The Humidifier Heater Base PMH7000 is intended to add moisture and to warm the breathing gases for administration to a patient. The Humidifier Heater Base, PMH7000 is intended for use on adult at hospital and home.
Device Story
The PMH7000 is a respiratory humidifier heater base used in hospital and home settings to warm and humidify breathing gases for adult patients. It operates by heating distilled water in a humidification chamber placed on the heater base. The device uses a dual servo control system with platinum resistance RTD temperature probes to monitor and regulate gas temperature at both the chamber and patient sides. It supports manual, IPPV, and NPPV modes, with settings adjusted via a front touch panel. The device provides visual LED alarms for temperature deviations or probe/heater wire abnormalities. Compared to the predicate, it utilizes a switching power supply, reducing total weight. It is operated by healthcare providers or patients/caregivers to mitigate airway dryness during ventilation.
Clinical Evidence
No clinical data provided. Substantial equivalence was established through bench testing, comparison of technological characteristics, and reproducibility studies.
Technological Characteristics
Heater base with dual servo control; platinum resistance RTD sensors; switching power supply; front touch panel interface; LED alarm indicators. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, ISO 8185, ISO 10993-1, and ISO 14971.
Indications for Use
Indicated for adult patients in hospital and home settings requiring respiratory gas humidification and warming via a breathing circuit.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with a stylized letter 'P' at the top and the word 'MEDICO' at the bottom. The letter 'P' is large and ornate, with thick and thin lines creating a sense of elegance. The word 'MEDICO' is in a simple, sans-serif font, and it is placed directly below the left side of the 'P'.
DEC 1 8 2008
Section 5.
#### 510(k) Summary
SUBMITTER
Pacific Medico Co., Ltd. 2-6-4 Hongo Bunkyo-ku, Tokyo Japan 113-0033 Tel: +81 3-3818-6127 Fax: +81 3-3818-6128
OFFICIAL CONTACT
Yukio Tomisawa
{1}------------------------------------------------
K080149
p. 2 of 9
Image /page/1/Picture/1 description: The image shows the text "510(k) Summary" in a large, bold font. Above the text is a logo that includes the word "MEDICO" in a smaller font. The logo also features a stylized letter "P" in a decorative font. The image appears to be a title or heading for a document related to a 510(k) submission, which is a type of premarket submission to the FDA.
CLASSIFICATION REFERENCE
21 CFR Subpart C 868.5450
PRODUCT CODE TRADE/DEVICE NAME
втт Humidifier Heater Base, PMH7000 K080149
#### CLASSIFICATION
Respiratory Gas Humidifier
NAME
CLASSIFICATION
SUBMISSION DATE
December 16, 2008
Class II
{2}------------------------------------------------
K080149
p. 3 of 9
Image /page/2/Picture/1 description: The image shows a logo with a stylized letter "P" above the word "MEDICO". Below the logo, the text "Description of Device:" is printed in a bold font. The logo and text are black and white, and the image appears to be a scan or photocopy.
The Humidifier Heater Base, PMH7000is a humidifier incorporating with ventilator and provides respiratory humidification through patient breathing circuit for adult at hospital and home.
Humidifier Heater Base, PMH7000 was designed to comply with applicable portions of the following standards.
| IEC 60601-1 | Medical Electrical Equipment - General Requirements for Safety<br>Requirements for Medical Electrical System |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Medical electrical Equipment-Part 1: General Requirements for<br>Safety-2. Collateral Standard: Electromagnetic Compatibility -<br>Requirements and Tests |
| IEC 60601-1-4 | Medical Electric Equipment Part 1: General Requirements for<br>Safety 4. Collateral Standard:<br>Programmable Electrical Medical Systems |
| ISO 8185 | Humidifiers for Medical Use-General Requirements for<br>Humidification Systems |
| ISO 10993-1 | Biological evaluation of medical devices Part 1: Evaluation and<br>Testing |
| ISO 14971 | Medical Devices-Application of Risk Management to Medical<br>Devices |
#### Intended Use
The Humidifier Heater Base PMH7000 is intended to add moisture and to warm the breathing gases for administration to a patient.
