K141353 · Teleflexmedical, Inc. · BTT · Aug 1, 2014 · Anesthesiology
Device Facts
Record ID
K141353
Device Name
HUDSON RCI AQUAPAK HUMIDIFIER ADAPTOR
Applicant
Teleflexmedical, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Aug 1, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.
Device Story
Device adds humidity to dry medical gas for patient inhalation; prevents respiratory mucosa drying/irritation. System consists of prefilled sterile water/saline reservoir (190ml-650ml) and adaptor component. Adaptor connects to flow-metered gas source (50 psi oxygen); gas bubbles through internal micro-diffuser to produce water vapor. Features audible pressure relief valve to indicate gas flow restriction. Used in hospitals, sub-acute/long-term care facilities, and home environments. Operated by clinicians or patients. Single-use device; provides therapeutic humidity output (at least 10mg H2O/L).
Clinical Evidence
Bench testing only. Tests included packaging integrity (ISTA 2A), ink adherence (ASTM F2252), cannula pop-off, bottle burst, occlusion, spitting, and humidification output (mg/L). Results verify performance is substantially equivalent to predicates.
Indicated for any patient connected to ventilation devices requiring humidification of breathing gases. No known contraindications.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2014
Teleflex Medical, Inc. Mr. James A. Cochie Sr. Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K141353
Trade/Device Name: Hudson RCI® AquaPak® Prefilled Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: June 30, 2014 Received: July 2, 2014
Dear Mr. Cochie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Cochie
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141353
Device Name Hudson RCI® AquaPak® Prefilled Humidifier
#### Indications for Use (Describe) The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/12 description: The image shows the FDA logo. The logo is a stylized version of the letters "FDA". The "F" is on the left, the "D" is in the middle, and the "A" is on the right. The logo is in a light gray color.
# Anya C. Harry -S 2014.07.31 12:36:04 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 24 - 510(k) Summary
## 510(k) SUMMARY
# Hudson RCI® AquaPak® Prefilled Humidifiers
## A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8083 919-433-4996 Fax:
#### B. Contact Person
James Cochie Sr. Regulatory Affairs Specialist
#### C. Date Prepared
July 31, 2014
#### D. Device Name
Trade Name: Hudson RCI® AquaPak® Prefilled Humidifier
Common Name: Humidifier
Classification Name: Humidifier, Respiratory Gas (Direct Patient Interface), CFR -868.5450, Class II
#### E. Device Description
Hudson RCI® AquaPak® Prefilled Humidifiers add water vapor (molecular H₂O) to a dry medical gas that is suspended in the gas to be inhaled by the patient.
The Hudson RCI® AquaPak® Prefilled Humidifiers are designed to add humidity to inspired gas, helping to control the drying and irritation of the respiratory mucosa. The humidifiers feature a micro-diffuser that produces smaller bubbles and greater surface agitation, allowing for a quiet operation and therapeutic humidity output.
Prefilled sterile reservoirs for AquaPak® Humidifiers come in four sizes; 190ml, 340ml, 540ml and 650ml. Each reservoir must be used with a suitable adaptor component, which connects the system to a flow-metered gas source and provides humidifier functionality.
{4}------------------------------------------------
#### Section 24 – 510(k) Summary
Adaptor types provided with Hudson RC1® AquaPak® Prefilled Humidifiers feature an audible pressure relief valve, which indicates a restriction of gas flow to the user.
## F. Indications for Use
The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.
#### G. Target Population
This device is intended for use on any patient connected to ventilation devices
#### H. Environments of Use
This device is intended for hospital, sub-acute facilities, long-term care facilities and in a home care environment.
This product is single use only.
#### I. Contraindications
There are no known contraindications.
