K163277 · Revolutionary Medical Devices, Inc. · BSJ · Oct 13, 2017 · Anesthesiology
Device Facts
Record ID
K163277
Device Name
SuperNO2VA Device
Applicant
Revolutionary Medical Devices, Inc.
Product Code
BSJ · Anesthesiology
Decision Date
Oct 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5550
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The SuperNO2VA™ Device is a mask that creates a seal when positioned over a patient's nose and mouth, or nose only, to direct anesthesia gas, air, and / or oxygen to the upper airway during the continuum of anesthesia care. To be used under clinical supervision with adequate alarms and safety systems for monitoring and treatment of ventilatory failure. The SuperNO2VA™ Device is intended for short-term (<24 hours) use on adult patients (>30 kg.) It is a single patient use, disposable. The SuperNO2VA™ Device is contraindicated for use in long-term ventilation conditions and treatment of sleep apnea.
Device Story
SuperNO2VA™ is a 2-piece anesthesia face mask (nasal and oral chambers) for delivering anesthesia gas, air, or oxygen to adult patients. Device features a rigid polypropylene shell and PVC seal; incorporates one-way duck-bill valves allowing the oral chamber to be removed while maintaining gas flow through the nasal portion. Used in hospital and out-patient surgery settings under clinical supervision; connected to anesthesia gas machines, manual resuscitators, or hyperinflation bags. Clinicians use the mask to establish a seal for positive pressure delivery; output is the directed flow of gases to the patient's upper airway. Benefits include flexibility for oral access during procedures while maintaining airway support. Not for use with CPAP/bi-level equipment.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included internal volume/dead space, resistance through connector, duck-bill valve seal integrity, drop testing, age/shelf-life, and biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993-1.
Indicated for adult patients (>30 kg) requiring delivery of anesthesia gas, air, or oxygen to the upper airway during the continuum of anesthesia care. Contraindicated for long-term ventilation and treatment of sleep apnea.
Regulatory Classification
Identification
An anesthetic gas mask is a device, usually made of conductive rubber, that is positioned over a patient's nose or mouth to direct anesthetic gases to the upper airway.
Predicate Devices
Medical Marketing Concepts – Anesthesia Respiratory Face Mask (K953107)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 13, 2017
Revolutionary Medical Devices, Inc. % Paul Dryden Consultant Promedic. LLC 24301 Woodsage Dr. Bonita Springs, Florida 34134
Re: K163277
Trade/Device Name: SuperNO2VATM Device Regulation Number: 21 CFR 868.5550 Regulation Name: Anesthetic Gas Mask Regulatory Class: Class I Product Code: BSJ Dated: September 8, 2017 Received: September 11, 2017
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
for Tina Kiang, PhD Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K163277
Device Name
#### SuperNO2VA™ Device
Indications for Use (Describe)
The SuperNOsY A™ Device is a mask that creates a seal when positioned over a patient's nose and mouth, or nose only, to direct anesthesia gas, air, and / or oxygen to the continum of anesthesia care. To be used under clinical supervision with adequate alarms and safety systems for monitoring and treatment of ventilatory failure.
The SuperNOsVA™ Device is intended for short-term (<24 hours) use on adult patients (>30 kg.) It is a single patient use, disposable.
The SuperNOsY A™ Device is contraindicated for use in long-term ventilation and treatment of sleep apnea.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (7/17)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
#### 510(k) Summary K163277 October 12, 2017
Revolutionary Medical Devices, Inc. 6363 N Swan Rd. Tucson AZ 85718 Tel - (520)-820-3749
| Official Contact: | Cassie Vollmer, Director of RA/QA |
|----------------------------|----------------------------------------------------------------------------|
| Proprietary or Trade Name: | SuperNO2VATM Device |
| Common/Usual Name: | Patient interface Nasal and Nasal / Oral mask |
| Classification Code/Name: | BSJ – Anesthesia face mask<br>21CFR 868.5550, Class 1 |
| Device: | SuperNO2VATM Device |
| Predicate Device: | K953107 - Medical Marketing Concepts – Anesthesia<br>Respiratory Face Mask |
#### Device Description:
The SuperNO>VA™ is a 2-piece construction with a nasal and an oral portion. When assembled together the oral portion is connected to the nasal portion (chamber) contains 2 one way (duck bill) valves which are normally closed until the oral chamber is connected. Then gas passes through each portion to the patient. If one desires to remove the oral chamber during a procedure which requires oral access, then gas can still be provided via the nasal portion. Depending upon the equipment and the clinician's desire one can provide non-invasive positive pressure ventilation as well.