The Humidifier Heater Base, PMH7000 is intended for use on adult at hospital and home.
| Prescription Use : | x | Over-The-Counter Use : | |
|-----------------------------|---|------------------------|--|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) | |
{3}------------------------------------------------
K080149
p. 4 of a
Substantial Equivalence to Predicate Device(s) (Part 1)
The comparison table is provided as a summary of the technological characteristics relative to the predicate devices. This is to demonstrate that Pacific Medico PMH7000 Humidifier Heater Base for respiratory therapy (heater base) has no difference from the predicate device that would adversely affect product safety and effectiveness.
| Comparison | Pegasus Research | Pacific Medico |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | PMH5000 | PMH7000 |
| | K020700 | |
| Intended Use | Respiratory Humidification | Respiratory Humidification |
| General | (1) Humidification:<br>Heat distilled water in the<br>humidification chamber placed on<br>the heater base.<br>(2) Heater Plate:<br>Using heater plate to heat distilled<br>water.<br>(3) Humidified Gas:<br>Humidified gas generated in the<br>chamber led to patient's breathing<br>circuit through inspiratory tube.<br>(4) Inhaling humidified gas:<br>Patient inhale ventilator supplied<br>gas together with humidified gas.<br>(5) Flow per minute:<br>5 liters | (1) Humidification:<br>Heat distilled water in the<br>humidification chamber placed on<br>the heater base.<br>(2) Heater Plate:<br>Using heater plate to heat distilled<br>water.<br>(3) Humidified Gas:<br>Humidified gas generated in the<br>chamber led to patient's<br>breathing circuit through<br>inspiratory tube.<br>(4) Inhaling humidified gas:<br>Patient inhale ventilator supplied<br>gas together with humidified gas.<br>(4) Flow per minute<br>5 liters |
| Mechanism<br>(Setting) Range | (1) Temperature - Heater Plate<br>100°C Thermostat regulated<br>(Normal Operation) | (1) Temperature - Heater Plate<br>100°C Thermostat regulated<br>(Normal Operation) |
| Sensor<br>Technology | 115°C Thermostat limited | 115°C Thermostat limited |
Table 1 Technological Comparison between PMH5000 (predicate device) and PMH7000
{4}------------------------------------------------
K080149
p 5 of 9
| | (AC power disabled) | (AC power disabled) |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Population | Adult | Adult |
| Power Source | AC115V | AC115V |
| Method of | Patient mask through patient | Patient mask through patient |
| Connection to<br>The Patient | breathing circuit | breathing circuit |
| Re-use | (1) Heater Base main unit:<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other applicable<br>regulations.<br>(2) Heater Wire - Inspiratory<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other applicable<br>regulations.<br>(3) Heater Wire - Expiratory<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other applicable<br>regulations.<br>(4) Temperature Probe<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other applicable<br>regulations. | (1) Heater Base main unit:<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other<br>applicable regulations.<br>(2) Heater Wire - Inspiratory<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other<br>applicable regulations.<br>(3) Heater Wire - Expiratory<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other<br>applicable regulations.<br>(4) Temperature Probe<br>Multiple uses with proper care and<br>maintenance in accordance with<br>manual and with all other<br>applicable regulations. |
From the above technological comparison, there is no difference in the technology applied in both PMH5000 and PMH7000 which may which may affect the safety and effectiveness of the subjective devices.
{5}------------------------------------------------
Image /page/5/Picture/8 description: The image shows a stylized letter "Q" in black, with a decorative flourish on the upper left. To the right of the "Q", there is a vertical arrangement of letters that appear to spell out "MEXICO" in a rotated orientation. The font used for "MEXICO" is a simple, sans-serif typeface, contrasting with the ornate style of the "Q".
## S ubstantial Equivalence to Predicate Device
### (Part 2)
The comparison table is provided as a summary of the technological on the predicate devices. This is to
demonstrate that Pacific Medico PMH7000 has no difference from the pr adversely affect product safety and effectiveness.
