DeVilbiss DV6WM Wireless Modem

K152810 · Devilbiss Healthcare, LLC · BZD · Jan 21, 2016 · Anesthesiology

Device Facts

Record IDK152810
Device NameDeVilbiss DV6WM Wireless Modem
ApplicantDevilbiss Healthcare, LLC
Product CodeBZD · Anesthesiology
Decision DateJan 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2

Indications for Use

The DeVilbiss Healthcare DV6WM Wireless Modem is intended to be used as a data collection tool for DeVilbiss IntelliPAP Series with SmartLink II Bluetooth module, IntelliPAP2, and DeVilbiss Blue Series CPAPs for patients in a home or healthcare environment. It is not intended to be used as a diagnostic tool.

Device Story

DV6WM Wireless Modem acts as data collection/transfer accessory for DeVilbiss CPAP devices; enables remote monitoring of usage/performance data. Device uses Bluetooth to read data from CPAP; transfers data via GSM cellular network to secure server. Healthcare providers access data via SmartLink Desktop software; allows remote prescription setting updates. Operated by patients in home or clinicians in healthcare environments. Features three operational modes: Wake (read CPAP), Wake (upload to server), and Low Power Sleep. Includes status LEDs and test call button. Benefits include improved compliance monitoring and remote therapy management without physical SD card transfer.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home use standards (IEC 60601-1-11), and human usability validation.

Technological Characteristics

Internally powered by rechargeable Lithium Polymer battery (IEC 62133, UN 38.3). Connectivity: Bluetooth (to CPAP) and GSM cellular (to server). Form factor: 11.2cm x 7.6cm x 2.5cm. Ingress protection: IP21. Standards: IEC 60601-1, EN ISO 17510-1, EN 60601-1-2, RTCA/DO-160.

