K183316 · Dr. Japan Co., Ltd. · BSP · Aug 12, 2019 · Anesthesiology
Device Facts
Record ID
K183316
Device Name
Dr J Spinal and Epidural Needles
Applicant
Dr. Japan Co., Ltd.
Product Code
BSP · Anesthesiology
Decision Date
Aug 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Dr. J Spinal and Epidural Needles are intended to be used for injection of local anesthetics into a patient to provide regional anesthesia
Device Story
Spinal and epidural needles used for regional anesthesia; spinal needles inject anesthetic into subarachnoid space; epidural needles inject into epidural space, often for catheter placement. Device consists of stainless-steel cannula, hub, stylet, and plunger; epidural needles include wing and curved tip. Used by clinicians in clinical settings. Output is delivery of anesthetic agent to block nerve signals, eliminating pain. Benefits include effective regional anesthesia for surgical or pain management procedures.
Clinical Evidence
Bench testing only. Testing included functional performance (bonding, bending, penetration force, leakage, axial load, unscrewing, overriding), packaging integrity (dye penetration, seal strength, transit test), sterility, EO residuals, and biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity) per ISO 10993 and ISO 80369-7 standards.
Indicated for injection of local anesthetics into adult and pediatric patients to provide regional anesthesia.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Predicate Devices
Dr. Japan's Phoenix Spinal & Epidural Needles (K990519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 12, 2019
Dr. Japan Co., Ltd. % Yolanda Smith Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
Re: K183316
Trade/Device Name: Dr J Spinal and Epidural Needles Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: July 8, 2019 Received: July 9, 2019
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K183316
Device Name Dr J Spinal and Epidural Needle
Indications for Use (Describe)
Dr. J Spinal and Epidural Needles are intended to be used for injection of local anesthetics into a patient to provide regional anesthesia
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
#### SPONSOR
| Company Name: | Dr. Japan Co., Ltd. |
|---------------------------|--------------------------------------------------------|
| Company Address: | 1-1 Kagurazaka, Shinjuku-ku<br>Tokyo<br>162-0825 Japan |
| Telephone: | 813-3513-8766 |
| Fax: | 813-3513-8226 |
| Contact Person: | Mitsuko Uchida |
| Summary Preparation Date: | August 2, 2019 |
#### DEVICE NAME
| Trade Name: | Dr. J Spinal and Epidural Needles |
|----------------------|--------------------------------------------------|
| Common/Usual Name: | Spinal and Epidural Needles |
| Classification Name: | Needle, Conduction, Anesthetic (W/Wo Introducer) |
| Regulation Number: | 21 CFR 868.5150 |
| Product Code: | BSP |
| Device Class: | Class II |
### PREDICATE DEVICE
Legally marketed Equivalent Device
| Company | Product | 510(k) # |
|-----------------------------|-----------------------------------------------|----------|
| Myco Medical Supplies, Inc. | Dr. Japan's Phoenix Spinal & Epidural Needles | K990519 |
#### DEVICE DESCRIPTION
Dr. J Spinal and Epidural Needles are intended to be used for injection of local anesthetics into a patient to provide regional anesthesia. The Dr. J Spinal and Epidural Needles are supplied sterile, intended to be single use.
Dr. J Spinal Needles are composed of a stainless-steel cannula, a hub, a stainless-steel stylet and plunger.
A guide needle may be attached to this product. The spinal needle is supplied in either the Quincke Point (=K-3 Point) or the Pencil Point.
The Dr. J Epidural Needle is composed of a stainless-steel cannula with depth marker, a hub, a stainlesssteel stylet, a plunger and a wing. The Dr J Epidural Needle is supplied in Detachable wing type which has a curved needle tip design to protect the catheter from damage during catheterization.
