DISCYPHOR DIRECT CATHETER SYSTEM, DISCYPHOR DIRECT NEEDLES, LONG, DISCYPHOR DIRECT NEEDLES, REGULAR
K073516 · Kyphon, Inc. · BSP · Feb 21, 2008 · Anesthesiology
Device Facts
Record ID
K073516
Device Name
DISCYPHOR DIRECT CATHETER SYSTEM, DISCYPHOR DIRECT NEEDLES, LONG, DISCYPHOR DIRECT NEEDLES, REGULAR
Applicant
Kyphon, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Feb 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Indications for Use
The Kyphon Discyphor Direct™ Catheter System for the Functional Anaesthetic Discography™ Procedure, and its components, are intended for use in delivering either a single dose or continuous administration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space. The Kyphon Discyphor Direct™ Outer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Direct™ Inner Needle and Discyphor Direct™ Catheter into the intradiscal space. The Discyphor Direct™ Outer Needle is intended to be used only with the Discyphor Direct™ Catheter System. The Kyphon Discyphor Direct™ Inner Needle is intended to access the nucleus of an intervertebral disc for the purpose of performing provocative discography and facilitating placement of the Discyphor Direct™ Catheter into the intradiscal space. The Discyphor Direct™ Inner Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Discyphor Direct™ Inner Needle is intended to be used only with the Discyphor Direct™ Catheter System.
Device Story
System includes micro-catheter with inflatable distal balloon, outer/inner needles, balloon inflation assembly, stopcock, injection assembly, and anchoring device. Used by clinicians to access intradiscal space for provocative discography and delivery of contrast, anesthetics, antibiotics, or saline. Outer needle accesses area adjacent to disc; inner needle accesses nucleus; catheter is then placed. Balloon inflation assembly (syringe/relief valve) and injection assemblies connect to catheter lumens. Anchoring device secures catheter at puncture site; adhesive clips secure catheter to patient. System facilitates diagnostic discography procedures; provides controlled delivery of fluids to intervertebral disc.
Clinical Evidence
No clinical data. Bench testing only, including dimensional verification, tensile testing, bending stiffness, corrosion verification, fatigue testing, functional testing, handling, biocompatibility, and sterilization validation.
Technological Characteristics
Stainless steel needles with polycarbonate wings; micro-catheter with double lumen shaft and inflatable distal balloon. Components include stopcocks, Touhy Borst adapters, and syringes. Single-use, sterile. No electronic or software components.
Indications for Use
Indicated for patients undergoing provocative discography or requiring administration of contrast, local anesthetics, antibiotics, or saline into the intradiscal space. Used for accessing the intervertebral disc nucleus and facilitating catheter placement.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Functional Anaesthetic Discography (F.A.D.) Catheter System (K061210)
Functional Anaesthetic Discography (F.A.D.) Catheter System (K043500)
Submission Summary (Full Text)
{0}------------------------------------------------
Kyphon Inc. Page 1 of 3, Tab 5
## Tab 5 PREMARKET NOTIFICATION [510(k)| SU
| Trade Name: | Kyphon Discyphor DirectTM Catheter System<br>Kyphon Discyphor DirectTM Outer Needle<br>Kyphon Discyphor DirectTM Inner Needle | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Common Name: | Anesthesia conduction catheter<br>Anesthesia conduction needle | |
| Classification<br>Name: | Class II, Anesthesia conduction needle, 21 CFR 868.5150 | |
| Device Code: | BSP | |
| Manufacturer's<br>Name:<br>Address: | Kyphon Inc.<br>1221 Crossman Avenue<br>Sunnyvale, CA 94089 | |
| Corresponding<br>Official: | Pamela Segale | |
| Title: | Director, Regulatory Affairs | |
| Address: | 1221 Crossman Avenue<br>Sunnyvale, CA 94089 | |
| Phone: | (408) 548-5235 | |
| Predicate<br>Device(s): | K063071 (cleared on April 13, 2007): | DiscyphorTM Catheter System<br>DiscyphorTM Introducer Needle<br>DiscyphorTM Spinal Needle |
| | K061210, (cleared on June 27, 2006) Functional Anaesthetic<br>Discography (F.A.D.) Catheter System. | |
| | K043500, (cleared on April 15, 2005) Functional Anaesthetic<br>Discography (F.A.D.) Catheter System. | |
| | The Kyphon Discyphor DirectTM Catheter System for the Functional<br>Anaesthetic DiscographyTM Procedure, and its components, are intended<br>for use in delivering either a single dose or continuous administration of<br>radiopaque contrast, local anaesthetics, and/or saline solution to the<br>intradiscal space. | |
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully market approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the froduct addressed in this submission, or its use, may be considered indistinct, from a patentiality perspective, from any of the other devices referenced in this filing.
{1}------------------------------------------------
The Kyphon Discyphor Direct™ Outer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Direct " Inner Needle and Discyphor Direct " Catheter into the intradiscal space. The Discyphor Direct ""Outer Needle is intended to be used only with the Discyphor Direct™ Catheter System.
