The SERVE-HF study results were used to support a labeling change, specifically adding a contraindication for the use of ASV therapy in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
SERVE-HF; Randomized, parallel, event-driven, international multicenter study
1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea; Sample Size: 1325; Number of Sites: International multicenter
Guideline-based medical management
Survival and cardiovascular outcomes
Indications for Use
The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The S9 VPAP Adapt is intended for home and hospital use. The VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The VPAP Adapt is intended for home and hospital use. The AirCurve 10 ASV device is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use. The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
Device Story
Flow generator devices providing adaptive servo-ventilation (ASV) therapy to stabilize ventilation during sleep; used in home and hospital settings. Devices measure instantaneous ventilation, calculate target ventilation based on recent averages, and adjust pressure support to servo-control patient ventilation. Therapy modes include CPAP, ASV, and ASVAuto. Systems comprise flow generator, patient tubing, mask, and integrated heated humidifier. Clinicians configure settings; devices operate automatically to maintain 'air splint' and address obstructive events. Benefits include stabilization of ventilation and treatment of sleep-disordered breathing. Contraindicated for specific heart failure populations per SERVE-HF study findings.
Clinical Evidence
No new clinical testing performed. Contraindication supported by the SERVE-HF study, a randomized, parallel, event-driven, international multicenter study of 1325 patients with chronic, symptomatic heart failure (NYHA 2-4), reduced LVEF (≤45%), and moderate to severe central sleep apnea, which showed increased cardiovascular mortality with ASV therapy.
Technological Characteristics
Blower-based positive pressure system (3-25 cmH2O). Integrated heated humidifier and heater controller. Micro-processor controlled. Power: 100-240V, 50-60Hz. Connectivity: Not specified. Software: Micro-processor controlled. Usability: IEC 62366 compliant.
Indications for Use
Indicated for patients >66 lb (30 kg) with obstructive sleep apnea (OSA), central/mixed apneas, or periodic breathing. Contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, often used as a symbol of medicine. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
ResMed Ltd c/o Ms. Larissa D'andrea Director, Government and Regulatory Affairs ResMed Corp 9001 Spectrum Center Boulevard San Diego, California 92123
Re: K160822
Trade/Device Name: S9 VPAP Adapt, VPAP Adapt, AirCurve 10 ASV Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: August 10, 2016 Received: August 11, 2016
Dear Ms. D'andrea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name S9 VPAP ADAPT
Indications for Use (Describe)
The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The S9 VPAP Adapt is intended for home and hospital use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name AirCurve 10 ASV
Indications for Use (Describe)
The AirCurve 10 ASV device is indicated for the treatment of patients weighing more than 66 Ib (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use. The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
Type of Use (Select one or both, as applicable)
| <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name VPAP Adapt
Indications for Use (Describe)
The VPAP Adapt is indicated for the treatment of patients weighing more than 66lb (30kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The VPAP Adapt is intended for home and hospital use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary – S9 VPAP Adapt
| Required | By Section 807.92 (c) |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | 18 March, 2016 |
| Owners Name | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia |
| Submitter | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au |
| Official Contact | Larissa D'Andrea<br>Director, Government & Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes | 73 BZD |
| Class<br>Classification Reference<br>Common/Usual Name<br>Proprietary Name | 21 CFR 868.5905, Product Code 73 BZD<br>Non continuous ventilator (IPPB)<br>S9 VPAP Adapt |
| Predicate device(s) | S9 VPAP Adapt (K102586) |
{6}------------------------------------------------
## Reason for submission
This 510(k) is being submitted as a labelling chanqe for the following contraindication:
ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to guideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.
## Intended Use
The S9 VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The S9 VPAP Adapt is intended for home and hospital use.
## Substantial Equivalence
The new device has the following similarities to the previously cleared predicate devices.
- Same intended use
- . Same operating principle
- Similar technologies
- . Same manufacturing process
ResMed has determined that the new device has not altered the safety and effectiveness of CPAP treatment for patients with Obstructive Sleep Apnoea (OSA), central and/or mixed appeas, or periodic breathing who weigh more than 66 lb (>30 kg). The new device complicable requirements referenced in the FDA guidance documents:
- . FDA Reviewer Guidance for Premarket Notification Submissions (November 1993)
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device S9 VPAP Adapt (K102586)
The new device is as safe and effective as the predicate device.
## Device description
The S9 VPAP Adapt is identical to the predicate device S9 VPAP Adapt (K102586), using a blower based positive pressure system with an integrated heated humidfier and heater controller. The device platform is identical to the S9 VPAP Adapt (K102586) and contains a Micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmHJO as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier.