Indications for Use

Indicated for patients using DeVilbiss IntelliPAP Series (with SmartLink II), IntelliPAP2, or DeVilbiss Blue Series CPAPs in home or healthcare settings. Not for diagnostic use.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 21, 2016 DeVilbiss Healthcare, LLC c/o Paul Dryden President, ProMedic, Inc. 100 Devilbiss Drive Somerset. Pennsylvania 15501 Re: K152810 Trade/Device Name: DeVilbiss DV6WM Wireless Modem Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: December 20, 2015 Received: December 22, 2015 Dear Paul Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, *Tejashri Purohit-Sheth, M.D.* Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith M.S. Director Division of Anesthesiology. General Hospital, Respiratory Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K152810 Device Name DeVilbiss DV6WM Wireless Modem #### Indications for Use (Describe) The DeVilbiss Healthcare DV6WM Wireless Modem is intended to be used as a data collection tool for DeVilbiss IntelliPAP Series with SmartLink II Bluetooth module. IntelliPAP2. and DeVilbiss Blue Series CPAPs for patients in a home or healthcare environment. It is not intended to be used as a diagnostic tool. Type of Use (Select one or both, as applicable) XX Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov "An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) {3}------------------------------------------------ # 510(k) Summary # Administrative Information and Device Identification | Name and address of the manufacturer and<br>sponsor of the 510(k) submission: | DeVilbiss Healthcare LLC<br>100 DeVilbiss Drive<br>Somerset, PA 15501 | |-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------| | Official contact person for all<br>correspondence: | Betty Miller<br>Regulatory/Compliance Manager<br>Phone: 814-443-7606<br>Fax: 814-443-7575 | | Date Prepared: | December 20, 2015 | | Device Name: | DeVilbiss DV6WM Wireless Modem | | Proprietary name of new device: | DeVilbiss DV6WM Wireless Modem | | Common or usual name of the device: | ventilator, non-continuous (respirator) | | DeVilbiss Model Number | DV6WM-NA, DV6WM-EU, DV6WM-UK | | Classification | Class II | | Panel Code: | 73 BZD | | CFR Regulation Number: | 21 CFR 868.5905 | | Predicate Device Name(s) and 510(k)<br>number(s): | K082209 – DeVilbiss DV5M Smartlink<br>System<br>K093684 Resmed EasyCare Online | {4}------------------------------------------------ # Description of Device: The proposed DeVilbiss DV6WM Wireless Modem is an accessory to a DeVilbiss CPAP machine. The proposed device provides automatic wireless telemetry of CPAP usage and performance data to a healthcare provider. Predicate devices DeVilbiss DV5M Smartlink System (K082209) and Resmed EasyCare Online (K093684) operate by the same basic principal of transferring CPAP data from a therapy device to a PC program for storing and reporting patient usage and machine performance. Predicate device Resmed EasyCare Online (K093684) is used for comparison of data transfer through a cellular data network. The proposed DeVilbiss DV6WM Wireless Modem is similar in size, shape and functionality to the predicate with the following modifications - Bluetooth wireless communication - Cellular wireless communication ● - Rechargeable Lithium Polymer battery power ### Indications for Use: The DeVilbiss Healthcare DV6WM Wireless Modem is intended to be used as a data collection tool for DeVilbiss IntelliPAP Series with SmartLink II Bluetooth module, IntelliPAP2, and DeVilbiss Blue Series CPAPs for patients in a home or healthcare environment. It is not intended to be used as a diagnostic tool. ### Comparison of Device Technological Characteristics to Predicate Devices: The proposed DeVilbiss DV6WM Wireless Modem has the following similarities to those which previously received 510(k) concurrence: - . Has the same intended use - Uses the same operating principle Modifications that were made: - Bluetooth wireless communication ● - Cellular wireless communication . - Rechargeable Lithium Polymer battery power ● {5}------------------------------------------------ # Comparison of Similarities and Differences: | Characteristics | Predicate Device:<br>Resmed EasyCare Online<br>(K093684) | Predicate Device:<br>DeVilbiss DV5M SmartLink<br>System (K082209) | Modified Device:<br>DeVilbiss DV6WM Wireless<br>Modem | Nature of<br>Change | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | Indications for<br>Use Statement | EasyCare Online is intended to<br>augment the standard follow-<br>up care of patients diagnosed<br>with obstructive sleep apnea<br>by displaying usage and<br>therapeutic information that<br>has been transmitted from<br>the patient's flow generator<br>located in the home to the<br>care giver. EasyCare Online<br>also provides remote settings<br>capabilities.