The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations.
Device Story
Continuous flow inhalation gas anesthesia system; delivers anesthetic vapor; provides automatic and manual ventilation modes. Incorporates O2, CO2, N2O, and anesthetic agent concentration monitoring (Desflurane, Isoflurane, Enflurane, Sevoflurane, Halothane). Main unit includes anesthetic ventilator and flow meter monitor assembly; detachable breathing system. Features 'Fresh Gas Flow Optimizer' to display recommended fresh gas values to assist clinicians. Used in health care facilities by licensed clinicians. Output displayed on monitor for clinician review to guide anesthetic delivery and ventilation management. Benefits include optimized gas usage and patient ventilation support.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-13, ISO 80601-2-55, and IEC 62304. Software verified and validated per FDA guidance.
Technological Characteristics
Continuous flow gas anesthesia system. Includes anesthetic ventilator, flow meter monitor, and detachable breathing circuit. Sensors: Paramagnetic O2 sensor, CO2/agent gas module. Connectivity: Integrated gas monitoring. Software: IEC 62304 compliant. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-13, ISO 80601-2-55.
Indications for Use
Indicated for adult and pediatric patients requiring general inhalation anesthesia and ventilation maintenance within a health care facility.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2015
Shenzhen Mindray Bio-Medical Electronics Co. LTD Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South High-tech Industrial Park, Nanshan 518057. P.R China
Re: K151954
Trade/Device Name: A7 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: Class II Product Code: BSZ, CCK, CBS, CBR, NHO, CBQ,NHQ,NHP, CCL, KDP Dated: September 23, 2015 Received: September 24, 2015
Dear Mr. Bai,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director. Division of Anesthesiology, Respiratory General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151954
Device Name A7 Anesthesia System
Indications for Use (Describe)
The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations.
Type of Use (Select one or both, as applicable):
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the A7 Anesthesia System is provided below.
| Device Common Name: | Gas-Machine, Anesthesia |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | A7 Anesthesia System |
| Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 |
| Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | July 15, 2015 |
| Panel: | Anesthesiology |
| Classification Regulation<br>and Product Code: | Primary:<br>868.5160- BSZ Anesthesia Gas Machine Class II<br>Secondary:<br>868.1400 - CCK - Carbon Dioxide Gas Analyzer<br>868.1500 – NHO/CBQ/NHQ/NHP - Enflurane gas analyzer<br>868.1620 - CBS - Halothane Gas Analyzer<br>868.1700 - CBR - Nitrous Oxide Gas Analyzer<br>868.1720 - CCL- Oxygen Gas Analyzer<br>880.6740 - KDP- Vacuum Regulator |
#### Predicate Devices:
K142552 - A7 Anesthesia System, ShenZhen Mindray Bio-Medical Electronics CO., LTD. K042607 - Primus US, Draeger Medical Inc.
{4}------------------------------------------------
### Indication for Use:
The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations.
### Device Description:
The A7 Anesthesia System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor and provides for automatic and manual modes of ventilation. The A7 incorporates O2, CO2, N2O and Agent concentration monitoring (Desflurane, Isoflurane, Enflurane, Sevoflurane and Halothane). The A7 consists of a main unit (includes an anesthetic ventilator and flow meter monitor assembly) and a detachable breathing system.
### Performance Data:
Software – The A7 Anesthesia System software has been fully verified and validated and documentation in accordance with FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: May 11, 2005" has been provided in this submission.
