The Clever TD-7001 Nebulizer is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize physician-prescribed solution for inhalation by the patient, except for Pentamidine. The Clever TD-7001 Nebulizer is for use with children and adult patients in the homecare environment.
Device Story
Clever TD-7001 is an ultrasonic vibrating mesh nebulizer. Device consists of a nebulizer unit, medication cap with metal alloy mesh, and medication cup. Powered by 2 AA alkaline batteries or electrical outlet. Operation: piezoelectric element vibrates, causing mesh to vibrate; medication solution in reservoir is aerosolized into particles for inhalation via mouthpiece or mask. Used in homecare environment by patients (adults/children). Healthcare providers prescribe the solution; patients self-administer. Device provides aerosolized medication for respiratory treatment.
Clinical Evidence
Bench testing only. Aerosol characterization via cascade impactor compared particle size distribution using Ipratropium bromide, Ventolin, and Cromolyn sodium; results showed consistency and equivalent performance to predicate (p>0.035). Gas path analysis confirmed no toxic emissions (CO, CO2, ozone, VOCs) and particulate matter conformed to EPA PM2.5 standards. Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 10993-6 (implantation), and ISO 10993-3 (genotoxicity). Electrical safety and EMC testing performed per EN 60601-1 and EN 60601-1-2.
Technological Characteristics
Ultrasonic vibrating mesh nebulizer. Components: nebulizer unit, medication cap with metal alloy mesh, medication cup. Power: 2 AA alkaline batteries or AC power. Biocompatibility: gas path materials tested per ISO 10993. Safety standards: EN 60601-1, EN 60601-1-2, CISPR 11.
Indications for Use
Indicated for children and adult patients in homecare settings requiring aerosolized physician-prescribed inhalation solutions, excluding Pentamidine.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Omron Micro Air Vibrating mesh nebulizer NE-U22 (K062263)
Submission Summary (Full Text)
{1}------------------------------------------------
### Re:K122060
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The Assigned 510(k) number is:
- 1. Submitter's Identification: TaiDoc Technology Corporation 3F,5F, No.127, Wugong 2nd Rd. City, 24888, Taiwan Correspondence: Pinjung Chen Regulatory Affairs S Tel: +886-2-6625-8 Fax: +886-2-6625-02 Email: pinjung.chen@taidoo Prepared date: Junes
- 2. Device name: Proprietary name: Clever TD-7001 Nebulizer
- Regulatory information:
- Regulation section: 21 CFR § 868.5630 A.
- B. Classification: Class II
- CAF, nebulizer (direct patient interface) C. Product Code:
- Panel: Anesthesiology D.
- 3. Indication for Use:
The Clever TD-7001·Nebulizer is an ultrasonic (vibrating mesh) nebulizer system
4- page lof 4
{2}------------------------------------------------
designed to aerosolize physician-prescribed solution for inhalation by the patient, except for Pentamidine.
The Clever TD-7001 Nebulizer is for use with children and adult patients in the homecare environment.
### 4. Device Description:
Clever TD-7001 Nebulizer is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize physician-prescribed solution for inhalation by the patient. It is for use with children and adult patients in the homecare environment.
Clever TD-7001 Nebulizer consists of a Nebulizer unit, a medication cap and a Jand Medication cup are single medication cup. The Nebulizer unit, me patient re-usable. The Nebulizer unit trol circuitry and is powered by using 2 AA alkaline batteries organ A onnects to electrical outlets. The patient interface comprises a mouthpiê medication cap contains the metal alloy mesh. When the device is turne ultrasonic piezoelectric element vibrate to cause the mesh vibrate. ation solution tin the reservoir of medication cup is aerosolized by the vibrating the aerosols can be inhaled through the mouthpiece or mask.
5. Substantial Equivalence Information:
- Predicate device name: Omron Micro Air Vibrating mesh nebulizer NE-U22 A.
- Predicate K number: K062263 B.
- C. Comparison with predicate:
The proposed Clever TD-7001 Nebulizer has the following similarities to the predicate device which make substantially equivalence:
- . Same intended use
- . Same patient population
- Same operating principle and working mechanism
- Same performance efficacy
- Same fundamental scientific technology
The differences encompass:
4- page 2of 4
{3}------------------------------------------------
- Physical appearance change 1
- Mesh cap and medication cup/bottle assemble method 1
- . The connection of electrode from the main unit to the transducer/vibrator
- D The material in fluid pathway
- Power requirement and power consumption
- 6. Performance Characteristics:
### 1) Aerosol Characterization Testing
The particle size distribution test via Cascade Impactor of Clever TD-7001 Nebulizer was performed in comparison to the predicate device, K062263 with three drugs (Ipratropium bromide, Ventolin, and Cromolyn sodium). The test has shown Clever TD-7001 Nebulizer consistent in repeatabil for each three classes of drug, and demonstrated equivalent performance? ate device K062263 that no significant difference (p>0.035) in the
Clever TD-7001 Nebulizerst as the ame performance characteristics as the predicate device and meets its product specification as well.
### 2) Airpath Testing
of Clever TD-7001 Nebulizer has performed and shown the Gas sample analysis device does not emit potential toxic gases that may cause harmful influences to human, including carbon monoxide, carbon dioxide, ozone, or volatile organic compounds (VOCs). The output of particulate matter conformed to EPA requirements of the PM2.5 standard.
#### 3) Materials
FDA authority considers the gas path contact components the external communicating components with tissue contact. Biocompatibility evaluations of cytotoxicity (ISO 10993-5), irritation (ISO 10993-10), sensitization (ISO 10993-10), implantation (ISO 10993-6) and genotoxicity (ISO 10993-3) tests have been conducted for the gas path contact materials.
4- page 3of 4
{4}------------------------------------------------
## 4) Safety and EMC
The electromagnetic compatibility and electric safety of the proposed device are tested to meet the following standards:
- EN 60601-1:1988+ A1:1991+ A2:1995 트
- EN 60601-1-1-2:2001+A1:2006, CISPR 11: 2003 ■
- 7. Conclusion:
Based on the information provided in this submission, Clever TD-7001 Nebulizer is substantially equivalent to the predicate device in safety and effectiveness.
Property of TaiD,
Confidential material
Duplication and copying prohibited
Without consent of TaiDoc
4- page 4of 4
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2013
Mr. Pinjung Chen Regulatory Affairs Specialist TaiDoc Technology Corporation 3F,5F, No.127, Wugong 2nd Road Wugu District, New Taipei City Taiwan 24888
Re: K122060
Trade/Device Name: Clever TD-7001 Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: January 31, 2013 Received: February 8, 2013
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Chen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Re:K122060
# Attachment 3. Indications for Use Statement
# Indications for Use
510(k) Number: K122060
Device Name: Clever TD-7001 Nebulizer
Indications for Use:
The Clever TD-7001 Nebulizer is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize physician-prescribed solution for inhalation by the patient, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... " except for Pentamidine. .:
ensepy. I chamburn.
The Clever TD-7001 Nebulizer is for use with children and adult patients in the : homecare environment.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR Part 801 Subpart D)
Over the Counter Use And/Or (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Albert E. Moyal=<br>c=US, o=U.S. Govemment, ou=HHS.<br>100.1 1=13000593331. | for LS |
|---------------------|-----------------------------------------------------------------------------------|--------|
| (Division Sign-Off) | Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | |
| | 1.100 mln | |
3-page lof 1
510(k) Number:_K122060
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.