K170112 · International Medical Development, Inc.(Imd) · BSP · Feb 2, 2018 · Anesthesiology
Device Facts
Record ID
K170112
Device Name
IMD's Lumbar Puncture Needle
Applicant
International Medical Development, Inc.(Imd)
Product Code
BSP · Anesthesiology
Decision Date
Feb 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170112 · Feb 2, 2018
IMD's Lumbar Puncture Needle
International Medical Development, Inc.(Imd)
FDA MAUDE database; FDA Safety Alert database; FDA MDR (Medical Device Reporting) database
The sponsor performed a retrospective review of FDA postmarket surveillance databases (MAUDE, Safety Alert, and MDR) to assess the safety profile of similar predicate and reference devices as part of the substantial equivalence justification.
Review of FDA Postmarket Databases; Retrospective database review
Patients reported in FDA postmarket databases for similar lumbar puncture and anesthesia needles; Number of Sites: Not applicable
Not applicable for this study
Incidence of unexpected patient problems or adverse reactions
Indications for Use
IMD's Lumbar Puncture Needle, a single patient use needle; it is intended to gain entry into or puncture the spinal cavity permitting injecting (including anesthesia) / withdrawl of fluids for purpose of diagnostic lumbar puncture myelography discography procedures. The device is intended for adult and pediatric patients. It is to be used only under the direction of a licensed clinician.
Device Story
Single-patient-use lumbar puncture needle; used to access spinal cavity for fluid withdrawal or injection of substances (e.g., anesthesia) during diagnostic procedures like myelography or discography. Device consists of introducer, stylet, and pencil-point needle with plastic hub and stainless steel cannula. Operated by licensed clinician in clinical setting. Provides physical access to spinal space; output is the successful puncture and subsequent fluid exchange or injection. Benefits include facilitating diagnostic and therapeutic spinal procedures.
Clinical Evidence
Bench testing only. Device sterility validated per ISO 11135 and ISO 11737-2. Biocompatibility confirmed per ISO 10993. No clinical data presented.
Indicated for adult and pediatric patients requiring entry into or puncture of the spinal cavity for diagnostic lumbar puncture, myelography, or discography procedures, or for the injection of substances (including anesthesia) and withdrawal of fluids. To be used only by a licensed clinician.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 2, 2018
International Medical Development, Inc. % John Lincoln Principal Consultant J.E. Lincoln and Associates P.O. Box 2786 St. George, Utah 84771
Re: K170112
Trade/Device Name: IMD's Lumbar Puncture Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: December 26, 2017 Received: January 2, 2018
Dear John Lincoln:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB | lo. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement be ow.
510(k) Number (if known)
#### K170112
Device Name IMD's Lumbar Puncture Needle
Indications for Use (Describe)
IMD's Lumbar Puncture Needle, a single patient use needle; it is intended to gain entry into or puncture the spinal cavity permitting injecting (including anesthesia) / withdrawl of fluids for purpose of diagnostic lumbar puncture myelography discography procedures.
The device is intended for adult and pediatric patients.