| Table 2 Discussions on Substantial Equivalence | | | | Temperature<br>Probe | Temperature<br>Control | Control Panel | Setting Function | Alarm | Humidity Control | Population | Power Source | Method of<br>Connection to |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------|------------------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------|---------------|---------------------------------|
| Comparison<br>Parameter | Pegasus Research<br>PMH5000<br>K020700 | Pacific Medico<br>PMH7000 | Discussion of<br>Differences and<br>Similarities | Platinum Resistance RTD | Dual Servo Control | Front Panel | Manual Mode:<br>Temperature can be set<br>manually. | All alarms : displayed by LED | Using heater wire | Adult | AC 115V | Using patient breathing circuit |
| Intended Use | The Humidifier Heater Base<br>PMH5000 is intended to add<br>moisture and to warm the<br>breathing gases for<br>administration to a patient.<br>The humidifier is intended for<br>use with flows of 5 liters per<br>Minute or more through the<br>humidifier. | The Humidifier Heater Base PMH7000 is<br>intended to add moisture and to warm the<br>breathing gases for administration to a patient.<br>The Humidifier Heater Base, PMH7000 is<br>intended for use on adult at hospital and home. | No difference | Platinum Resistance RTD | Dual Servo Control | Front Panel | Manual Mode: same as in PMH5000<br>IPPV Mode: Patient side & Chamber side<br>temperature can be pre-set.<br>Patient side: 40°C<br>Chamber side: 37°C<br>NPPV Mode: Patient side & Chamber side<br>temperature can be pre-set.<br>Patient side: 34°C<br>Chamber side: 37°C<br>Both IPPV and NPPV mode are controlled on<br>front touch panel. | All alarms: displayed by LED at the illustrated<br>portion on the front touch panel.<br>* Refer below note for alarm details | Using heater wire | Adult | AC 115V | Using patient breathing circuit |
| Temperature<br>Setting | Patient Side : 30 to 40 °C<br>Chamber Side: 23 to 34°C | Patient Side : 30 to 40 °C<br>Chamber Side: 23 to 34°C | No difference | No difference | No Difference | No difference | Differences do<br>not affect safety<br>or effectiveness | Differences do<br>not affect safety<br>or effectiveness | No difference | No difference | No difference | No difference |
F
2080149
p. 6 of 9
{6}------------------------------------------------
Image /page/6/Picture/3 description: The image shows a black and white logo. The logo features a stylized letter "Q" with a flourish at the bottom. To the right of the "Q", the word "MEDICO" is written vertically, with the letters stacked on top of each other.
KOBONA
KOBOTY
{7}------------------------------------------------
Marrier
| The Patient | | | |
|---------------|---------------------------------|--------------------------------------|-------------------|
| Re-use | Multiple Use | Multiple Use | No difference |
| Heater Wire | Controlled by an internal | Controlled by Switching Power Supply | Differences do |
| Control Power | Controlled by an<br>transformer | Total weight reduced almost half. | not affect safety |
| Supply | | | or effectiveness |
# NOTE: * Details of alarm display on PMH7000 front touch panel at the illustrated porti
Chamber side temperature over-decreasing: color of LED becomes Orange Patient side temperature over-decreasing: color of LED becomes Orange Chamber side temperature over-increasing; color of LED becomes Red Abnormal temperature of Thermo-probe: color of LED becomes Red Patient side temperature over-increasing: color of LED becomes Re Abnormal temperature of Heater Wire: color of LED becomes Red
K080149
p. 8 of 9
{8}------------------------------------------------
K080149
f 7 of 9
Image /page/8/Picture/1 description: The image shows a stylized letter P in a bold, serif font. The letter is large and dominates the image. Below the letter, in a smaller font size, is the word "MEDICO". The overall impression is that of a logo or emblem, possibly for a medical-related entity.
#### Conclusion:
- 1) Based on the results of the comparison and reproducibility studies described in this 510(k) submission, it is concluded that the Humidifier Heater Base, PMH7000 is as safe and effective (therefore substantially equivalent) as the predicate device as an aid for mitigation of dryness of airway.
- 2) Under the above technological comparison, discussion on substantial equivalence, it is demonstrated that the subjective device is as safe and effective, performs as well as the predicate device.
- 3) It is determined that the Performance Testing Clinical is not required because of the above comparison.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, which is the department's emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### DEC 1 8 2008
Mr. Yoshio Toyama Manager of Regulatory Affairs Pacific Medico Company, Limited 2-6-4 Hongo, Bunkyo-ku Tokyo JAPAN 113-0033
Re: K080149
Trade/Device Name: Humidifier Heater Base: PMH7000 Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: November 26, 2008 Received: December 1, 2008
Dear Mr. Toyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
Page 2 - Mr. Toyama
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sautge Y. Michels OMD.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### Indications for Use
K080149 510(k) Number : Humidifier Heater Base: PMH7000 Device Name :
#### Indication for Use:
The Humidifier Heater Base PMH7000 is intended to add moisture and to warm the breathing gases for administration to a patient.
The Humidifier Heater Base, PMH7000 is intended for use on adult at hospital and home.
Prescription Use : × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use : (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Smitty J. Michie Dmd
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 长 080149
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.