Nitronox Plus 0-70, Nitronox 0-50, Nitronox Plus 50/50
Applicant
Parker Hannifin
Product Code
BZR · Anesthesiology
Decision Date
Apr 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5330
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The Nitronox Plus is intended to provide a mixture of N20/02, on demand, to a conscious, spontaneously breathing patient. The Nitronox Plus is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates.
Device Story
Nitronox Plus is a portable, pneumatic, on-demand analgesia system; delivers nitrous oxide/oxygen mixture to conscious, spontaneously breathing patients. Used in healthcare settings by professionals. Inputs: compressed oxygen and nitrous oxide via flexible hoses. Operation: oxygen-piloted dual diaphragm regulator equalizes inlet pressures; demand valve triggers gas delivery upon patient inhalation through face mask or mouthpiece. Adjustable models allow healthcare providers to set N2O concentration (0-70% or 0-50%); fixed model provides 50/50% mix. Output: low-pressure mixed gas. Safety features: pneumatic whistle alarms for low inlet pressure; emergency air intake valve for oxygen loss; key lock to prevent unauthorized use. Benefits: provides controlled analgesia for procedures; enhances safety via adjustable ratios and fail-safe mechanisms.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional, operational, and performance verification, biocompatibility testing, reprocessing validation, and usability assessment.
Technological Characteristics
Pneumatic system; force-balanced diaphragms, springs, and poppet valves; Bourdon tube gauges; DISS gas fittings. Gas mixing via tapered valve stems and orifices. No electronic components or software; purely mechanical/pneumatic. Sterilization/biocompatibility addressed via standard materials.
Indications for Use
Indicated for conscious, spontaneously breathing adult and pediatric patients requiring N2O/O2 analgesia. Not for use with infants or neonates.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
Predicate Devices
Nitronox Pre-Set 50/50 Nitrous Oxide/Oxygen Gas Mixer (K760766)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 22, 2021
Parker Hannifin % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K202480
Trade/Device Name: Nitronox Plus Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing gas mixer Regulatory Class: Class II Product Code: BZR Dated: August 26, 2020 Received: August 28, 2020
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202480
Device Name Nitronox Plus
Indications for Use (Describe)
The Nitronox Plus is intended to provide a mixture of N20/02, on demand, to a conscious, spontaneously breathing patient. The Nitronox Plus is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### న్. 510(k) Summary
# Parker Hannifin Nitronox Plus (per 21CFR 807.92)
#### 5.1. Submitter/510(k) Holder
| Parker Hannifin Corporation<br>Precision Fluidics Division<br>245 Township Line Road<br>Hatfield, Pennsylvania 19440<br>Phone: 215-723-4000 | Contact: Andrew Ellinger,<br>Product Development Engineer<br>Contact Phone: 215-660-8175<br>Contact Email: andrew.ellinger@parker.com<br>Date Prepared: May 1, 2020 |
|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### 5.2. Device Name
| Proprietary Name: | Nitronox Plus |
|----------------------------|-----------------------------------------------|
| Common Name: | Breathing gas mixer |
| Classification Name: | Mixer, breathing gases, anesthesia inhalation |
| Classification Regulation: | 21 CFR 868.5330 |
| Classification Panel: | Anesthesiology |
| Product Code: | BZR |
| Device Class: | Class II |
#### 5.3. Predicate Device
Nitronox Pre-Set 50/50 Nitrous Oxide/Oxygen Gas Mixer (K760766)
#### 5.4. Device Description
The Nitronox Plus is a portable "on demand" analgesia system used to deliver a mixture of nitrous oxide and oxygen gases to a conscious, spontaneously breathing patient. The device is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates. It is only for use in a healthcare setting by healthcare professionals. The Nitronox Plus is available in three models which differ only in adjustability and maximum ratio of nitrous oxide that can be delivered. The Nitronox Plus features a 0-70% N2O adjustable model, 0-50% N2O adjustable model, and 50% N2O/50% O2 fixed model.
The Nitronox Plus is a pneumatic system. Oxygen and nitrous oxide are supplied to the device through flexible hoses. Inlet gas is regulated and equalized through the use of an oxygen piloted dual diaphragm regulator. The inlet pressure of both gases and the mixed gas pressure are displayed on the front panel of the device through the use of pressure gauges. The device is designed to function with a medical breathing circuit and face mask or mouthpiece accessory.