{4}------------------------------------------------
#### DEVICE INDICATIONS FOR USE
Dr. J Spinal and Epidural Needles are intended to be used for injection of local anesthetics into a patient to provide regional anesthesia.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Dr. Japan Co., Ltd. is the manufacturer of the predicate device. This premarket notification represents a modification to the outer box only. The technological characteristics of the predicate device are not changed. The subject device is manufactured using the same inert, biocompatible material as the predicate device. In addition, bench testing was performed (see Section 18) to assure the performance and safety of subject device.
| | Subject New Device | Predicate Device | Discussion |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| | Dr. J Spinal and Epidural Needles | Myco Medical Supplies, Inc. | |
| K Number | | K990519 | |
| Manufacturer | Dr. Japan Co., Ltd.<br>Tianjin Hanaco Medical Co., Ltd. | Dr. Japan Co., Ltd.<br>Tianjin Hanaco Medical Co., | Same |
| Brand Name | Dr. J Spinal and Epidural Needles | Dr. Japan's Phoenix Spinal and<br>Epidural Needles | |
| Regulation<br>Description | Anesthesia conduction needle | Anesthesia conduction needle | Same |
| Regulation No. | 21 CFR 868.5150 | 21 CFR 868.5150 | Same |
| Product Code | BSP | BSP | Same |
| Indications for<br>Use | Intended to be used for injection of<br>local anesthetics into a patient to<br>provide regional anesthesia | Intended to be used to inject local<br>anesthetics into a patient to<br>provide regional anesthesia | Same |
| Principle of<br>Operation –<br>Spinal Needle | Spinal anesthesia is a form of<br>regional anesthesia involving the<br>injection of a local anesthetic into<br>the subarachnoid space through a<br>spinal needle.<br>Regardless of the anesthetic agent<br>used, the desired effect is to block<br>the transmission of afferent nerve<br>signals from peripheral<br>nociceptors. Sensory signals from<br>the site are blocked, thereby<br>eliminating pain. | Spinal anesthesia is a form of<br>regional anesthesia involving the<br>injection of a local anesthetic into<br>the subarachnoid space through<br>a spinal needle.<br>Regardless of the anesthetic<br>agent used, the desired effect is<br>to block the transmission of<br>afferent nerve signals from<br>peripheral nociceptors. Sensory<br>signals from the site are blocked,<br>thereby eliminating pain. | Same |
| Principle of<br>Operation –<br>Epidural<br>Needle | Epidural anesthesia is a regional<br>anesthesia that blocks pain in a<br>particular region of the body. An<br>epidural needle is used to inject<br>local anesthetic agent into the<br>epidural space of the spinal cord<br>to block the pain. To provide<br>continuous epidural anesthesia, a<br>small hollow catheter is threaded | Epidural anesthesia is a regional<br>anesthesia that blocks pain in a<br>particular region of the body. An<br>epidural needle is used to inject<br>local anesthetic agent into the<br>epidural space of the spinal cord<br>to block the pain. To provide<br>continuous epidural anesthesia, a<br>small hollow catheter is threaded | Same |
| | through the epidural needle into<br>the epidural space and left there<br>while the needle is removed. | through the epidural needle into<br>the epidural space and left there<br>while the needle is removed. | |
| Intended<br>Patient<br>Population | Adult and child | Adult and child | Same |
| Material Specifications | | | |
| Needle/cannula<br>material | Cold-rolled stainless steel<br>(JIS G4305-SUS304) | Cold-rolled stainless steel<br>(JIS G4305-SUS304) | Same |
| Cannula depth<br>markings | The system to cause a chemical<br>change of the metal ion on the<br>surface to a black color by<br>electrolysis without changing the<br>material of cannula. | The system to cause a chemical<br>change of the metal ion on the<br>surface to a black color by<br>electrolysis without changing the<br>material of cannula. | Same |
