The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the percutaneous entry (Seldinger) technique via the cricothyroid membrane. The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the surgical technique via the cricothyroid membrane. The Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing either the percutaneous entry (Seldinger) technique or the surgical technique via the cricothyroid membrane.
Device Story
Emergency airway access kits for patients where endotracheal intubation is not feasible. Sets include airway catheter, loading dilator, wire guide, and needles; surgical sets include additional tools like tracheal hooks and dilator forceps. Percutaneous sets use Seldinger technique; surgical sets use surgical incision. Used in emergency clinical settings by trained medical professionals. Catheters are 22 French, 9 cm length, 5 mm ID. Provides immediate airway establishment to prevent patient hypoxia. MR Conditional status confirmed via testing.
Clinical Evidence
Bench testing only. Evaluations included cuff durability (zero-time and 3-year accelerated aging), dilator shaft tensile strength, and MR compatibility testing per ASTM F2052 (displacement force), ASTM F2213 (torque), and ASTM F2119 (image artifact). All acceptance criteria were met.
Technological Characteristics
Materials: Polyvinylchloride (catheter/dilator), Polyethylene (cuff), Stainless steel (wire guide/needle), Fluorinated ethylene propylene (catheter needle sheath). Dimensions: 22 Fr, 9 cm length, 5 mm ID catheter. MR Conditional. Sterilization: EtO. No software or active energy source.
Indications for Use
Indicated for patients requiring emergency airway access when endotracheal intubation cannot be performed. Access is achieved via the cricothyroid membrane using percutaneous (Seldinger) or surgical techniques.
Regulatory Classification
Identification
An emergency airway needle is a device intended to puncture a patient's cricothyroid membrane to provide an emergency airway during upper airway obstruction.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 17, 2017
Cook Incorporated Kotei Aoki Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K160200
Trade/Device Name: Melker Cuffed Emergency Cricothyrotomy Catheter Set -Percutaneous, Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical, Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set Regulation Number: 21 CFR 868.5090 Regulation Name: Emergency Airway Needle Regulatory Class: II Product Code: BWC, OGP, JOH Dated: January 13, 2017 Received: January 17, 2017
Dear Kotei Aoki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160200
#### Device Name
Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set
#### Indications for Use (Describe)
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the percutaneous entry (Seldinger) technique via the cricothyroid membrane.
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical is intended to establish emergency arrway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the surgical technique via the cricothyroid membrane.
The Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set is intended to establish emergency arrway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing either the percutaneous entry (Seldinger) technique or the surgical technique via the cricothyroid membrane.
| Type of Use (Select one or both, as applicable) | | | | | | | | | | | | | | | | | | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | のバイト 【 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 | | | | | | | | | | | | | | | | | |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The red background is a rectangle with a diagonal cut on the bottom right corner.
# 510(k) Summary
## Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set 21 CFR §868.5090 Date Prepared: 15 February 2017
#### Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|------------------------------------------|
| Contact: | Kotei Aoki |
| Applicant Address: | 750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x102630 |
| Contact Fax Number: | (812) 332-0281 |
#### Device Information
| Trade Name: | Melker Cuffed Emergency Cricothyrotomy Catheter Set – Percutaneous<br>Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical<br>Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Emergency airway needle |
| Classification Name: | Emergency airway needle; BWC (21 CFR §868.5090)<br>Cricothyrotomy Kit; OGP<br>Tracheostomy tube and tube cuff; JOH |
## Predicate Device:
- K010016, Melker Cuffed Emergency Cricothyrotomy Catheter
## Device Description:
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous, Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical, and the Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set consist of an airway catheter, a loading dilator, a wire guide, a catheter needle, and an introducer needle. In addition to these components, standard preparation components, such as a syringe, scalpel, dilator forceps, gauze, drape, tracheostomy tape, and tracheal hook are included. The sets consist of any combination of components appropriate for either percutaneous or surgical placement of a cricothyrotomy tube, or both.