The S9 VPAP ADAPT is a flow generator device designed to provide adaptive servo-ventilation therapy (ASV) to stabilize a patient's ventilation during sleep. The device continually measures instantaneous ventilation,
{7}------------------------------------------------
and calculates a target ventilation based on to the patient's recent average. It then adjusts the degree of pressure support to servo-control the patient's ventilation to at least equal the target ventilation. Therapy modes contained in the S9 VPAP Adapt are:
- . CPAP mode;
- CPAP with EPR .
- Auto Servo Ventilation (ASV) ●
Therapy modes come from the predicate S9 VPAP Adapt (K102586).
The functional characteristics of the S9 VPAP Adapt system include all the clinician and user friendly features of the predicate device.
| Characteristic | S9 VPAP Adapt (K102586) | New Device (S9 VPAP Adapt ) | Comments |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Indication for<br>use | The S9 VPAP Adapt is indicated<br>for the treatment of patients<br>weighing more than 66 lb (30kg)<br>with obstructive sleep apnea<br>(OSA), central and/or mixed<br>apneas, or periodic breathing.<br>The S9 VPAP Adapt is intended<br>for home and hospital use. | The S9 VPAP Adapt is indicated<br>for the treatment of patients<br>weighing more than 66 lb (30kg)<br>with obstructive sleep apnea<br>(OSA), central and/or mixed<br>apneas, or periodic breathing.<br>The S9 VPAP Adapt is intended<br>for home and hospital use. | Equivalent<br>Only labeling change<br>to include<br>contraindication |
| Location of use | Hospital/Home | Hospital/Home | Equivalent |
| Pressure Range and Treatment Modes | | | |
| | 4-20 cm H2O (CPAP)<br>3-25 cm H2O (ASV) | 4-20 cm H2O (CPAP)<br>3-25 cm H2O (ASV) | Equivalent: |
| RAMP Settings | > User selected as "Off" to 45<br>minutes in 5 minute<br>increments<br>> Max Ramp time set at<br>clinician's discretion | > User selected as "Off" to 45<br>minutes in 5 minute<br>increments<br>> Max Ramp time set at<br>clinician's discretion | Equivalent |
| System<br>Components | > Flow generator<br>> Integrated humidifier (5i)<br>> Mask, air tubing and heated<br>tubing | > Flow generator<br>> Integrated humidifier (5i)<br>> Mask, air tubing and heated<br>tubing | Equivalent |
| Power supply | 100-240V, 50-60Hz | 100-240V, 50-60Hz | Equivalent |
| Flow generator<br>weight | 1.7lb | 1.7lb | Equivalent |
| Dimensions H x<br>W x D (inches) | Flow generator unit:<br>3.4 x 5.5 x 6.0 | Flow generator unit:<br>3.4 x 5.5 x 6.0 | Equivalent |
{8}------------------------------------------------
| Characteristic | S9 VPAP Adapt (K102586) | New Device (S9 VPAP Adapt ) | Comments |
|------------------------|---------------------------------------------|---------------------------------------------|------------|
| Supplemental<br>oxygen | Labeled for use with<br>supplemental oxygen | Labeled for use with supplemental<br>oxygen | Equivalent |
## Conclusion
The S9 VPAP Adapt is substantially equivalent to the predicate device, S9 VPAP Adapt (K102586).
{9}------------------------------------------------
# 510(k) Summary – VPAP Adapt
| Required | By Section 807.92 (c) |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | 18 March, 2016 |
| Owners Name | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia |
| Submitter | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au |
| Official Contact | Larissa D'Andrea<br>Director, Government & Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes | 73 BZD |
| Class<br>Classification Reference<br>Common/Usual Name<br>Proprietary Name | 21 CFR 868.5905, Product Code 73 BZD<br>Non continuous ventilator (IPPB)<br>VPAP Adapt |
| Predicate device(s) | VPAP Adapt (K113801) |
{10}------------------------------------------------
## Reason for submission
This 510(k) is being submitted as a labelling chanqe for the following contraindication:
ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to guideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.
## Indication for Use
The VPAP Adapt is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. The VPAP Adapt is intended for home and hospital use
## Substantial Equivalence
The new device has the following similarities to the previously cleared predicate devices.
- . Same intended use
- Same operating principle
- Similar technologies
- Same manufacturing process ●
ResMed has determined that the new device has not altered the safety and effectiveness of CPAP treatment for patients with Obstructive Sleep Apnoea (OSA), central and/or mixed apneas, or periodic breathing who weigh more than 66 lb (>30 kg). The new device complicable requirements referenced in the FDA guidance documents:
- FDA Reviewer Guidance for Premarket Notification Submissions (November 1993) ●
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May ● 11, 2005)
As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device VPAP Adapt (K113801).
The new device is as safe and effective as the predicate device.
