The TRACOE® smart Cuff Manager is intended to maintain the HVLP (high volume low pressure) cuff pressure, of an endotracheal tube or tracheostomy tube, within the range of 20 to 30 cm H20 through passive control. It is for single patient use, and indicated for mechanically ventilated or spontaneously breathing patients. It is used under medical supervision in hospital (EMS), extended care facilities or outpatient clinics and is suitable for inter- or intra-facility transport.
Device Story
Accessory for HVLP endotracheal/tracheostomy tubes; maintains cuff pressure between 20-30 cm H2O. Device consists of transparent hard-shell casing with internal blue elastic balloon and male/female luer connectors. Operator connects device to tube inflation line and cuff inflation device (syringe/monitor); inflates inner balloon to 2/3-3/4 diameter. Passive operation: inner balloon acts as pressure reservoir. Integrated damping function allows surplus air to flow into balloon during pressure spikes (e.g., coughing/motion) to prevent tracheal damage; air flows back to cuff when pressure drops to maintain target range. Used in hospitals, EMS, extended care, and clinics; suitable for transport. Benefits patient by preventing cuff pressure fluctuations, reducing risk of tracheal injury or aspiration.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including sterilization/packaging validation (ISO 1135-1, ISO 10993-7), mechanical performance testing (luer connections, damping function, pressure maintenance), and comparative testing against predicates under simulated clinical conditions (shock, vibration, temperature, extended duration).
Technological Characteristics
Passive pressure control device. Materials: transparent hard-shell casing, elastic inner balloon. Connectivity: male/female luer connectors for integration with standard inflation lines and syringes. Sterile, single-patient use. Dimensions/form factor: spherical shell with internal reservoir. No electronic components or software.
Indications for Use
Indicated for mechanically ventilated or spontaneously breathing patients requiring high volume low pressure (HVLP) cuffed endotracheal or tracheostomy tubes. Used under medical supervision in hospitals, EMS, extended care facilities, or outpatient clinics.
Regulatory Classification
Identification
An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2016
TRACOE medical GmbH Eva Schaeffer Head of Quality Reichelsheimer Straße 1/3 55268 Nieder-Olm Germany
Re: K152865
Trade/Device Name: TRACOE® smart Cuff Manager Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: Class II Product Code: BSK Dated: August 21, 2015 Received: August 26, 2016
Dear Eva Schaeffer:
This letter corrects our substantially equivalent letter of September 27, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152865
Device Name TRACOE® smart Cuff Manager
#### Indications for Use (Describe)
The TRACOE® smart Cuff Manager is intended to maintain the HVLP (high volume low pressure) cuff pressure, of an endotracheal tube or tracheostomy tube, within the range of 20 to 30 cm H20 through passive control.
It is for single patient use, and indically ventilated or spontaneously breathing patients. It is used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities or outpatient clinics and is suitable for interor intra-facility transport.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY as required by section 21 CFR 807.92
| Date: | September 26, 2016 |
|------------------------------------|-----------------------------------------------------------------------|
| Submitter of 510(k): | |
| Company name: | TRACOE medical GmbH |
| Establishment Registration number: | 8010485 |
| Address: | Reichelsheimer Str. 1/3 |
| | 55268 Nieder-Olm |
| | Germany |