The curved cricothyrotomy catheter is manufactured as a 22 French catheter with a length of 9 centimeters and an inner diameter of 5 millimeters. The loading dilators are either 11
{4}------------------------------------------------
Image /page/4/Picture/11 description: The image is the logo for Cook Medical. The logo is set on a red background. The word "COOK" is in white, bold, sans-serif letters. Below the word "COOK" is the word "MEDICAL" in white, sans-serif letters. The word "MEDICAL" is set on a red background that is slightly darker than the background for the word "COOK".
centimeters or 12 centimeters long. The stainless steel wire guide has a 0.038-inch diameter and a length of 40 centimeters. The catheter needle is a two-part needle comprised of a small hubbed catheter sheath covering a hubbed needle. The sheath of the catheter needle is manufactured from fluorinated ethylene propylene tubing. The catheter needle is manufactured using an 18 gage cannula with a length of 7 centimeters. The introducer needle is also manufactured from 18 gage stainless steel and has a length of either 5 centimeters or 7 centimeters.
# Intended Use:
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the percutaneous entry (Seldinger) technique via the cricothyroid membrane.
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the surgical technique via the cricothyroid membrane.
The Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing either the percutaneous entry (Seldinger) technique or the surgical technique via the cricothyroid membrane.
The intended uses are same as that of the predicate device, the Melker Cuffed Emergency Cricothyrotomy Catheter (K010016).
# Comparison to Predicate:
The Melker Cuffed Emergency Cricothyrotomy Catheter Set -Percutaneous, Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical, and the Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set are substantially equivalent to the predicate device, the Melker Cuffed Emergency Cricothyrotomy Catheter (K010016), in that these devices have the same intended use, technological characteristics, materials of construction, and are similar in design. The subject device sets offer both percutaneous (Seldinger) and surgical techniques of establishing the airway, and contain sets of accessory components to support both surgical and percutaneous methods of performing emergency cricothyrotomy, as circumstances require.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font. Below "COOK" is a smaller red rectangle with the word "MEDICAL" in white, sans-serif font.
Table 1, Table 2, and Table 3 provide the detailed comparison of each subject device to the predicate device.
| Catheter Set - Percutaneous | | |
|-----------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | Subject Device |
| | Melker Cuffed Emergency<br>Cricothyrotomy Catheter<br>K010016 | Melker Cuffed Emergency<br>Cricothyrotomy Catheter Set –<br>Percutaneous |
| Regulation Number<br>Product Code | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>JOH | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>OGP<br>3) 21 CFR §868.5800<br>JOH |
| Classification Name | 1) Emergency airway needle<br>2) Tracheostomy tube and tube cuff | Identical |
| Class | II | Identical |
| Intended Use | Used for emergency airway access<br>when endotracheal intubation<br>cannot be performed | Intended to establish emergency<br>airway access when endotracheal<br>intubation cannot be performed.<br>Airway access is achieved utilizing<br>the percutaneous entry (Seldinger)<br>technique via the cricothyroid<br>membrane.[1] |
| Product Illustrations | Image: Predicate Device Product Illustrations | Image: Subject Device Product Illustrations |
Table 1 Substantial Equivalence Comparison Chart for Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous
[1] Both the subject device and the predicate device establish the airway by the percutaneous access. The subject device specifies the access method clearly.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background. The word "MEDICAL" is smaller than the word "COOK".
