SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle
K151069 · Spectra Medical Devices, Inc. · BSP · Feb 5, 2016 · Anesthesiology
Device Facts
Record ID
K151069
Device Name
SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle
Applicant
Spectra Medical Devices, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Feb 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle is intended for use in injecting a single dose of local anesthetic or analgesic into a patient for regional anesthesia or pain therapy with the echogenic reflective pattern providing visualization of the needle tip using Ultrasound Imaging. The needles are intended to mate with the male connector of a Luer Lock or Luer Slip Syringe or extension set.
Device Story
Disposable single-lumen stainless steel needle; used for regional anesthesia or pain therapy; injects local anesthetic/analgesic. Features mechanical dimpling on needle tip for echogenic visualization under ultrasound imaging. Mates with standard Luer Lock/Slip syringes or extension sets. Used in hospitals, clinics, and ambulances by clinicians. Provides visual confirmation of needle tip location during ultrasound-guided procedures; improves accuracy of anesthetic placement; reduces risk of misdirected injections.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included bond strength of hub and cannula, Luer compatibility (ISO 594-1), and hub/needle strength (ISO 7864:11). Biocompatibility testing included cytotoxicity (ISO 10993-5), in vitro hemolysis (ISO 10993-4), systemic toxicity (ISO 10993-11), intracutaneous study (ISO 10993-10), and sensitization (ISO 10993-10).
Technological Characteristics
Cannula: 304 stainless steel. Hubs: Polypropylene (round) or K-Resin (square). Protector: LDPE/Polypropylene. Lubrication: Silicone. Echogenic feature: Mechanical dimpling. Dimensions: 16-30 gauge, 2.0-6.0 inch lengths. Sterilization: Ethylene Oxide (EO). No energy source. Standalone device.
Indications for Use
Indicated for adult and pediatric patients requiring regional anesthesia or pain therapy via injection of local anesthetic or analgesic.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2016
Spectra Medical Devices, Inc. % Tanya O'Brien Clinical Affairs Specialist AJW Technology Consultants, Inc. 445 Apollo Beach Blvd Apollo Beach, Florida 33572
Re: K151069
Trade/Device Name: SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: November 23, 2015 Received: December 8, 2015
Dear Ms. O'Brien:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151069
Device Name
SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle
## Indications for Use (Describe)
The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle is intended for use in injecting a single dose of local anesthetic or analgesic into a patient for regional anesthesia or pain the echnogenic reflective pattern providing visualization of the needle tip using Ultrasound Imaging. The needles are intended to mate with the male connector of a Luer Lock or Luer Slip Syringe or extension set.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
(as required by 807.92)
#### -SUBMITTER
Spectra Medical Devices, Inc. 260 F/H Fordham Road Wilmington, MA 01887
Phone: 757-220-8917 Fax: 757-220-8919
Contact Person: Scott Henderson Date Prepared: March 26, 2015
## REGULATORY CORRESPONDENT
AJW Technology Consultants, Inc. 445 Apollo Beach, Blvd. Apollo Beach, FL 33572
Contact Person: Tanya O'Brien Email: tmobrien@ajwtech.com
Phone: 813-645-2855 x102 Fax: 813-645-2856
#### 11. DEVICE
Name of Device: SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle Common or Usual Name: Conduction Needle Classification Name: Needle, Anesthesia Conduction Device Panel: Anesthesiology Regulatory Class: 2 Product Code: BSP Contact Type: Limited Exposure Device
#### lll. PREDICATE DEVICE
● Havel's Inc. - K063380 ECHOSTIM FACIT TIP These predicates have not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle consists of a stainless steel needle and a clear or colored (translucent) standard female Luer lock square or round Hub locking connector for rapid (flashback) visualization. The stainless steel needles are available with and without an echogenic feature (i.e. mechanical dimpling treatment to the needle point surface), and with or without lubrication. The needles are available in a range of wall thicknesses, gauges and lengths to match the end-user need (16 to 30 gauges).
### V. INIDCATIONS FOR USE
The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle is intended for use in injecting a single dose of local anesthetic or analgesic into a patient for regional anesthesia or pain therapy with the echogenic reflective pattern providing visualization of the needle tip using Ultrasound Imaging. The needles are intended to mate with the male connector of a Luer Lock or Luer Slip Syringe or an extension set.
