K173280 · Teleflexmedical, Inc. · BZE · Aug 2, 2018 · Anesthesiology
Device Facts
Record ID
K173280
Device Name
Neonatal ConchaSmart Breathing Circuits
Applicant
Teleflexmedical, Inc.
Product Code
BZE · Anesthesiology
Decision Date
Aug 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and include heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.
Device Story
Neonatal ConchaSmart Breathing Circuits serve as conduits for respiratory gas between neonatal/infant patients and ventilators or Infant Flow Generators; includes heated wires for use with Hudson RCI Neptune Heated Humidifier. Device consists of 10mm corrugated tubing housing heated wires, kitted with adaptors/connectors. Used in professional healthcare environments by respiratory clinicians. Heated wires maintain set patient temperature and minimize condensation in tubing. Output is delivered humidified gas to patient via nasal prongs (single limb) or endotracheal tube (dual limb). Benefits include reduced airway irritation and prevention of secretion desiccation. Disposable device with 21-day useful life.
Clinical Evidence
Bench testing only. Performance verified against ISO 5367:2014, ISO 8185:2007, ISO 5356-1:2015, and IEC 60601-1. Biocompatibility confirmed per ISO 10993-1, -5, -10, -17, and -18. Shelf life validated via ASTM F1980. Useful life of 21 days validated under worst-case scenarios.
Technological Characteristics
Corrugated tubing (10mm diameter); Nickel Chromium Iron or Nickel Chromium heated wires; max power 21-30W; max 2.5W/linear foot. Connectivity: compatible with Hudson RCI Neptune Heated Humidifier. Standards: ISO 5356-1:2015, ISO 5367:2014, ISO 8185:2007, IEC 60601-1. Non-sterile, disposable.
Indications for Use
Indicated for neonatal and infant patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases in professional healthcare environments.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
Predicate Devices
RT265 and RT266 Dual Heated Infant Breathing Circuits (K103767)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
August 2, 2018
Teleflex Medical, Inc Brian Gall Senior Regulatory Affairs Specialist, Respiratory Division 3015 Carrington Mill Blvd Morrisville, North Carolina 27560
Re: K173280
Trade/Device Name: Neonatal ConchaSmart Breathing Circuits Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: June 29, 2018 Received: July 2, 2018
Dear Brian Gall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173280
#### Device Name
Neonatal ConchaSmart Breathing Circuit
#### Indications for Use (Describe)
The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and include heated wire(s) for use with the Hudson RCI Neptune Heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
#### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 Phone: 919-228-4350
#### B. Contact Person
Brian Gall Regulatory Affairs Group Lead, Respiratory Division
#### C. Date Prepared
31 July 2018
#### D. Device Name
| Trade Name: | Neonatal ConchaSmart Breathing Circuits |
|-----------------------|-----------------------------------------|
| Common Name: | Breathing system heater |
| Product Code: | BZE |
| Regulation Number: | 868.5270 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
### E. Predicate Device
The Neonatal ConchaSmart Breathing Circuits are substantially equivalent to the following devices:
- K103767 RT265 and RT266 Dual Heated Infant Breathing Circuits .
In addition, the following devices are used as reference devices as described:
K140556 has been added as a reference 510(k). The primary reason for referencing K140556 is that this is the adult version of the proposed device. The adult breathing circuits were added as a reference for the following primary reasons:
- . Different technological characteristics (single vs. dual limb) do not raise different questions of safety and effectiveness from the predicate device since the adult circuit is available in both configurations.
- . The materials used in the proposed neonatal breathing circuit and the reference are similar and well understood.
- . The useful life testing strategy of the reference device was accepted. It is an equivalent test method to the proposed device.
- . The heated wires used in the reference device are similar to the proposed device. The intent of using the adult circuits as a reference was to show that Teleflex has market knowledge of the interaction of this type of heated wire being used with similar materials with the Hudson RCI Neptune Heated Humidifier system.
{4}------------------------------------------------
K151959 has been added as a reference 510(k). The primary reason for referencing K151959 the following:
- . It is a single limb breathing circuit cleared to be used with a heated humidifier for neonatal and infant patient populations.
- . The testing related to the single limb configuration was similar (biocompatibility to ISO 10993; circuit construction and performance to ISO 5356-1, ISO 5367, and ISO 8185; and electrical testing to IEC 60601-1).
