
Trudell Medical International 725 Third Street London, Ontario N5V 5G4, Canada
Contact: Marianne Tanton Director, Quality and Requlatory Affairs Phone: 1-519-455-7060 Email: mtanton@trudellmed.com
### 2. Device Name
Trade Name: Common Name: Classification Name: Regulatory Class: Product Code:
MC 300*R Nebulizer Small Volume Nebulizer Nebulizer 21 CFR 868.5630 = CAF
### 3. Predicate Device
AeroEclipse* Durable (XL) Nebulizer - K080926 Trudell Medical International
The predicate device has not been subject to a recall.
### 4. Reference Device
MC 300* Nebulizer - K173367 Trudell Medical International
The predicate device has not been subject to a recall.
### 5. Device Description
The MC 300*R Nebulizer is a small volume jet nebulizer designed to deliver aerosolized medications for inhalation to the respiratory system. The device is intended to be used by pediatric (ages 2 years and above) and adult patients in hospital, clinic or home settings. The MC 300*R Nebulizer is a single patient use device and may be used for multiple treatments. This device is not used with a specific drug nor is it distributed with such drugs. The MC 300*R Nebulizer consists of four components: mouthpiece, nebulizer top, nozzle cover, and nebulizer bottom. It is marketed with oxygen tubing. The MC 300*R Nebulizer is not packaged with a mask, however the Disposable Aerosol Mask Assembly can be ordered per the reorder information on the IFU.
{4}------------------------------------------------
### 6. Principle of Operation
This device operates on the Venturi principle. Compressed air is driven through a converging nozzle, where it accelerates and emerges at a high velocity, creating a vacuum (Venturi effect). The vacuum draws a liquid residing in a reservoir up through a cylindrical channel and into the emerging airstream formed by the nozzle, to mix with air and impact upon a rigid surface. This process uses energy from the airstream to convert liquid into small droplets called aerosol. Upon reaching the user aerosol is suitably refined to enter the lungs effectively.
### 7. Indications for Use
The nebulizer is intended to be used with pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician. The device is designed to aerosolize prescribed medication for inhalation by a patient in the hospital, clinic or home care environment. The nebulizer is a single patient use device.
### 8. Comparison to predicate device
The MC 300*R Nebulizer and AeroEclipse* Durable (XL) Nebulizer (K080926), are identical in purpose, function, core technology and method of operation. Only minor differences exist between the MC 300*R Nebulizer and predicate, which do not affect the safety or effectiveness of the subject device. Table 1 provides a comparison of the subject and predicate devices.
{5}------------------------------------------------
#### Table 1: Comparison to Predicate Device
| Element of Comparison | MC 300*R Nebulizer<br>(Subject Device) | AeroEclipse* Durable (XL)<br>Nebulizer<br>(Predicate Device - K080926) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The MC 300*R Nebulizer is intended<br>to be used with pediatric (ages 2<br>years and above) and adult patients, who are under the care of a licensed<br>healthcare provider or physician. The device is designed to aerosolize<br>prescribed medication for inhalation<br>by a patient in the hospital, clinic or<br>home care environment. The<br>nebulizer is a single patient use device. | The AeroEclipse* Durable (XL)<br>nebulizer is a single patient, reusable<br>device, intended to be used by<br>patients who are under the care or<br>treatment of a licensed health care<br>provider or physician. The device is<br>intended to be used by these patients<br>to administer aerosolized medication<br>prescribed by a physician or health<br>care professional. The intended<br>environments for use include the<br>home, hospitals and clinics. |
| Principle of Operation | Pneumatic Jet Nebulizer | |
| Environment of use | Hospital, Clinic or Home | |
| Patient population | Adult and pediatric patients<br>(ages 2 years and above) | All |
| Single Patient Use | | Yes |
| Aerosolization | Continuous during inhalation and<br>exhalation | Built-in mode selector for breath<br>actuated or continuous mode |
| Type of device | | Reusable Device for single patient use, prescription only, non-sterile |
| Manufacturing process | Plastic molding | |
| Type of gas source | Compressed air or oxygen | |
| Flow rate | 4-8 LPM | 2.75 – 8 LPM |
| Maximum Fill Volume | | 6 ml |
### 9. Performance Data
### 9.1 Aerosol Characterization
Aerosol characterization testing for the subject (Mouthpiece and Mask) and predicate device was conducted in accordance with the relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH - 1993). Testing was performed at both low and high supplied air flow rates. The table below reflects data at 8 l/min supplied air flow rate to the nebulizer and 15 l/min flow rate through the cascade impactor. The table below also includes data from testing conducted with the medium mask, however, the nebulizer was also tested with small and large masks, which demonstrated similar performance to the medium mask.
