The SleepNet Mojo and Veraseal masks are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 3 cm H2O. The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. Veraseal - Disposable single use (hospital/institutional) Single patient, multi-use up to 7 days (hospital/institutional) Mojo - Single patient, multi-use (home or hospital/institutional)
Device Story
Mojo and Veraseal are full-face masks for delivery of positive airway pressure (CPAP/bi-level) to adult patients. Devices interface between pressure generator and patient airway; function via mask shell, cushion, and exhaust elbow. Modifications from predicate include updated exhaust elbow design and varied durations of use: Veraseal (disposable single-use or single-patient multi-use up to 7 days in hospital); Mojo (single-patient multi-use for home/hospital). Operated by clinicians or patients; masks channel airflow to patient. Benefit includes non-invasive respiratory support. Performance validated via bench testing of dead space, exhaust flow, pressure drop, anti-asphyxia valve function, and CO2 washout.
Clinical Evidence
Bench testing only. No clinical data. Testing included dead space, exhaust flow, pressure drop, anti-asphyxia valve pressure, CO2 washout (ISO 17510-2), biocompatibility (ISO 10993), cleaning validation, and storage/aging. CO2 washout results at 3, 5, 10 cm H2O and occluded states were compared to predicates, showing similar performance.
Technological Characteristics
Full face mask with headgear, shell, cushion, and swivel exhaust elbow. Materials tested per ISO 10993. Connectivity: interfaces with standard CPAP/bi-level tubing. Sterilization: not specified (disposable/multi-use cleaning protocols provided). Technology is identical to predicate Mojo mask.
Indications for Use
Indicated for adult patients (>30kg) requiring positive airway pressure therapy (CPAP or bi-level) at or above 3 cm H2O. Used in home or hospital/institutional settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
JUL 2 7 2012
#### 4120463 510(k) Summary Page I of 6 6-Jul-12
| SleepNet Corporation<br>5 Merrill Industrial Drive<br>Hampton, NH 03842 | Tel - 603-758-6625<br>Fax - 603-758-6699 |
|-------------------------------------------------------------------------|------------------------------------------|
|-------------------------------------------------------------------------|------------------------------------------|
| Official Contact: | Jennifer Kennedy - Director of Quality |
|----------------------------|------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | Mojo Full Face Mask<br>Veraseal Full Face mask |
| Common/Usual Name: | Patient interface |
| Classification Code/Name: | BZD - non-continuous ventilator (IPPB)<br>CFR 868.5905<br>Class II |
| Device: | Mojo Full Face Mask<br>Veraseal Full Face mask |
| Predicate Devices: | K060273 - SleepNet Mojo Full Face mask<br>K063122 - ResMed Mirage Quattro Full Face mask |
| Device Description: | |
The SleepNet Mojo and Veraseal Full Face mask are modifications of the SleepNet Mojo Full Face mask cleared under K060273.
The modifications to the cleared Mojo Full Face mask are:
- different durations of use
- o single use, disposable (Hospital/Institutional) Veraseal
- o single patient, multi-use up to 7 days (Hospital/Institutional) Veraseal
- o single patient, multi-use (Home or Hospital/Institutional) MoJo
- a change in materials .
- updated design of the exhaust elbow used with the mask
As a result of these modifications we have established two (2) product tradenames to help distinguish them.
Veraseal model has 2 durations of use - single use, disposable or single patient, multi-use up to 7 days in the hospital.
Mojo model is single patient, multi-use (typically more than 7 days) for use in the home or hospital.
### Indications for Use:
●
The SleepNet Mojo and Veraseal masks are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 3 cm H2O.
The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.
{1}------------------------------------------------
# 510(k) Summary Page 2 of 6 6-Jul-12
Veraseal -
Disposable single use (hospital/institutional)
Single patient, multi-use up to 7 days (hospital/institutional)
Mojo -
Single patient, multi-use (home or hospital/institutional)
Patient Population: For adults (>30 kg)
Environment of Use: Home or hospital / institutional environments
The SleepNet Veraseal and Mojo Full face mask are viewed as substantially equivalent to the predicates device because:
### Indications -
- The Veraseal and Mojo masks are intended to be used with positive airway pressure . devices, such as CPAP or bi-level operating at or above 3 cm H2O.
