The Independence™ Medical Air Compressor is intended to provide compressed air to ventilators. The device is intended for use in hospitals and hospital-type facilities. The device is not to be used in the presence of flammable anesthetics. The device is intended for sale by or on the order of a physician only. The device is intended for operation by trained and qualified personnel
Device Story
Independence™ Medical Air Compressor provides filtered, oil-free, dehumidified compressed air to ventilators in intensive care units and recovery rooms. Device connects between existing air supply network and ventilator; features integrated backup function; automatically activates to supply air if primary network fails. Operated by trained/qualified personnel in hospitals/clinics. Output is compressed air; enables continuous ventilator function during supply failure; benefits patient by ensuring uninterrupted respiratory support. Not for use with flammable anesthetics.
Clinical Evidence
Bench testing only. No clinical data. Performance testing conducted per internal requirements; environmental and electromagnetic compatibility testing completed per FDA draft Reviewer Guidance (Nov '93). Device compliant with IEC 60601-1:1990, IEC 60601-1-2:1993, and 93/42/EEC.
Technological Characteristics
Portable air compressor; Class I, Type B equipment per IEC 60601-1. Delivers filtered, oil-free, dehumidified air. Standards: IEC 60601-1:1990, IEC 60601-1-2:1993, 93/42/EEC. Standalone operation.
Indications for Use
Indicated for use as a medical air compressor providing compressed air to ventilators in hospital and hospital-type facilities. Prescription use only.
Regulatory Classification
Identification
A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.
Predicate Devices
Drager Inc. Medical Air Compressor, Model 98 (K982789)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a handwritten alphanumeric string. The string appears to be "K040220". The characters are written in a cursive style, with some connections between the letters and numbers.
# SUMMARY OF SAFETY AND EFFECTIVENESS
| SUBMITTER: | eVent Medical Ltd. |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE: | January 29, 2004 |
| COMMON NAME: | Medical Air Compressor |
| PROPRIETARY NAME: | Independence™ Medical Air Compressor |
| CONTACT: | Robbie Walsh, VP Quality Assurance & Regulatory Affairs<br>eVent Medical Ltd.<br>6A Lisoban Business Park<br>Tuam Road<br>Galway,<br>Ireland.<br>Tel: + 353 91 764472<br>Fax: + 353 91 764379 |
| CLASSIFICATION: | Class II per 21 CFR 868.6250<br>Portable Air Compressor. |
### PREDICATE DEVICES:
eVent Medical Ltd.. is claiming substantial equivalence to the following predicate medical device:
| Predicate Device | 510(k) Number | Classification |
|-------------------------------------------------|---------------|-------------------------------------------------------------|
| Drager Inc. Medical Air<br>Compressor, Model 98 | K982789 | Class II,<br>Portable Air Compressor<br>per 21 CFR 868.6250 |
{2}------------------------------------------------
#### Device Description: A
The Independence™ Medical Air Compressor is a Class II device, "Portable Air Compressor'', per 21 CFR Part 868.6250. This device is designated a Class I. Type B equipment per IEC 60601-1.
The Independence™ Medical Air Compressor is an air compressor for use in intensive care units or the recovery rooms of clinics, hospitals and hospital-type facilities which delivers filtered, oil-free and dehumidified compressed air to ventilators. Using the integrated backup function, the Independence™ Medical Air Compressor can be connected between an existing air supply network and a ventilator. If the compressed air from the air supply network fails, the Independence™ Medical Air Compressor automatically starts and supplies compressed air to the ventilator.
#### Intended Use: B
The Independence™ Medical Air Compressor is intended to provide compressed air to ventilators.
The device is intended for use in hospitals and hospital-type facilities.
The device is not to be used in the presence of flammable anesthetics.
The device is intended for sale by or on the order of a physician only. The device is intended for operation by trained and qualified personnel
#### C Substantial Equivalence:
The intended use of the Independence™ Medical Air Compressor is the same as that for standard, currently marketed medical air compressors. The materials and design of this device are similar to those of the predicate device (Drager Inc. Medical Air Compressor, Model 98, K982789). The technical characteristics of the Independence™ Medical Air Compressor do not introduce new questions regarding safety or effectiveness. Furthermore, the labeling associated with the Independence™ Medical Air Compressor provides similar information as the predicate device.
Information provided in the 510(k) submission supports the determination of substantial equivalence. Performance testing was conducted per internal, company requirements while environmental and electromagnetic compatibility testing has been successfully completed in accordance with the requirements of the FDA's draft Reviewer Guidance for PreMarket Notification Submissions document (Nov '93). The Independence™ Compressor device is also compliant with various
{3}------------------------------------------------
voluntary and international standards including: IEC 60601-1:1990, IEC 60601-1-2:1993 and 93/42/EEC Medical Device Directive.
The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use.
In summary eVent Medical Ltd. has demonstrated the Independence™ Medical Air Compressor to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, featuring three horizontal lines above a wavy line, representing health and well-being.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 9 2004
Mr. Robbie Walsh Vice President, Regulatory Affairs & Quality Assurance Event Medical, Ltd. 6A Liosban Business Park, Tuam Road, Galway, Ireland
Re: K040220
Trade/Device Name: Independence Medical Air Compressor Regulation Number: 21 CFR 868.6250 Regulation Name: Compressor, Air, Portable Regulatory Class: II Product Code: BTI Dated: May 14, 2004 Received: May 17, 2004
Dear Mr. Walsh:
We have reviewed your Section 510(k) premarket notification of intent to market the device w one reviews and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce ass suited in also 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic nave been toot do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general thereore, manier are as a vice include requirements for annual registration, listing of devices, good controller provisions or ulabeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be may of subjoct to cler. adam Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); an the Ace o FFR Part 801); good manufacturing practice requirements as set forth in the quality
{5}------------------------------------------------
Page 2 – Mr. Robbie Walsh
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation in
ternation of the country of the Arth 21 CFR 1000-1050 systems (QS) regulation (21 CFR-1 at 626), and S20), 21 CFR 1000-1050.
control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. control provisions (Sections 31 t-34 of the Act, 21 cm described in your Section 510(k)
This letter will allow you to begin marketing your device to a legall This letter will allow you to begin marketung your actives of your device of your device of your device to a legally
premarket notification. The FDA finding of substantial e premarket notification. The FDA finding of substantial equil and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r and the many of the many of the Also, also, also, also, also, also, also If you desire specific advice for your device on our mooning mease note the regulation entitled,
contact the Office of Compliance at (301) 594-4646. Also, please note You contact the Office of Compliance at (301) 594-400. 71037 Part 807.97). You may obtain other
"Misbranding by reference to premarket notification of Small "Misbranding by reference to premailities under the Act from the Division of Small
general information on your responsibilities under the est its a pumber (800) 6 general information on your responsibilities uncern and the mumber (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-free number (800). Manufacturers, International and Collsumer Assistance at is teach .
(301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
| 510(k) Number: | K040220 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Independence™ Medical Air Compressor |
| Indications for Use: | The Independence™ Medical Air Compressor is indicated for use as a Medical Air Compressor providing compressed air to ventilators. |
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (21 CFR 807 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quis Secom
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K040220
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.