{4}------------------------------------------------
Delivery of mixed gas starts with a patient's demand; upon inhalation through a face mask or mouthpiece, a demand valve opens and supplies mixed gas to the patient through the connected Medical Breathing Circuit. In adjustable models, the healthcare professional is able to set the desired concentration through the use of a control knob on the front panel. Mixed gas is supplied through the breathing circuit until inhalation ceases.
The Nitronox Plus contains several safety features. Pneumatically powered alarms for both gas inlets alert the user when either gas supply begins to deplete. The device will also maintain delivery of the set concentration with depleting oxygen pressure. Once the oxygen pressure drops below a certain threshold, the device will allow ambient air to be delivered through an Emergency Air Intake valve. The device also features a key lock to control the nitrous oxide gas supply, allowing healthcare professionals to prevent unauthorized use of the device.
#### 5.5. Indications for Use
The Nitronox Plus is intended to provide a mixture of N20/O2, on demand, to a conscious, spontaneously breathing patient. The Nitronox Plus is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates.
The Indications for Use statement for the Nitronox Plus is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device. Modifications made to enhance gas delivery functionality do not significantly change the key functional technology and operational characteristics of the device, nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices are intended to deliver mixtures of N20/02, on demand, to a conscious, spontaneously breathing patient.
#### Summary of Technological Characteristics Compared to the Predicate Device 5.6.
The Nitronox Plus (subject device) is a redesigned version of the Nitronox HD originally cleared by FDA under submission K760766 (predicate device). The key technological characteristics of the Nitronox Plus are based on well-established technology that uses springs, diaphragms, valves, orifices, and other pneumatic mechanisms to regulate the pressure and flow of gases. The subject device is similar to the predicate device regarding:
- Force-balancing system consisting of diaphragms, springs, and poppet valves ●
- Gas mixing through tapered valve stems and orifices ●
- Pneumatic "whistle" alarms
- Bourdon tube gauges ●
- Demand valve to only deliver gas upon patient inhalation
{5}------------------------------------------------
The major technological differences between the subject device and predicate device include:
- Adjustable mixture ratio .
- . Low inlet pressure alarms
- Emergency air intake valve .
- . Key lock feature
A summary comparison of the subject and predicate devices is provided below:
| Characteristic | Subject:<br>Nitronox Plus | Predicate:<br>Nitronox (K760766) | Comparison |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Intended Use | Provide a mixture of N2O/O2, on<br>demand, to a conscious,<br>spontaneously breathing patient. The<br>Nitronox Plus is designed for use<br>with adult and pediatric patients and<br>is not intended to be used with<br>infants or neonates. | Provide 50/50% N2O/O2 mixture, on<br>demand, to a conscious,<br>spontaneously breathing patient. | Similar |
| Target<br>Population | Adult and pediatric patients. Not<br>intended to be used with infants or<br>neonates | Spontaneous breathing patients<br>requiring pain relief. | Similar |
| Prescription or<br>Over-the-<br>Counter Use | Prescription Use | Prescription Use | Same |
| Anatomical Site | Gas is to be inhaled through the<br>mouth and nose region. | Gas is to be inhaled through the<br>mouth and nose region. | Same |
| Where used | To be used by healthcare<br>professionals in a healthcare setting. | Dentistry, Emergency Departments,<br>Ambulances, Obstetrics,<br>Physiotherapy. Nursing. Specialist<br>Procedures, and Podiatry. | Similar |
| Operating<br>principle | Pneumatic, demand flow system. | Pneumatic, demand flow system. | Same |
| Energy used | Compressed oxygen/nitrous oxide<br>gas. | Compressed oxygen/nitrous oxide<br>gas. | Same |
| Energy<br>delivered | Low pressure mixed gas. | Low pressure mixed gas. | Same |
| Gas input<br>pressure | 45-60 psi. | 45-60 psi. | Same |
| Gas regulation<br>mechanism | Oxygen Piloted Regulator with dual<br>diaphragms. | Oxygen Piloted Regulator with dual<br>diaphragms. | Same |
| Gas delivery | Proportional to patient's demand. | Proportional to patient's demand. | Same |