## VI. COMPMARISOIN OF TECHNOLIGICAL CHARACTERISTICS IWHT THE PREDICATE DEVICE
To inject fluids (drug solutions), or withdraw fluids from parts of the body below the surface of the skin is the technological principle for both the subject and predicate devices. It is based on the use of modules and accessories that are either connected internally or externally to the monitors in order to monitor the specific physiological parameter. At a high level, the subject and predicate devices are based on the following same technological characteristics:
- . Design
- . Materials of the needle
- Sterility
- . No thermal source
- Indications for use ●
The following technological differences exist between the subject and predicate devices:
- . Additional or different sizes of needles are offered
- Some additional/different materials are used for the Spectra hub designs
- Physical dimensions
- VII. DEVICE COMPARISON CHART See Appendix A
{5}------------------------------------------------
#### VIII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
Mechanical Testing
- Bond strength of hub and cannula
- ISO 594-1
- ISO 7864:11 ●
- Luer compatibility ●
- Hub and Needle Strength
#### BIOCOMPATIBILITY TESTING וא.
The product contact materials utilized in the Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needles have been well characterized chemically and physically. The following table indicates the individual biocompatibility tests performed on the Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needles.
| Test | Description |
|--------------|------------------------------------------------------------|
| ISO 10993-5 | Cytotoxicity Study Using the ISO Elution Method |
| ISO 10993-4 | In Vitro Hemolysis Study (Modified ASTM-Extraction Method) |
| ISO 10993-11 | USP and ISO Systemic Toxicity Study - Extracted |
| ISO 10993-10 | ISO Intracutaneous Study - Extract |
| ISO 10993-10 | ISO Maximization Sensitization Study - Extract |
#### X. CONCLUSIONS
The testing completed demonstrates that the Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needles have exhibited comparable mechanical and functional characteristics to the predicate device in addition to being biocompatible acceptable. Based on these characteristics, the Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needles, ARE substantially equivalent to the predicate device in addition to being intended for the same uses.
{6}------------------------------------------------
# APPENDIX A
{7}------------------------------------------------
| SPECTRA DEVICE | HAVEL'S | SIMILARITIES | DIFFERENCES |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pending<br>21 CFR 868.5150<br>BSP | K063380<br>21 CFR 868.5150<br>BSP | N/A | N/A |
| II | II | Same | None |
| The SPECTRA Disposable<br>Single Lumen Echogenic<br>Injection Anesthesia<br>Conduction Needle is<br>intended for use in injecting a<br>single dose of local anesthetic<br>or analgesic into a patient for<br>regional anesthesia or pain<br>therapy with the echogenic<br>reflective pattern providing<br>visualization of the needle tip<br>using Ultrasound Imaging.<br>The needles are intended to<br>mate with the male<br>connector of a Luer Lock or<br>Luer Slip Syringe or an<br>extension set. | Havel's Electrically insulated<br>anesthesia conduction<br>needles are used to puncture<br>the tissue in order to gain<br>entry and locally inject<br>anesthetics to induce regional<br>anesthesia. An electrical<br>stimulus may be applied to<br>the needle via a cable and<br>connector to assist the<br>physician to pinpoint the area<br>of application. | Same<br>The Indications for Use are<br>basically the same, in that<br>they are both conduction<br>needles intended to<br>administer regional<br>anesthesia, and both are used<br>as a non-insulated needle. | None for Non-Insulated<br>Havel's Needle Configuration |
