MEDINET INCENTIVE SPIROMETER – RESPIPROGRAM is intended as an inspiratory deep breathing positive exerciser. MEDINET INCENTIVE SPIROMETER - RESPIPROGRAM is intended for single-patient, single use in a hospital or home care setting.
Device Story
Inspiratory flow exerciser; stimulates deep breathing. Patient inhales through mouthpiece connected to three chambers containing colored balls (red, yellow, green); balls indicate approximate flow rates of 600, 900, and 1200 ml/second. Device provides visual feedback on inhalation capacity; no specific measurement function. Used in hospital or home care settings; operated by patient under physician prescription. Single-patient, single-use disposable device. Benefits patient by encouraging deep inspiration to improve lung function.
Clinical Evidence
Bench testing only. Comparative performance testing against predicate device included flow accuracy tests (600, 900, 1200 cc/sec) and drop tests. Accuracy verified within ±5% across various humidity conditions and product ages. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).
Indicated for patients requiring inspiratory exercise to stimulate inhalation under physician prescription. Suitable for hospital or home care settings.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which consists of a stylized caduceus symbol with three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 23, 2017
Medinet SRL % Maurizio Pantaleoni C.E.O. Isemed Srl Via A.bonetti 3/A Imola, 40026 ITALY
Re: K160837
Trade/Device Name: Medinet Incentive Spirometers - Respiprogram Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: II Product Code: BWF Dated: January 23, 2017 Received: January 25, 2017
Dear Maurizio Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K160837
Device Name
MEDINET INCENTIVE SPIROMETER - RESPIPROGRAM
Indications for Use (Describe)
MEDINET INCENTIVE SPIROMETER – RESPIPROGRAM is intended as an inspiratory deep breathing positive exerciser.
MEDINET INCENTIVE SPIROMETER - RESPIPROGRAM is intended for single use in a hospital or home care setting.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
| | <span> <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
# 510(k) Summary for the MEDINET INCENTIVE SPIROMETER
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
| 1. | General Information<br>Submitter: | MEDINET s.r.l. is located at:<br>Via Londonio, 12<br>MILANO<br>ITALY |
|----|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Maurizio Pantaleoni<br>Isemed s.r.l.<br>Via A. Altobelli Bonetti 3/A<br>40026 Imola (BO)<br>Mob.phone: +39 3484435155<br>Telephone: +39 0542 683803<br>Fax: +39 0542 698456<br>Email: regulatory@isemed.eu |
Summary Preparation Date:
February 13, 2016
#### 2. Names
| Device Name: | MEDINET INCENTIVE SPIROMETER- RESPIPROGRAM |
|----------------------|--------------------------------------------|
| Classification Name: | Incentive spirometer |
| Product Code: | BWF |
| Regulation number: | 868.5690 |
| CLASS | II |
### 3. Predicate Devices
The MEDINET INCENTIVE SPIROMETER RESPIPROGRAM is substantially equivalent to the following device:
| Applicant | Device name | 510(k)<br>Number | Product<br>code |
|---------------------------------|------------------------------|------------------|-----------------|
| BESMED Health Business<br>Corp. | TriBall incentive spirometer | K133873 | BWF |
{4}------------------------------------------------
## 4. Device Description
The RESPIPROGRAM is an inspiratory device used as breathing exerciser; RESPIPROGRAM is a FLOW exerciser
The following model is available:
- RESPIPROGRAM – code 15 10 002 600
The RESPIPROGRAM is intended as an inspiratory deep breathing positive exerciser. It is intended for single-patient, single use both in a hospital or for home care therapy.
RESPIPROGRAM is a single patient, single use, disposable, non-sterile device.
RESPIPROGRAM is a single use device that allows the patient to indicate visually the capacity of inhaling, even though RESPIPROGRAM doesn't have a specific measuring function. The device is a stimulator for inspiration; therefore it is indicated in all situations for when it is necessary to stimulate the patient's inhalation, under the prescription of a physician. The patient inspires through a mouthpiece that is connected to 3 chambers that have 3 balls of different colors (red, yellow and green) each one indicating a different approximate flow rate (respectively 600, 900 and 1200 ml per second).
### 5. Indications for Use
MEDINET INCENTIVE SPIROMETER – RESPIPROGRAM is intended as an inspiratory deep breathing positive exerciser.
MEDINET INCENTIVE SPIROMETER - RESPIPROGRAM is intended for single-patient, single use in a hospital or home care setting.
## 6. Performance Data
The following comparative tests have been performed for the MEDINET incentive spirometer RESPIPROGRAM in comparison with its predicate device, to support the substantial equivalence
FLOW ACCURACY TESTS between
- RESPIPROGRAM and BESMED TRIBALL ●
The tests have been performed using new and aged samples and considering various humidity conditions. The results showed that all the samples tested, met the pre-determined acceptance criteria defined in the performance test protocol.
DROP TEST between:
- RESPIPROGRAM and BESMED TRIBALL
The results showed that the samples tested met the pre-determined acceptance criteria without any breakage or crack.