| Gas mixing | Oxygen and nitrous oxide are<br>equalized and mixed at variable<br>ratios. | Oxygen and nitrous oxide are<br>equalized and mixed at a fixed 50%<br>ratio. | Different |
| Alarms | Audible alarm sounds when:<br>- N2O input is below 35 psi;<br>- O2 input is below 35 psi. | Audible alarm is triggered if oxygen<br>mixture pressure rises above 40 psi. | Different |
| Security feature | Key Lock allows healthcare<br>provider to turn device ON/OFF; in<br>the "OFF" position device will not<br>deliver nitrous oxide gas; with the<br>key turned in the "ON" position,<br>mixed gas can be supplied. | None. | New Feature |
| Characteristic | Subject:<br>Nitronox Plus | Predicate:<br>Nitronox (K760766) | Comparison |
| Accessories | - Medical Breathing Circuit;<br>- Face mask;<br>- Mouth Piece;<br>- Oxygen & nitrous oxide supply<br>hose;<br>- Oxygen & nitrous oxide;<br>- E-stand mobile cart;<br>- Two Cylinder Cart;<br>- Wall arm mount; | - Medical Breathing Circuit;<br>- Face mask;<br>- Mouth Piece;<br>- Oxygen & nitrous oxide supply<br>hose;<br>- Oxygen & nitrous oxide;<br>- E-stand mobile cart;<br>- Two Cylinder Cart;<br>- Wall arm mount; | Same |
| Pressure<br>regulation | Force-balanced system of<br>diaphragms/poppets/springs | Force-balanced system of<br>diaphragms/poppets/springs | Same |
| Gas mixing<br>mechanism | Adjustable dual stem/orifice | Factory set dual stem/orifice | Different |
| Demand<br>mechanism | Demand Valve | Demand Valve | Same |
| Gauge type | Bourdon Tube | Bourdon Tube | Same |
| Alarm type | Pneumatic whistle | Pneumatic whistle | Same |
| Safety Features | O2 Piloted System: Shut off mixer<br>output if O2 pressure drops. | O2 Piloted System: Shut off mixer<br>output if O2 pressure drops. | Same |
| | DISS fittings to prevent<br>misconnection of gas supply. | DISS fittings to prevent<br>misconnection of gas supply. | Same |
| | Activated only via patient<br>inspiratory effort (self-<br>administration principle). | Activated only via patient<br>inspiratory effort (self-<br>administration principle). | Same |
| | In the event of N2O supply failure,<br>100% O2 will be supplied. | In the event of N2O supply failure,<br>100% O2 will be supplied. | Same |
| | Dual diaphragms prevent reverse<br>gas flow between supplies. | Dual diaphragms prevent reverse<br>gas flow between supplies. | Same |
| | Audible alarm for low inlet pressure. | Audible alarm for increased mix<br>pressure. | Different |
| | Emergency Air Intake Valve, if O2<br>pressure is lost patient can breathe<br>ambient air. | None. | New Feature |
| | Key lock system prevents<br>unauthorized use of device. | None. | New Feature |
Side-by-Side Comparison of the Subject and Predicate Devices
{6}------------------------------------------------
### Summary of Non-clinical Performance Testing as Basis for Substantial 5.7. Equivalence
The results of performance testing demonstrate that the Nitronox Plus meets all design input requirements, including functional, operational, and performance requirements. Performance testing was conducted to verify performance characteristics, safety features, capabilities compared to the predicate device, and biocompatibility; validate reprocessing activities; and ensure usability of the device to meet patient needs when used as intended. The results of the performance testing support the conclusion that the subject device is substantially equivalent to the predicate device.
Risk management processes were applied throughout product development. Results of the comprehensive risk analysis conclude that the benefits associated with the design modifications for new and enhanced safety systems exceed any residual risk. The remaining risks are acceptable when weighed against the benefits provided to patients when using the device. None of the residual risks outweigh the benefits of using the Nitronox Plus system for its intended use.
{7}------------------------------------------------
#### 5.8. Conclusions
The indications for use, principles of operation, and technological characteristics of the subject device are similar to the predicate device. Differences between the subject and predicate devices are limited to minor differences in the technology used to adjust the mixture ratio, triggering conditions of audible alarms, and addition of the emergency air intake and key lock safety features. Performance testing supports the conclusion that the subject device is as safe and effective as the predicate device, and therefore the two devices are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.