The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a combined Continuous Positive Airway Pressure Blower and Heated Respiratory Humidifier (as per 73 BZD, 21 CFR §868.5905 and 73 BTT, 21 CFR §868.5450). The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction. The addition of heated respiratory humidification to the device (normally included as an accessory to CPAP therapy) relieves the drying and irritating effects on the patient airways which may arise from use of a CPAP system. The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.
Device Story
Device is a combined CPAP flow generator and heated respiratory humidifier. Input: ambient air via rear filter. Transformation: motorized fan provides positive pressure; air passes over heated water chamber to add moisture/warmth. Output: humidified pressurized air delivered to patient via tube/mask. Used in home/clinical settings by adult patients. Physician prescribes pressure settings; user adjusts heaterplate temperature. Software monitors CPAP generator, humidifier, power supply, and microprocessor; includes thermal cut-out. Benefits: treats OSA by preventing airway obstruction; humidification reduces airway drying/irritation.
Clinical Evidence
Bench testing only. Performance testing demonstrated equivalent pressure drop vs. flowrate and dynamic pressure variation characteristics compared to predicate CPAP systems. Humidification performance (temperature and humidity levels) verified as equivalent to predicate humidifier across varying flowrates and heaterplate temperatures.
Technological Characteristics
Materials: polycarbonate thermoplastic (case), cast aluminum (heaterplate/chamber base). Sensing: pressure and temperature monitoring. Energy: AC mains power. Connectivity: standalone. Software: microprocessor-based control for pressure and thermal regulation. Safety: thermal cut-out, electronic monitoring. Standards: IEC 601-1.
Indications for Use
Indicated for adult, spontaneously breathing (non-ventilator-dependent) patients with Obstructive Sleep Apnea (OSA) requiring CPAP therapy to prevent airway obstruction.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
JUL 10 1998
Image /page/0/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line in a bold, sans-serif font. Underneath the top line is a horizontal line, and below that is the word "HEALTHCARE" in a thinner, sans-serif font.
Fisher & Paykel Electronic Box 14-348 Panmure. 4-9-574-0100
# K973161
7 August, 1997
1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
## 510(k) Summary of Safety and Effectiveness Information
Model Number / Name: HC200 CPAP Humidifier
Ventilator, Non-Continuous (Respirator) - 73 BZD Classification Name: Anesthesiology Devices, 21 CFR §868.5905 (Class II) Respironics Inc, Solo™ CPAP System, K961626 Predicate Devices: ResCare Ltd, Sullivan® III Nasal CPAP System, K930656 Fisher & Paykel, HC100 Respiratory Humidifier, K915460
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a Continuous Positive Airway Pressure flow generator, as per 73 BZD, 21 CFR §868.5905. It includes a Heated Respiratory Humidifier, as per 73 BTT, 21 CFR §868.5450.
The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction. The addition of heated respiratory humidification relieves the drying and irritating effects on patient airways which may arise from use of a CPAP system. The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.
A motorized fan assembly provides positive air pressure, which can be adjusted as prescribed by a physician. The fan speed is directly related to air pressure, and is controlled by software. An air filter is located at the back of the device leading to the input of the fan assembly. Warmth and moisture are added to the air as it passes over a heated body of water in a humidification chamber, which slides into place on a heaterplate at the front of the unit. The chamber connects directly to the CPAP generator section via a port at the back of the chamber. The temperature of the heaterplate can be adjusted by the user. Temperature is controlled by software, with the level of humidification delivered dependent on heaterplate temperature, rate of air flow, and ambient conditions of temperature and humidity.
Safety features for the device consist of software monitoring of CPAP generator. humidifier and power supply parameters, electronic monitoring of the microprocessor and a thermal cut-out on the heaterplate.
{1}------------------------------------------------
## 510(k) Summary continued - Fisher & Paykel HC200 CPAP Humidifier
Controls include a pressure set potentiometer and mains power switch, on the right side of the device, with the mains power cord. A heaterplate adjustment potentiometer, mask test mode button and display LED are located on the front control panel.
The case enclosure components are a polycarbonate thermoplastic, and the heaterplate is cast aluminum. The humidification chamber is a clear thermoplastic top with an aluminum base. Accessories for the HC200 consist of a delivery tube and replacement humidification chambers.
The technological characteristics of the HC200 are equivalent to those of the predicate devices. The devices operate in the same CPAP operating pressure and heaterplate temperature ranges, and have similar features available to the user in both CPAP and humidification sections. The size, materials and technology used in the devices are equivalent, including the power supply, motor, control technology, enclosure components and air filter sections. The CPAP devices use microprocessor control with associated software safety features, and the humidifier devices use the same electronic and thermal safety features. They comply with international device safety standards for electromedical devices including IEC 601-1.
Testing carried out on the HC200 demonstrates that the device has equivalent performance in provision of both constant positive airway pressure and humidity to the two predicate device types.
The HC200 has equivalent pressure drop vs flowrate and dynamic pressure variation characteristics to the predicate CPAP systems. It provides the same levels of temperature and humidity over varying flowrates as the predicate humidifier, at the same heaterplate temperatures.
signed:
Chris Mander Fisher & Paykel Healthcare
te: 7 August 1997
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 0 1998
Mr. Chris Mander Fisher & Paykel Electronics Limited 25 Carbine Road Panmure, Auckland NEW ZEALAND
K973161 Re: HC200 CPAP Humidifier Requlatory Class: II (two) Product Code: 73 BZD Dated: May 20, 1998 Received: May 22, 1998
Dear Mr. Mander:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Chris Mander
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Cillehan
Thomas J. Callahan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a large, bold font on the top line, with a line underneath. The word "HEALTHCARE" is in a smaller, non-bold font on the bottom line.
[510(k)] Number:
K97316 1
Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158
6 August, 1997
#### Fisher & Paykel HC200 CPAP Humidifier
### PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT
The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a combined Continuous Positive Airway Pressure Blower and Heated Respiratory Humidifier (as per 73 BZD, 21 CFR §868.5905 and 73 BTT, 21 CFR §868.5450).
The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction.
The addition of heated respiratory humidification to the device (normally included as an accessory to CPAP therapy) relieves the drying and irritating effects on the patient airways which may arise from use of a CPAP system.
The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark Kramer
Division of Cardiovascular, Respiratory, and Neurological Devic
510(k) Number K973161
Prescription Use
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.