It is to be used only under the direction of a licensed clinician.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Su part C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOWI*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other ispect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
- SUBMITTED BY: IMD INC. 560 Hwy 39 PO Box 510 Huntsville UT 84317 (800)-824-8223, or (801) 745-4700
#### 01. DEVICE NAME (Trade/common, and classification):
Common Name: Anesthesia Conduction Needle
Trade Name: IMD's Lumbar Puncture Needle
Product Code: BSP
Regulation Nos .: 868.5150
#### 02. PREDICATE AND REFERENCE DEVICES:
- 2.1 Predicate Device:
Pajunct's Lumber Puncture Needle, K160294, Regulation Number 868.5150, Class II, Product Code BSP, Panel: Anesthesiology, cleared 10/21/2016; Same Intended Use, materials and method of manufacture are similar; see matrix below;
#### 2.2 Reference Devices:
This submission is identical to the following reference devices in materials, method of assembly / testing, and packaging / sterilization; but one with the addition of side fenestrations (K113662), and all with different Indications for Use:
IMD's Fenestrated Nerve Block Needle (Closed-end Tip / Pencil Point), K113662 Regulation Number 868.5140, Class II, Product Code CAZ, Panel: Anesthesiology, cleared 04/20/2012;
IMD's Anesthesia Needles (Touhy, Quincke, and Pencil Point), K070354 Regulation Number 868.5150, Class II, Product Code BSP, Panel: Anesthesiology, cleared10/05/2007; and
IMD's Anesthetic Needle (Gertie Marx), K931644, Regulation Number 868.5150,
{4}------------------------------------------------
Class II, Product Code BSP, Panel: Anesthesiology, cleared 09/22/1993;
Device Comparison Table:
| | IMD's Lumbar<br>Puncture Needle –<br>K170112 | Pujunk Lumbar<br>Puncture Needle –<br>K160294 | IMD's Fenestrated<br>Nerve Block<br>Needle – K113662 | IMD's, Anesthetic<br>Needles: Touhy,<br>Quincke, Pencil<br>Point – K070354 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use: | Single patient use<br>lumbar puncture<br>needle used gain<br>entry into or<br>puncture the spinal<br>cavity permitting<br>injecting (including<br>anesthesia) /<br>withdrawal of fluids<br>for purposed of<br>diagnostic lumbar<br>puncture,<br>myelography<br>discography<br>procedures. It is to<br>be used only under<br>the direction of a<br>licensed clinician. | Single patient use<br>lumbar puncture<br>needle used gain<br>entry into or<br>puncture the spinal<br>cavity permitting<br>injecting (including<br>anesthesia) /<br>withdrawal of fluids<br>for purposed of<br>diagnostic lumbar<br>puncture,<br>myelography<br>discography<br>procedures. It is to<br>be used only under<br>the direction of a<br>licensed clinician. | Single patient use<br>peripheral nerve<br>block needle used to<br>provide local or<br>regional nerve<br>blocking by the<br>injecting of a local<br>anesthesia. It is to<br>be used only under<br>the direction of a<br>licensed clinician. | Single patient use<br>needles intended for<br>the transient delivery<br>of anesthetics to<br>provide regional<br>anesthesia ...<br>placement of an<br>epidural catheter...<br>used only under the<br>direction of a<br>licensed clinician. |
| Length: | 3.5 in. (90 mm)<br>- 5.0 in (150 mm) | Various (90 – 150<br>mm) | 3.5 in. (90 mm) and<br>5.0 in. (150 mm) | Various (22 ga. = 3.5<br>in. / 90 mm) |
| Gauge: | Various (18 - 27 ga.) | Various (18 - 20 ga.) | 20 ga. | Various (18, 22, 25<br>ga.) |
| Tip<br>Configuration: | Pencil Point | Sprotte and Quincke | Pencil Point | Pencil Point, et al |
| Materials: | | | | |
| Needle /<br>Cannula | 304 Stainless steel | 304 Stainless steel | 304 Stainless steel | 304 Stainless steel |
| Hub | K-Resin (SBC); or<br>Cyrolite acrylic | Polycarbonate | K-Resin (SBC); or<br>Cyrolite acrylic | K-Resin (SBC); or<br>Cyrolite acrylic |
| Stylet | Stainless steel | Stainless steel | Stainless steel | Stainless steel |
| Methods of<br>Manufacture: | Identical to the<br>reference IMD<br>Needles; using<br>same needle<br>manufacturer | Similar to IMD | Identical to the IMD<br>Needles; using<br>same needle<br>manufacturer | Identical to the IMD<br>Needles; using<br>same needle<br>manufacturer |
| Cannula: | One (side)<br>fenestration | One fenestration | Multiple side<br>fenestrations | One fenestration |
| Biocompat-<br>ibility: | Yes, per ISO 10993 | Yes, per ISO 10993 | Yes, per ISO 10993 | Yes, per ISO 10993 |
| Packaging: | Same header bag /<br>pouches; using<br>same contract | Header bag /<br>pouches. | Same header bag /<br>pouches; using<br>same contract | Same header bag /<br>pouches; using<br>same contract |
# COMPARATIVE (SE) INFORMATION MATRIX
{5}------------------------------------------------
| | packaging company<br>(IPM) with same<br>validated sealers | packaging company<br>(IPM) with same<br>validated sealers | packaging company<br>(IPM) with same<br>validated sealers |
|----------------------|-----------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| Product<br>Labeling: | Single- and 10-unit | Single- and multiple | Single- and 10-unit |
| Sterilization: | 100% EtO | 100% EtO | 100% EtO |
| | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| | Sterigenics, SLC<br>UT: EO steriliz'n,<br>cycle 870 | Unknown | Sterigenics, SLC<br>UT: EO steriliz'n,<br>cycle 870 |
| | Nelson Labs, SLC<br>UT: BI testing | Unknown | Nelson Labs, SLC<br>UT: BI testing |
| | Nelson Labs: EO<br>validation (ISO<br>11135) | Unknown | Nelson Labs: EO<br>validation (ISO<br>11135) |
| | Sterigenics, SLC<br>UT: EO sterilization,<br>cycle 870 | Nelson Labs, SLC<br>UT: BI testing | Nelson Labs: EO<br>validation (ISO<br>11135) |
Note: Only the needle / cannula comes into contact with the patient in routine use.
