This mask is intended to provide an interface for application of non-invasive ventilation to patients. The mask is for single patient use in the home and multi-patient use in the hospital/institutional environment. The mask is to be used on patients greater than 10 kg/22 lbs (>10 kg/22 lbs) for whom non-invasive ventilation has been prescribed. Use of this mask is limited by the indications for use of the compatible therapy device with respect to patient weight.
Device Story
Simple T Pediatric Nasal Mask provides interface for non-invasive ventilation; delivers positive airway pressure to patient nose. Device comprises fabric frame, adjustable headgear, and silicone cushion with integrated adjustment dial to reduce minor leaks. Used in home (single-patient) or hospital/institutional (multi-patient) settings; operated by clinicians or patients/caregivers. Mask connects to positive pressure therapy devices via 15mm to 22mm swivel connector. Healthcare providers use mask to facilitate ventilation therapy; adjustment dial allows optimization of seal. Benefits include effective delivery of prescribed respiratory support for pediatric patients >10kg.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via intentional leak, total mask leak, deadspace, pressure drop, CO2 rebreathing, and testing post-reprocessing (cleaning/disinfection).
Technological Characteristics
Materials: Silicone cushion, Polyurethane Foam/Nylon frame, Polycarbonate elbow/swivel, Thermoplastic Elastomer tubing. Design: Nasal mask with integrated exhalation ports and cushion adjustment dial. Connectivity: 15-22mm swivel connector for therapy devices. Standards: ISO 17510-2, ISO 10993-1, ISO 14971. Reprocessing: High-level disinfection for multi-patient use.
Indications for Use
Indicated for patients >10 kg/22 lbs requiring non-invasive ventilation. Compatible with therapy devices; patient weight limits of the therapy device apply.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2016
Respironics, Inc. Shaylee Masilunas Regulatory Affairs Engineer 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668
Re: K151120
Trade/Device Name: Simple T Pediatric Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: March 2, 2015 Received: March 4, 2015
Dear Ms. Masilunas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Indications for Use | Expiration Date: January 31, 2017<br>See PRA Statement below. |
| 510(k) Number (if known) | |
| Device Name | Simple T Pediatric Nasal Mask |
| Indications for Use (Describe) | |
| This mask is intended to provide an interface for application of non-invasive ventilation to patients. The mask is for single patient use in the home and multi-patient use in the hospital/institutional environment. The mask is to be used on patients greater than 10 kg/22 lbs (>10 kg/22 lbs) for whom non-invasive ventilation has been prescribed. Use of this mask is limited by the indications for use of the compatible therapy device with respect to patient weight. | |
| Type of Use (Select one or both, as applicable) | |
| <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect | |
of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Psblishing Survices (JUI) 443-6740 ED
{3}------------------------------------------------
### TAB 6
# 510(K) SUMMARY
| 510(k) Owner | Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668 |
|---------------------------------------|---------------------------------------------------------------------------------------------------|
| Official Contact | Shaylee Masilunas<br>Regulatory Affairs Engineer, Patient Interface |
| Establishment Registration # | 2518422 |
| Proprietary Name | Simple T Pediatric Nasal Mask |
| Common/Usual Name | Nasal Mask |
| Classification Name /<br>Product Code | 21 CFR 868.5905, Product Code BZD – Ventilator, non-continuous<br>(respirator) |
| Predicate Device(s) | Respironics Simple T Youth Nasal Mask (K140268)<br>Respironics Profile Lite Small Child (K093416) |
| Reference Device | Respironics Trilogy 200 (K093416) |
## Indications for Use
This mask is intended to provide an interface for application of non-invasive ventilation to patients. The mask is for single patient use in the home and multi-patient use in the hospital/institutional environment. The mask is to be used on patients greater than 10 kg/22 lbs) for whom non-invasive ventilation has been prescribed. Use of this mask is limited by the indications for use of the compatible therapy device with respect to patient weight.
