The MiniMe 2 Nasal mask is intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old. The mask is designed and labeled for different durations of use: - Single use, disposable (hospital or institutional settings) . - Single patient, multi-use (home setting) . - Single patient, short-term use up to 7 days (hospital or institutional settings) .
Device Story
MiniMe 2 Nasal Mask serves as patient interface for CPAP or Bi-level positive pressure therapy. Device delivers pressurized air to pediatric patients (>2 to <12 years old). Components include headgear, shell/cushion, swivel elbow, and tubing. Features include exhalation port with multiple holes for CO2 washout and non-vented elbow configuration for circuits with integrated exhalation valves. Used in home or hospital/institutional settings; operated by clinician or patient/caregiver. Output is delivery of prescribed positive airway pressure to patient. Benefits include non-invasive respiratory support for pediatric patients. Device is disposable (single use) or multi-use depending on setting.
Clinical Evidence
Bench testing only. Comparative performance testing included exhaust flow (pressure vs. flow), pressure drop/resistance to flow, internal volume/dead space, CO2 washout (ISO 17510-2), cleaning validation, environmental testing, and mechanical drop testing. Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Nasal mask interface. Materials: Gel bladder, bladder film, shell, elbow, headgear, tubing. Biocompatibility per ISO 10993. Features: Exhalation port with multiple holes, swivel elbow, 15mm connector. Pressure range: 4-20 cm H2O. Dead space: 25-40 ml. Connectivity: None (mechanical accessory).
Indications for Use
Indicated for patients >2 years old and <12 years old requiring positive pressure therapy via CPAP or Bi-level systems. Used in home or hospital/institutional settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{1}------------------------------------------------
KI21692
NOV
510(k) Summary Page 1 of 6
| SleepNet Corporation<br>5 Merrill Industrial Drive<br>Hampton, NH 03842 | Tel - 603-758-6625<br>Fax - 603-758-6699 |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Official Contact: | Jennifer Kennedy – Director of Quality |
| Proprietary or Trade Name: | MiniMe 2 Nasal Mask |
| Common/Usual Name: | Patient interface |
| Classification Code/Name: | BZD – non-continuous ventilator (IPPB)<br>21CFR 868.5905<br>Class 2 |
| Device: | MiniMe 2 Nasal Mask |
| Predicate Devices: | K090935 – SleepNet MiniMe pediatric nasal mask<br>K060105 – ResMed Kidsta Pediatric Mask |
#### Device Description:
.
The MiniMe 2 Nasal mask is intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old. The mask is designed and labeled for different durations of use:
- Single use, disposable (hospital or institutional settings) .
- Single patient, multi-use (home setting) .
- Single patient, short-term use up to 7 days (hospital or institutional settings) .
These indications for use are identical to the predicate SleepNet MiniMe Pediatric Nasal mask, K090935.
The durations for use are identical to
- SleepNet MiniMe Pediatric Nasal mask, K090935 for
- Single patient, multi-use (home) o
- Single patient, multi-use < 7 days is a subset of the single patient, multi-use o (hospital / institutional)
- Single patient, disposable (hospital / institutional) is a shorter duration of use and a . labeling claim only.
- The design, materials, and performance are identical to the predicates and would O not be considered a new indication that raises new safety or effectiveness concerns.
5 2012
{2}------------------------------------------------
#### 510(k) Summary Page 2 of 6 5-Nov-12
### Indications for Use:
The MiniMe 2 Nasal masks are intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old. The mask is designed and labeled for different durations of use:
- . Single use, disposable (hospital or institutional settings)
- Single patient, multi-use (home setting) .
- . Single patient, short-term use up to 7 days (hospital or institutional settings)
Patient Population: Patients >2 years old and <12 years old
Environment of Use: Home or hospital / institutional environments
#### Predicate Device Comparison:
The MiniMe 2 Nasal masks are viewed as substantially equivalent to the predicate device because:
### Indications -
- . The MiniMe 2 Nasal masks are intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory for use with CPAP or Bi-level positive pressure systems intended to provide an interface for application of CPAP or bi-level therapy. Identical to SleepNet MiniMe Pediatric Nasal mask (K090935).
- . Discussion - The indications for use are identical to the predicate, SleepNet MiniMe Pediatric Nasal mask (K090935).
#### Patient Population -
- The masks are for patients > 2 yo and < 12 yo for whom positive airway pressure therapy . has been prescribed. Identical to SleepNet MiniMe (K090935).
- Discussion The population is identical to the predicate, SleepNet MiniMe Pediatric . Nasal mask (K090935).