## F. Device Description
The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and includes heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.
These devices are made of corrugated tubing (10mm in diameter), which houses the heated wires and are kitted with various adaptors and connectors to aid the respiratory care clinician in system configuration. In general these heated wire circuits are connected to both a ventilator and a Hudson RCI Neptune Heated Humidifier. The single limb configuration is compatible with the CareFusion Infant Flow system.
All Neonatal ConchaSmart Breathing Circuits have a useful life of 21 days. Neonatal ConchaSmart Breathing Circuits are sold non-sterile.
### G. Indications for Use
The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and include heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.
### H. Technological Characteristics - Comparison to the Predicate
The proposed Heated Wire Breathing Circuits are substantially equivalent to the predicate device as the intended use, fundamental scientific technology, and operating principles are the same. The circuits differ with regards to limb configurations, tubing dimensions, and materials, however through testing and the introduction of reference predicates, the proposed devices are substantially equivalent to the predicate as indicated below:
{5}------------------------------------------------
| Comparative<br>Characteristics | RT265 and RT266 Dual Heated<br>Infant Breathing Circuits<br>Predicate (K103767) | Neonatal ConchaSmart Breathing<br>Circuits<br>Proposed | |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Indications for<br>Use | The dual-heated breathing circuits<br>are intended as conduits of<br>breathing gas for ventilation of infant<br>patients, and to maintain the<br>temperature of humidified inspired<br>gas. The RT265 is used for flow<br>rates greater than 4LPM, and the<br>RT266 is for flow rates between<br>0.3LPM and 4LPM. | The Neonatal ConchaSmart<br>Breathing Circuits are intended for<br>neonatal and infant patients in<br>professional healthcare<br>environments as a conduit for<br>respiratory gas between a patient<br>and a ventilator or Infant Flow<br>Generator (for single limb<br>configuration) and include heated<br>wire(s) for use with the Hudson RCI<br>Neptune Heated Humidifier. The<br>heated wires are intended to aid in<br>maintaining the set patient<br>temperature and minimize<br>condensation in the ventilator<br>tubing. | |
| Intended Use | The RT265 and RT266 infant<br>breathing circuits are intended to<br>deliver humidified breathing gases<br>for administration to an infant<br>patient. Gases available for medical<br>use do not contain sufficient<br>moisture and may damage or irritate<br>the respiratory tract, or desiccate<br>secretions of patients whose<br>supraglottic airways have been<br>bypassed. Thus humidified gases<br>via heated breathing circuit may be<br>indicated for patients requiring<br>mechanical ventilation, positive<br>pressure breathing assistance, or<br>general medical gases. These<br>gases may be delivered by<br>facemask or through bypassing<br>upper airways, for example use of<br>an endotracheal tube. | The proposed Neonatal<br>ConchaSmart breathing circuits are<br>intended to deliver humidified<br>breathing gases for administration<br>to an infant/neonate patient. Gases<br>available for medical use do not<br>contain sufficient moisture and may<br>damage or irritate the respiratory<br>tract, or desiccate secretions of<br>patients whose supraglottic airways<br>have been bypassed. Thus<br>humidified gases via heated<br>breathing circuit may be indicated<br>for patients requiring mechanical<br>ventilation, positive pressure<br>breathing assistance, or general<br>medical gases. These gases may<br>be delivered by nasal prongs using<br>the CareFusion Infant Flow system<br>and the single limb circuit, or<br>through bypassing upper airways,<br>for example through the use of an<br>endotracheal tube with the dual<br>limb configuration. | |
| Environment of<br>Use | Professional Healthcare<br>Environment | Professional Healthcare<br>Environment | |