| Phone: | (+49)61369169-0 |
| Fax: | (+49)61369169-200 |
| Correspondent: | Eva Schaeffer |
| | Head of Quality |
| Device Name: | |
| Trade/Proprietary Name: | TRACOE® smart Cuff Manager |
| Common/Usual Name: | Cuff, tracheal tube, inflatable |
| Classification | Class II |
| Classification Name: | Inflatable tracheal tube cuff<br>(21 CFR 868.5750, Product Code: BSK) |
### Legally Marketed Devices
Our device is based on the legally marketed devices cited in the table below:
| Manufacturer | Device | Product Code | 510(k) # |
|--------------------|---------------------------------------|--------------|----------|
| Smith Medical Inc. | PressureEasy Cuff Pressure Controller | BSK | K833327 |
| Covidien | Model GV-10 Lanz Pressure Valve | BTR | K791045 |
#### Device Description:
The TRACOE® smart Cuff Manager is an accessory device for high volume low pressure (HVLP) cuffed endotracheal/tracheostomy tubes to maintain the cuff pressure within a range of 20 cm H₂O to 30 cm H₂O. This device includes a spherical, hard shell (transparent) outer casing with an elastic inner balloon (blue), that is attached to a base housing with both a male and female luer connector. The male connector is inserted into the inflation line (pilot balloon valve of the endotracheal/tracheostomy tube HVLP cuff) and the female connector is attached to a cuff inflation device (syringe or cuff pressure monitor). When the TRACOE® smart Cuff Manager is connected to the cuff inflation line the inner balloon is inflated with a cuff inflation device until the balloon diameter is between 2/3 - 3/4 of the outer casing. The cuff inflation device is then removed from the TRACOE® smart Cuff Manager, and after a few minutes the cuff pressure is rechecked to confirm the cuff is properly inflated inner balloon of the TRACOE® smart Cuff Manager will act as a pressure reservoir, allowing the air to flow between the balloon and the endotracheal/tracheostomy cuff, through passive control, in order to maintain the pressure range.
{4}------------------------------------------------
When the endotracheal/tracheostomy tube cuff pressure increases, due to compression through patient motion, coughing etc., the surplus air flows slowly into the inner balloon of the TRACOE® smart Cuff Manager through the integrated damping function. The damping function prevents sudden changes in cuff pressure which could result in a gap between the trachea. When the cuff pressure is reduced (following the patient motion) the inner balloon will allow the stored air to freely flow (without damping) back into the endotracheal/tracheostomy tube cuff in order to reinstate the cuff pressure to the prescribed range of 20 cm H₂O to 30 cm H₂O.
The TRACOE® smart Cuff Manager is for single patient use, and provided sterile for ease of use when connecting in a sterile environment. The device is applicable for mechanically ventilated or spontaneously breathing patients under medical supervision in hospital (EMS), extended care facilities or outpatient clinics and is suitable for inter- or intra-facility transport.
# Intended use:
The TRACOE® smart Cuff Manager is intended to maintain the HVLP (high volume low pressure) cuff pressure, of an endotracheal tube or tracheostomy tube, within the range of 20 to 30 cm H20 through passive control.
It is for single patient use, and indicated for mechanically ventilated or spontaneously breathing patients. It is used under medical supervision in hospital (EMS), extended care facilities or outpatient clinics and is suitable for inter- or intra-facility transport.
# Summary of the Technical Characteristics
For substantial equivalence, the predicate devices K833327 PressureEasy and K791045 Lanz were selected based on their intended use and similarity in design and functionality.