| Adult and Pediatric<br>Single Use Disposable Device<br>Echogenic Injection<br>Anesthesia Conduction<br>Needle with Standard Round<br>Luer Style Hub or Square<br>Hubs and a Stainless Steel<br>Needle. The needle protector<br>is an extracted tube or a<br>molded protector. The<br>square clear hub has a<br>colored printed bevel<br>indicator and the round luer<br>hub is colored with no bevel<br>indicator. | Adult and Pediatric<br>Single Use Disposable Device<br>Echogenic Injection<br>Anesthesia Conduction<br>Needle with a molded Square<br>Hub and a Stainless Steel<br>Needle. The needle protector<br>is an extruded tube. The Sq.<br>needle hub has a colored<br>molded insert as a bevel<br>indicator. | Same<br>The molded Hub for both<br>products is functionally<br>equivalent, as are the<br>stainless steel needle and<br>needle protector. | None<br>The Havel's Hub Assembly has<br>a metal Bushing and is Glued;<br>the Spectra Sq. Hub Assembly<br>is over molded and the Round<br>Luer Hub Assembly is glued.<br>The Havel's Hub has a molded<br>bevel indicator; the Spectra<br>hub has a printed bevel<br>indicator. The Havel's needle<br>has depth marks on the<br>needle and the Spectra<br>needles do not. |
| B = 17° & C = 30°<br>21G to 25G; 2.0" to 6.0" | B = 17° & C= 30°<br>21G to 22G; 2.0" to 6.00" | Same<br>Similar Gauge Range | None<br>Spectra has a 25G Needle |
{8}------------------------------------------------
| Design - Echogenic Treatment of Needle Tip | Echogenic Treatment [dimpling] of Needle Tip for Ultrasound Visualization | Echogenic Treatment [4 x 4 Corner Cube indentations] of Needle Tip for Ultrasound Visualization | Similar; both are functional | Echogenic Treatment methods are different and shape of reflective surface is different. |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------------------------------------------------------------|
| Materials | Cannula = 304 Stainless Steel<br>Round Hub = Polypropylene<br>Square Hub = K-Resin<br>Protector = LDPE / Polypropylene (Molded)<br>Lubrication = Silicone | Cannula = 304 Stainless Steel<br>Square Hub = N/A (PP)<br>Protector = N/A | Equivalent Materials of Construction | None |
| Performance | Hub/Cannula Tensile Strength<br>Penetration Force<br>Luer Taper | Hub/Cannula Tensile Strength<br>Luer Taper<br>Separation from Luer | Same | None |
| Sterility | Separation from Luer | Supplied Packaged Sterile | Same | None |
| Sterilization Method(s) | Supplied Packaged Sterile | EO (Ethylene Oxide) | Same | None |
| Chemical safety | EO (Ethylene Oxide) | Assumed; Device does not pose a chemical hazard and is compatible with chemicals commonly used in the areas designated for use. | Same | None |
| Anatomical Sites | Device does not pose a chemical hazard and is compatible with chemicals commonly used in the areas designated for use. | Various, including: Arm, Hand, Neck, Shoulder, Groin, Buttock, Leg, Knee, Foot, etc. | Same | None |
| Energy used and /or delivered | Various, including: Arm, Hand, Neck, Shoulder, Groin, Buttock, Leg, Knee, Foot, etc. | No Energy used or delivered by the Device | Same | None |
| Compatibility with environment and other devices | No Energy used or delivered by the Device | Assumed; Device does not contain any materials considered to be dangerous to the environment. | Same | None |
| Where used: hospital, home, ambulance, etc. | Device does not contain any materials considered to be dangerous to the environment. | Device is designed to connect to a piston syringe (L/S and L/L) or extension set with a male luer connector. | Same | None |
| Thermal safety | Device is designed to connect to a piston syringe (L/S and L/L) or extension set with a male luer connector. | Used in Hospital, Clinics, Research Centers, Ambulance, and others<br>Device does not contain a thermal source | Same | None |
{9}------------------------------------------------
| Radiation safety | Device does not admit any<br>form of radiation | Device does not admit any<br>form of radiation | Same | None |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------|------|
| Performance testing | Hub/Cannula Tensile Strength<br>Penetration Force | Hub/Cannula Tensile Strength | Same | None |
| Electrical safety | Device does not contain an<br>electrical source or connect to<br>an electrical source | Device does not contain an<br>electrical source or connect to<br>an electrical source | Same | None |
| Biocompatibility | Biocompatibility Testing has<br>been conducted for each<br>product configuration with<br>test results confirming<br>compliance with ISO 10993<br>requirements. | Assumed; Biocompatibility<br>Testing has been conducted<br>and included with 510(K)<br>Submission to FDA. | Same | None |
| Mechanical safety | Device does not include a<br>separate safety feature and<br>does not claim it is a safety<br>device. | Device does not include a<br>separate safety feature and<br>does not claim it is a safety<br>device. | Same | None |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.