The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.
- Identical to ResMed Mirage Quattro (K063122) with pressure minimum identical to . SleepNet Mojo (K060273)
### Patient Population -
- The masks are to be used on adult patients (>30kg) for whom positive airway pressure . therapy has been prescribed.
- Identical to SleepNet MoJo (K060273) and ResMed Mirage Quattro (K063122) .
### Technology -
- Identical technology to SleepNet Mojo Full Face mask (K060273) .
#### Materials -
- The materials in patient contact are identical to predicate devices or have been tested in . accordance with ISO 10993.
#### Environment of Use -
- The masks are intended for use in the home or hospital/institutional environment .
- Identical to predicates SleepNet MoJo (K060273) and ResMed Mirage Quattro . (K063122)
### Duration of Use and Cleaning Method
- The duration of use and cleaning methods for single patient Less 7 days, single patient, . multi-use, disposable are identical to the predicate - SleepNet MoJo (K060273)
### Differences -
- There are no differences between the predicates and the proposed modified devices which . would raise any new safety or risks and thus can be found to be substantially equivalent.
{2}------------------------------------------------
510(k) Summary Page 3 of 6 6-Jul-12
# Comparative Performance
We have performed comparative performance testing pre- and post- cleaning or conditioning that included:
- Dead space .
- Exhaust Flow ●
- Pressure drop ●
- Anti-asphyxia valve (AAV) pressure testing .
- Exhaust Elbow Change Performance .
- CO2 washout per ISO 17510-2 .
- ISO 10993 for biocompatibility .
- Cleaning validation ●
- Storage and aging .
The results demonstrated that the devices are substantially equivalent.
{3}------------------------------------------------
510(k) Summary
Table of Comnarison to Predicates
| ante of Compal Boll | | | | Attributes | Modified<br>Mojo and Veraseal<br>Full Face Mask | Mojo Full Face Mask<br>K060273 | ResMed Mirage Quattro<br>K063122 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------|
| ttributes | | | ResMed Mirage Quat K063122 | Features | | | |
| | Modified Modernets Marka Morith Mojio masks are The United Manika Movich positi as The United Maska Movith positi as announds and research are are ar Sboray pressure de | Mojo Full Face M. K060273 | | Available sizes | 3 | 3 | 4 |
| | | | | Shape | Similar | Similar | Similar |
| ications for | | | e Mirage Quattro channels airflow ninvasively to a patient from a possive way pressure device such as comments of the many pressure airway pressure airway pressure airway pre | Shell | Veraseal<br>Rigid for single patient models | Soft only | |
| | | | | Materials ISO 10993<br>CO₂ washout profile<br>Tested per ISO 17510-2 | Mojo<br>Soft<br>Yes | Yes | N/A |
| | | | | | Pressure | Pressure | Not tested as this predicate for duration of<br>use, environment of use, and cleaning<br>method only |
| | | | | | ETCO₂% at mask<br>(% increase) | ETCO₂% at mask<br>(% increase) | |
| | | | evel system | | 3 cm H₂O<br>6.1 (17%) | 3 cm H₂O<br>5.9 (15%) | |
| | | | | | 5 cm H₂O<br>5.9 (14%) | 5 cm H₂O<br>5.8 (14%) | |
| | | | | | 10 cm H₂O<br>5.8 (12%) | 10 cm H₂O<br>5.6 (9%) | |
| | The masks are to be used on adult patients (230kg) for whom positive airway pressure therapy has been prescribed | The Sin Bir More More Maria Massal Massal Massa Short All Properti Andress Marin Marian Managing Marian Managiri Marian Manakar Manakar Manakar Manakari Marin Marian Manakari | The Mirage Quattro is to be used by adult patients (> 66 lbs. /> 30 Kg) for whom positive airway pressure has been | | Occluded<br>6.3 (21%) | Occluded<br>6.0 (17%) | |
| | | | | Deadspace | Mojo | | Varies according to cushion size |