<br>It is intended to be used by<br>Clinicians in conjunction with<br>ResMed compatible flow<br>generators. | The DeVilbiss DV5M SmartLink<br>System in conjunction with a<br>DeVilbiss CPAP Unit is intended<br>for use in treating obstructive<br>sleep apnea in patients 30 Kg<br>and above. The system is not<br>intended to be used as a tool for<br>diagnosis of obstructive sleep<br>apnea (OSA) or related<br>disorders. It is intended to be<br>used to follow up the progress<br>of a patients previously<br>diagnosed with OSA or related<br>disorders treated with nasal<br>CPAP. The system is to be<br>used in home and clinical<br>environments. | The DeVilbiss Healthcare<br>DV6WM Wireless Modem is<br>intended for use as an<br>accessory to DeVilbiss<br>IntelliPAP Series with SmartLink<br>II Bluetooth module, IntelliPAP2,<br>and DeVilbiss Blue Series<br>CPAPs for patients in a home or<br>healthcare environment. | Similar to<br>predicate.<br>no impact on<br>safety or<br>effectiveness | | Operating<br>Principle | Device reads data recorded<br>and stored in the Resmed S9<br>or higher Series CPAP,<br>encodes the data and transfers<br>data to a secure data server.<br>Data is viewed using an on-<br>line data management and<br>report generator web portal. | Device records data from a<br>DeVilbiss DV5 Series CPAP<br>while the CPAP is operating.<br>Internal storage of recorded<br>data is transferred to SmartLink<br>Desktop software via SD card<br>and stored in a local database.<br>SmartLink Desktop software<br>creates reports of CPAP usage<br>and performance. | Device reads data recorded and<br>stored in the DV6 Series CPAP,<br>encodes the data and sends to<br>a secure data server. Data is<br>downloaded into SmartLink<br>Desktop software from secure<br>data server and stored in a local<br>database. SmartLink Desktop<br>software creates reports of<br>CPAP usage and performance. | Similar to<br>predicate.<br>no impact on<br>safety or<br>effectiveness | | Operating<br>Modes | Acts as a data transfer device,<br>transfers data from Resmed<br>S9 CPAP to a secure data<br>server. | Records data from CPAP,<br>1. Transfers data to<br>2. SmartLink Desktop<br>software via SD card. | Acts as a data transfer device.<br>Three modes of operation:<br>1. Wake, talking to CPAP,<br>reads recorded usage data<br>from CPAP<br>2. Wake, talking to server,<br>transfers CPAP data to a<br>secure data server<br>3. Sleep, low power mode,<br>waiting for next scheduled<br>call time | Similar to<br>predicate.<br>no impact on<br>safety or<br>effectiveness | | 510(k) Product<br>Code | BZD (Ventilator, non-<br>continuous (respirator)) | BZD (Ventilator, non-continuous<br>(respirator)) | BZD (Ventilator, non-continuous<br>(respirator)) | Same as<br>predicate | | Power<br>Requirements | Powered from host CPAP | Powered from host CPAP | Powered by internal<br>rechargeable Lithium Polymer<br>battery | Change to<br>internally<br>powered<br>device | | Compliance<br>Monitoring | Transfers data recorded and<br>stored in the CPAP wirelessly<br>to a secure data server | Records 3 years of "while<br>breathing" ON/OFF compliance,<br>120 days of Daily Performance,<br>Rx change, approximately 72<br>hours of Oximetry, and Fault<br>logs | Transfers data recorded and<br>stored in the CPAP wirelessly to<br>a secure data server | Similar to<br>predicate.<br>no impact on<br>safety or<br>effectiveness | | User interface | Shows info on host CPAP LCD<br>display | Shows info on host CPAP LCD<br>display | Shows status info on host CPAP<br>display<br>On proposed device: three LED<br>status indicators for battery,<br>Bluetooth and GSM (cellular);<br>One pushbutton to initiate a test<br>call | Change to<br>indicate new<br>functions.<br>no impact on<br>safety or<br>effectiveness | | Characteristics | Predicate Device: | Predicate Device: | Modified Device: | Nature of Change | | | Resmed EasyCare Online<br>(K093684) | DeVilbiss DV5M SmartLink<br>System (K082209) | DeVilbiss DV6WM Wireless<br>Modem | | | Remote<br>settings change<br>function | Can change CPAP<br>prescription settings through<br>Wireless Modem from<br>EasyCare Online Software | Can change CPAP prescription<br>settings through SD Card from<br>SmartLink Desktop Software | Can change CPAP prescription<br>settings through Wireless<br>Modem from SmartLink Desktop<br>Software | Similar to<br>predicate.