#### Technology -
- . Identical technology to - SleepNet MiniMe Pediatric Nasal mask - K090935
- Discussion The technology, shape, design, configuration of the mask, head strap as . well as the manufacturing methods are identical to the predicate, SleepNet MiniMe Pediatric Nasal mask - K090935.
The use of multiple ports in the elbow for exhalation and CO2 washout is substantially equivalent as demonstrated by the comparative Pressure vs. Flow curves and the CO2 washout testing vs. the predicate ResMed Kidsta (K060105).
{3}------------------------------------------------
# 510(k) Summary Page 3 of 6 5-Nov-12
### Materials –
- The materials in patient contact are identical or have been tested per ISO 10993 to our . own predicate devices
- . Discussion - The materials utilized in the MiniMe 2 are commonly used in medical devices and are either identical to the predicate or have been tested per ISO 10993 and found to be acceptable for the intended use.
# Environment of Use -
- The masks are intended for use in the home or hospital/institutional environment. .
- Identical to predicate SleepNet MiniMe Pediatric Nasal mask (K090935) .
- Discussion The environments of use are identical to the predicate. .
See Table 1 for comparison.
### Non-Clinical Testing Summary:
### Comparative Performance -
We have performed comparative performance testing that included:
- Exhaust Flow (Pressure vs. Flow) .
- Pressure Drop / Resistance to Flow .
- Internal Volume / Dead space .
- CO2 washout per ISO 17510-2 .
- Cleaning validation .
- Environmental testing .
- Mechanical testing Drop Test .
Discussion - The comparative performance and specifications demonstrate that the MiniMe 2 is equivalent in performance to the predicates.
# Biocompatibility of Materials -
Materials listed in the following table are either identical to predict ate SleepNet mask or were evaluated per ISO 10993-1.
Parts which are Externally communicating, Tissue - gas pathway contact, permanent duration. Pass / fail criteria was for the each respective ISO 10993 test
Parts which are Surface communicating, Skin, permanent duration. Pass / fail criteria was for the each respective ISO 10993 test
Discussion - The listed materials are either identical to predicate SleepNet mask which have the same patient contact and duration of use or were tested per ISO 10993-1.
See Table 2 for a list of materials, patient contact and duration of use.
{4}------------------------------------------------
510(k) Summary
Table 1 Comparison to Predicates
| | MiniMe 2 Nasal Mask | SleepNet MiniMe Nasal Mask<br>K090935 | ResMed Kidsta<br>K060105 |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Attributes | | | |
| Indications for Use | Intended to provide a patient interface<br>for application of positive pressure<br>therapy. The mask is to be used as an<br>accessory to CPAP or Bi-level positive<br>pressure systems. | Intended to provide an interface of<br>CPAP or bi-level therapy to patients. | Intended to provide non-invasive<br>ventilation with RI or OSA |
| Patient Population | >2 yo to <12 yo | >2 yo to <12 yo | > 7 yo and up |
| Environment of Use | The masks are intended for use in the<br>home or hospital/institutional<br>environment. | The masks are intended for use in the<br>home or hospital/institutional<br>environment. | The masks are intended for use in the<br>home or hospital/institutional<br>environment. |
| Duration of Use | Single patient, disposable (hospital or<br>institutional settings) | Single patient, multi-use (home setting) | Single patient, multi-use (home setting) |
| | | Single patient, short-term use up to 7<br>days (hospital or institutional settings) | Single patient, short-term use up to 7<br>days (hospital or institutional settings) |
| Prescriptive | Yes | Yes | Yes |
| Cleaning | Soap and water | Soap and water | Soap and water |
| Incorporates Exhalation port<br>with multiple holes | Yes | Yes | Yes |
| Non-vented elbow<br>configuration used with circuit<br>with integrated exhalation valve | Yes | Yes | N/A |
| Delivered Pressure range | $4-20 \text{ cm H}_2\text{O}$ | $3-20 \text{ cm H}_2\text{O}$ | $>4 \text{ cm H}_2\text{O}$ |
| Attributes | MiniMe 2 Nasal Mask | SleepNet MiniMe Nasal Mask<br>K090935 | ResMed Kidsta<br>K060105 |