| Stylet material | Spinal Needle: Stainless steel wire<br>for spring<br>Epidural Needle: Stainless steel<br>wire for spring | Stainless steel wire for spring<br>Epidural Needle: Stainless steel<br>wire for spring | Same |
| Stylet hub<br>material | Polycarbonate | Polycarbonate | Same |
| Retaining plate<br>(Plunger) | Polycarbonate | Polycarbonate | Same |
| Final Needle<br>assembly<br>protection | Protector (polyethylene) | Protector (polyethylene or<br>polypropylene) | Similar<br>Minor<br>difference<br>(*1) |
| Sterilization<br>Method | Sterile EO (=ETO) | Sterile ETO | Same |
| Sterile package | Peel open type package using<br>laminate film of polyester and<br>polyethylene / Tyvek high density<br>polyethylene or medical grade<br>craft paper | Peel open type package using<br>laminate film of polypropylene or<br>polyethylene / polyethylene of<br>high density or medical grade<br>craft paper | Similar<br>Minor<br>difference<br>(*2) |
#### COMPARISON OF TECHNICAL CHARACTERISTICS
{5}------------------------------------------------
## Comparison Table of devices - Subject device
| Dr.J Brand | Item no | Length | Gauge | Point |
|---------------|-----------------|--------|-------|-------|
| Spinal Needle | 18Gx90mm (3.5") | 3.5" | 18G | K-3 |
| Spinal Needle | 19Gx90mm (3.5") | 3.5" | 19G | K-3 |
{6}------------------------------------------------
| Spinal Needle | 20Gx90mm (3.5") | 3.5" | 20G | K-3 |
|-----------------|-----------------|------|-----|-------------|
| Spinal Needle | 22Gx90mm (3.5") | 3.5" | 22G | K-3 |
| Spinal Needle | 25Gx90mm (3.5") | 3.5" | 25G | K-3/ Pencil |
| Spinal Needle | 26Gx90mm (3.5") | 3.5" | 26G | K-3/ Pencil |
| Spinal Needle | 27Gx90mm (3.5") | 3.5" | 27G | K-3/ Pencil |
| Epidural Needle | 16Gx90mm (3.5") | 3.5" | 16G | Tuohy |
| Epidural Needle | 17Gx90mm (3.5") | 3.5" | 17G | Tuohy |
| Epidural Needle | 18Gx90mm (3.5") | 3.5" | 18G | Tuohy |
| Epidural Needle | 20Gx90mm (3.5") | 3.5" | 20G | Tuohy |
#### Comparison Table of devices – Predicate device
| Phoenix Brand | Item no | Length | Gauge | Point |
|-----------------|-----------------|--------|-------|--------------|
| Spinal Needle | 18Gx90mm (3.5") | 3.5" | 18G | K-3 |
| Spinal Needle | 19Gx90mm (3.5") | 3.5" | 19G | K-3 |
| Spinal Needle | 20Gx90mm (3.5") | 3.5" | 20G | K-3 |
| Spinal Needle | 22Gx90mm (3.5") | 3.5" | 22G | K-3 / Pencil |
| Spinal Needle | 25Gx90mm (3.5") | 3.5" | 25G | K-3/ Pencil |
| Spinal Needle | 26Gx90mm (3.5") | 3.5" | 26G | K-3/ Pencil |
| Spinal Needle | 27Gx90mm (3.5") | 3.5" | 27G | K-3/ Pencil |
| Epidural Needle | 16Gx90mm (3.5") | 3.5" | 16G | Tuohy |
| Epidural Needle | 17Gx90mm (3.5") | 3.5" | 17G | Tuohy |
| Epidural Needle | 18Gx90mm (3.5") | 3.5" | 18G | Tuohy |
| Epidural Needle | 20Gx90mm (3.5") | 3.5" | 20G | Tuohy |
#### Discussion of Similarities
The subject device and the predicate device have the same intended use, design, technological characteristics and similar materials.
#### Discussion of Differences
- *1: Predicate device used 2 types of protector, made in either polyethylene or polypropylene while subject device used only 1 type in polyethylene (in same material and specification as predicate device).
- *2: Tyvek high density polyethylene used in subject device showed similar safety and stability as polyethylene of high density used in predicate device. See Section 14 for test results.
- *3: Models of subject device (needle specification, gauge size and length etc.) are all included in the family of predicate device.
{7}------------------------------------------------
### SAFETY AND EFFECTIVENESS CONCLUSIONS:
This submission represents only a modification to the outer box for the currently marketed predicate. Therefore, the similarities between the subject device and predicate device are numerous (as shown in the comparison table). These equivalencies combine to justify a substantially equivalent determination.
Shelf life – Packaging testing, Shelf life- Functionality testing and Sterilization (See Section 14 and Section 18) were conducted to show that the safety, integrity or functional characteristics of the subject device is assured.