<br><br>no impact on<br>safety or<br>effectiveness | | Operating<br>Temperature<br>Range | Operating Temperature:<br>+5° to +35° C<br>(+41° to +95° F) | Operating Temperature:<br>+5 to +40 °C.<br>(+41°F to +104°F) | Operating Temperature:<br>+5 °C to +40 °C<br>(+41°F to +104°F) | Same as<br>predicate. | | Operating<br>Humidity<br>Range | Relative Humidity:<br>15 to 95% (non-condensing) | Relative Humidity:<br>0 to 95% (non-condensing) | Relative Humidity:<br>15 % to 93 % (non-condensing) | | | Storage<br>Temperature<br>Range | Storage Temperature:<br>-20° to 60°C<br>(-4° to 140° F) | Storage Temperature:<br>- 40°C to + 70°C<br>(-40 to +158 °F) | Storage Temperature:<br>- 25°C to + 70°C<br>(-13 to +158 °F) | | | | Relative Humidity:<br>15 to 95% (non-condensing) | Relative Humidity:<br>0% to 95% (non-condensing) | Relative Humidity:<br>15% to 93% (non-condensing) | | | Operating<br>atmospheric<br>pressure range | 101 to 77 kPa<br>(0 - 2286 m / 0 - 7500 ft) | 101 kPa to 74 kPa<br>Sea level to 9,000 feet<br>(0 to 2743 m) | 700 hPa to 1060 hPa<br>(~9800 ft to ~ 1400ft below sea<br>level) | | | Shock and<br>Vibration | Not published | Unit shall be able to withstand a<br>drop from a height of 30"<br>IEC 68 Environmental Testing<br>ASTM D-4169 Testing of<br>Shipping Containers | Unit shall be able to withstand a<br>drop from a height of 30"<br>IEC 68 Environmental Testing<br>ASTM D-4169 Testing of<br>Shipping Containers | Same as<br>predicate. | | IEC 60601-1<br>Classification | Type of Protection Against<br>Electric Shock: Class II<br>Equipment | Type of Protection Against<br>Electric Shock: Class II<br>Equipment | Type of Protection Against<br>Electric Shock: Internally<br>Powered ME Equipment | Change in<br>classification<br>to internally<br>powered<br>device, no<br>applied parts | | | Degree of Protection Against<br>Electric Shock: Type BF<br>Applied Part | Degree of Protection Against<br>Electric Shock: Type BF Applied<br>Part | Degree of Protection Against<br>Electric Shock: N/A, No applied<br>part | no impact on<br>safety or<br>effectiveness | | | Degree of Protection against<br>Ingress of Water (device & AC<br>power supply): Drip Proof,<br>IPX1 | Degree of Protection against<br>Ingress of Water (device & AC<br>power supply): Drip Proof, IPX1 | Degree of Protection against<br>Ingress of Water (device & AC<br>power supply): Drip Proof, IP21 | | | | Mode of Operation:<br>Continuous Electrical | Mode of Operation: Continuous<br>Electrical | Mode of Operation:<br>Non-Continuous | | | Standards | | | IEC 60601-1<br>EN ISO 17510-1<br>EN 60601-1-2<br>RTCA/DO-160 section 21,<br>category M | | | Dimensions | 39.6mm x 54.1mm x 125.0mm | 3.9" H x 3.1" W x 1.4" D<br>(9.9cm H x 7.9cm W x 3.6cm D) | 4.4"H x 3"W x 1"D<br>(11.2cm x 7.6cm x 2.5cm) | | | Weight | < 100 g | 0.30 lbs. (0.14 Kg) | 0.35 lbs. (158.8 g) | | {6}------------------------------------------------ # Discussion of Substantial Equivalence and Differences: ### Indications for Use Similar to predicate device DeVilbiss DV5M SmartLink System (K082209). The proposed DeVilbiss DV6WM Wireless Modem is intended to be used as a data collection tool for DeVilbiss IntelliPAP Series with SmartLink II Bluetooth module, IntelliPAP2, and DeVilbiss Blue Series CPAPs and is not intended to be used as a diagnostic tool. Although the method of data retrieval is different, the purpose of collecting data is the same. {7}------------------------------------------------ # Operating Principle Similar to predicate device Resmed EasyCare Online (K093684). The proposed DeVilbiss DV6WM Wireless Modem operates as a wireless data transfer device, similar to the predicate Resmed EasyCare Online (K093684). Both devices use GSM cellular telephone technology to send data recorded by the patient's CPAP to a remote data server for access by a healthcare provider. The proposed DeVilbiss DV6WM Wireless Modem uses a wireless link to the patient's CPAP device whereas the predicate Resmed EasyCare Online (K093684) physically and electrically attaches to the patient's CPAP device. The physical separation of the proposed DeVilbiss DV6WM Wireless Modem allows the user to locate the modem where it can get a sufficient cellular signal to connect to the remote data server, and still be in Bluetooth range of the patient's CPAP device. The proposed DeVilbiss DV6WM Wireless Modem completes a data transfer once per day. ## Operating Modes Similar to predicate device Resmed EasyCare Online (K093684). The proposed DeVilbiss DV6WM Wireless Modem has three modes of operation; 1) Wake, reading data from a CPAP, 