| | TRACOE®<br>smart Cuff Manager | Smith Medical Inc.<br>PressureEasy Cuff<br>Pressure Controller<br>(Radionics) | Covidien Model GV-10<br>Lanz Pressure Valve<br>(Lanz) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Intended Use | To maintain the HVLP<br>(high volume low<br>pressure) cuff<br>pressure, of an<br>endotracheal tube or<br>tracheostomy tube,<br>within the range of 20<br>to 30 cm H2O through<br>passive control. | To maintain the<br>HVLP (high volume<br>low pressure) cuff<br>pressure, of an<br>endotracheal tube<br>within the range of<br>20 to 30 cm H2O<br>through passive<br>control. | To maintain the HVLP<br>(high volume low<br>pressure) cuff<br>pressure, of an<br>endotracheal tube or<br>tracheostomy tube,<br>within the range of 25<br>to 33 cm H2O through<br>passive control. | |
| Device Classification | 2 | 2 | 2 | |
| 510k Number | K152865 | K833327 | K791045 | |
| Classification Product Code | BSK | BSK | BTR | |
| Regulation Number | 868.5750 | 868.5750 | 868.5730 | |
| Patient Population | | | | |
| Patients requiring a High<br>volume low pressure cuffed<br>endotracheal tube | Yes | Yes | Yes | |
| Patients requiring a High<br>volume low pressure cuffed<br>tracheostomy tube | Yes | No | Yes | |
| Patients requiring mechanical<br>ventilation | Yes | Yes | Yes | |
| Spontaneous breathing patients | Yes | Yes | Yes | |
| Patients under medical<br>supervision in hospitals, pre-<br>hospitals, emergency rooms<br>(ER), extended care facilities,<br>outpatient clinics | Yes | Yes | Yes | |
| Technical Characteristics | | | | |
| Type of control | Passive | Passive | Passive | |
| Cuff pressure range | 20 cm H2O to 30 cm<br>H2O | 20 cm H2O to 30 cm<br>H2O | 25 cm H2O to 33 cm<br>H2O | |
| Internal pressure reservoir<br>(balloon) | Elastic inner balloon<br>that expands and<br>contracts as air flows<br>to and from the cuff | Elastic inner balloon<br>that expands and<br>contracts as air<br>flows to and from<br>the cuff | Elastic inner balloon<br>that expands and<br>contracts as air flows<br>to and from the cuff | |
| Visual pressure range | Blue inner balloon<br>diameter | Green stripe | Beige inner balloon<br>diameter | |
| Damping function of air flow<br>from cuff to the balloon | Yes | No | Yes | |
| Free flow of air from the balloon<br>to the cuff | Yes | Yes | Yes | |
| Cuff pressure is maintained<br>when the device is<br>disconnected. | Yes | Yes | No | |
| Inflation via female luer<br>connector | Yes | Yes | Yes | |
| Inflation by syringe or cuff<br>pressure monitor | Yes | Yes | Yes | |
| Connection to inflation line of<br>cuff | Yes | Yes | Yes | |
| Single patient use | Yes | Yes | Yes | |
| Patient contact: | Skin surface | Skin surface | Skin surface | |
| Provided sterile | Yes | No | Yes | |
{5}------------------------------------------------
The TRACOE® smart Cuff Manager intended use represents a combination of the predicate devices intended use, therefore any differences in the intended use are not considered critical and do not affect the safety and effectiveness of the device. The similarities in design and technology are the basis and reason for substantial equivalence of the TRACOE® smart Cuff Manager to the legally marketed predicate devices.
### Summary of Non- clinical testing
Performance and mechanical testing was performed with clear acceptance criteria that addressed design, performance, and functionality of the device. In addition, predicate device comparison testing was performed which demonstrated substantial equivalence between the TRACOE® smart Cuff Manager to the legally marketed predicate devices. This testing included:
{6}------------------------------------------------
- Sterilization and Packaging Validation in accordance with ISO 1135-1 and ISO 10993-7.
- . Intra and inter-facility transportation testing includes shock and vibration testing, external pressure and temperature testing while maintaining the cuff pressure range (20 cm H₂O to 30 cm H2O).
- . Mechanical testing confirmed luer connections, pressure reservoir filling, damping function and maintaining the cuff pressure range (20 cm H₂O to 30 cm H₂O) under multiple situations (e.g. extended time periods, cuff compression).
- . Comparison testing with the predicate devices PressureEasy (K833327) and Lanz (K791045) included maintaining the cuff pressure during volume/pressure changes and over an extended period of time.
The results of this testing demonstrated that the TRACOE® smart Cuff Manager is substantially equivalent to the legally cleared predicate devices.
### Summary of Clinical testing
Clinical testing was not required to demonstrate substantial equivalence.
### Conclusion
The TRACOE® smart Cuff Manager has passed all defined criteria. The device has performed as well or better than the predicate devices and is therefore considered substantially equivalent to the cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.