## Device Description
The VPAP Adapt is identical to the predicate device (K113801), using a blower based positive pressure system with an integrated heated humidifier and heater controller. The device platform is identical to the VPAP Adapt (K113801) and contains a Micro-processor controlled blower system that qenerates controlled positive airway pressure between 3-25 cmH2O as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier.
The VPAP Adapt is a flow generator device designed to provide adaptive servo-ventilation therapy (ASV) to stabilize a patient's ventilation during sleep. The device continually measures the patient's instantaneous ventilation, and
{11}------------------------------------------------
calculates a target ventilation based on to the patient's recent average. It then adjusts the degree of pressure support to servo-control the patient's ventilation to at least equal the target ventilation.
Therapy modes contained in the VPAP Adapt are
- СРАР .
- ASV ●
- ASVAuto .
The CPAP and ASV therapy modes come from the VPAP Adapt (K113801).
The functional characteristics of the VPAP Adapt system includes all the clinician and user friendly features of the predicate device.
| Characteristic | VPAP Adapt (K113801) | New Device (VPAP Adapt) | Comments |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indication for<br>use | The VPAP Adapt is indicated for<br>the treatment of patients<br>weighing more than 66 lb (30<br>kg) with obstructive sleep apnea<br>(OSA), central and/or mixed<br>apneas, or periodic breathing. It<br>is intended for home and<br>hospital use. | The VPAP Adapt is indicated for<br>the treatment of patients weighing<br>more than 66 lb (30 kg) with<br>obstructive sleep apnea (OSA),<br>central and/or mixed apneas, or<br>periodic breathing. It is intended for<br>home and hospital use. | Equivalent<br>Only labeling change to<br>include contraindication |
| Location of use | Hospital/Home | Hospital/Home | Equivalent |
| Pressure Range and Treatment Modes | | | |
| | 4-20 cm H2O (CPAP)<br>3-25 cm H2O (ASV)<br>3-25 cm H2O (ASVAuto) | 4-20 cm H2O (CPAP)<br>3-25 cm H2O (ASV)<br>3-25 cm H2O (ASVAuto) | Equivalent: |
| RAMP Settings | > User selected as "Off" to 45<br>minutes in 5 minute<br>increments<br>> Max Ramp time set at<br>clinician's discretion | > User selected as "Off" to 45<br>minutes in 5 minute<br>increments<br>> Max Ramp time set at<br>clinician's discretion | Equivalent |
| System<br>Components | > Flow generator<br>> Integrated humidifier (5i)<br>> Mask, air tubing and heated<br>tubing | > Flow generator<br>> Integrated humidifier (5i)<br>> Mask, air tubing and heated<br>tubing | Equivalent |
| Power supply | 100-240V, 50-60Hz | 100-240V, 50-60Hz | Equivalent |
| Flow generator<br>weight | 1.7lb | 1.7lb | Equivalent |
| Dimensions H x<br>W x D (inches) | Flow generator unit:<br>3.4 x 5.5 x 6.0 | Flow generator unit:<br>3.4 x 5.5 x 6.0 | Equivalent |
| Characteristic | VPAP Adapt (K113801) | New Device (VPAP Adapt) | Comments |
| Supplemental<br>oxygen | Labeled for use with<br>supplemental oxygen | Labeled for use with supplemental<br>oxygen | Equivalent |
{12}------------------------------------------------
# Conclusion
The VPAP Adapt is substantially equivalent to the predicate devices, VPAP Adapt (K113801).
{13}------------------------------------------------
# 510(k) Summary – AirCurve 10 ASV
| Required | By Section 807.92 (c) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | 18 March, 2016 |
| Owners Name | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia |
| Submitter | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au |
| Official Contact | Larissa D'Andrea<br>Director, Government & Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes | 73 BZD |
| Class | II |
| Classification Reference | 21 CFR 868.5905, Product Code 73 BZD |
| Common/Usual Name | Non continuous ventilator (IPPB) |
| Proprietary Name | AirCurve 10 ASV |
| Predicate device(s) | S9 Greenhills (K140279) |
{14}------------------------------------------------
## Reason for submission
This 510(k) is being submitted as a labelling chanqe for the following contraindication:
ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to guideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.
### Indication for Use
The AirCurve 10 ASV device is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing. It is intended for home and hospital use. The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
### Substantial Equivalence
The new device has the following similarities to the previously cleared predicate devices.
- . Same intended use
- . Same operating principle
- . Similar technologies
- Same manufacturing process ●
ResMed has determined that the new device has not altered the safety and effectiveness of CPAP/Bilevel treatment for patients with Obstructive Sleep Apnoea (OSA), central and/or mixed apneas, or periodic breathing who weigh more than 66 lb (>30 kg).