Performance Testing - The A7 Anesthesia system has been tested to comply with its product specifications and intended use. The following standards were adhered to for performance testing submitted in this 510(k):
- ANSI/AAMI ES60601-1:2005/(R) 2012 and C1:2009/(R) 2012 and, A2:2010/(R) ● 2012 Medical electrical equipment - Part 1: General Requirements for basic safety and essential performance
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements . for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- . IEC 60601-1-8:2006 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-2-13:Edition 3.1 2009-08 Medical electrical equipment Part 2-13: ● Particular requirements for the safety and essential performance of anesthetic systems
- ISO 80601-2-55:2011 Medical electrical equipment Part 2-55: Particular ● requirements for the basic safety and essential performance of respiratory gas monitors
- IEC 62304: 2006 Medical device software Software life cycle processes ●
{5}------------------------------------------------
## Substantial Equivalence:
The A7 that is the subject of this 510(k) submission has been modified from the A7 that was cleared in K142552 to include the Fresh Gas Flow Optimizer feature. The Fresh Gas Flow Optimizer displays a recommended value for fresh gas to assist the clinician in determining a value that will meet the needs of the patient. The predicate device as cleared by FDA did not provide this feature. This feature does not change the fresh gas flow setting nor does it supply an alarm if the user does not follow the recommended fresh gas flow setting.
A technical comparison of the subject device to the primary predicate device as cleared in K142552 is provided in the table below.
| Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System<br>ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(Subject Device) | Predicate Device<br>A7 Anesthesia System<br>ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(K142552) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Vaporizers | Two or Three, variable bypass | Two, variable bypass | |
| Agent - Sevoflurane | Yes | Yes | |
| Agent - Isoflurane | Yes | Yes | |
| Agent - Desflurane | Yes | Yes | |
| Agent - Halothane | Yes | Yes | |
| Agent - Enflurane | Yes | Yes | |
| Automatic Ventilator | Yes | Yes | |
| Bellows | Yes | Yes | |
| Bellows Volume | 1500mL | 1500mL | |
| Ventilation Modes | | | |
| VCV | Yes | Yes | |
| PCV | Yes | Yes | |
| PCV – VG | Yes | Yes | |
| SIMV - VC | Yes | Yes | |
| SIMV – PC | Yes | Yes | |
| PS | Yes | Yes | |
| Tidal Volume | Yes | Yes | |
| Specifications | | | |
| Range, ml | 20 - 1500 | 20 - 1500 | |
| Minute Volume | Yes | Yes | |
| Rate, bpm | 4-100 bpm | 4-100 bpm | |
| Inspiratory Flow, L/min | 110 L/min + fresh gas flow | 110 L/min + fresh gas flow | |
| | Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System | Predicate Device<br>A7 Anesthesia System |
| | | ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(Subject Device) | ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(K142552) |
| | I:E Ratio | 4:1 to 1:8 with 0.5 increment | 4:1 to 1:8 with 0.5 increment |
| | Inspiratory Pause | Off, 5 - 60% of insp. Period | Off, 5 - 60% of insp. Period |
| | Air Flow Range | 0~15 L/min | 0~15 L/min |
| Fresh Gas | N2O Flow Range | 0~12 L/min | 0~12 L/min |
| | O2 Flow Range | 0~15 L/min | 0~15 L/min |
| | Individual Gas<br>Flow Accuracy | $±50$ ml/min or $±5%$ of setting value,<br>whichever is greater | $±120$ ml/min or $±10%$ of setting<br>value, whichever is greater |
| | Pressure Limit, cm H2O | 0 - 100 | 0 - 100 |
| | PEEP, cm H2O | Off, 3-30, 1 cmH2O increment | Off, 3-30, 1 cmH2O increment |
| | System Checks | Auto at start | Auto at start |
| | Airway Pressure Measured at | Inspiratory | Inspiratory |
| | High/Low Airway Pressure Alarm | Yes | Yes |
| | Pressure Limiting Alarm | Yes | Yes |
| | Sub Atmospheric Pressure Alarm | Yes | Yes |
| | Continuous Press Alarm | Yes | Yes |
| | Apnea >2 Minute Alarm | Yes | Yes |
| | Apnea Alarm | Yes | Yes |