#### Indications for Use:
The SuperNO2VA™ Device is a mask that creates a seal when positioned over a patient's nose and mouth, or nose only, to direct anesthesia gas, air, and / or oxygen to the upper airway during the continuum of anesthesia care. To be used under clinical supervision with adequate alarms and safety systems for monitoring and treatment of ventilatory failure.
The SuperNO VA™ Device is intended for short-term (<24 hours) use on adult patients (>30 kg.) It is a single patient use, disposable.
The SuperNO2VA™ Device is contraindicated for use in long-term ventilation conditions and treatment of sleep apnea.
Patient Population: Adult patients (>30 kg)
Environment of Use: Hospital and Out-patient surgery settings
{4}------------------------------------------------
# 510(k) Summary
| Feature | Predicate<br>K953107<br>Anesthesia mask | Proposed<br>SuperNO2VA™ |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Classification | BSJ – anesthetic face mask<br>21 CFR 868.5550 | BSJ – anesthetic face mask<br>21 CFR 868.5550 |
| Indications for Use | An Anesthesia / Respiratory face mask that<br>is positioned over a patient's nose or mouth<br>to direct anesthetic gases, air, and oxygen<br>to the upper airway. | The SuperNO2VA™ Device is a mask that<br>creates a seal when positioned over a<br>patient's nose and mouth, or nose only, to<br>direct anesthesia gas, air, and / or oxygen to<br>the upper airway during the continuum of<br>anesthesia care. To be used under clinical<br>supervision with adequate alarms and safety<br>systems for monitoring and treatment of<br>ventilatory failure.<br><br>The SuperNO2VA™ Device is intended for<br>short-term (<24 hours) use on adult patients<br>(>30 kg.) It is a single patient use,<br>disposable.<br><br>The SuperNO2VA™ Device is<br>contraindicated for use in long-term<br>ventilation conditions and treatment of sleep<br>apnea. |
| Patient Population | All age patients | Adults > 30 kg |
| Environment of Use | Hospital<br>Sub-acute | Hospital<br>Sub-acute |
| Duration of Use | Single patient use, disposable | Single patient use, disposable |
| Equipment | Anesthesia gas machine<br>Manual resuscitators<br>Not intended to be used with CPAP or bi-<br>level equipment | Anesthesia gas machine<br>Manual resuscitators<br>Not intended to be used with long-term<br>ventilations conditions and with CPAP<br>equipment intended for long-term use. |
| Anatomical seal | Nasal and Oral | Nasal and Oral<br>Nasal Only |
| Components | Single part<br>Non-vented connector<br>Headstrap | 2 parts – nasal and oral chamber<br>Non-vented connector<br>Headstrap |
| Other feature | | Can be separated to be nasal only<br>Contains duck-bill valves which seal when<br>the oral chamber is removed |
| Oxygen or pressure port | Yes | Yes |
| Monitoring of airway / circuit<br>pressures | Yes | Yes |
| Sizes | Multiple - more than 2 | 2 |
| 510(k) Summary | | |
| K163277 | | |
| October 12, 2017 | | |
| Feature | Predicate | Proposed |
| | K953107 | SuperNO2VA™ |
| | Anesthesia mask | |
| Biocompatibility | ISO 10993-1<br>Externally communicating / Tissue<br>Surface Contact<br>Skin / Mucosal<br>Limited duration (< 24 hours) | ISO 10993-1<br>Externally communicating / Tissue<br>Surface Contact<br>Skin / Mucosal<br>Limited duration (< 24 hours) |
| Materials | Shell - Flexible PVC<br>Cushion - PVC | Shell - polypropylene<br>Seal – PVC |
| Provided sterile | No | No |
| Performance testing | Not available | Comparative - Internal Volume |
| | | Comparative - Pressure Drop Testing |
| | | Comparative - Pressure and Leak Rate Test |
| | | Comparative - Strap and Anchor |
| | | Connector Testing (unaged and aged) |
| | | Drop Test |
| | | Age / Shelf-life |
| Internal Volume | ~ 150 ml - medium | Medium –<br>Nasal - 53 ml<br>Full mask – 84 ml<br>Large –<br>Nasal - 87 ml<br>Full mask - 137 ml |
| Therapy Pressure Range | This mask does not have a defined<br>pressure limitation | The device does not provide therapy but the<br>maximum pressure range is recommended to<br>be 30 cmH2O in order to maintain a seal. |
| Connectors and<br>Resistance at connector | 15 mm and 22 mm<br>< $1 cmH_{2}O$ @ 60 lpm | 15 mm<br>$0.46 cmH_{2}O$ @ 50 lpm<br>$1.8 cmH_{2}O$ @ 100 lpm |
| Per ISO 17501 acceptance is a<br>reported value. | | Note subject device will not be used with<br>CPAP / bi-level equipment. |
| Operating /<br>Storage Conditions | No stated conditions are available | -28 to + 60°C<br>15 to 85% RH |
{5}------------------------------------------------
The above table compares the key features of the proposed SuperNO2VA™ mask with the identified predicate device to demonstrate that the proposed device can be found to be substantially equivalent.