The new device complies with the applicable requirements referenced in the FDA quidance documents:
- . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- FDA Draft Guidance for Industry and FDA Staff - Design Considerations for Devices Intended for Home Use- Document Issued on: December 12, 2012
- FDA Draft Guidance for Industry and FDA Staff Radio Frequency Wireless Technology in Medical Devices ● - Document Issued on: August 13, 2013
- Reviewer Guidance for Premarket Notification Submissions, ARDB, CDRH, FDA, November 1993. .
As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device S9 Greenhills (K140279).
The new device is as safe and effective as the predicate device.
{15}------------------------------------------------
### Device Description
The AirCurve 10 ASV is identica to the predicate device, S9 Greenhills (K140279), using a blower based positive pressure system with an integrated heated humidifier and heater controller. The device platform is identical to the S9 Greenhills (K140279) and contains a Micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmH2O. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier.
The AirCurve 10 ASV is a flow generator device designed to provide adaptive servo-ventilation therapy (ASV mode) or (ASVAuto mode) to stabilize a patient's ventilation during sleep. The device continually measures the patient's instantaneous ventilation, and calculates a target ventilation based on to the patient's recent average minute ventilation. It then adjusts the degree of pressure support to servo-control the patilation to at least equal the target ventilation. The same is true for ASVAuto mode except the EPAP is adjusted to address any obstructive apneas detected.
Therapy modes contained in the AirCurve 10 ASV are CPAP, ASV, and ASVAuto. They are:
- . CPAP mode - the device delivers a continuous positive airway pressure throughout the entire therapy session;
- ASV mode the device automatically adjusts pressure support in response to the patient's recent average . minute ventilation; and
- ASVAuto mode the device automatically adjusts pressure support in response to the patient's recent ● average minute ventilation and EPAP level for OSA events.
The functional characteristics of the AirCurve 10 ASV system includes all the clinician and user friendly features of the predicate device which have been verified during usability studies in accordance with IEC 62366 Medical devices - Application of usability engineering to medical devices.
The functional characteristics of the system include all the clinician and user friendly features of the predicate device S9 Greenhills (K140279).
| Characteristic | S9 Greenhills (K140279) | New Device (AirCurve 10 ASV) | Comments |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indication for<br>use | The S9 Greenhills is indicated<br>for the treatment of patients<br>weighing more than 66 lb (30<br>kg) with obstructive sleep<br>apnea (OSA), central and/or<br>mixed apneas, or periodic<br>breathing. It is intended for<br>home and hospital use. The<br>humidifier is intended for single<br>patient use in the home<br>environment and re-use in a<br>hospital/institutional<br>environment. | The AirCurve 10 ASV device is<br>indicated for the treatment of<br>patients weighing more than 66 lb<br>(30 kg) with obstructive sleep apnea<br>(OSA), central and/or mixed<br>apneas, or periodic breathing. It is<br>intended for home and hospital use.<br>The humidifier is intended for single<br>patient use in the home<br>environment and re-use in a<br>hospital/institutional environment. | Equivalent<br>Marketed name change<br>only<br>Only labeling change to<br>include contraindication |
| Location of use | Hospital/Home | Hospital/Home | Equivalent |
{16}------------------------------------------------
| Characteristic | S9 Greenhills (K140279) | New Device (AirCurve 10 ASV) | Comments |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-------------|
| <b>Pressure Range and Treatment Modes</b> | | | |
| | 4-20 cm H2O (CPAP) | 4-20 cm H2O (CPAP) | Equivalent: |
| | 3-25 cm H2O (ASV) | 3-25 cm H2O (ASV) | |
| | 3-25 cm H2O (ASVAuto) | 3-25 cm H2O (ASVAuto) | |
| RAMP Settings | User selected as "Off" to<br>45 minutes in 5 minute<br>increments<br>Max Ramp time set at<br>clinician's discretion | User selected as "Off" to 45<br>minutes in 5 minute increments<br>Max Ramp time set at<br>clinician's discretion | Equivalent |
| System Components | Flow generator<br>Humidifier<br>Mask, air tubing and<br>heated tubing | Flow generator<br>Humidifier<br>Mask, air tubing and heated<br>tubing | Equivalent |
| Power supply | 100-240V, 50-60Hz | 100-240V, 50-60Hz | Equivalent |
| Flow generator<br>weight | 2.5lb | 2.5lb | Equivalent |
| Dimensions H x<br>W x D (inches) | Flow generator unit:<br>4.5 x 9.6 x 6.0 | Flow generator unit:<br>4.5 x 9.6 x 6.0 | Equivalent |
| Supplemental<br>oxygen | Labeled for use with<br>supplemental oxygen | Labeled for use with supplemental<br>oxygen | Equivalent |
## Conclusion
The AirCurve 10 ASV is substantially equivalent to the predicate device, S9 Greenhills (K140279).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.