K100749 · Dymedso, Inc. · BYI · Aug 31, 2010 · Anesthesiology
Device Facts
Record ID
K100749
Device Name
FREQUENCER MODEL V2X
Applicant
Dymedso, Inc.
Product Code
BYI · Anesthesiology
Decision Date
Aug 31, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Frequencer™ v2x provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. This device is intended to be a component of chest physiotherapy by providing a convenient method of external thorax manipulation. It is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma, muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis and thoracic wall defects. Indications for this form of therapy are described in the Clinical Practice Guidelines for Postural Drainage Therapy of the American Association for Respiratory Care (AARC) published in 1991. This particular device provides a gentler, less painful form of therapy than the traditional "clapping" method of postural drainage therapy, allowing it to be used on patients who cannot be treated by clapping. The Frequencer™ v2x is only intended for use in hospitals and clinics.
Device Story
Frequencer™ v2x is a powered percussor for airway clearance therapy. Device uses electro-acoustical transducer (Power Head) placed on patient's chest to induce oscillatory sound waves (20-65 Hz) via frequency generator. Vibrations loosen mucus deposits to promote bronchial drainage. Used in hospital/clinic settings by healthcare providers. Provides gentler alternative to manual 'clapping' (postural drainage). Output is mechanical vibration; provider monitors patient response to therapy. Benefits include improved mucus clearance for patients with respiratory conditions.
Clinical Evidence
No new clinical data provided for v2x. Relies on clinical data from original Frequencer™ demonstrating comparable sputum weights to Conventional Chest Physiotherapy (CCPT) in cystic fibrosis subjects. Bench testing showed v2x produces slightly lower maximum force than predicate, both lower than CCPT. In vitro capillary tube testing showed increased mucus flow rates compared to no treatment.
Technological Characteristics
Powered percussor; electro-acoustical transducer; 20-65 Hz frequency range. Housing/adapters: polycarbonate/ABS blend. User interface: touch screen. Electrical safety: IEC 60601-1; EMC: IEC 60601-1-2. Multi-patient use supported by disposable filters.
Indications for Use
Indicated for patients with respiratory ailments involving defective mucociliary clearance, including cystic fibrosis, chronic bronchitis, bronchiectasis, ciliary dyskinesia, asthma, muscular dystrophy, neuromuscular degenerative disorders, postoperative atelectasis, and thoracic wall defects.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{0}------------------------------------------------
KY00749
## l.510(k) Summary
This summary of safety and effectiveness is being submitted in accordance AUG 3 1 2010 with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Date of Submission:<br>Submitter:<br>Address: | August 31, 2010<br>DYMEDSO, Inc.<br>901A, Boul. Du Curé Boivin<br>Boisbriand, Québec J7G 2S8<br>CANADA |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Establishment Registration #: | 3004365906 |
| Contact: | Yvon Robert, President |
| Phone: (450) 437-9601 | Fax: (450) 437-2063 |
| Proprietary Device Name: | Frequencer™ v2x |
| Common Name of Device: | Airway Clearance Device |
| Classification Name: | Percussor, Powered-Electric |
| Device Class: | 2 |
| Product Code: | BYI |
Predicate Device: K063645 - Frequencer™ 1001
### Description:
The Frequencer™ v2x provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. It induces oscillatory sound waves in the chest by means of an electro-acoustical transducer (hereafter referred to as the "Power Head") which is placed externally on the patient's chest. The Power Head is connected to a frequency generator which is capable of producing frequencies between 20 and 65 Hz. The Power Head induces sound waves in the patient's chest which are effective in loosening mucus deposits. This model is intended for use in hospital and clinic settings only and disposable filters are available for multi-patient use.
### Indications for Use:
The Frequencer™ v2x provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. This device is intended to be a component of chest physiotherapy by providing a convenient method of external thorax manipulation.
{1}------------------------------------------------
# Dymedso Frequencer v2x™ Powered Percussor
It is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma, muscular dystrophy, neuromuscular degenerative disorders, postoperative atelectasis and thoracic wall defects. Indications for this form of therapy are described in the Clinical Practice Guidelines for Postural Drainage Therapy of the American Association for Respiratory Care (AARC) published in 1991. This particular device provides a gentler, less painful form of therapy than the traditional "clapping" method of postural drainage therapy, allowing it to be used on patients who cannot be treated by clapping.
The Frequencer™ v2x is only intended for use in hospitals and clinics.
### Comparison of Technological Characteristics:
The Frequencer™ v2x is identical to the predicate device in terms of indications for use, contra-indications, acoustic operating principle and basic electronics. The polymer material used for the housing and adapters has been changed from a polyurethane to a polycarbonate/ABS blend. In addition, the user interface no longer consists of touch pad buttons and an LCD display but is now a touch screen. The device is only intended for use in hospitals or clinics as opposed to the predicate, which was intended for home or residential environments.
### Performance Testing
The device was tested in accordance with IEC 60601-1 for electrical safety and tested in accordance with IEC 60601-1-2 for electromagnetic radiation. with acceptance criteria consistent with a Type A device for non-domestic environments such as hospitals or clinics. Software verification and validation testing has also been performed on the device.
Performance bench testing was conducted to determine the maximum force applied by the device in comparison to the predicate device. The testing demonstrated a similar, though slightly lower maximum applied force in comparison to the predicate. Both devices still provide considerably lower forces than applied during Conventional Chest Physiotherapy (CCPT). The lower forces produced by the Frequencer™ may be able to provide a gentler treatment compared to CCPT. Clinical testing conducted on the original Frequencer™ have demonstrated comparable sputum weights produced after treatment with the Frequencer™ in comparison to Conventional Chest Physiotherapy (CCPT) in subjects with Cystic Fibrosis (CF). In vitro flow rate measurements of mucus through a capillary tube also demonstrated higher flow rates of mucus when using the Frequencer compared to no treatment.
{2}------------------------------------------------
# Dymedso Frequencer v2x™ Powered Percussor
## Conclusions:
The proposed Frequencer™ v2x has the identical indications, operating principle and electronics as the predicate device Frequencer™ 1001. The only differences are the polymer housing material, user interface, and intended environments. In addition, disposable filters are available for use in multi-patient settings. The performance testing provides validation of new features of this device and demonstrates similar performance characteristics to the predicate device. Thus, the proposed Frequencer™ v2x is substantially equivalent to the predicate device.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human figures connected at the shoulders, with their arms raised in a gesture of support or unity. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Dymedso, Incorporated C/O Ms. Jean Bigoney Springfield Metallurgical Services, Incorporated 127 Main Street Springfield, Vermont 05156-0826
AUG 3 1 2010
Re: K100749
Trade/Device Name: Frenquencer" v2x Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: II Product Code: BYI Dated: August 2, 2010 Received: August 5, 2010
Dear Ms. Bigoney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
### Page 2- Ms. Bigoney
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
hh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K100749
510(k) Number (if known):
Device Name:
Frequencer™ v2x
Indications for Use:
The Frequencer™ v2x provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. This device is intended to be a component of chest physiotherapy by providing a convenient method of external thorax manipulation.
lt is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma, muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis and thoracic wall defects. Indications for this form of therapy are described in the Clinical Practice Guidelines for Postural Drainage Therapy of the American Association for Respiratory Care (AARC) published in 1991. This particular device provides a gentler, less painful form of therapy than the traditional "clapping" method of postural drainage therapy, allowing it to be used on patients who cannot be treated by clapping.
The Frequencer™ v2x is only intended for use in hospitals and clinics.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-theCounter Use (Part 21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device EvaJuation (ODE)
f Device Evaluation (ODE)
L. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K/00749 Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.