PharmaCaribe inhaled saline solutions are used in conjunction with a nebulizer. The contents of these vials are for the induction of sputum production where sputum production is indicated. Concentrations of 0.9%, 3%, 3.5%, 6%, 7%, and 10% with a pH of 7.4
Device Story
PulmoSal is a sterile, preservative-free, aqueous sodium chloride solution (0.9%, 3%, 3.5%, 6%, 7%, 10%) with a pH of 7.4. Packaged in 4 ml LDPE blow-fill-sealed vials. Used in hospital, sub-acute care, or home settings by patients requiring sputum induction. Device is administered via a nebulizer; saline solution acts as an irritant to induce sputum production. Healthcare providers use the induced sputum for diagnostic purposes. Benefits include facilitating sample collection for clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included USP monograph compliance, salinity, pH (USP <791>), identification of sodium (USP <191>), sodium chloride assay, and sterility (USP <71>).
Technological Characteristics
Aqueous sodium chloride solution in concentrations of 0.9%, 3%, 3.5%, 6%, 7%, and 10%. pH 7.4. Materials: Blow-fill-sealed low density polyethylene (LDPE) vials. Preservative-free. Sterile. No software or electronic components.
Indications for Use
Indicated for patients requiring induction of sputum production. Used in conjunction with a nebulizer.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
### 510(k) Summary Page 1 of 4 21-Feb-13
# MAR 1 3 2013
| PharmaCaribe<br>3513 Di Leuca St.<br>Punta Gorda, FL 33950 | Tel – (941) 505-0793<br>Fax – (941) 505-0718 |
|------------------------------------------------------------|--------------------------------------------------------------------------------|
| Official Contact: | W. Randolph Warner, Managing Partner |
| Trade Name: | PulmoSalTM inhaled saline solution |
| Common/Usual Name: | Saline solution |
| Classification Name: | CAF – Nebulizer (Direct Patient Interface)<br>CFR – 868.5630, Class 2 |
| Device: | Inhaled saline solutions – 0.9%, 3%, 3.5%, 6%, 7%, and 10%<br>with a pH of 7.4 |
| Predicate Device: | PharmaCaribe – Sodium Inhalation solutions - K101424 |
#### Device Description:
The proposed inhaled saline solutions are in 0.9%, 3%, 3.5%, 6%, 7%, and 10% concentrations with a pH of 7.4. They are packaged sterile in 4 ml vials for use as indicated.
#### Indications for Use:
PharmaCaribe inhaled saline solutions are used in conjunction with a nebulizer. The contents of these vials are for the induction of sputum production when used in conjunction with a nebulizer.
Concentrations of 0.9%, 3%, 3.5%, 6%, 7%, and 10% with a pH of 7.4
#### Patient Population:
Any patient requiring induction of sputum production.
#### Environment of Use:
Hospital, sub-acute care or home
## Contraindications:
None
{1}------------------------------------------------
#### 510(k) Summary Page 2 of 4 21-Feb-13
#### Predicate Device Comparison:
The PulmoSal™ inhaled saline solution is viewed as substantially equivalent to the predicate devices because:
#### Indications -
- Intended to be in conjunction with a nebulizer for sputum production is identical to the . predicate PharmaCaribe Inhaled Saline Solutions (K101424)
Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K 101424) except the addition of 0.9% saline and a pH of 7.4.
#### Technology of Manufacturing-
A homogeneous aqueous mixtures of sterile water and saline . Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K 101424)
#### Mode of Action -
- Used with a nebulizer to increase sputum production .
Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K101424)
#### Formulation -
- Based upon USP specification except pH of 7.4 which matches the pH of the lungs .
- 3%, 3.5%, 6%, 7% and 10% solutions identical to the predicate PharmaCaribe Inhaled . Saline Solutions (K 101424)
- 0.9% solution is identical to the predicate Nephron Pharma Sodium Chloride Inhalation . solution (K113033)
Discussion - the concentrations are identical to the predicates PharmaCaribe Inhaled Saline Solutions (K101424) and Nephron Pharma – Sodium Chloride Inhalation solution (K11303). The pH of 7.4 is identical pH Pulmicort Respules" (budesonide) AstraZeneca, which has a 7.4 pH, was approved under (NDA # N020929 an inhaled drug.