#### J. Comparative Characteristics
The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the predicate device:
| Comparative<br>Characteristics | Proposed Device:<br>Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier | Predicate Device:<br>Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier | Predicate Device:<br>CareFusion Airlife®<br>Prefilled Humidifier |
|--------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Teleflex Medical,<br>Inc. | Teleflex Medical,<br>Inc. | CareFusion |
| 510(k) Number | TBD | K780562 – Prefilled<br>Humidifier<br>K833974 –<br>Humidifier Adaptor | K853146 |
| Indications for Use | The Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier adds<br>humidity to a<br>patient's breathing<br>gases. | The Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier adds<br>humidity to a<br>patient's breathing<br>gases. | The CareFusion<br>Airlife® Prefilled<br>Humidifiers provide<br>a cost-effective<br>solution for sterile<br>water inhalation. |
| | | | (www.caefusion.com) |
| Principle of Operation | Gas bubbles through built-in diffuser in reservoir | Gas bubbles through built-in diffuser in reservoir | Gas bubbles through reservoir |
| Gas source | 50 Psi oxygen regulated via a flow meter | 50 Psi oxygen regulated via a flow meter | 50 Psi oxygen regulated via a flow meter |
| Technology and Features | Flow rate: 1.5 – 10 LPM<br>Audible notification of occlusion: 5 – 30 psi<br>Humidity output: At least 10mg H2O/L | Flow rate: 1.5 – 10 LPM<br>Audible notification of occlusion: 5 – 30 psi<br>Humidity output: At least 10mg H2O/L | Flow rate: 4 LPM for audible notification<br>Audible notification of occlusion: Not specified<br>Humidity output: Not specified |
| Sterilization | Adaptor – Non-sterile<br>Sterile Water/Saline Reservoirs - Reverse osmosis, distillation, and aseptic fill | Adaptor – Gamma Irradiation<br>Sterile Water/Saline Reservoirs - Reverse osmosis, distillation, and aseptic fill | Adaptor – Non-sterile<br>Sterile Water/Saline Reservoirs – Steam or dry heat |
| Single Use | Yes | Yes | Adaptor - Single Patient Use<br>Sterile Water/Saline Reservoirs – Single Use |
| Shelf Life | Adaptors – N/A Sterile Water/Saline – 2 years from date of manufacture | Adaptors – 5 years from date of manufacture<br>Sterile Water/Saline – 2 years from date of manufacture | Adaptors - N/A Sterile Water/Saline – 2 years from date of manufacture |
| Packaging | Adaptor Top web: Coated High Density Polyethylene (HDPE)<br>Adaptor Bottom web: Surlyn Amcor C-060 | Adaptor Top web: Tyvek<br>Adaptor Bottom film: Surlyn Amcor C-060<br>Bottle – Sealed | Adaptor - Heat sealed polybag<br>Bottle - Sealed polypropylene |
| | Bottle - Sealed<br>polypropylene | polypropylene | |
Teleflex Medical, Inc.
{5}------------------------------------------------
# Section 24 – 510(k) Summary
{6}------------------------------------------------
## Section 24 - 510(k) Summary
## K. Non-clinical Comparative Performance Testing
Bench testing has been performed to verify that the performance of the proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the predicate device, and that the Hudson RC1® AquaPak® Prefilled Humidifier will perform as intended.
| Test<br>Performed | Reference to Standard (if<br>applicable) | Principle of Test |
|---------------------------------------------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging<br>Integrity Tests | ISTA 2A | Determines the integrity of the packaged<br>device against known shipping tests |
| Ink Adherence<br>Tests | ASTM F2252 | Determines ink adherence properties of<br>the printed film used in packaging the<br>device |
| Cannula Pop-<br>Off, Bottle<br>Burst,<br>Occlusion Test | N/A | Determines strength of the tubing<br>connector, the reservoir does not burst<br>when nares are occluded and the relieve<br>valve performs as expected to relieve<br>pressure |
| Spitting Test | N/A | Determines the expected outcome that<br>the humidifier will not spit or flood the<br>tubing at the maximum flow rate of 10<br>LPM |
| Humidification<br>Output Test | N/A | Determines the rate of humidification at<br>which the liquid is humidified and<br>emitted in milligrams per liter |
#### L. Substantial Equivalence
The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent in intended use, design, performance and principles of operation to the identified predicate devices cleared under 510(k) K780562 and K833974. The differences between the proposed Hudson RCI® AquaPak® Prefilled Humidifier and the predicate devices are minor (sterile to non-sterile packaging) and raise no new issues of safety and efficacy. The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.