The following biocompatibility tests were performed:
- · Cytotoxicity
- · Irritation
- · Sensitization
{5}------------------------------------------------
## 7. Substantial Equivalence Discussion & Summary of Technological Characteristics
The indications for use of the MEDINET Incentive Spirometer can be considered substantially equivalent to the identified predicate devices.
The design features, the technological characteristics and the performances are equivalent to that of the predicate devices as shown in the table below:
| | Subject Device | Predicate Devices | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MEDINET INCENTIVE SPIROMETER<br>RESPIPROGRAM | BESMED TriBall - K133873 | Substantially<br>Equivalent |
| Regulation<br>Number: | 868.5690 | 868.5690 | Same |
| Classification | II | II | Same |
| Product Code | BWF | BWF | Same |
| Indications for<br>use | MEDINET INCENTIVE SPIROMETER<br>RESPIPROGRAM is intended as an<br>inspiratory deep breathing positive<br>exerciser.<br>MEDINET INCENTIVE SPIROMETER is<br>intended for single-patient, single use in a<br>hospital or home care setting. | The Besmed Incentive Spirometer is<br>intended as an inspiratory deep breathing<br>positive exerciser.<br>Intended for single-patient, multi-use in a<br>hospital or home care setting. | Equivalent. The only<br>difference is that<br>Medinet Incentive<br>spirometer<br>Respiprogram is single<br>use, while predicate<br>device is multi-use |
| Users | | | |
| Prescription<br>devices | YES | YES | Same |
| Population | Patients requiring inspiratory exercise | Patients requiring inspiratory exercise | Same |
| Single Patient,<br>Single Use | YES | No, Single patient, Multi Use | Equivalent. The only<br>difference is that<br>Medinet Incentive<br>spirometer<br>Respiprogram is single<br>use, while predicate<br>device is multi-use |
| Environment of<br>use | Home care settings and hospitals | Home care settings and hospitals | Same |
| Design features | | | |
| Basic<br>components | Housing<br>3 balls<br>Tubing<br>Mouthpiece | Housing<br>3 balls<br>Tubing<br>Mouthpiece | Same |
| Patient Interface | Mouthpiece | Mouthpiece | Same |
| Available<br>Models | One model with three balls | One model with 3 balls | Same |
| Biocompatibility | | | |
| Cytotoxicity | Conforming to ISO 10993 testing<br>(ISO 10993-5) | Conforming to ISO 10993 testing | Same |
| Sensitization | Conforming to ISO 10993 testing<br>(ISO 10993-10) | Conforming to ISO 10993 testing | Same |
| Intracutaneous<br>irritation | Conforming to ISO 10993 testing<br>(ISO 10993-10) | Conforming to ISO 10993 testing | Same |
| Performances | | | |
| Flow | 600/900/1200 cc/sec | 600/900/1200 cc/sec | Same |
| Accuracy | +/- 5% | +/- 5% | Same |
| Resistance | Drop Test results: no breakage or crack at the<br>end of the test | Drop Test | Same |
| | Subject Device | Predicate Devices | |
| | MEDINET INCENTIVE SPIROMETER<br>RESPIPROGRAM | BESMED TriBall - K133873 | Substantially<br>Equivalent |
| Condition of<br>testing and<br>results | ±5% accuracy is maintained in the<br>following test conditions:<br>- Age Testing (products recently<br>manufactured and expired products)<br>- Low Humidity Testing (products recently<br>manufactured and expired products tested<br>with humidity < 40%)<br>- High Humidity Testing (products recently<br>manufactured and expired products tested<br>with humidity > 80%).<br>Low humidity test results (cc/sec):<br>Test at 600 cc/sec: 604,6±5,5<br>Test at 900 cc/sec: 925,2±3,28<br>Test at 1200 cc/sec:1222,8±10,0<br>High humidity test results (cc/sec)<br>Test at 600 cc/sec: 593,5±4,9<br>Test at 900 cc/sec: 887,3±3,7<br>Test at 1200 cc/sec:1177,3±7,5 | Accuracy specification ±5% verified in the<br>following test conditions:<br>Age Testing<br>High / Low Humidity conditions<br>Inspiratory rate / Volume accuracy<br>declared :<br>600 - 0,5%<br>900 - 0,3%<br>1200 - 0,1%<br>Test results using the MEDINET test<br>method - Low humidity (cc/sec):<br>Test at 600 cc/sec: 624,0±10,1<br>Test at 900 cc/sec: 886,8±10,2<br>Test at 1200 cc/sec:1210,5±12,89 | Same |
{6}------------------------------------------------
## Summary of Technological Characteristics:
Both subject and predicate device are incentive spirometers. The fundamental technology of the subject and predicate device are based on the following same technological elements:
- Housing containing 3 balls
- Mouthpiece as a patient interface ●
- Flow rate of 600, 900 and 1200 ml/second
- Non Sterile Device
The following technological difference exists between the subject and the predicate device:
- Respiprogram is single use, while predicate device is multi-use.
### 8. Conclusion
Based on technological characteristics (intended use, dimensions and features) and performance data (comparative tests and biocompatibility tests) included in this submission, the MEDINET INCENTIVE SPIROMETER RESPIPROGRAM has been shown to be substantially equivalent to the listed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.