| Table 1 Substantial Equivalence Comparison Chart for Melker Cuffed Emergency Cricothyrotomy | | |
|---------------------------------------------------------------------------------------------|--|--|
| Catheter Set - Percutaneous (Continued) | | |
| | | Predicate Device | Subject Device |
|-----------------------------------------|-------------------|---------------------------------------------------------------|--------------------------------------------------------------------------|
| | | Melker Cuffed Emergency<br>Cricothyrotomy Catheter<br>K010016 | Melker Cuffed Emergency<br>Cricothyrotomy Catheter Set -<br>Percutaneous |
| | | | |
| Airway<br>Catheter | Material | Polyvinylchloride with<br>Polyethylene Cuff | Identical |
| | Inner Diameter | 5 mm | Identical |
| | Outer Diameter | 22 French | Identical |
| | Length | 8 cm | 9 cm[2] |
| Loading<br>Dilator | Material | Polyvinylchloride | Identical |
| | Outer Diameter | 14 French | Identical |
| | Length | 14 cm | 12 cm[2] |
| | Tip Shape | Tapered | Identical |
| Wire Guide | Material | Stainless Steel | Identical |
| | Diameter | 0.038 in | Identical |
| | Length | 40 cm | Identical |
| Catheter<br>Needle (if<br>applicable) | Catheter Material | Fluorinated Ethylene Propylene | Identical |
| | Needle Material | Stainless Steel | Identical |
| | Gage | 18 gage | Identical |
| | Length | 7 cm | Identical |
| | | | |
| Introducer<br>Needle (if<br>applicable) | Hub Material | Polycarbonate | Identical |
| | Needle Material | Stainless Steel | Identical |
| | Gage | 18 gage | Identical |
| | Length | 7 cm | 5, 7 cm[2] |
| | Needle Coating | Silicone | Identical |
| Accessory Components | | Syringe, Scalpel | Syringe, Scalpel, Tracheostomy<br>Tape[3] |
| MR Compatibility | | Not indicated | MR Conditional[4] |
| Packaging | | Tyvek/polyethylene film pouch | Identical |
| Sterilization | | EtO | Identical |
[2] The dimensional variations are within the tolerances on the specifications of the predicate device.
Furthermore, the materials of construction in the respective components are identical.
[3] The Tracheostomy Tape is a legally marketed component and is an accessory for use with cricothyrotomy procedures.
[4] MR testing was not previously done for the predicate device and has been performed, as per FDA Guidance, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. Testing to support the MR Conditional status for this device has been provided.
In conclusion, the differences noted in the table above do not raise different questions of safety or effectiveness of the subject device, as compared to the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Cook Medical. The logo is white text on a red background. The word "COOK" is in large, bold letters at the top. Below that, the word "MEDICAL" is in smaller, bold letters.
| Table 2 Substantial Equivalence Comparison Chart for Melker Cuffed Emergency Cricothyrotomy | | | |
|---------------------------------------------------------------------------------------------|--|--|--|
| Catheter Set - Surgical | | | |
| | Predicate Device | Subject Device |
|-----------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Melker Cuffed Emergency | Melker Cuffed Emergency |
| | Cricothyrotomy Catheter<br>K010016 | Cricothyrotomy Catheter Set -<br>Surgical |
| Regulation Number<br>Product Code | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>JOH | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>OGP<br>3) 21 CFR §868.5800<br>JOH |
| Classification Name | 1) Emergency airway needle<br>2) Tracheostomy tube and tube cuff | Identical |
| Class | II | Identical |
| Intended Use | Used for emergency airway access<br>when endotracheal intubation<br>cannot be performed | Intended to establish emergency<br>airway access when endotracheal<br>intubation cannot be performed.<br>Airway access is achieved utilizing<br>the surgical technique via the<br>cricothyroid membrane.[1] |
| Product Illustrations | Image: Predicate Device Product Illustrations | Image: Subject Device Product Illustrations |
| Method of airway access | Percutaneous | Surgical[1] |
[1]The subject device establishes the airway by the surgical access. The appropriate accessory components are provided with the subject device to establish the airway surgically. The Airway Catheter and the Loading Dilator, used in the surgical access, function in the same manner as those used in the percutaneous access. The Airway Catheter is as capable of being placed via the surgical access as via the percutaneous access.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol next to it. Below the word "COOK" is the word "MEDICAL" in white, set against a red background. The red background extends behind the word "COOK".