#### Materials -
- Blow-fill-sealed low density polyethylene (LDPE) vial containing sterile, preservative-. free, clear, colorless, aqueous solution of sodium chloride in various concentrations by weight / volume. The formulation contains no additives.
Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K 101424)
#### Environment of Use -
- Hospital, home, sub-acute care settings .
Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K 101424)
#### Patient Population -
Any patient requiring induction of sputum production. ● Discussion - identical to the predicate PharmaCaribe Inhaled Saline Solutions (K 101424)
{2}------------------------------------------------
## 510(k) Summary Page 3 of 4 21-Feb-13
.
| Features | Proposed<br><i>PulmoSal</i> ™ Inhaled saline<br>solution | Predicate<br>PharmaCaribe K101424 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | PharmaCaribe inhaled saline solutions<br>are used in conjunction with a<br>nebulizer.<br><br>The contents of these vials are for the<br>induction of sputum production. | PharmaCaribe inhaled saline<br>solutions are used in conjunction<br>with a nebulizer.<br><br>The contents of these vials are for<br>the induction of sputum production<br>when used in conjunction with a<br>nebulizer. |
| Prescription | Yes | Yes |
| Environment of Use | Home, Hospital, Sub-acute<br>Institutions | Home, Hospital, Sub-acute<br>Institutions |
| Patient Population | Any patient requiring induction of<br>sputum production. | Any patient requiring induction of<br>sputum production. |
| Contraindications | None | None |
| Materials in fluid contact | Vial LDPE | Vial LDPE |
| Sterile water | Yes | Yes |
| Solutions | 3%, 3.5%, 6%, 7% and 10%<br>0.9% | 3%, 3.5%, 6%, 7% and 10%<br>0.9% - K113033 - Nephron Pharma<br>- USP pH |
| pH | 7.4 | USP 4.5 to 7.0 |
| Manufactured | Per USP monograph except pH | Per USP monograph |
| Supplied | Sterile | Sterile |
| Used with a nebulizer | Yes | Yes |
## Non-clinical Testing Summary -
Asept Pak will perform Lot Release testing. The concentrations meet the USP Specifications with exception having a pH of 7.4, which is a normal physiological value. Additional performance testing includes:
In Process testing includes:
- Incoming resin certification and testing .
- Current USP water testing .
- pH as per USP <791> .
- Appearance .
- Salinity .
- Environmental monitoring per Asept SOPs .
- Container Fill weight monitoring ●
- Visual 100% inspection .
{3}------------------------------------------------
## 510(k) Summary Page 4 of 4 21-Feb-13
Finished Product Tests
- . Salinity
- . Identification of Sodium per USP (191)
- . Sodium Chloride Assay per USP Monograph
- Sterility as per USP <71> .
The above demonstrates that the proposed devices are substantially equivalent to the predicate devices.
## Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2013
PharmaCaribe C/O Mr. Paul Dryden President ProMedic. Incorporated 24301 Woodsage Drive BONITA SPRINGS FL 34134
Re: K130091
Trade/Device Name: PulmoSalTM Inhaled Saline Solution Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: February 21, 2013 Received: February 22, 2013
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
| 510(k) Number: | K130091 (To be assigned) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | PulmoSal <sup>TM</sup> Inhaled saline solution |
| Indications for Use: | PharmaCaribe inhaled saline solutions are used in conjunction with a nebulizer. The contents of these vials are for the induction of sputum production where sputum production is indicated. |
| | Concentrations of 0.9%, 3%, 3.5%, 6%, 7%, and 10% with a pH of 7.4 |
| Prescription Use XX<br>(Part 21 CFR 801 Subpart D) | or | Over-the-counter use __<br>(21 CFR 807 Subpart C) |
|----------------------------------------------------|----|---------------------------------------------------|
|----------------------------------------------------|----|---------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Albert E
| c=US, o=U.S. Government, |
|--------------------------------|
| ou=HHS; ou=FDA, ou=People, |
| 0.9.2342.19200300.100.1.1=130 |
| 0059331; cn=Albert E. Moyal -S |
| 2013.03.07 16:55:28 -05'00' |
Moyal S L for LS
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
**510(k) Number:** k130091
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.