2) Wake, sending data to secure data server, 3) Low Power Sleep. The proposed DeVilbiss DV6WM Wireless Modem is normally in a low power sleep mode. The proposed DeVilbiss DV6WM Wireless Modem wakes from low power sleep mode by either user pressing the Test Call button or at a scheduled call time. When the proposed DeVilbiss DV6WM Wireless Modem wakes to begin a call cycle, it searches for new Bluetooth devices in range, searches for the target device and processes any new pair requests. Data is read from any paired CPAP devices in range, up to ten devices. After all CPAP data has been read, the Bluetooth radio is switched off and the GSM cellular radio is switched on. The proposed DeVilbiss DV6WM Wireless Modem proceeds to find cellular service, contact the secure data server and upload data read from the CPAPs to the secure data server. The secure data server receives and checks the data integrity and responds to the proposed DeVilbiss DV6WM Wireless Modem that the data was received or has errors. If the data was received without errors, the modem clears data on the CPAP. If the data was received with errors, the modem does not clear data on the CPAP so it is available for the next call. The proposed DeVilbiss DV6WM Wireless Modem cannot be contacted through the cellular phone system, it only makes outgoing calls, and it does not receive calls. The predicate device Resmed EasyCare Online (K093684) wireless modem device attached to a host Resmed CPAP and is powered by the host CPAP. The wireless modem sends CPAP data to a secure data server. The data is stored in an online data base and viewed online using a web browser and login credentials. {8}------------------------------------------------ #### Power Requirements The proposed DeVilbiss DV6WM Wireless Modem is internally powered by a rechargeable Lithium Polymer (LiPo) battery. The internal LiPo battery is certified to IEC 62133 and UN 38.3, as required by IEC 60601-1 3rd Edition. The LiPo battery allows placement of the modem where it can place cellular calls while in Bluetooth range of the target CPAP. The predicate devices Resmed EasyCare Online (K093684) and DeVilbiss DV5M SmartLink System (K082209) are physically and electrically attached to the host CPAP device, receiving power from the host CPAP device. #### Compliance Monitoring Similar to predicate device Resmed EasyCare Online (K093684). The proposed DeVilbiss DV6WM Wireless Modem reads data recorded by the host CPAP and transfers the data to a secure data server. The proposed DeVilbiss DV6WM Wireless Modem does not record data and only stores data temporarily during transfer to the secure data server. The predicate device Resmed EasyCare Online (K093684) performs the same function, transferring data recorded by the host CPAP to a secure data server. #### User interface Changed to indicate new functions on proposed device. The proposed DeVilbiss DV6WM Wireless Modem user interface consists of three status LEDs and one pushbutton. The status LEDs indicate status of battery and activity of Bluetooth and GSM (cellular) radio. The proposed DeVilbiss DV6WM Wireless Modem sends status information to a DeVilbiss IntelliPAP2/DeVilbiss Blue Series CPAP during a call sequence. The status information tells the user the battery charge level, Bluetooth signal strength and cellular signal strength. It also indicates the modem serial number and number of days since the last time data was sent to the server. DeVilbiss IntelliPAP Series CPAPs do not display modem status information, this information is available in patient records and reports in DeVilbiss SmartLink Desktop Software. Functions of the predicate device DeVilbiss DV5M SmartLink System (K082209) that display usage summary information are built into the DV6x Series CPAP. ### Remote settings change function Similar to predicate device Resmed EasyCare Online (K093684) and predicate device DeVilbiss DV5M SmartLink System (K082209). The predicate device Resmed EasyCare Online (K093684) accommodates a similar method of updating CPAP settings. A healthcare provider can modify the CPAP settings on the EasyCare Online web interface. The Resmed wireless modem receives the setting update from the EasyCare Online secure data server and applies them to the patient's CPAP. {9}------------------------------------------------ The proposed DeVilbiss DV6WM Wireless Modem accommodates wireless prescription settings updates in the same manner that the predicate device DeVilbiss DV5M SmartLink System (K082209) accommodates prescription setting updates via SD card. The healthcare provider