{8}------------------------------------------------
## SUMMARY OF NON-CLINICAL TESTING
Nonclinical testing as shown in the summary below was conducted to ensure the safety and effectiveness of subject device. Subject device passed all the testing and successfully demonstrate that subject device is as safe, as effective, and performs as well as legally marketed device predicate.
| Category | Evaluation | Test Criteria |
|---------------------------|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional<br>Performance | Stability test bonding to<br>hub | ISO 7864 Sterile hypodermic needles for single use --<br>Requirements and test methods |
| | Stability test bending<br>rigidity | ISO 9626 Stainless steel needle tubing for the<br>manufacture of medical devices -- Requirements and test<br>methods |
| | Penetration force and drag<br>force for needles | ISO 7864 Sterile hypodermic needles for single use --<br>Requirements and test methods |
| | Breakage test | ISO 9626 Stainless steel needle tubing for the<br>manufacture of medical devices -- Requirements and test<br>methods |
| | Fluid leakage test | ISO 80369-7 Small-bore connectors for liquids and<br>gases in healthcare applications -- Part 7: Connectors for<br>intravascular or hypodermic applications |
| | Separation axial load test | ISO 80369-7 Small-bore connectors for liquids and<br>gases in healthcare applications -- Part 7: Connectors for<br>intravascular or hypodermic applications |
| | Separation unscrewing<br>test | ISO 80369-7 Small-bore connectors for liquids and<br>gases in healthcare applications -- Part 7: Connectors for<br>intravascular or hypodermic applications |
| | Overriding test | ISO 80369-7 Small-bore connectors for liquids and<br>gases in healthcare applications -- Part 7: Connectors for<br>intravascular or hypodermic applications |
| | Stability test bonding to<br>hub | ISO 7864 Sterile hypodermic needles for single use --<br>Requirements and test methods |
| | Stability test bending<br>rigidity | ISO 9626 Stainless steel needle tubing for the<br>manufacture of medical devices -- Requirements and test<br>methods |
| | Dye Penetration Test | ASTM F 1929-15 Standard Test Method for Detecting<br>Seal Leaks in Porous Medical Packaging by Dye<br>Penetration Method A |
| | Seal Strength: | EN 868-5:18 Packaging for terminally sterilized medical<br>devices. Sealable pouches and reels of porous and<br>plastic film construction. Requirements and test methods<br>Annex D<br>ASTM D4169-16 Standard Practice for Performance<br>Testing of Shipping Containers and Systems DC13<br>ASTM F1886/ F1886M-09 Standard Test Method for<br>Determining Integrity of Seals for Flexible Packaging by<br>Visual Inspection. |
| | Packaging | Transit Test |
| Sterility | | <China> GB/T14233 2-2005 Test methods for medical<br>infusions, blood transfusions and injection device - Part 2:<br>Biological test methods |
| Residuals | | <Japan> ISO10993-7 Biological evaluation of medical<br>devices -- Part 7: Ethylene oxide sterilization residuals<br><China> EO: GB/T14233.1 Test methods for medical<br>infusions, blood transfusions and injection device - Part 1:<br>Chemical analysis method<br>ECH: ISO10993-7 Biological evaluation of medical<br>devices -- Part 7: Ethylene oxide sterilization residuals |
| Sterilization | | |
| | Bioburden | ISO11737-1 Sterilization of health care products --<br>Microbiological methods -- Part 1: Determination of a<br>population of microorganisms on products |
| Biocompatibility | Cytotoxicity | ISO 10993-5 Biological evaluation of medical devices --<br>Part 5: Tests for in vitro cytotoxicity |
| | Intracutaneous Reactivity | ISO 10993-10 Biological evaluation of medical devices -<br>- Part 10: Tests for irritation and skin sensitization |
| | Irritation | ISO 10993-10 Biological evaluation of medical devices -<br>- Part 10: Tests for irritation and skin sensitization |
| | Sensitization | ISO 10993-10 Biological evaluation of medical devices -<br>- Part 10: Tests for irritation and skin sensitization |
| | Acute Systemic Toxicity | ISO 10993-11 Biological evaluation of medical devices -<br>- Part 11: Tests for systemic toxicity |
| | Hemolytic | ISO 10993-4 Biological evaluation of medical devices --<br>Part 4: Selection of tests for interactions with blood |
| | Pyrogen | ISO 10993-11 Biological evaluation of medical devices -<br>- Part 11: Tests for systemic toxicity |
#### Summary of Non-Clinical Testing
{9}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE CONCLUSION
The evaluation of subject device classification, indications for use, and technological characteristics demonstrate substantial equivalence to the predicate device. The only difference between marketed predicate device and subject device is a modification to the outer box.
No new issues of safety and effectiveness questions are raised by the subject device when compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.