| Features | | | |
| Available sizes | 2 | 1 | Multiple |
| Shape | Similar triangular shape | Similar triangular shape | Similar triangular shape |
| Shell | Hard | Soft | Hard |
| Materials | Soft | This is a small mask so we compared<br>to equivalent size of the Kidsta | |
| CO2 washout profile<br>Tested per ISO 17510-2<br>(largest size tested, which is<br>worst case) | Identical to other SleepNet masks or<br>tested per ISO 10993 | | |
| ETCO2% at mask<br>(% increase) | Pressure | Pressure | Pressure |
| | 4 cm H2O<br>5.3 (6%) | 4 cm H2O<br>5.3 (6%) | 4 cm H2O<br>5.8 (14%) |
| | 5 cm H2O<br>5.2 (4%) | 5 cm H2O<br>5.2 (4%) | 5 cm H2O<br>5.7 (12%) |
| | 10 cm H2O<br>5.1 (3%)<br>Occluded<br>5.3 (7%) | 10 cm H2O<br>5.1 (3%)<br>Occluded<br>5.3 (7%) | 10 cm H2O<br>5.5 (7%)<br>Occluded<br>6.8 (34%) |
| Dead space | Small<br>25 ml | Small<br>21 ml | Small<br>81 ml |
| | Large<br>40 ml | Large | Large<br>81 ml |
| Exhaust - pressure / flow | Pressure (cmH2O) Flow (lpm) | Pressure (cmH2O) Flow (lpm) | Pressure (cmH2O) Flow (lpm) |
| | 4 17.1<br>12 33.5<br>20 43.6 | 4 19.1<br>12 34.8<br>20 45.8 | 4 19.5<br>12 36.2<br>20 48.7 |
| | | | With 15mm tubing |
| Resistance to Flow<br>(Flow and pressure) | 30 lpm - 0.22 cm H2O<br>50 lpm - 0.72 cm H2O<br>60 lpm - 1.03 cm H2O<br>100 lpm - 3.09 cm H2O | 30 lpm - 0.10 cm H2O<br>50 lpm - 0.36 cm H2O<br>60 lpm - 0.54 cm H2O<br>100 lpm - 1.71 cm H2O | 30 lpm - 0.12 cm H2O<br>50 lpm - 0.44 cm H2O<br>60 lpm - 0.67 cm H2O<br>100 lpm - 2.10 cm H2O |
| | | | |
| | | | |
| Components | Headgear<br>Shell / Cushion<br>Swivel elbow<br>Tubing | Headgear<br>Shell / Cushion<br>Swivel elbow<br>Tubing | Headgear<br>Shell / Cushion<br>Swivel elbow<br>Tubing |
| Component | Patient Contact / Duration<br>(all are permanent) | Biocompatibility | |
| Gel Bladder | Skin and Gas Pathway | Identical Sleepnet MiniMe Pediatric mask K090935 | |
| Bladder film | Skin and Gas Pathway | Identical | |
| | | Sleepnet MiniMe Pediatric mask K090935 | |
| Shell | Gas Pathway | Identical Sleepnet IQ Ventilation mask K102317 | |
| | | Sleepnet MiniMe Pediatric mask K090935 | |
| Elbow assembly | Gas Pathway | Identical Sleepnet MiniMe Pediatric mask K090935 | |
| Headgear | Skin | Identical Sleepnet IQ Ventilation mask K102317 | |
| Tubing | Gas Pathway | Cytotoxicity | |
| | | Sensitization | |
| | | Genotoxicity | |
| | | Implantation | |
| | | Intracutaneous | |
| | | Systemic Toxicity | |
| Swivel Connector | Gas Pathway | Identical Sleepnet IQ Nasal mask K993269 | |
| Tubing Connector | Gas Pathway | Identical Sleepnet IQ Nasal mask K993269 | |
| 15 mm Connector | Gas Pathway | Identical Sleepnet IQ Nasal mask K021534 | |
{5}------------------------------------------------
510(k) Summary
{6}------------------------------------------------
# 510(k) Summary Page 6 of 6 5-Nov-12
# Table 2
# Materials in the Patient Contact
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 5, 2012
SleepNet Corporation C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
Re: K121692
Trade/Device Name: MiniMe 2 Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: September 18, 2012 Received: September 19, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
#### Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Kwame O. Ulme
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
K121692
Device Name:
MiniMe 2 Nasal Mask
Indications for Use:
The MiniMe 2 Nasal mask is intended to provide a patient interface for application of positive pressure therapy. The mask is to be used as an accessory to CPAP or Bi-level positive pressure systems. It is intended for patients >2 years old and <12 years old.
The mask is designed and labeled for different durations of use:
- Single use, disposable (hospital or institutional settings) .
- Single patient, multi-use (home setting) .
- Single patient, short-term use up to 7 days (hospital or institutional settings) .
| Prescription Use XX | or | Over-the-counter use<br>_ |
|-----------------------------|----|---------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hm
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number;
Page 24 of 72
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.