| | | | prescribed | Small | 160 ml | 200 ml | |
| tient Populati | Adult (≥30 kg) The masks are intended for use in the home or hospital/institutional environment. | dult (>30 kg | dult (>30 kg | Medium | 185 ml | 245 ml | |
| vironment of | | | | Large | 205 ml | 295 ml | 242 ml |
| | | The masks are intended for use in the nome or hospital/institutional | he masks are intended for use ome or hospital/institutional | Exhaust - pressure / flow | Mojo | | |
| | | environment | environment | | Pressure (cmH₂O) Flow (lpm) | Pressure (cmH₂O) Flow (lpm) | Pressure (cmH₂O) Flow (lpm) |
| | Yes | Yes | Y es | | 3<br>21.20 | 3<br>32.00 | 4<br>22 |
| escriptive uration of U | | | | | 10<br>37.20 | 10<br>43.00 | 8<br>32 |
| | | | | | 20<br>53.10 | 20<br>55.00 | 12<br>41 |
| | | | | | Veraseal | | 20<br>54 |
| | Veraseal Disposable, single patient use (hospital) Single patient, multi-use up to 7 days (hospital) | Single patient, multi-use (home | Single patient, multi-use but does not the length of re-use (home / hospital) | | Pressure (cmH₂O) Flow (lpm) | | |
| | | hospital | | | 3<br>21.10 | | |
| | | | | | 10<br>36.40 | | |
| | patient, multi-use (hon<br>្រទេសចរនា ជា ជាចំនួនដែន ស្ថិតនៅក្នុងស្រុក និង ស្រីបន្ត ស្រុកស្រីប ស | ingle patient, multi-use (hom | | | 20<br>53.00 | | |
| | | hospital) | | | | | |
| eaning metho | Soap and water | oap and wate | | | | | |
| | Isopropyl alcoho Yes | sopropyl alcoh | OPA | | | | |
| corporates an Exhaust elbo | | Yes | Yes | | | | |
:
Page 16
:
{4}------------------------------------------------
510(k) Summary
:
Page 17
T
{5}------------------------------------------------
510(k) Summary
| Attributes | Modified Mojo and Veraseal Full Face Mask | Mojo Full Face Mask<br>K060273 | ResMed Mirage Quattro<br>K063122 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------|
| Pressure AAV<br>Opening / closing | Mojo<br>Opening - 1.10 cm H₂O<br>Closing - 1.60 cm H₂O<br><br>Veraseal<br>Opening - 1.29 cm H₂O<br>Closing - 1.45 cm H₂O | Opening - 0.65 cm H₂O<br>Closing - 1.38 cm H₂O | Not disclosed |
| Resistance to Flow | Pass / Fail criteria is <3 cm H₂O<br><br>Mojo<br>50 lpm - 0.40 cm H₂O<br>100 lpm - 0.65 cm H₂O<br><br>Veraseal<br>50 lpm - 0.49 cm H₂O<br>100 lpm - 0.89 cm H₂O | Pass/Fail criteria is < 3 cm H₂O<br><br>50 lpm - 0.38 cm H₂O<br>100 lpm - 0.80 cm H₂O | 50 lpm - 0.5 cm H₂O<br>100 lpm - 1.6 cm H₂O |
| Components | Headgear<br>Shell / Cushion<br>Swivel elbow | Headgear<br>Shell / Cushion<br>Swivel elbow | Headgear<br>Shell / Cushion<br>Swivel elbow |
Page 18
ﺷﺮ ﺍﻟ
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 2 7 2012
Sleepnet Corporation C/O Mr. Paul Dryden President Promedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134-2958
Re: K120463
Trade/Device Name: Veraseal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: July 6, 2012 Received: July 9, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
# Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's'requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm_for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{8}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
K120463
Device Name:
Mojo Mask Veraseal Mask
## Indications for Use:
The SleepNet Mojo and Veraseal masks are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 3 cm H2O.
The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed.
Veraseal -
Disposable single use (hospital/institutional)
Single patient, multi-use up to 7 days (hospital/institutional)
Mojo -
Single patient, multi-use (home or hospital/institutional)
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulten
(Division Sign-Off) Division of Anestheslology, General Hospital Intection Control, Dental Devices
510(k) Number: K120463
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.