The INOmax® DSIR Plus delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with ventilators and respiratory care devices that the INOmax DSIR Plus has been validated with. The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breath. It may also use the INOblender® for backup. The INOmax® DSIR Plus must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. INOmax DSIR Plus is indicated for a maximum of 14 days of use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
Device Story
Device delivers nitric oxide (NO) gas to neonates in NICU or during transport; operates by injecting NO into ventilator inspiratory limb; tracks ventilator waveforms for synchronized, proportional dosing. Dual-channel architecture: channel 1 manages delivery (CPU, flow controller, injector module); channel 2 manages monitoring (separate CPU, gas cells for NO, NO2, O2, user interface). Monitoring system can shut down delivery if NO concentration exceeds 100 ppm. Includes backup delivery mode (fixed 250 mL/min NO flow with 10 L/min O2). Clinicians set NO concentration; device provides continuous monitoring and alarms. Battery supports 6 hours of operation. Benefits include precise, synchronized NO delivery to treat pulmonary hypertension while maintaining safety via independent monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including risk analysis, design reviews, module verification, system integration testing, and safety testing (IEC 60601-1, IEC 60601-1-2). Biocompatibility testing performed per ISO 18562 and ISO 10993.
Technological Characteristics
Dual-channel delivery/monitoring system. Components: delivery unit, blender, stand, gas tanks. Materials: PTFE (0.2 micron) and glass fiber (1.0 micron) in 2-stage filter; indium tin oxide glass display. Connectivity: Ethernet. Power: Battery (6-hour capacity) and external source. Software: Embedded firmware with updated drivers for new sound/Ethernet chips.
Indications for Use
Indicated for term and near-term (>34 weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. Maximum 14 days of use.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 20, 2023
Mallinckrodt Manufacturing LLC % David Trueblood Senior Director, Regulatory Affairs, Devices 6603 Femrite Drive Madison, Wisconsin 53718
Re: K211153
Trade/Device Name: INOmax DSIR® Plus Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: Class II Product Code: MRN. MRO. MRP Dated: February 16, 2023 Received: February 21, 2023
Dear Jamie Yieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
for James J. Lee, PhD Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211153
Device Name INOmax® DSIR Plus
#### Indications for Use (Describe)
The INOmax® DSIR Plus delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant on of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with ventilators and respiratory care devices that the INOmax DSIR Plus has been validated with.
The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.
The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breath. It may also use the INOblender® for backup.
The INOmax® DSIR Plus must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. INOmax DSIR Plus is indicated for a maximum of 14 days of use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(K) SUMMARY
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
# Submitter Information
| Date: | March 30, 2023 |
|---------------------------|-----------------------------------------------------------------------------------|
| Company: | Mallinckrodt Manufacturing, LLC<br>6603 Femrite Drive<br>Madison, Wisconsin 53718 |
| Contact Person: | David M. Trueblood<br>Senior Director, Regulatory Affairs, Devices |
| Email: | David.trueblood@sbiopharma.com |
| Telephone: | 608.320.2755 |
| FAX: | 908.238.6402 |
| Secondary Contact Person: | Erica Mullaney |
| Email: | Erica.mullaney@sbiopharma.com |
| Telephone: | 353.87.1193048 |
## Identification of the Device
| Device Trade Name: | INOmax DSIR® Plus (Delivery System) |
|------------------------|-------------------------------------------------------------------------------------------------------|
| Common Name: | Nitric Oxide Administration Apparatus (primary)<br>Nitric Oxide Analyzer<br>Nitrogen Dioxide Analyzer |
| Classification Name: | Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric<br>Oxide Backup Delivery |
| Device Classification: | Class II - 21 CFR 868.5165 |
| Product Code: | MRN (Primary), MRQ, MRP |
| Predicate Device(s) | K200389 |
#### Description of Device The INOmax DSR® Plus uses a "dual-channel" design to ensure the safe delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX®
{4}------------------------------------------------
delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.
Intended Use The INOmax® DSIR Plus delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with ventilators and respiratory care devices that the INOmax DSIR Plus has been validated with.
> The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.
> The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.
The INOmax® DSIR Plus must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. INOmax DSIR Plus is indicated for a maximum of 14 days of use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
All revisions of INOmax DSm® Plus utilize component technology to Technology deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. In this revision of the INOmax DSIR® Plus, the significant changes to the device include the labeling and main circuit board.
{5}------------------------------------------------
# Determination of Substantial Equivalence
The modified INOmax DSn® Plus has the same intended use as the previously cleared INOmax DSm®. All features are identical except those described in the table below.
| Feature /<br>Specification | INOmax DSiR® Plus-<br>K131686 | INOmax DSiR® Plus Device Modification |
|----------------------------|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DSIR Head | DSIR Head | Modified DSIR head with RoHS compliant<br>electrical components, revised mounting<br>method for pneumatic pressure switch, supplier<br>change to LCD component in touchscreen<br>display, revised shroud around display for<br>improved ferrite fit, monitoring board changes<br>to eliminate potential ADC override, and<br>changed to an equivalent indium tin oxide<br>glass display |
| Injector Module | Injector Module | Modified injector module with updated<br>exterior color and design and RoHS compliant<br>electrical components |
| Power Supply | Power Supply | Modified power supply due to obsolescence<br>and added protection components |
| Disk Filter | Patient Sample Gas Line<br>Disk Filter, 0.5 mircon | 2-stage filter (coalescing media is glass fiber<br>1.0 micron and hydrophobic media is PTFE 0.2<br>micron) |
| Software | Software | Modified brand name to Mallinckrodt, added<br>transmission of timestamp when cylinder is<br>open to INOmeter, minor modifications to<br>troubleshooting help, and minor sustaining<br>changes to resolve anomalies |
| Main circuit board | Main circuit board | The INOmax DSIR main board has been<br>redesigned due to the obsolescence of the<br>sound and Ethernet chips. This includes<br>inclusion of new components, board layout<br>updates and an update in software drivers to<br>support the new chips. |
| Sustaining Changes | N/A | Part obsolescence and minor sustaining<br>changes to improve reliability and<br>manufacturability. |
# Comparison to Predicate Device
# Summary of Nonclinical Tests
The following quality assurance measures were applied to the modification of the system:
- Risk Analysis .
- Requirements Review .
{6}------------------------------------------------
- . Design Reviews
- Testing on unit level (module verification)
- Integration testing (system verification)
- Performance testing (verification)
- Safety testing (verification) ●
- Biocompatibility testing of new materials in accordance with the FDA recognized standards in the ISO 18562 and ISO 10993 series.
Support for the substantial equivalence of the INOmax DSIR Plus was provided as a result of risk management and testing which included electrical safety, performance and software tests. This testing includes conformity to FDA recognized consensus standards as follows:
- IEC 60601-1:2005 Medical electrical equipment -- Part 1: General requirements for basic . safety and essential performance
- . IEC 60601-1-2:2014 General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
# Summary of Clinical Tests
The subject of this premarket submission, INOmax DSm® Plus, did not require clinical studies to support substantial equivalence.
# Conclusion
Mallinckrodt Manufacturing, LLC considers the INOmax DSm® Plus to be as safe and as effective as the predicate device, with performance substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.