Indications for Use - SuperNO2VA™ indications for use of the predicate that is Class 1 exempt. The subject device is an accessory that can be used with anesthesia gas machine, manual resuscitators, or Hyperinflation bags. It is intended to deliver anesthesia gases, air, or oxygen to the patient's upper airway. In order to do so, the mask must provide a positive seal so that the clinician can deliver the gases to the patient. The delivery of these gases is under positive pressure.
Discussion - The proposed indications are an expansion of the anesthesia face mask predicate. The expanded indications for use the anesthesia face mask is within the current practice and use of the device. The proposed device is contraindicated for use in long-term ventilation and with CPAP / bilevel equipment for long-term use.
Technology and construction – SuperNO2VA™ is a 2-part mask with a nasal and oral chamber but each must provide a proper seal for the device to be effective for the intended use. Like the predicate devices it incorporates a rigid shell with a soft sealing surface which contacts the patient's
{6}------------------------------------------------
#### 510(k) Summary K163277 October 12, 2017
face. They contain a standard fitting to connect to a circuit. The subject device is connected to devices which operate as significantly lower flow rates and the equipment has one-way valves and flow which prevent rebreathing.
Discussion - The difference of a 2-part configuration is that it allows the proposed device to represent 2 separate configurations, nasal and full face. Whether they are a full face mask or a nasal only mask the SuperNO2VATM can serve as either one. The use of connecting channels to allow for gas to pass to the oral chamber yet is sealed off when the oral chamber is removed has been evaluated for its effectiveness of the seal.
Environment of Use - The environment of use is similar to the predicate device. Discussion - An Anesthesia / Respiratory face mask can and is used in the hospital and sub-acute settings where the equipment to which it is attached may be used.
#### Discussion of Overall Differences -
We have identified differences which have been tested and compared to the predicate devices and they do not raise new risk concerns for the proposed indications for use.
## Non-Clinical Testing Summary -
We performed testing which evaluated the comparative performance and some tests specific to the subject device. These tests included:
- . Internal volume / Dead space
- Resistance through the connector
- Age and shelf-life
- . Seal of duck-bill valves (only the subject device)
- Drop testing .
- Biocompatibility testing ●
#### Differences -
Since the proposed device is limited in the equipment to which it would be used, namely, anesthesia gas machines with appropriate alarms and safety features; manual resuscitators, and Hyperinflation bags, which do not include an exhalation port.
Anesthesia gas machines provide one-way flow with directional valves and the flow rates have builtin pressure relieve features. This same one-way directional flow is in the manual resuscitators and Hyperinflation bags.
#### Biocompatibility
SuperNO2VATM based upon ISO 10993-1 is considered:
Surface Contact Mucosal membrane Limited Duration (< 24 hours)
And
Externally Communicating Tissue Limited Duration (< 24 hours)
{7}------------------------------------------------
Testing would include:
- Cytotoxicity ●
- . Sensitization
- . Irritation
## Substantial Equivalence Conclusion -
Based upon the presented information the sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent and there are no new concerns raised.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.