K153297 · B.Braun Medical, Inc. · BSO · Aug 12, 2016 · Anesthesiology
Device Facts
Record ID
K153297
Device Name
Perifix and Contiplex Catheter
Applicant
B.Braun Medical, Inc.
Product Code
BSO · Anesthesiology
Decision Date
Aug 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The B. Braun Perifix catheters are regional anesthesia catheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative, perioperative and postoperative periods associated with general and orthopedic surgery as well as labor and delivery. Routes of administration are epidural. The catheters may remain indwelling for up to 72 hours. The B. Braun Contiplex catheters are regional anesthesia catheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative, perioperative and postoperative periods associated with general and orthopedic surgery. Routes of administration are perineural (peripheral nerve block). The catheters may remain indwelling for up to 72 hours.
Device Story
Perifix and Contiplex catheters are flexible, hollow polyamide tubes used for percutaneous delivery of local anesthetics and analgesics. Perifix is indicated for epidural use; Contiplex is indicated for peripheral nerve block. Available in 19 and 20 Ga. diameters with open-tip or closed-tip (three side ports) configurations. Catheters feature depth markings for insertion guidance. Used by clinicians in preoperative, perioperative, and postoperative settings. Device provides continuous or intermittent infusion to manage pain; indwelling for up to 72 hours. Benefits include targeted regional anesthesia and pain management. No software or electronic components.
Clinical Evidence
Bench testing only. Testing included dimensional specifications, functional performance (flexibility, flow rate, kink resistance, tensile strength, connector integrity), and associate device compatibility (needle/guide passage). Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, systemic toxicity, genotoxicity, and implantation. Sterilization validated to SAL 10^-6.
Technological Characteristics
Polyamide tube construction; 19 and 20 Ga. diameters; open or closed-tip (3 side ports) configurations. Depth markings thermo-diffused into tubing. Sterilization via ethylene oxide (EO) to SAL 10^-6. Biocompatibility testing per ISO 10993-1; endotoxin levels per USP <161> and <85>.
Indications for Use
Indicated for patients undergoing general or orthopedic surgery, or labor and delivery, requiring continuous or intermittent infusion of local anesthetics/analgesics for regional anesthesia and pain management. Perifix catheters are for epidural administration; Contiplex catheters are for perineural (peripheral nerve block) administration. Indwelling time up to 72 hours.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a person's profile. The symbol is composed of three overlapping shapes, creating a sense of depth and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 12, 2016
B. Braun Medical Inc. Ms. Lisa Giaquinto Sr. Specialist, Regulatory Affairs 901 Marcon Boulevard Allentown, PA 18109
Re: K153297
Trade/Device Name: Perifix and Contiplex Catheters Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: July 8, 2016 Received: July 13, 2016
Dear Ms. Giaquinto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) Unknown K153297
Device Name Perifix and Contiplex Catheters
## Indications for Use (Describe)
The B. Braun Perifix catheters are regional anestheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative and postoperative periods associated with general and orthopedic surgery as well as labor and delivery. Routes of administration are epidural. The catheters may remain indwelling for up to 72 hours.
The B. Braun Contiplex catheters are regional anesthesia catheters intended to provide, via percutaneous administration. continuous and/or intermittent infusion of local anesthetics and anerve for regional anesthesia and pain management during the preoperative and postoperative periods associated with general and orthopedic surgery. Routes of administration are perineural (peripheral nerve block). The catheters may remain indwelling for up to 72 hours.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Perifix™ and Contiplex TM Catheters August 12, 2016
#### 5. 510(k) SUMMARY
| SUBMITTER: | B. Braun Medical Inc.<br>901 Marcon Boulevard<br>Allentown, PA 18109-9341<br>610-266-0500 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact: Lisa Giaquinto, MS, RAC<br>Sr. Specialist, Regulatory Affairs<br>Phone: (610) 596-2354<br>Fax: (610) 266-4962<br>E-mail: lisa.giaquinto@bbraun.com |
| DEVICE NAME: | Perifix and Contiplex Catheters |
| COMMON OR USUAL NAME:<br>DEVICE | Catheter, Conduction, Anesthetic<br>(An anesthesia conduction catheter is a flexible tubular device used<br>to inject local anesthetics into a patient and to provide continuous<br>regional anesthesia.) |
| CLASSIFICATION: | Class II, Product Code BSO, 868.5120 |
| PREDICATE DEVICE: | B. Braun Regional Anesthesia Catheter,<br>Braun<br>Medical Inc., K042488, Class II, BSO, 868.5120. |
# DESCRIPTION
The B. Braun Perifix and Contiplex catheters are regional anesthesia catheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management. Routes of administration include epidural and perineural (peripheral nerve block). There is no difference in the catheter design of the Perifix catheter when compared to the Contiplex catheter (i.e. the catheters are the same). The only difference between the Perifix and Contiplex catheter is the intended route of administration; Perifix catheters are marketed for use in epidural anesthesia, while Contiplex catheters are marketed for use in peripheral nerve block procedures.
