K150731 · Medline Industries, Inc. · BSO · Mar 7, 2016 · Anesthesiology
Device Facts
Record ID
K150731
Device Name
Medline Epidural Catheter
Applicant
Medline Industries, Inc.
Product Code
BSO · Anesthesiology
Decision Date
Mar 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5120
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Medline Epidural Catheter is indicated for the injection of local anesthetics into the epidural space for patients over 22 lbs. (10kgs.), for a duration of use of up to 72 hours.
Device Story
Medline Epidural Catheter is a single-use, flexible nylon tube designed for epidural anesthesia delivery. Device features a closed tip with 3 eyes for anesthetic distribution and a black radiopaque stripe for visualization. Catheter includes depth markings (1cm increments up to 25cm) and a wide warning mark indicating exit from the needle. Used in clinical settings by healthcare professionals; intended for use with a cleared epidural catheter connector. Device facilitates continuous or intermittent anesthetic administration, potentially benefiting patients requiring regional anesthesia or pain management.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -11), hemolysis (ASTM F 756-13), pyrogenicity (USP <151>), flow rate (ASTM F 623), tensile strength, and radiopacity.
Technological Characteristics
Flexible nylon tubing; 19G diameter; 800mm length; closed tip with 3 eyes; black radiopaque stripe; depth markings. Biocompatibility per ISO 10993; flow rate per ASTM F 623. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for injection of local anesthetics into the epidural space for patients > 22 lbs (10 kg) for up to 72 hours.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
Predicate Devices
B. Braun Medical, Inc. - Perifix Catheter and Contiplex Catheter (K042488)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
Medline Industries, Inc. Matt Clausen Sr. Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060
Re: K150731
Trade/Device Name: Medline Epidural Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: Class II Product Code: BSO Dated: October 27, 2015 Received: October 30, 2015
Dear Matt Clausen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150731
Device Name Medline Epidural Catheter
Indications for Use (Describe)
The Medline Epidural Catheter is indicated for the injection of local anesthetics into the epidural space for patients over 22 lbs. (10kgs.), for a duration of use of up to 72 hours.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medline. The word "MEDLINE" is written in blue, stylized, sans-serif font. Below the text is a blue, four-pointed star-like shape, with the bottom point extending further down than the other three. The logo is simple and clean, using only the color blue.
# 510(k) Summary (as required per 21 CFR 807.92)
# Summary Preparation Date
March 4, 2016
### Submitter / 510(k) Sponsor
Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060
## Contact Person
Matt Clausen Sr. Regulatory Affairs specialist Phone: 847-643-4785 Fax: 847-643-4482
#### Device Name / Classification
Device Name: Epidural Catheter Proprietary Name: Medline Epidural Catheter Common Name: Epidural Catheter or Anesthesia Conduction Catheter Classification Name: Catheter, Conduction, Anesthetic (21 CFR 868.5120, product code - BSO)
### Predicate Device
#### B. BRAUN MEDICAL, INC. - PERIFIX CATHETER AND CONTIPLEX CATHETER, K042488
### Device Description
The Medline Epidural Catheter is a single use device made of flexible, nylon tubing which is sold as sterile individually packaged and sterile packaged inside a kit/procedure tray. The epidural catheter is designed as a closed tip device with 3 eyes for distribution of anesthetic agents. The catheters have a black radiopaque stripe. The catheter is available in 19G and is 800 millimeters long. The catheter has a marked tip, with 1cm increments up to 25cm. The 10cm mark is indicated by two marks, 15cm by three marks, 20cm by four marks and 25cm by five marks. As an added safety feature, the solid wide warning mark indicates exit of catheter from needle when using a Medline Epidural Needle. This device is intended to be used in conjunction with an epidural catheter connector cleared under K051171.
### Indications for Use
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The word "MEDLINE" is written in blue, with a stylized star-like shape extending above and below the text. The star shape is also in blue and appears to be a compass rose.
The Medline Epidural Catheter is indicated for the injection of local anesthetics into the epidural space for patients over 22 lbs. (10kgs.), for a duration of use of up to 72 hours.
## Summary of Technological Characteristics
Information within this Premarket Notification demonstrates that there are no significant differences in technological characteristics between Medline's Epidural Catheter and the cited predicate device.
A summary of the technological characteristics of this subject device compared to the predicate device is provided below.
| Attribute | Subject Device | Predicate Device | Comparison |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product | Medline's Epidural Catheter | PERIFIX CATHETER AND<br>CONTIPLEX CATHETER | n/a |
| 510(k) | To be assigned | K042488 | n/a |
| Intended Use | The Medline Epidural<br>Catheter is indicated for the<br>injection of local anesthetics<br>into the epidural space for<br>patients over 22 lbs. (10kgs.) | The B. Braun Regional Anesthesia<br>Catheter is a device intended to<br>provide, via percutaneous<br>administration, continuous and/or<br>intermittent infusion of local<br>anesthetics and analgesics in the<br>epidural space or near a nerve for<br>regional anesthesia and pain<br>management during the pre-<br>operative, perioperative and<br>postoperative periods associated<br>with general and orthopedic surgery<br>as well as labor and delivery. Routes<br>of administration include epidural and<br>perineural (peripheral nerve block). | Same |
| Usage | Prescription use only | Prescription use only | Same |
| Materials | Polymer nylon | Polyamide/polyurethane | Similar |
| How supplied | Sterile | Sterile | Same |
| Diameter | 19 and 20 gauge | 19 and 20 gauge | Same |
| Length | 800 millimeters | 400-1010 millimeters | Similar |
| Type | Closed tip / 3 eyelits | Closed, open tips / 3-6 eyelits | Similar |
| Catheter<br>markings | 1-5 rings (black ink) | 1-4 rings (blue ink) | Similar |
Based on the differences in the table above the following tests were performed to mitigate risks:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a bold, sans-serif font, with the letters in red. Below the word is a blue star-like symbol with four points. The logo is simple and recognizable, and it is likely used to represent the Medline company or its products.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060-4486
# Summary of Non-Clinical Testing
The substantial equivalence of Medline's Epidural Catheter is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. Functional performance testing of the Medline Epidural Catheter demonstrates device substantial equivalence in accordance with relevant test methods.
| Standard or Reference | Test Method | Data Generated | Test Result |
|---------------------------------------------------------|-------------------------------------|----------------------------------------------------|-----------------------|
| ISO 109993-5 | Cytotoxicity | Cytotoxicity | Pass |
| ISO 109993-10 | Intracutaneous reactivity | Dermal irritation | Pass |
| ISO 109993-10 | Maximization sensitization | Dermal sensitization | Pass |
| ISO 109993-11 | Subacute toxicity | Subacute toxicity | Pass |
| ISO 109993-11 | Systemic injection | Systemic injection | Pass |
| ASTM F 756-13 | Hemolysis assay | Hemolytic potential | Pass |
| USP Pyrogen Test<br>Procedure, Section <151><br>(USP37) | Pyrogen | Non-pyrogenic | Pass |
| ASTM F 623 | Flow rate | Flow rate | Pass |
| | Extractables analysis | Extractables analysis | No toxicological risk |
| | Appearance / ink / x-ray detection | Visual appearance / ink tenacity / x-ray detection | Pass |
| | Tensile strength | Tensile properties | Pass |
| | Particulate matter | Particulate size | Pass |
| | Kink / dimensional analysis testing | Flow rate / marking accuracy | Pass |
# Summary of Clinical Testing
Not applicable.
### Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Epidural Catheter is as safe, as effective and substantially equivalent to the predicate device [K042488] as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.