| | High/Low Minute Volume Alarm | Yes | Yes |
| | High/Low O2 Concentration<br>Alarm | Yes | Yes |
| | Type of O2 Sensor | Paramagnetic | Paramagnetic or Galvanic (The<br>Paramagnetic sensor is utilized<br>when using the integrated gas<br>module. The Galvanic sensor is<br>utilized when the gas module is<br>not connected.) |
| | Heated Breathing Circuit | Yes | Yes |
| | Spirometry: Pressure-Volume and<br>Flow Volume loops | Yes | Yes |
| | Subject Device<br>A7 Anesthesia System | Predicate Device<br>A7 Anesthesia System | |
| Technical<br>Characteristics | ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(Subject Device) | ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(K142552) | |
| Anesthetic Gas Module Sampling<br>Rate | P/N 9200-10-10530 water trap: 120,<br>150, 200mL/min<br>P/N 9200-10-10574 water trap: 70,<br>90, 120mL/min | P/N 9200-10-10530 water trap:<br>120, 150, 200mL/min<br>P/N 9200-10-10574 water trap:<br>70, 90, 120mL/min | |
| Anesthetic Gas Module Sampling<br>Delay Time: | <4 seconds | <4 seconds | |
| Anesthetic Gas Module Refresh<br>Rate: | 1 second | 1 second | |
| Anesthetic Gas Module Warm-up<br>Time: | 45 seconds to warm-up status 10<br>minutes to ready-to-measure status | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status | |
| Anesthetic Gas Module<br>Accuracy CO2: | 0-1%: +/-.1%<br>1-5%: +/-.2%<br>5-7%: +/-.3%<br>7-10%: +/-.5%<br>>10%: unspecified | 0-1%: +/-.1%<br>1-5%: +/-.2%<br>5-7%: +/-.3%<br>7-10%: +/-.5%<br>>10%: unspecified | |
| Anesthetic Gas Module<br>Accuracy N2O: | 0-20%: +/-2%<br>20-100%: +/-3% | 0-20%: +/-2%<br>20-100%: +/-3% | |
| Anesthetic Gas Module<br>Accuracy Desflurane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-10%: +/-.4%<br>10-15%: +/-.6%<br>15-18%: +/-1%<br>>18%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-10%: +/-.4%<br>10-15%: +/-.6%<br>15-18%: +/-1%<br>>18%: unspecified | |
| Anesthetic Gas Module<br>Accuracy Sevflurane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-8%: +/-.4%<br>>8%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-8%: +/-.4%<br>>8%: unspecified | |
| Anesthetic Gas Module<br>Accuracy Enflurane/Isoflurane/<br>Halothane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>>5%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>>5%: unspecified | |
| Anesthetic Gas Module<br>Accuracy O2: | 0-25%: +/-1%<br>25-80%: +/-2%<br>80-100%: +/-3% | 0-25%: +/-1%<br>25-80%: +/-2%<br>80-100%: +/-3% | |
| Anesthetic Gas Module<br>Accuracy awRR: | 2-60rpm: +/-1rpm<br>>60rpm: unspecified | 2-60rpm: +/-1rpm<br>>60rpm: unspecified | |
| Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System<br>ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(Subject Device) | Predicate Device<br>A7 Anesthesia System<br>ShenZhen Mindray Bio-Medical<br>Electronics CO., LTD.<br>(K142552) | |
| Anesthetic Gas Module<br>Measurement Rise Time: | CO2: ≤250ms<br>N2O: ≤250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: ≤350ms | CO2: ≤250ms<br>N2O: ≤250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: ≤350ms | |
| Measurement Range CO2 | 0-30% | 0-30% | |
| Measurement Range N2O | 0-100% | 0-100% | |
| Measurement Range Des | 0-30% | 0-30% | |
| Measurement Range Sev | 0-30% | 0-30% | |
| Measurement Range Enf/Iso/Hal | 0-30% | 0-30% | |
| Measurement Range O2 | 0-100% | 0-100% | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Substantial Equivalence Conclusion
The subject A7 Anesthesia System and the primary predicate are both gas anesthesia machines used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. These systems may be used in adult and pediatric populations. Both of the devices are prescription use, and are not suitable for use in an MRI environment.
Based on the technical comparison, the identical indications for use and the complete system level testing, the modifications to the A7 Anesthesia System do not raise different questions of safety and effectiveness and can be found substantially equivalent to the predicate device cleared in K142552.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.