The B. Braun Perifix and Contiplex catheters consists of a hollow polyamide tube with depth markings thermo-diffused into the polyamide tube. The catheters are available in both 19 and 20 Ga. diameters. The Perifix/Contiplex catheters are available in an open-tip configuration (no side ports) as well as a closed-tip configuration (with side ports). The closed-tip catheters have a rounded closed-tip with three side ports for distribution of anesthetics and analgesics.
{4}------------------------------------------------
The catheters include depth markings to identify the depth of insertion during use. The tip of the catheter (proximal to the patient) is also marked to signify the end of the catheter.
# INTENDED USE
The B. Braun Perifix catheters are regional anesthesia catheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative, perioperative and postoperative periods associated with general and orthopedic surgery as well as labor and delivery. Routes of administration are epidural. The catheters may remain indwelling for up to 72 hours.
The B. Braun Contiplex catheters are regional anesthesia catheters intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative, perioperative and postoperative periods associated with general and orthopedic surgery. Routes of administration are perineural (peripheral nerve block). The catheters may remain indwelling for up to 72 hours.
# SUBSTANTIAL EQUIVALENCE
Predicate Device – B. Braun Regional Anesthesia Catheter (K042488)
# Device Design
The proposed Perifix and Contiplex catheters are indicated for the same use as the predicate device. In addition, the proposed devices are available in the same gauges and configurations as the predicate device (19-20 Ga., open and closed-tip). The primary differences between the proposed catheters and the predicate catheters are the number of side ports included on the closed-tip version (six in the predicate vs. three in the proposed device), minor dimensional differences, and the materials of construction. The predicate device consists of a co-extruded catheter body, while the proposed device consists of a single polyamide extrusion. In addition, the depth marking ink on the proposed device differs from the predicate device. Despite the differences in design and materials, the proposed catheters meet or exceed the minimum performance requirements of the predicate catheters and do not raise different questions of safety or effectiveness.
{5}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Perifix™ and Contiplex™ Catheters August 12, 2016
# Intended Use
The predicate catheters cleared in K042488 are intended to provide, via percutaneous administration, continuous and/or intermittent infusion of local anesthetics and analgesics in the epidural space or near a nerve for regional anesthesia and pain management during the pre-operative, perioperative and postoperative periods associated with general and orthopedic surgery as well as labor and delivery. Routes of administration include epidural and perineural (peripheral nerve block). The proposed catheters are indicated for the same use as the predicate device, however, the intended routes of administration are distinguished by the respective proprietary names: 'Perifix' for the epidural route and 'Contiplex' for the peripheral nerve block route. In addition, the proposed catheters may remain indwelling for up to 72 hours. The indications for use of the predicate catheters cleared in K042488 does not include a limit for indwell time.
# Performance Testing
Dimensional and functional performance, including associate device testing (summarized in Table 1) was completed with the proposed catheters. Results of testing demonstrate that the Perifix and Contiplex catheters are substantially equivalent to the predicate device.
| Requirement | Test |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensional Specifications | Catheters meet dimensional requirements for outer diameter, inner<br>diameter, overall length and depth markings |
| Functional Testing | Flexibility<br>Markings are legible<br>Catheter tubing is translucent<br>Catheter achieves flow rate requirements<br>Catheter resists kinking<br>Catheter permits flow of fluid when attached to catheter connector<br>Catheter withstands fluid pressure without leakage<br>Catheter withstands minimum tensile force when attached to<br>connector<br>Catheter body withstands minimal tensile force<br>Catheter has sufficient elongation properties |
| Associate Device Testing | Catheter passes through appropriate size procedural<br>needle<br>without being damaged<br>Catheter passes freely through the threading assist guide<br>Catheter passes freely through sideport valve assembly |
| Sterilization | -SAL 10-6<br>-EO residual levels met per ISO 10993-7<br>-LAL bacterial endotoxin levels met per USP <161> an USP<br><85> |
Table 1: Performance Test Requirements
{6}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Perifix™ and Contiplex™ Catheters August 12, 2016
# Biocompatibility
The materials of construction of the Perifix and Contiplex catheters are either 1) used in another legally marketed device with the same type and duration of patient contact, and/or 2) are patient contacting and have been evaluated for biocompatibility through testing, which meets and/or exceeds the tests recommended for consideration in ISO 10993-1: 2009, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The tests completed with the Perifix and Contiplex catheters are provided in Table 2: Biocompatibility Tests.
| Requirement | Test |
|------------------------------------------|---------------------------------------------------------------------------------------------------|
| Biocompatibility and<br>Chemical Testing | Cytoxicity |
| | Intracutaneous Reactivity |
| | Sensitization |
| | Rabbit Pyrogenicity |
| | Systemic Toxicity |
| | Subchronic Systemic Toxicity |
| | Genotoxicity (bacterial reverse mutation, mouse lymphoma,<br>mouse peripheral blood micronucleus) |
| | Implantation |
| | Chemical Characterization of Materials (Exaggerated Extraction<br>and Simulated Use) |
Table 2: Biocompatibility Tests
Based on test results, the Perifix and Contiplex catheter materials of construction are considered substantially equivalent to the predicate device.
# CONCLUSION:
Results of functional performance and biocompatibility/chemical testing demonstrate that the Perifix and Contiplex catheters do not raise different questions of safety or effectiveness when compared to the predicate device. The Perifix and Contiplex catheters are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.