## Device Description
The Simple T Pediatric Nasal Mask is intended to be used with positive airway pressure devices. For use of this mask with a CPAP therapy device, the patient population is limited by the intended use of the device (e.g. > 30 kg). In other words, this mask can be used with a variety of therapy devices which may have their own limitations on patient population. If the therapy device itself is limited to patients
{4}------------------------------------------------
greater than 10 kg (e.g. 30 kg), then the use of this mask does not expand the intended use of that therapy device.
The mask provides a seal such that positive pressure from the positive pressure source is directed into the patient's nose. It is held in place with a fabric frame and an adjustable headgear. The cushion contains an adjustment dial that can be engaged to reduce minor leaks around the nose. The mask may be cleaned in the home (single-patient use) or reprocessed by the professional in the hospital/institutional environment through high-level disinfection processes (multi-patient use).
## Substantial Equivalence
The Simple T Pediatric Nasal Mask has the following similarities to the previously cleared predicate devices Respironics Simple T Youth Nasal Mask (K140268) and Respironics Profile Lite Small Child mask (K093416):
- Same operating principle
- Similar intended use
- . Similar materials
- Similar design
- Similar manufacturing process
The following modifications have been made to the previously cleared predicate devices Respironics Simple T Youth Nasal Mask (K140268) and Respironics Profile Lite Small Child mask (K093416) for the Simple T Pediatric Nasal Mask:
- . The intended use population was modified to include patients greater than 10kg/22lbs
- . An existing warning was converted to a contraindication to align with the definition of a contraindication. The language of that statement was revised to reflect standardized terminology.
- . The mask components were modified for sizing. The cushion was modified to include a leak adjustment dial.
{5}------------------------------------------------
### Table 1 – Technological characteristics comparison of the Simple T Pediatric Nasal Mask and its predicate devices
| Intended Use | Primary Predicate:<br>Device:<br>Simple T Youth<br>K140268<br>Manufacturer:<br>Respironics, Inc.<br>The Simple T Youth Nasal Mask | Secondary Predicate:<br>Device:<br>Profile Lite Small Child K093416<br>Manufacturer:<br>Respironics, Inc.<br>The Small Child Profile Lite | Subject Device:<br>Device:<br>Simple T Pediatric Nasal Mask<br>Manufacturer:<br>Respironics, Inc.<br>This mask is intended to provide an |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | is intended to provide an<br>interface for Continuous Positive<br>Airway Pressure (CPAP) or bi-<br>level therapy. | Nasal Mask and Softcap are<br>intended to provide an<br>interface when used with CPAP<br>or bi-level therapy. | interface for application of non-<br>invasive ventilation to patients. |
| Patient<br>Population | Patients 7 years or older (>40<br>lbs/18kg) | Profile Lite Small Child Mask:<br>Patients 1 year or older (>7 kg)<br>Trilogy 200 Ventilator: Patients<br>at least 5kg/11lbs | Patients greater than 10kg/22lbs<br>(>10kg/22lbs) |
| Environment of<br>Use | Home or Hospital/<br>Institutional Environment | Home or Hospital/<br>Institutional Environment | Unchanged from K140268,<br>K093416 |
| Patient Usage<br>Type | Single patient reuse or multi-<br>patient use | Single Patient Use | Unchanged from K140268 |
| Provided Sterile<br>or Non-Sterile | Non-Sterile | Non-Sterile | Unchanged from K140268,<br>K093416 |
| Anatomical Sites | Nose | Nose | Unchanged from K140268,<br>K093416 |
| Device Design | 1. Nasal cushion<br>2. Frame<br>3. Elbow with built-in exhalation<br>device<br>4. Tubing with 15mm to 22mm<br>swivel connector<br>5. Headgear | 1. Nasal cushion<br>2. Faceplate<br>3. Elbow with built-in<br>exhalation device<br>4. Tubing with 15mm to 22mm<br>swivel connector.<br>5. Headgear | Unchanged from K140268<br>1. Nasal Cushion with Dial<br>2. Frame<br>3. Elbow with built-in exhalation<br>device<br>4. Tubing with 15 mm to 22 mm<br>swivel connector.<br>5. Headgear |