{6}------------------------------------------------
#### Table 2: Performance Data
| Aerosol<br>Characteristics | Particle Characterization | |
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| | Continuous Nebulizer,<br>with Mouthpiece<br>(Subject Device)<br>(Job # 1668) | Continuous Nebulizer,<br>with Mask*<br>(Subject Device)<br>(Job # 1655) |
| Total Mass (µg) | 1494.4 ± 38.4 Albuterol†<br>602.5 ± 34.5 Ipratropium<br>Bromide‡‡<br>338.9 ± 29.9<br>Budesonide‡‡‡ | 1604.3 ± 28.1 Albuterol†<br>689.5 ± 16.3 Ipratropium<br>Bromide‡‡<br>357 ± 19.3 Budesonide‡‡‡ |
| Total Output Rate (µg/s) | 7.1 ± 0.1 Albuterol†<br>1.9 ± 0.2 Ipratropium<br>Bromide‡‡<br>0.9 ± 0.1 Budesonide‡‡‡ | 3.7 ± 0.2 Albuterol†<br>1.1 ± 0.1 Ipratropium<br>Bromide‡‡<br>0.5 ± 0.0 Budesonide‡‡‡ |
| Fine Particle Fraction<br>(0.98-5.39 µm<br>aerodynamic diameter)<br>(%) | 72.4 ± 0.4 Albuterol†<br>72.0 ± 0.4 Ipratropium<br>Bromide‡‡<br>64.6 ± 1.4 Budesonide‡‡‡ | 71.1 ± 0.9 Albuterol†<br>69.3 ± 2.8 Ipratropium<br>Bromide‡‡<br>65.9 ± 2.2 Budesonide‡‡‡ |
| Fine Particle Mass (µg) | 1081.2 ± 30.2 Albuterol†<br>433.9 ± 23.3 Ipratropium<br>Bromide‡‡<br>218.8 ± 17.0<br>Budesonide‡‡‡ | 1150.1 ± 19.7 Albuterol†<br>478.1 ± 22.3 Ipratropium<br>Bromide‡‡<br>235.0 ± 6.6 Budesonide‡‡‡ |
| Fine Particle Output<br>Rate (µg/s) | 5.1 ± 0.1 Albuterol†<br>1.4 ± 0.1 Ipratropium<br>Bromide‡‡<br>0.6 ± 0.1 Budesonide‡‡‡ | 2.7 ± 0.1 Albuterol†<br>0.8 ± 0.1 Ipratropium<br>Bromide‡‡<br>0.3 ± 0.0 Budesonide‡‡‡ |
| Particle Size (MMAD) | 2.6 µg Albuterol†<br>2.7 µg Ipratropium<br>Bromide‡‡<br>4.2 µg Budesonide‡‡‡ | 2.4 µg Albuterol†<br>2.5 µg Ipratropium<br>Bromide‡‡<br>4.1 µg Budesonide‡‡‡ |
| GSD | 2.1 Albuterol†<br>2.1 Ipratropium Bromide‡‡<br>1.9 Budesonide‡‡‡ | 2.2 Albuterol†<br>2.3 Ipratropium Bromide‡‡<br>1.9 Budesonide‡‡‡ |
† Albuterol Sulfate Inhalation Solution, 833µg/ml
†† Ipratropium Bromide Inhalation Solution 250µg/ml
††† Budesonide Suspension for Inhalation 0.25mg/ml
* Disposable Aerosol Mask Assembly – Medium Mask
- Disposable Aerosol Mask Assembly – Medium Mask
{7}------------------------------------------------
### 9.2 Biocompatibility Testing
Biological endpoints applicable to an externally communicating device, tissue contact by way of gas pathway with permanent duration (> 30 days) are listed below. All in vitro and in vivo studies were performed by an independent source and included the following battery of tests: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Genotoxicity and Extractables/Leachables with a Biological Risk Assessment.
#### Summary of Biocompatibility Testing Conducted
| ISO Standard | Biological Endpoint |
|--------------|---------------------------------------------------------------------------------------|
| 10993-5 | Tests for In Vitro Cytotoxicity |
| 10993-10 | Tests for Irritation and Skin Sensitization |
| 10993-11 | Tests for systemic toxicity (Acute Toxicity) |
| 10993-3 | Tests for genotoxicity (Bacterial Reverse Mutation Study and Mouse<br>Lymphoma Assay) |
| 10993-12 | Sample preparation and reference materials |
| 10993-17 | Establishment of allowable limits for leachable substances |
| 10993-18 | Chemical characterization of materials |
### 9.3 Dry Gas Pathway Testing
Testing pertaining to the dry gas pathway and associated risk assessments/conclusions were conducted by an independent source. Testing included the following assessments:
- · Emissions of volatile organic compounds (VOCs)
- · Fine particles (particulate matter PM2.5)
- · Inorganic gases (ozone, CO2, and CO)
### 10. Clinical Performance Summary
Not applicable, the determination of substantial equivalence is not based on Clinical Performance data.
### 11. Conclusion
The non-clinical data demonstrates that the MC 300*R Nebulizer is as safe and as effective as the predicate and therefore substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.