The Kyphon Discyphor Direct" Inner Needle is intended to access the nucleus of an intervertebral disc for the purpose of performing provocative discography and facilitating placement of the Discyphor Direct™ Catheter into the intradiscal space. The Discyphor Direct™ Inner Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Discyphor Direct™ Inner Needle is intended to be used only with the Discyphor Direct " Catheter System.
The Discyphor Direct™ Catheter System is comprised of a Discyphor Device Description: Direct" Catheter, a Balloon Inflation Assembly, a Stopcock Assembly, an Injection Assembly, an EPI-Guard Catheter Anchoring Device, Adhesive Clips, a Sheet of Labels, 1-cc Syringes and a Discyphor Direct™ Outer Needle (with stylet) and a Discyphor Direct™ Inner Needle (with stylet). The Discyphor Direct™ Catheter is a micro-catheter with a double lumen
shaft and an inflatable balloon at its distal end.
The Balloon Inflation Assembly, a Stopcock Assembly, an Injection Assembly are intended to be connected to the lumens of the catheter and to provide access to the lumens. The Balloon Inflation Assembly consists of an a 3-cc Syringe and a pressure relief valve. The Stopcock Assembly consists of a Stopcock, a Touhy Borst adapter and a Wing Nut. The Injection Assembly consists of a Touhy Borst adapter, a One-way valve and a Wing Nut. The EPI-Guard Catheter Anchoring Devices secures the catheter close to the needle puncture site. The Adhesive Clips are used to secure the catheter to the patient.
The 1-cc Syringes can be used for injection of contrast, antibiotic, and/or saline into the disc space.
The labels provided on the Label Sheet are used to identify the solutions used during the procedure or to denote the level at which the Discyphor Catheter has been placed.
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
{2}------------------------------------------------
The Discyphor Direct " Outer Needle is manufactured from stainless steel and has a stainless steel stylet with a beveled tip and a polycarbonate wing.
The Discyphor Direct™ Inner Needle is manufactured from stainless steel and has a stainless steel stylet.
Testing of the Discyphor Direct™ Catheter System and the Discyphor Testing Direct™ Needles was completed to demonstrate that the devices meet the specifications and performance characteristics, and were substantially equivalent to the predicate devices. This included dimensional verification, tensile testing, bending stiffnes, corrosion verification, fatigue testing, functional testing and testing of the handling. In addition, biocompatibility testing and sterilization validation were completed.
- The Discyphor Direct™ Catheter System and the Discyphor Direct™ Sterilization Outer and Inner Needles will be provided sterile and are intended for single use only.
The Discyphor Direct"" Catheter System is packaged in a Tyvek® sealed Packaging and PETG tray and a cardboard box. The 1-cc syringes are packaged Labeling separately in Tyvek® pouches and are provided with the Discyphor Direct™ Catheter System package. The Discyphor Direct™ Outer and Inner Needles are packaged separately in Tyvek® pouches and are provided in a cardboard box.
The information submitted in this pre-market notification supports a Substantial determination that the Discyphor Direct™ Catheter System and Equivalence: Discyphor Direct" Outer and Inner Needles are substantially equivalent to the predicate devices cleared under K063071. The results of testing demonstrate that the Discyphor Direct™ Catheter System and Discyphor Direct " Outer and Inner Needles are safe and effective for their intended uses. The changes described in this submission do not represent a change to the intended usees and they do not represent a change in the fundamental scientific technology of the device. Therefore, the devices included in this submission are substantially equivalent to the predicate devices already cleared under K063071.
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully market approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or fo whether the product addressed in this submission, or its use, may be considered indistinct, from a pateriability perspective, from any of the other devices referenced in this filing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three strokes forming its body and wings. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a sans-serif font.
FEB 2 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Pamela Segale Director, Regulatory Affairs Kyphon, Incorporated 1221 Crossman Avenue Sunnyvale, California 94089
Re: K073516
Trade/Device Name: Kyphon Discyphor™ Catheter System Kyphon Discyphor™ Outer Needle Kyphon Discyphor™ Inner Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: December 13, 2007 Received: December 14, 2007
Dear Ms. Segale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Segale
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Owes
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K073516
Device Name: -
Kyphon Discyphor Direct™ Catheter System Kyphon Discyphor Direct™ Outer Needle Kyphon Discyphor Direct™ Inner Needle
Indications for Use:
The Kyphon Discyphor Direct™ Catheter System for the Functional Anaesthetic Discography™ Procedure, and its components, are intended for use in delivering either a single dose or continuous administration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space.
The Kyphon Discyphor Direct™ Outer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Direct™ Inner Needle and Discyphor Direct™ Catheter into the intradiscal space. The Discyphor Direct™ Outer Needle is intended to be used only with the Discyphor Direct " Catheter System.
The Kyphon Discyphor Direct™ Inner Needle is intended to access the nucleus of an intervertebral disc for the purpose of performing provocative discography and facilitating placement of the Discyphor Direct™ Catheter into the intradiscal space. The Discyphor Direct " Inner Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Discyphor Direct" voner Needle is intended to be used only with the Discyphor Direct " Catheter System.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K073616 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.