# 03. DESCRIPTION:
The IMD Inc. Lumbar Puncture Needle shares major similarities with the predicate device(s), in as much as the configuration, materials, area of use, and are identical to the IMD reference devices.
It consists of an introducer, stylet, and pencil-point needle, with plastic hubs, and stainless steel cannula. Offered in 18-27 gauges, and 3.5 in. (90 mm) and 5.0 in (124 mm) lengths.
# 04. INDICATIONS FOR USE / INTENDED USE:
IMD's Lumbar Puncture Needle, a single patient use needle; it is intended to gain entry into or puncture the spinal cavity permitting injecting (including anesthesia) / withdrawl of fluids for purpose of diagnostic lumbar puncture, myelography discography procedures.
The device is intended for adult and pediatric patients.
It is to be used only under the direction of a licensed clinician.
# 05. SUBSTANTIAL EQUIVALENCE (SE) RATIONALE:
The IMD's Lumbar Puncture shares the identical configuration, features, materials, methods of manufacture / manufacturer, testing, packaging / packager, and sterilization / sterilizer as the reference needles (IMD's). Biocompatibility and Performance testing has been provided in the reference device submissions, which support substantial equivalence (SE).
It's Indications for Use is similar to the predicate device and the intended use is the same. The indications differ in minor semantics and include the prescription device statement. Therefore, the different semantics and prescription statement included in the IFU does not change the intended use, and does not affect the safety and effectiveness of the device when used as labeled
{6}------------------------------------------------
It it is therefore substantially equivalent to the predicate device listed above.
In addition:
- No record of unexpected patient problems or adverse reactions were found in our O review of the FDA's MAUDE, Safety Alert, and MDR databases;
| Test: | References: |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Sterility | Cleanroom, Direct transfer method, Nelson<br>Labs PSC 110, STP0077 Rev 07, USP General<br>Chapter 71, ANSI/AAMI/ISO 11737-2, EP 2.6.1,<br>JP and 21 CFR 610.12 |
| Bacteriostasis and Fungistasis | Method II: Direct Inoculation, Nelson Labs BFS<br>120, USP <71>, EP 2.6.1 and <jp 54="" xv,="">;<br/>and AAMI/ISO 11737-2.</jp> |
- The device and its packaging was tested by an independent lab for the following: O
Production Lots are also subjected to inspection / testing by Incoming QC, in-process OC, and finished goods QC, and monitored in the field by means of our CAPA system.
### 06. CONCLUSION:
There are no substantive differences between the device defined in this 510(k) submission and the predicate and reference devices. It is identical or similar in configuration, material and manufacturing / sterilization technologies that are currently used in other similar medical devices. It was developed and documented under IMD's Quality Management System, under the Quality System Regulation, 21 CFR Part 820, including design / change control, and is verified / validated to applicable standards / guidance documents, including vendors' and our SOPs. It is designed and manufactured to be as safe and as effective as the predicate device when used as intended, under a licensed clinician's supervision. IMD's Lumbar Puncture Needle share similar or identical Indications for Use,and characteristics, materials,
manufacturer, sterilization, and functional features, and thus are substantially equivalent to the currently marketed predicate and reference devices, cited above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.