modifies CPAP settings for the patient's assigned CPAP device within SmartLink Desktop software. When saving the setting changes, the SmartLink Desktop software provides 3 ways to update settings on the patient's CPAP device; 1) direct wired connection of CPAP to PC, 2) SD card transfer, 3) Wireless Modem. Options 1) and 2) are part of predicate device DeVilbiss DV5M SmartLink System (K082209). Option 3) is new for the wireless modem, settings are sent to the secure data server and relayed to the wireless modem that is associated with the patient's CPAP serial number. When the modem makes a connection to the CPAP, the settings are updated by the modem and verified. New settings are sent back to the secure data server and a notification is posted for the patient's CPAP serial number that the new settings were applied. The setting update is acknowledged in SmartLink Desktop software on the next download form the secure data server for the healthcare provider to verify that the patient's device has been updated. #### Operating Humidity Range Similar to predicate device DeVilbiss DV5M SmartLink System (K082209). The proposed DeVilbiss DV6WM Wireless Modem complies with requirements IEC 60601-1-11 Medical Devices intended for Home Use. The operating humidity range is more narrow and within the range of the predicate device DeVilbiss DV5M SmartLink System (K082209) ### Storage Temperature Range Changed storage temperature range to comply with IEC 60601-1-11. The proposed DeVilbiss DV6WM Wireless Modem complies with requirements IEC 60601-1-11 Medical Devices intended for Home Use. The storage temperature range is more narrow and within the range of the predicate device DeVilbiss DV5M SmartLink System (K082209) #### Operating atmospheric pressure range Changed atmospheric pressure range to comply with IEC 60601-1-11. The proposed DeVilbiss DV6WM Wireless Modem complies with requirements IEC 60601-1-11 Medical Devices intended for Home Use. The atmospheric pressure range is broader than the predicate device DeVilbiss DV5M SmartLink System (K082209). #### Dimensions The proposed DeVilbiss DV6WM Wireless Modem is similar in size to the predicate device DeVilbiss DV5M SmartLink System (K082209). #### Weight The proposed DeVilbiss DV6WM Wireless Modem is similar in weight to the predicate device DeVilbiss DV5M SmartLink System (K082209). {10}------------------------------------------------ # Performance Testing # Non-Clinical Testing: This device has been tested to appropriate ISO. ASTM. and IEC standards and other applicable requirements passing all test protocols. The proposed DeVilbiss DV6WM Wireless Modem was designed and tested to demonstrate compliance with the applicable sections of the following standards: - 1. - 2. IEC 60601-1-2:2007 Ed. 3.0, - 3. IEC 60601-1-11:2010 Ed. 1.0 + Corrigendum 1.0 (4-2011) - 4. FCC Regulation 47 CFR Part 15 Radio Frequency Devices, Class B - 5. Wireless Standards - FCC Part 15 . - Industry Canada RSP-100 - European wireless standard RTTE directive 1999/5/EC ● Human usability validation was conducted with lay users and clinical users to focus on critical and essential tasks as per analysis of user tasks, and evaluation of use reports for similar devices. Any harm that could result from task failures is described in the usability report. Subjective assessments from all participants, particularly in the case of task failures, were collected and analyzed as part of the usability validation. ### Clinical Testing: No clinical testing was necessary to demonstrate the safety and effectiveness of the proposed DV6WM Wireless Modem. ### Statement of Safety and Effectiveness: Analysis of comparison of design, function and features of the proposed DV6WM Wireless Modem to the predicates DV5M SmartLink System (K082209) and Resmed EasyCare Online (K093684), together with the results of testing demonstrate the device to be substantially equivalent to the predicate devices in terms of meeting performance criteria and functioning as intended. ### Conclusion: The test methods used are the similar to those submitted in the original predicate device submission, updated to reflect new versions of standards and FDA quidance. The proposed DeVilbiss DV6WM Wireless Modem meets all applicable FDA recognized consensus standards. Based on these results, the proposed device, as changed, is substantially equivalent to the following predicate devices: DeVilbiss DV5M SmartLink System (K082209) and Resmed EasyCare Online (K093684) and does not raise any new issues of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...