| Table 2 Substantial Equivalence Comparison Chart for Melker Cuffed Emergency Cricothyrotomy | | | |
|---------------------------------------------------------------------------------------------|-------------------|---------------------------------------------|------------------------------------------------------------------------------|
| Catheter Set - Surgical (Continued) | | | |
| | | Predicate Device | Subject Device |
| | | Melker Cuffed Emergency | Melker Cuffed Emergency |
| | | Cricothyrotomy Catheter | Cricothyrotomy Catheter Set - |
| | | K010016 | Surgical |
| Airway<br>catheter | Material | Polyvinylchloride with<br>Polyethylene Cuff | Identical |
| | Inner Diameter | 5 mm | Identical |
| | Outer Diameter | 22 French | Identical |
| | Length | 8 cm | 9 cm[2] |
| Loading<br>dilator | Material | Polyvinylchloride | Identical |
| | Outer Diameter | 14 French | Identical |
| | Length | 14 cm | 11 cm[2] |
| | Tip Shape | Tapered | Blunt[3] |
| Wire Guide | Material | Stainless Steel | N/A |
| | Diameter | 0.038 in | N/A |
| | Length | 40 cm | N/A |
| Catheter<br>Needle (if<br>applicable) | Catheter Material | Fluorinated Ethylene Propylene | N/A |
| | Needle Material | Stainless Steel | N/A |
| | Gage | 18 gage | N/A |
| | Length | 7 cm | N/A |
| Introducer<br>Needle (if<br>applicable) | Hub Material | Polycarbonate | N/A |
| | Needle Material | Stainless Steel | N/A |
| | Gage | 18 gage | N/A |
| | Length | 7 cm | N/A |
| | Needle Coating | Silicone | N/A |
| Accessory Components | | | Syringe, Scalpel, Tracheostomy<br>Tape, Tracheal Hook, Dilator<br>Forceps[4] |
| MR Compatibility | | Not indicated | MR Conditional[5] |
| Packaging | | Tyvek/polyethylene film pouch | Identical |
| Sterilization | | EtO | Identical |
[2] The dimensional variations are within the specifications of the predicate device. Furthermore, the materials of construction in the respective components are identical.
[3] As the surgical access to the airway does not utilize the wire guide, it also does not require a tapered Loading Dilator. The dilator tip is left blunt to help mitigate the risk of perforating the posterior tracheal wall (in the percutaneous access, the same risk is mitigated as the wire guide diverts the protruding tapered dilator tip).
- [4] The Tracheostomy Tape, the Tracheal Hook, and the Dilator Forceps are all legally marketed components and are accessories for use with the cricothyrotomy procedures using the surgical access.
- [5] MR testing was not previously done for the predicate device and has been performed, as per FDA Guidance, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. Testing to support the MR Conditional status for this device has been provided.
In conclusion, the differences noted in the table above do not raise different questions of safety or effectiveness of the subject device, as compared to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background.
Table 3 Substantial Equivalence Comparison Chart for Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set
| | Predicate Device | Subject Device |
|-----------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Melker Cuffed Emergency<br>Cricothyrotomy Catheter<br>K010016 | Melker Universal Cuffed<br>Emergency Cricothyrotomy<br>Catheter Set |
| Regulation Number<br>Product Code | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>JOH | 1) 21 CFR §868.5090<br>BWC<br>2) 21 CFR §868.5800<br>OGP<br>3) 21 CFR §868.5800<br>JOH |
| Classification Name | 1) Emergency airway needle<br>2) Tracheostomy tube and tube cuff | Identical |
| Class | II | Identical |
| Intended Use | Used for emergency airway access<br>when endotracheal intubation<br>cannot be performed | Intended to establish emergency<br>airway access when endotracheal<br>intubation cannot be performed.<br>Airway access is achieved utilizing<br>either the percutaneous entry<br>(Seldinger) technique or the<br>surgical technique via the<br>cricothyroid membrane.[1] |
| Product Illustrations | Image: Predicate Device Product Illustration | Image: Subject Device Product Illustration |
| Method of airway access | Percutaneous | Percutaneous, Surgical[1] |
[1] Both the subject device and the predicate device establish the airway by the percutaneous access. The subject device also allows for the surgical method of establishing the airway. The appropriate accessory components are provided with the subject device to establish the airway both percutaneously and surgically. The Airway Catheter and the Loading Dilator, used in the surgical access, function in the same manner as those used in the percutaneous access. The Airway Catheter is as capable of being placed via the surgical access as via the percutaneous access. Each Loading Dilator is easily identified by the tip shape.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif letters on a red background that is a slightly darker shade than the background for the word "COOK".