| Cushion | Silicone sealing cushion that fits<br>around the nose. | Encapsulated gel cushion that<br>fits around the nose. | Similar to K140268<br>The cushion contains an<br>adjustment dial. |
| Cushion Sizes | 3 sizes | 1 size | 2 sizes |
| Frame Design | The frame stabilizes the cushion<br>on the nose and allows<br>clearance of facial features.<br>Four slots exist for insertion of<br>headgear tabs. | The faceplate and forehead<br>arm secure the cushion to the<br>face. Four slots exist for<br>insertion of headgear tabs. | Unchanged from K140268 |
| Exhalation device<br>design | No separate exhalation device is<br>required. 18 exhalation ports<br>are integrated. | No separate exhalation device<br>is required. Two exhalation<br>ports are integrated. | Unchanged from K140268 |
| Tubing | 15 mm tubing | 15 to 22 mm tubing | Unchanged from K140268 |
{6}------------------------------------------------
| | Primary Predicate: | Secondary Predicate: | Subject Device: |
|-------------------------------|---------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------|
| | Device: | Device: | Device: |
| | Simple T Youth<br>K140268 | Profile Lite Small Child K093416 | Simple T Pediatric Nasal Mask |
| | Manufacturer:<br>Respironics, Inc. | Manufacturer:<br>Respironics, Inc. | Manufacturer:<br>Respironics, Inc. |
| General<br>headgear design | Four point headgear with hook<br>and loop tabs for attaching to<br>frame. | Four point bonnet with hook<br>and loop tabs for attaching to<br>frame. | Unchanged from K140268 |
| Patient Circuit<br>Connection | 22 mm swivel connector | 22 mm swivel connector | Unchanged from K140268 and<br>K093416 |
{7}------------------------------------------------
| | Primary Predicate: | Secondary Predicate: | Subject Device: |
|-----------|-------------------------------------|----------------------------------|-------------------------------------|
| | Device: | Device: | Device: |
| | Simple T Youth | Profile Lite Small Child K093416 | Simple T Pediatric Nasal Mask |
| | K140268 | | |
| | | Manufacturer: | Manufacturer: |
| | Manufacturer: | Respironics, Inc. | Respironics, Inc. |
| Component | Respironics, Inc. | | |
| Cushion | Silicone | Gel encapsulated in urethane | Silicone |
| Frame | Polyurethane Foam, Nylon | Polycarbonate | Polyurethane Foam, Nylon |
| Elbow and | Polycarbonate | Polycarbonate | Polycarbonate |
| Swivel | | | |
| Tubing | Thermoplastic Elastomer | Ethylene Vinyl Acetate | Thermoplastic Elastomer |
| Headgear | Polyurethane Foam,<br>Nylon/Spandex | Nylon and Lycra | Polyurethane Foam,<br>Nylon/Spandex |
#### Table 2 – Material comparison for the Simple T Pediatric Nasal Mask and its predicate devices
Design verification tests were performed on the Simple T Pediatric Nasal Mask. All tests were verified to meet the required acceptance criteria. Respironics has determined that the modifications have no impact on the safety and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate devices.
## Non-Clinical Tests
Performance testing was performed before and after reprocessing treatments to verify that the device modifications did not affect the safety and effectiveness of the subject device. Performance testing included:
- Intentional Leak
- . Total Mask Leak
- . Mask Deadspace
- . Pressure Drop
- CO2 Rebreathing
- Performance Post Cleaning, Disinfection and Sterilization
- Cleaning and Disinfection Validation Testing
- Storage
The Simple T Pediatric Nasal Mask has been designed to meet the requirements of the following standards:
{8}------------------------------------------------
- ISO 17510-2 Sleep Apnoea Devices Part 2: Masks and Application Accessories ●
- . ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 14971 Medical devices – Application of risk management to medical devices
## Clinical Tests
Clinical tests were not required to demonstrate the safety and effectiveness of the Simple T Pediatric Nasal Mask. Product functionality has been adequately assessed by non-clinical tests.
## Conclusion
The performance and technological characteristics of the Simple T Pediatric Nasal Mask are substantially equivalent to those of the Simple T Youth Nasal Mask (K140268) and the Profile Lite Small Child (K093416). The differences described above do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.