| Patient<br>Population | Infant | Neonate/Infant | |
| Compatible<br>Humidifiers | MR850 Heated Humidifier | Hudson RCI Neptune Heated<br>Humidifier | |
| Comparative<br>Characteristics | RT265 and RT266 Dual Heated<br>Infant Breathing Circuits | Neonatal ConchaSmart Breathing<br>Circuits | |
| | Predicate (K103767) | Proposed | |
| Circuit<br>Configurations | Dual Limb Heated Wire Circuit | Dual Limb Heated Wire Circuit<br>Single Limb Heated Wire Circuit | |
| Flow Range | RT265 - flow rates greater than<br>4LPM<br>RT266 - 0.3LPM and 4LPM" | 1-15 LPM | |
| Kit components | Remote temperature port tubing<br>Water feed tube<br>Tubing adaptor set<br>15 mm adaptors<br>Luer tee connector<br>Step down adaptor<br>Humidifier limb | ConchaSmart Column, Non-sterile<br>Airway adaptors and connectors<br>Humidifier Limb tubing with<br>connectors<br>Remote Temp Port tubing with<br>connectors<br>Pressure Line tubing with<br>connectors | |
| Disposable vs.<br>Reusable | Disposable | Disposable | |
| Circuit Tubing<br>Diameter | 11mm | 10mm | |
| Design | Corrugated | Corrugated | |
| Wire Maximum<br>Power | Unknown | Single Limb: 21 Watts<br>Dual Limb: 21 to 30 Watts | |
| Min Circuit<br>Resistance | Unknown | Single Limb: 3 Ohms<br>Dual Limb: 0.8 to 4.7 Ohms | |
| Wire Material | Unknown | Single Limb: Nickel Chromium Iron<br>Dual Limb: Nickel Chromium Iron<br>and Nickel Chromium | |
| Wire Length | Unknown | Single Limb: 116 inches<br>Dual Limb: 94 to 141 inches | |
| Heated Wire<br>Specification | Unknown | Maximum of 2.5W per linear foot | |
| Circuit Length | 1.5 meters | Single Limb: 1.8 meters<br>Dual Limib: 1.5 meters to 1.8<br>meters | |
| Resistance to<br>Flow<br>(inspiratory<br>limb) | 2 cmH2O @ 13LPM | Less than 1.89 cmH2O @ 2.5LPM<br>(per ISO 5367:2014) | |
| Resistance to<br>Flow (expiratory<br>limb) | 2 cmH2O @ 13LPM | Less than 1.89 cmH₂O @ 2.5LPM<br>(per ISO 5367:2014) | |
| Compliance | 0.81 ml/cmH20 | Less than 1.5 ml/cmH₂O @ 60<br>cmH2O<br>(per ISO 5367:2014) | |
| Leakage | Maximum Circuit Gas Leakage:<br>75ml/min @ 60 cmH2O | Less than 30 ml/min @ 60 cmH2O<br>(per ISO 5367:2014) | |
| Useful Life | 7 days | 21 days | |
| Comparative<br>Characteristics | RT265 and RT266 Dual Heated<br>Infant Breathing Circuits<br>Predicate (K103767) | Neonatal ConchaSmart Breathing<br>Circuits<br>Proposed | |
| Standards | Interface Connection: ISO 5356-1<br>(no other standards cited in 510(k) summary or on labeling) | Interface Connection: ISO 5356-1:2015<br>Compliant to ISO 5367:2014<br>Tested per ISO 10993 and<br>Guidance for Industry and Food<br>and<br>Drug Administration Staff: Use of<br>International Standard ISO<br>10993-1, "Biological evaluation of<br>medical devices - Part 1: Evaluation<br>and testing within a risk<br>management process" | |
| Heated Wire<br>Breathing<br>Circuit Materials | Not disclosed in 510(k) Summary | | |
{6}------------------------------------------------
{7}------------------------------------------------
## Similarities
The proposed device and predicate are similar in that they have the same intended use and same indications for use. They are used for the same patient population in the dual limb configuration and were subjected to and passed testing to the same performance (ISO 5367, ISO 8185, ISO 5356-1), biological (ISO 10993), and electrical (IEC 60601-1 and IEC 60601-1-2) standards.
#### Differences
The main differences are the limb configuration of single limb, the flow rates, and the use of the device with the Neptune Heated Humidifier. With respect to the limb configurations, a reference predicate was introduced that is a single limb configuration cleared for the neonatal / infant population for use with a heated humidifier. The physical and electrical characteristics of the proposed device were testing according to IEC 60601-1 and IEC 60601-1-2 and were appropriate for the indicated Heated Humidifier (Hudson RCI Neptune). For the tubing dimensions, testing to ISO 5367 showed that the proposed device and the predicate were substantially equivalent.
#### I. Performance Data
The following testing was performed on the proposed devices.
- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for . Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995, applicable sections
- ISO 8185:2007 - Respiratory tract humidifiers for medical use -- Particular requirements for respiratory humidification systems, applicable sections
- . ISO 5356-1:2015 - Anaesthetic and respiratory equipment -- Conical connectors --Part 1: Cones and sockets, applicable sections
- ISO 5367:2014 Anaesthetic and respiratory equipment -- Breathing sets and . connectors, applicable sections
- Biocompatibility .
- o ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- o ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5:
{8}------------------------------------------------
- Tests for in vitro cytotoxicity o
- ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: o Tests for irritation and skin sensitization
- ISO 10993-17:2002 Biological evaluation of medical devices -- Part 17: o Establishment of allowable limits for leachable substances
- o ISO 10993-18:2002 Biological evaluation of medical devices -- Part 18: Chemical characterization of materials
- Shelf Life (accelerated aging) ●
- . Useful life
The proposed devices were tested to the neonatal requirements of ISO 5367:2014. In addition, testing was performed to ensure compatibility with the Hudson RCI Neptune Heated Humidifier and a useful life of 21 days. Cytotoxicity, sensitization, irritation, and Extractable and Leachable testing were performed to demonstrate biocompatibility of the patient contacting materials.
| Test Description | Test Objective | Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Shelf Life testing | The proposed Neonatal ConchaSmart Breathing<br>Circuits were aged according to ASTM F1980:<br>Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices for the<br>equivalent of one year. | N/A<br>(for test setup<br>only) |
| Useful Life testing | The proposed Neonatal ConchaSmart Breathing<br>Circuits were tested for a 21 day useful life after<br>accelerated aging equivalent to one year using worst<br>case use scenarios. During and after the useful life<br>testing, the circuits were tested for applicable sections<br>of ISO 5367 | PASS |
| All testing below was conducted on the proposed devices after the one year accelerated shelf<br>life and 21 day useful life testing. This was considered the worst case. | | |
| Design Verification<br>testing:<br>ISO 8185:2007 Testing | The proposed Neonatal ConchaSmart Breathing<br>Circuit, when used with the Hudson RCI Neptune<br>Heated Humidifier, must perform as intended post<br>accelerated aging and useful life testing when subject<br>to the test methods from applicable sections of ISO<br>8185:2007 Respiratory tract humidifiers for medical<br>use -- Particular requirements for respiratory<br>humidification systems | PASS |
| Design Verification<br>testing:<br>IEC 60601-1:1988 +<br>A1:1991 + A2:1995<br>Testing | The proposed Neonatal ConchaSmart Breathing<br>Circuit, when used with the Hudson RCI Neptune<br>Heated Humidifier, must perform as intended post<br>accelerated aging and useful life testing when subject<br>to the test methods from applicable sections of IEC<br>60601-1 Medical electrical equipment - Part 1:<br>General requirements for safety: | PASS |
#### J. Summary on Non-Clinical Testing
{9}------------------------------------------------
| Test Description | Test Objective | Results |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Design Verification<br>testing:<br>ISO 5367:2014 Testing | The proposed Neonatal ConchaSmart Breathing<br>Circuit, when used with the Hudson RCI Neptune<br>Heated Humidifier, must perform as intended post<br>accelerated aging and useful life testing when subject<br>to the test methods from applicable sections of ISO<br>5367:2014 - Anaesthetic and respiratory equipment --<br>Breathing sets and connectors | PASS |
| Design Verification<br>testing:<br>ISO 5356-1:2015<br>Testing | The proposed Neonatal ConchaSmart Breathing<br>Circuit must perform as intended post accelerated<br>aging and useful life testing when subject to the test<br>methods from the following applicable sections of ISO<br>5356-1:2015 – Anaesthetic and respiratory equipment<br>-- Conical connectors -- Part 1: Cones and sockets: | PASS |
| Design Verification<br>testing:<br>Additional Design<br>Testing | The proposed Neonatal ConchaSmart Breathing<br>Circuit must perform as intended post accelerated<br>aging and useful life testing when subject to the test<br>methods from additional design testing including<br>component break and slip testing. | PASS |
## K. Conclusion
As with the predicate, all of the acceptance criteria required through the use of the harmonized standards (including performance, biocompatibility, and electrical as discussed above) were met for the proposed device. In addition, the circuit was found to function as intended when used with the Neptune Heated Humidifier, for which it is intended to use. The device data and test results demonstrate that the devices meet the applicable standards for breathing circuits and are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.