Table 3 Substantial Equivalence Comparison Chart for Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set (Continued)
| | | Predicate Device | Subject Device | |
|----------------------|-------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------|
| | | Melker Cuffed Emergency<br>Cricothyrotomy Catheter<br>K010016 | Melker Universal Cuffed<br>Emergency Cricothyrotomy<br>Catheter Set | |
| Airway<br>catheter | Material | Polyvinylchloride with<br>Polyethylene Cuff | Identical | |
| | Inner Diameter | 5 mm | Identical | |
| | Outer Diameter | 22 French | Identical | |
| | Length | 8 cm | 9 cm[2] | |
| Loading<br>dilator | Material | Polyvinylchloride | Identical | |
| | Outer Diameter | 14 French | Identical | |
| | Length | 14 cm | 11, 12 cm[2] | |
| | | Tip Shape | Tapered | Tapered, Blunt[3] |
| | | Material | Stainless Steel | Identical |
| Wire Guide | Diameter | 0.038 in | Identical | |
| | Length | 40 cm | Identical | |
| Catheter<br>Needle | Catheter Material | Fluorinated Ethylene Propylene | Identical | |
| | Needle Material | Stainless Steel | Identical | |
| | | Gage | 18 gage | Identical |
| | | Length | 7 cm | Identical |
| | | Hub Material | Polycarbonate | Identical |
| Introducer<br>Needle | Needle Material | Stainless Steel | Identical | |
| | Gage | 18 gage | Identical | |
| | Length | 7 cm | Identical | |
| | | Needle Coating | Silicone | Identical |
| Accessory Components | | | Syringe, Scalpel, Tracheostomy<br>Tape, Tracheal Hook, Dilator<br>Forceps, Gauze, Drape[4] | |
| MR Compatibility | | Not indicated | MR Conditional[5] | |
| Packaging | | Tyvek/polyethylene film pouch | Identical | |
| Sterilization | | EtO | Identical | |
[2] The dimensional variations are within the specifications of the predicate device. Furthermore, the materials of construction in the respective components are identical.
[3] As the subject device allows both the percutaneous access and the surgical access to the airway, it provides all the appropriate accessory components required for both methods of airway access. Therefore, the subject device provides both the tapered tip dilator and the blunt tip dilator.
[4] The Tracheostomy Tape, the Tracheal Hook, the Dilator Forceps, the Gauze, and the Drape are all legally marketed components and are accessories for use with the cricothyrotomy procedures using either method of airway access.
[5] MR testing was not previously done for the predicate device and has been performed, as per FDA Guidance, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. Testing to support the MR Conditional status for this device has been provided.
In conclusion, the differences noted in the table above do not raise different questions of safety or effectiveness of the subject device, as compared to the predicate device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in large, bold letters, and the word "MEDICAL" is in smaller letters below it.
## Technological Characteristics:
The following tests have been conducted to ensure reliable design and performance under the specified design requirements. These tests include:
- . Evaluation of Durability of the Cuff (Zero-Time, Three-Year Accelerated Aging) – The acceptance criterion was that the cuff must remain fully intact and without any leaks following insertion. The predetermined acceptance criterion was met.
- . Tensile Test of the Shaft of the Dilator (Three-Year Real-Time) – The acceptance criterion was that the dilator shaft must withstand a minimum force at break of 15 N. The predetermined acceptance criterion was met.
- Magnetically Induced Displacement Force Per ASTM F2052, the displacement . force shall be less than or equal to the worst case displacement force due to the earth's gravity. The predetermined acceptance criterion was met.
- Magnetically Induced Torque Per ASTM F2213, the magnetically induced . torque shall be less than or equal to the worst case torque due to earth's gravity. The predetermined acceptance criterion was met.
- Radiofrequency Induced Heating The electrical conductivity of the Melker . Emergency Cricothyromtomy Catheter must be < 1 S/m. The predetermined acceptance criterion was met.
- Magnetic Resonance Image Artifact Image artifact was assessed according to . ASTM F2119.
The results of these tests provide reasonable assurance that the device is as safe and effective as the predicate device for its intended use.
## Conclusion:
The results of these tests support a conclusion that the subject devices, Melker Cuffed Emergency Cricothyrotomy Catheter Set -Percutaneous, Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical, and the Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set, met the design input requirements based on the intended use and support the conclusion that these devices do not raise different questions of safety or effectiveness as compared to the predicate device. Therefore, the subject devices are substantially equivalent to the predicate device, the Melker Cuffed Emergency Cricothyrotomy Catheter (Cook Incorporated, K010016).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.